646 Podiatry Procedures Chair

Midmark 646 Podiatry Procedures Chair, 647 Barrier-Free® Podiatry Procedures Chair Installation guide

  • Hello! I am an AI chatbot trained to assist you with the Midmark 646 Podiatry Procedures Chair Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
1
MA7829i
Podiatry Chair Installation
Applies to Models:
646 (all)
647 (all)
Special Tools:
level
003-1895-00 Rev. C 5/24/16
Step 1: Remove table from skid.
A) Remove shipping bolts.
B) Remove shipping brace. (Rotational models only)
C) Carfefully remove table from skid.
Shipping Bracket
(Rotational models only)
Shipping Bolts
NextNext
2
MA7830i
Step 4: Connect foot control.
A) Slide narrow portion of cord connector into bracket.
B) Connect cord to inlet (see Note).
C) Tighten locking ring.
Note: The cord connector is “keyed” and must be aligned with inlet.
Locking Ring
Step 2: Connect power cord.
Note: 646 models have a dual power cord. Both
cords must be connected for proper operation.
647 models have a detachable power cord.
Step 3: Install magnetic base covers.
Step 5: Level the table.
A) Adjust four leveling feet as necessary.
Note: Table must be level to ensure proper operation.
647 only
Before connecting the Midmark device power cord into an electrical power
source outlet, it is recommended to contact a local licensed electrician or a
local Authority Having Jurisdiction over the internal building power distribution
to verify that the device when powered up will comply with all local electrical codes.
Use the electrical power rating marked on the device when determining the
appropriate electrical branch circuit and outlet requirements. For 115V rated device
models with a current rating at or above 12 Amps, a dedicated branch circuit may be
required if the branch circuit protection is provided by a 15 Amp breaker.
BackBack
/