2 PrioCHECK™ Porcine Trichinella Ab Strip Kit Instructions for Use
Additional material required
Laboratory equipment according to national safety regulations.
• Demineralized or water of equal quality must be used
• Dummy plate, used for sample dilution (e.g. clear colorless
round bottom 96 well plates) or equivalent
• Single channel pipette (10–100 µL)
• Single channel pipette (20–200 µL)
• Multichannel pipette (5–50 µL)
• Multichannel pipette (50–300 µL)
• Pipette tips (as recommended by pipette manufacturer)
• Solution reservoirs
• Dedicated blood collection tubes
• Dedicated meat juice collection tubes
Plate Reader, e.g. Tecan Sunrise or equivalent. The reader has to
have an appropriate filter set to read the plates at 450 nm.
(1) Unless otherwise indicated, all materials are available through thermofisher.com.
Test procedure
Precautions
• National Safety Regulations must be strictly followed.
• The PrioCHECK™ Porcine Trichinella Ab Strip Kit must be performed
in laboratories suited for this purpose.
• Samples should be considered as potentially infectious and all items
which contact the samples as potentially contaminated.
Notes
To achieve optimal results with the PrioCHECK™ Porcine Trichinella Ab
Strip Kit, the following aspects must be considered:
• The Test Procedure protocol must be strictly followed.
• Pipette tips have to be changed for every pipetting step.
• Separate solution reservoirs must be used for each reagent.
• Kit components must not be used after their expiry date or if changes
in their appearance are observed.
• Kit components of different kit lot numbers must not be used together.
• Demineralized water or equal must be used for the test.
Sample preparation
• Serum can be obtained using standard methods.
• If meat juice is tested, a piece of muscle tissue e.g. 10 g (preferable tongue,
diaphragm or masseter) is either freeze-thawed in a dedicated device or
alternatively the piece of meat can be squeezed to obtain meat juice.
Reconstitution of control samples
1. Reconstitute the control sample by adding 150 µL of Demineralized
Water (delivered with the kit).
2. Mix by vortexing thoroughly and inverting the vial several times or
use already reconstituted control samples that have been stored at
−20°C to −80°C.
Sample dilution for serum samples
1. Use a Dummy Plate or equivalent for first sample dilution (see Table 1).
2. Add 10 µL of Positive Control to wells A1 and B1 of the Dummy Plate.
3. Add 10 µL of Negative Control to wells C1 and D1 of the Dummy Plate.
4. (Optional) Add 10 µL of Weak Positive Control to wells E1 and F1 of the
Dummy Plate.
5. Add 10 µL of serum samples to the remaining wells of the Dummy Plate.
6. Add 90 µL of Sample Diluent to each well of the Dummy Plate and
mix by pipetting up and down 5 times.
7. Add 80 µL of Sample Diluent to each well of the Test Plate.
8. Transfer 20 µL of the diluted samples and controls from the Dummy
Plate to the corresponding wells of the Test Plate and mix by pipetting
up and down 5 times.
Sample dilution for meat juice samples
1. Use a Dummy Plate or equivalent for first sample dilution (see Table 1).
2. Add 10 µL of Positive Control to wells A1 and B1 of the Dummy Plate.
3. Add 10 µL of Negative Control to wells C1 and D1 of the Dummy Plate.
4. (Optional) Add 10 µL of Weak Positive Control to wells E1 and F1 of
the Dummy Plate.
5. Add 90 µL of Sample Diluent to each control in the Dummy Plate.
6. Add 100 µL of meat juice sample to the remaining wells of the
Dummy Plate.
7. Add 80 µL of Sample Diluent to each well of the Test Plate.
8. Transfer 20 µL of the samples and diluted controls from the
Dummy Plate to the corresponding wells of the Test Plate.
Sample incubation
1. Incubate the samples on the Test Plate for 30±1 minutes at room
temperature (22±3°C).
2. Wash the Test Plate four times with 300 µL 1x Wash Fluid working
solution.
Remark: If you use a plate washer be sure that no needles are clogged.
Conjugate incubation
Preparatory steps
• Dilute the needed amount of Conjugate (30x) 30 fold in Conjugate
Diluent (e.g. add 400 µL Conjugate to 11.6 mL Conjugate Diluent for
one full plate.
Conjugate incubation
1. Add 100 µL of the diluted Conjugate to each well of the Test Plate.
2. Incubate the Test Plate for 30±1 minutes at 22±3°C.
3. Wash the Test Plate four times with 300 µL 1x Wash Fluid working
solution.
Detection
Substrate reaction
1. Add 100 µL of the Chromogen (TMB) substrate to each well on the
Test Plate.
2. Incubate the Test Plate for 15±1 minutes at 22±3°C.
3. Add 100 µL of the Stop Solution to each well of the Test Plate.
Remark: Start the addition of stop solution 15±1 minutes after the first well
was filled with Chromogen (TMB) Substrate solution. Add the Stop Solution in
the same order as the Chromogen (TMB) Substrate solution was dispensed.
Note: The color of the Positive Controls will change from blue to yellow.
Detection
1. Shake the Test Plate shortly (5–10 s) either on an orbital shaker
(~300 rpm) or manually on the working bench.
2. Read the Test Plate in the ELISA reader at 450 nm within 15 minutes.
Recommendation: Use a reference filter at 620 nm.
Result interpretation
Calculation of results
PP = (OD450 Sample / OD450 Positive Control) × 100
Validation criteria
1. The mean OD450 of the Positive Controls must be >1.0.
2. The mean OD450 of the Negative Controls must be <0.2.
3. If the Weak Positive Control was used, the mean percentage of
positivity (PP) of the Weak Positive Controls must be >35%.
If these criteria are not met, the results are invalid and the samples have to
be retested.
Interpretation of results
• Results obtained above or equal the cut-off of 15 PP are considered
positive.
• Results obtained below the cut-off of 15 PP are negative.
Recommended plate layouts
The following plate layouts allow for efficient transfer of pre-diluted samples
and controls from the Dummy Plate to the Test Plate (P — Positive Control;
N — Negative Control; W — Weak Positive Control (optional); S — Sample).
Table 1 Plate layouts: Dummy Plate and Test Plate