  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments AN310 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
USERGUIDE
LargeVaneCFM/CMM
Anemometer/Psychrometer

ModelAN310
AN310-en-GB_V1.5 11/15
2
Introduction
CongratulationsonyourpurchaseoftheExtechAN310VaneAirflowAnemometerPsychrometer.
ThisinstrumentmeasuresAirVelocity,AirFlow(volume),AirTemperature,RelativeHumidity,Wet
BulbTemperature,andDewPointTemperature.Thelarge,easytoreadbacklitLCDincludes
primaryandsecondarydisplaysplusnumerousstatusindicators.Theoptionalroundandsquare
airflowconeallowsforquickvolumemeasurements.Thismeterisshippedfullytestedand
calibratedand,withproperuse,willprovideyearsofreliableservice.Pleasevisitourwebsite
(www.extech.com)tocheckforthelatestversionofthisUserGuide,ProductUpdates,and
CustomerSupport.
CAUTIONS
Improperuseofthismetercancausedamagetothemeterandpersonalinjury.Readand
understandthisusermanualbeforeoperatingthemeter.
Inspecttheconditionoftheprobeandthemeterforanydamagebeforeoperatingthe
meter.Repairorreplacedamagebeforeuse.
Iftheequipmentisusedinamannernotspecifiedbythemanufacturer,theprotection
providedbytheequipmentmaybeimpaired.
Thisdeviceshouldnotbemadeavailabletochildren.Itcontainshazardousobjectsaswell
assmallpartsthatcanbeaccidentallyswallowed.Themeter’sbatteriesandpacking
material
canalsobedangeroustochildren.
Intheeventthatthemeteristobeunusedforanextendedperiodoftime,removethe
batteriestoprotectagainstbatteryleakage.
Expiredordamagedbatteriescanbehazardousifallowedtocomeincontactwithskin.Use
suitablehandprotectioninsuchcases.
Donotshortcircuitbatteriesorputbatteriesinfire.
AN310-en-GB_V1.5 11/15
3
MeterDescription
MeterDescription(front)
1. ToplineLCDdisplayreading
2. BottomlineLCDdisplayreading
3. U/modeAVGanduparrowbutton
4. L/modeanddownarrowbutton
5. ONOFF(Power)andSETbutton
6. HOLD,MAXMIN,andleftarrow(Enter)button
7. Unitofmeasuresymbolforlowerdisplayrow
8.
Unitofmeasuresymbolforupperdisplayrow
9. AirVelocityVane
MeterDescription(rear)
10. Airtemperatureandhumiditysensors
11. Batterycompartment
AN310-en-GB_V1.5 11/15
4
SET
HOLD
MAX/MIN
U/mode
AVG
L/mode
U/mode
AVG
BriefKeypadDescription
PressmomentarilytoswitchthemeterONorOFF.Pressandholdthisbuttonfor2
secondstoenterorexittheSetupmode(refertoSetupmodesectionfordetails).
o Pressmomentarilytofreezethedisplayedreading(HOLDiconswitchesON);press
againmomentarilytoexittheHOLDmode(HOLDiconswitchesOFF)
o Pressandholdfor2secondstodisplaytheMIN(minimum)reading;pressandhold
againtodisplaytheMAX(maximum)reading;Pressandholdathirdtimetoexit
o ThisbuttonisalsousedintheAVERAGEmode(forrecordingreadingsoneatatime
orfora‘timed’recordingsession)andintheSETUPmode(forconfiguringthe
meter).PleasereviewtheAVERAGE
andSETUPmodesectionsfordetails
PressmomentarilytotoggleRHandAirVelocityreadings.Pressandholdfor2seconds
toaccesstheMultiPointAverageandTimedAveragemodes(theAVGiconswitches
ON).Detailedinstructionsforthesemodesareprovidedlaterinthisguide.Inthe
Setupmodepresstoselectacategory
ortoincreaseavalue(refertoSetupmodelater
inthisguide)
Presstotogglethelowerdisplaybetweenairtemperature,airvolume,wetbulb
temperature,anddewpointtemperaturereadings;AlsousedintheMultiPoint
Average,TimedAverage,andSetupmodes;refertotheAVERAGEandSETUPmodes
sectionslaterinthisguide

+ WiththemeterswitchedOFF,pressandholdthesetwobuttonstodisable
AutoPowerOFF.DetailedAUTOPOWEROFF(Sleepmode)instructionsare
providedlaterinthisguide
AN310-en-GB_V1.5 11/15
5
DisplayLayout
1. LowerDisplaydigits(airvolume,temperature,timer,ormenuparameterdisplays)
2. UpperDisplaydigits(airvelocity,humidity,ormenudisplayparameters)
3. OptionalConeorFunnelattachmenticon
4. Unitsofmeasureforductareaprogramming
5. Vaneindicator(flashesverybrieflyonpowerup)
6. DisplayHOLDicon
7. MAXmodeicon
8. MINmodeicon
9. Recordingicon
10. ElapsedTimericon
11. AVGmodeicon
12. Batterystatusicon
13. Unitofmeasureicons*
*Fulllistshown.Availableunitsofmeasureiconsvarybymodel.
AN310-en-GB_V1.5 11/15
6
Operation
AirVelocityandAirVolumeMeasurements
1. SwitchthemeterONbypressingthePOWER buttonmomentarily.PressthePOWER
buttonagaintoswitchthemeterOFF.
2. Notethatthemeterwillautomaticallyenterthesleepmodeafter20minutesofinactivity.
Todisablethisfeature:withthemeterOFF,pressandholdthePOWERandtheU/mode
buttonsfor2seconds.TheLCDwilldisplay'n'asitswitchesON.Nowthesleepmodeis
disabledandtheusermustpressthePOWERbuttonmomentarilytoswitchthemeterOFF.
3. WhenthemeterisON,theupperdisplaycanindicate
airvelocityorrelativehumidity(usetheU/modebutton
totoggle).The
lowerdisplaydigitscanindicateairflow,
dewpointtemperature,wetbulbtemperature,orair
temperature(usetheL/modebuttontoscroll).
4. Placethesensorintheairstream.Theairmustenter
thevanefromtherearofthemeter.TheAir
TemperatureandRelativehumiditysensorsare
locatedatthecenterofthevaneontherear
ofthemeter.
5. ReadthemeasurementdataontheLCD.Airvelocityreadingsareindicatedinmetersper
second(m/s)orfeetperminute(fpm).Temperaturereadings(air,dewpoint,orwetbulb)
areindicatedin°C/°F.RelativeHumidityis
indicatedin%.Airflow(volume)isindicatedin
CFM(cubicfeetperminute)orCMM(cubicmetersperminute).UsetheL/modebutton(for
lowerdisplay)andtheU/modebutton(forupperdisplay)toscrollthroughthedisplayed
readingoptions.
6. TochangebetweenImperialandMetricunitsof
measurerefertotheSETUPModesection.
7. TomeasuretheAirVolumeofaduct,firstmeasuretheduct’sarea(seetheAppendixfor
areacalculations)andthenentertheareavalueinthemeter’sSETUPmodeinin
2
orcm
2
.
Aftertheareavaluehasbeenentered,usetheL/modebuttontoscrolltotheairvolume
display(CMM/CFM)onthelowerdigits.Placethemeterinairstreamtomeasurevolume.
8. Optional:measureAirVolumewithanairflowConeorSquareadaptorbyfirstattachingone
ofthe
adaptorstothemeter(theoptionalconeadaptorkitincludesaroundandasquare
cone;kitpartnumberAN300C).Themeterwillautomaticallyrecognizethattheadaptoris
attachedtothemeterandaconeiconwillappearonthe display.
Note:regardlessoftheAREAsettingin
theSETUPmode,themeterwilldefaulttothe
dimensionsofthevaneopening.Ensurethattheairflowadaptoriswellmountedandlocked
beforetakingmeasurements.
LCDbacklight
PresstheU/modebuttontoswitchONtheLCDbacklight.Thebacklightwillautomaticallyswitch
OFFafterapproximately10secondstoconservebatterypower.Notethatthebacklighting
switchesONautomaticallywhenthemeterisswitchedONbytheuser.
AN310-en-GB_V1.5 11/15
7
RelativeHumidity,WetBulb,andDewPointMeasurements
Therelativehumiditysensorisbuiltintothecenterofthevane(rearofmeter)andcanbeusedto
measurerelativehumidity,dewpointandwetbulbtemperatures.
1. PressU/modetoswitchtothe%RHdisplayifne cessary.Themeasuredhumidityreadings
willbedisplayedontheupper
LCDdigits.
2. UsetheL/modebuttontoscrollthroughairtemperature,dewpointtemperature,wetbulb
temperature,andairvolumereadings.
3. Tomeasurethehumidityusinganoffsetreferencevalue,pleaseaccesstheSetupmodeand
inputtheoffsetvalueasdescribedintheSetupmodesection.The
LCDwilldisplaythe
measuredvalueminustheoffsetvalueoncetheoffsetisprogrammedintheSetupmode.
DataHoldMode
1. PresstheHOLDbuttonfromthenormaloperatingmodetofreezethecurrentmeasurement.
2. The‘HOLD’iconwillappearatthetopoftheLCDdisplay.
3. PressHOLDagaintoreturntonormaloperation.The‘HOLD’iconwillswitchOFFandthe
meterwillreturntodisplayingreadings
inrealtime.
MAX/MINMode
TheMAXMINfeaturerecordsthehighest(MAX)andlowest(MIN)readingscapturedsincethe
MAXMINmodewasaccessed.
1. PressandholdtheMAX/MINbuttonfor2secondstobegindisplayingthelowest(minimum)
readingsencounteredsincethismodewasaccessed.The‘MIN’iconwillappearatthetopof
theLCDdisplayindicatingthatthereadingsshownarethelowestmeasured.
2. PressandholdtheMAX/MINbuttonagainfor2secondstodisplaythemaximumreadings
encountered.The‘MAX’iconwillappearontheLCD.
3. Pressandholdthisbuttonagainfor2secondstoreturn
tothenormalmodeofoperation
(the‘MIN’and‘MAX’iconsshouldswitchOFF).
4. WhileviewingtheMAXorMINdata,usetheL/modebuttontoswitchthedisplaysothatthe
AirVolumeMINandMAXreadingscanbeviewedalso.
5. PressandholdMAX/MINfor2seconds
toreturntothenormaloperationmode.
MultiPointAverageMode
1. FromthenormaloperatingmodepresstheU/modebuttonfor2secondstoenterMulti
PointAveragemode(theAVGdisplayiconandtheassociatedblackdotwillswitchON).
2. PresstheHOLDbuttonmomentarilytorecordonereading.Theupperdisplaydigitswill
showthereading’smemory
locationnumberandthelowerdisplaywillshowthemeasured
value.PresstheHOLDbuttonagaintotakeandstoreanotherreading;thememorylocation
numberwillincrementbyoneandthecurrentreadingwillbestoredinthisnewmemory
location.Themetercanstoreupto19,999pointsinmemory.
3. PressU/modemomentarilytoseethemultipointmean(average)forallofthereadings
recordedduringthesession(the'•AVG'displayiconwillflashduringthisprocess).
4. PressL/modetoviewtheaveragereadingfortheotheravailableparameters.
5. PressU/modemomentarilytoreturnto
thenormaloperatingmode.
AN310-en-GB_V1.5 11/15
8
TimedAverageMode
1. Fromthenormaloperatingmode,pressandholdtheU/modebuttonfor2secondstoenter
theMultiPointAveragemodefirstandthenpressU/modeonemoretimemomentarilyto
entertheTimedAverageMode(theAVGandtheelapsedtimer iconswillswitchON).
2. PressHOLDtobeginrecording.Theelapsedtimerstarts(asshownontheupperdisplay
digits)andtheclockiconflashesonandoff.Thelongestamountoftimethemetercan
operateinTimedAverageModeis19,999seconds.
3. PressU/modetostop
theclockandtocalculatethetimedaverage.TheupperLCDwill
displaytheelapsedtimeandthelowerLCDwillshowtheaveragedreadings.TheLCDwill
flashthe'AVG'andtheelapsedtimericons.
4. PressL/modetoviewtheaverageforotheravailablemeasurements.
5. PressU/modetoreturntothenormaloperatingmode.
SetupMode
EnteringSetupMode
PressandholdthePOWERSETbuttonfor2secondstoenterSetupmode.Twooptionsare
available:P1.0:UnitandP2.0:AREA.
Usetheup/downbuttonstotogglebetweenP1.0UNITandP2.0AREAoptions.
P1.0:UnitSelection
1. Onceinsetupmode,attheP1.0UNITdisplay,momentarilypress
theHOLDbuttontoenter
theP1.0UNITmenu.
2. Useeitherarrowbuttontoswitchbetweenmetricandimperialunits.Theavailableunits
are:airvelocity(m/s,fpm),temperature(C,F),airvolume(CMM,CFM),andareasize(cm²,
inch²).
3. PressHOLDmomentarilytoreturntotheP1.0
display.
4. PresseitherarrowbuttontomovetotheP2.0AREAoptionandfollowthestepsbelow.
P2.0:AreaSizeSelection(entersquareinchesorsquarecentimeters)
1. Insetupmode,fromtheP2.0AREAdisplay,pressHOLDtoentertheAREAsettingmenu.The
LCDwillshow‘99999’in
thelowerdisplaywiththefirstdigitflashing.
2. Usetheuparrowbuttonchangetheflashingdigit(from0through9).
3. Usethedownarrow buttontomovetothenextdigit.Usetheuparrowbuttonagainto
changetheflashingdigit.Continueinthiswayuntil
thedesiredAREAvalueisdisplayed.
4. PresstheHoldbuttonagainmomentarilytostorethe valueandtoreturntotheP2.0display.
5. PresseitherarrowbuttontomovetotheP3.0RHOFFSET optionandfollowthestepsbelow.
P3.0:RelativeHumidity(RH)OffsetSelection
1. WheninsetupmodeP3.0pressHOLDtoentertheRHoffsetsetting.TheLCDwillshow0.0
RHintheupperdisplay.
2. PressU/modetoincreaseoffsetvalueandL/modetodecreaseoffset.
3. PresHOLDoncemoretoreturntoP3.0.
4. PressandholdthePOWERSETbutton
for2secondstoreturntothenormaloperating
mode.
AN310-en-GB_V1.5 11/15
9
BatteryReplacement
WhenappearsontheLCD,theAAAbatteriesmustbereplaced.Thebatterycoverislocated
onthebackofthemeter.
1. Opentherearbatterycompartmentbycarefullypullingdownonthecompartment’slatch.
2. Thebatterycompartmentcovershouldbecompletelyremovedbeforeproceeding.
3. Replacethe
four(4)1.5V‘AAA’batteriesensuringproperpolarity.
4. Closethebatterycompartmentbeforeattemptingtousethemeter.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatteries
werepurchased,orwherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuseris
obligatedtotakeendoflifedevicestoadesignatedcollectionpointforthedisposal
ofelectricalandelectronicequipment.
OtherBatterySafetyReminders
o Neverdisposeofbatteriesin
afire.Batteriesmayexplodeorleak.
o Nevermixbatterytypes.Alwaysinstallnewbatteriesofthesametype.
AN310-en-GB_V1.5 11/15
10
Specifications
AirVelocity Ranges Resolution Accuracy
m/s(meterspersec) 0.230m/s 0.01m/s ±(1.5%rdg+0.3m/s)
fpm(feetperminute) 405900fpm 1fpm ±(1.5%rdg+59fpm)
RelativeHumidity Range Resolution Accuracy
RH 0.1%99.9%RH 0.1RH ±3%RH(1090%RH)
±5%RH(<10%or>90%)
AirFlow Ranges Resolution AreaRange
CMM(cubicmeters/min) 099999m
3
/min
0.1upto9999.9
then1.0
0to99999cm
2
CFM(cubicft/min) 099999ft
3
/min
0.1upto9999.9
then1.0
0to99999in
2
Temperature Ranges Resolution Accuracy
o
C/
o
F
AirandWetBulb:
20to60
o
C(4to140
o
F)
DewPoint:
5to60
o
C(23to140
o
F)
0.1
o
C/
o
F
±0.6°C(20to50°C)
±1.2°C(51to60°C)
±1.1°F(5to122°F)
±2.2°F(123to140°F)
Circuit CustomLSImicroprocessor
Display Dualfunction13mm(0.5")4digitLCD
Samplingrate 1readingpersecondapprox.
Airvelocity/flowsensor Conventionalangledvanearmswithlowfrictionballbearing
Sensors NTCtypeprecisionthermistors
AutomaticPoweroff Autoshutoffafter20minutestopreservebatterylife
(sleepmode)
OperatingTemperature 0°Cto50°C(32°Fto122°F)
StorageTemperature 10to50
o
C(14to122
o
F)
OperatingHumidity <80%RH
StorageHumidity <90%RH
OperatingAltitude 2000meters(7000ft)maximum
BatteryPower Four(4)‘AAA’1.5Vbatteries
BatteryLife >40hours
Batteryconsumption 8.3mADC(approx.)
Weight 725g(1.6lbs.)includingbatteryandprobe
Dimensions
Maininstrument:269x106x51mm(10.6x4.2x2”)
Vaneinnerdiameter:100mm/3.94inches
OptionalAirFlowAdaptors:
Round:210mm(8.3”)Diameter
Square:346x346mm(13.6x13.6”)
AN310-en-GB_V1.5 11/15
11
ErrorMessages
Errorcodesandsolutions
AirTemperature
E02:Temperaturereadingislowerthanthelowrangelimit.Placethemeterinroom
temperaturefor30minutes.IfE02continues,sendtheunittoExtechforrepair.
E03:Temperatureexceedsthehighrangelimit.Placethemeterinroomtemperaturefor30
minutes.IfE03continuestoappear,sendtheunittoExtechforrepair.
E31:TheAtoDconverterrequiresreplacing.SendthemetertoExtechInstrumentsforrepair.
RelativeHumidity(RH)
E04:Temperatureerrorcorruptedthehumidityreading.Checkthetemperatureerrorsolution
section.
E11:RHcalibrationerror.SendtoExtechInstrumentsforrepair.
E33:RHmeasurementcircuiterror.SendtoExtechInstrumentsforrepair.
DewPoint(DP)andWetBulb(WBT)Temperature
E04:TemperatureorRHerror.ChecktemperatureorRHerrorsolutions.
Airvelocity
E03:Velocityexceedsthehighrangelimit.Testthemeterusinganairvelocitythatisknown
tobewithinthespecifiedrangelimits.SendtheunittoExtechforrepairiftheE03error
messagepersists.
Airvolume
E03:Readingexceedsthemeter’sdisplaylimit.CheckthattheAREAsettingiscorrectinthe
SetupMode.
E04:Airvelocityerror.ReturnmetertoExtechforrepair.
OtherErrors
E32:MemoryICerror.Rebootthemeterandcheckitagain;sendittoExtechInstrumentsfor
repairiftheerrormessagepersists.
Nodisplay
1. Checkthatthebatteriesaremakinggoodcontactandarealignedforcorrectpolarity.
2. Replacebatteries.
DisplayswitchesOFFautomatically
Thissymptomcouldbethenormal‘SleepMode’whichswitchesthemeterOFFafter20
minutesofinactivity.Ifthisisnotthe‘SleepMode’,checkwhetherthelowbatteryindicator
appearsbeforetheLCDattemptstoswitchON;ifyes,replacethebatteries.
AN310-en-GB_V1.5 11/15
12
Appendix:UsefulEquationsandConversions
Areaequationforrectangularorsquareducts

Areaequationforcircularducts
Cubicequations

NOTE:Measurementsmadeininchesorcentimetersmustbeconvertedtofeetormetersbeforeusingtheseformulae.
UnitofMeasureConversionTable
m/s ft/min knots km/h MPH
1m/s 1 196.87 1.944 3.6 2.24
1ft/min 0.00508 1 0.00987 0.01829 0.01138
1knot 0.5144 101.27 1 1.8519 1.1523
1km/h 0.2778 54.69 0.54 1 0.6222
1MPH 0.4464 87.89 0.8679 1.6071 1
Copyright©20142015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
Width(W)
Height(H) Area(A)=Width(W)xHeight(H)
CFM(ft
3
/min)=AirVelocity(ft/min)xArea(ft
2
)
CMM(m
3
/min)=AirVelocity(m/sec)xArea(m
2
)x60
Radius
Area(A)=pi xr
2
Wherepi=3.14andr
2
=radiusxradius
/