2 PrioCHECK™ Porcine HEV Ab Strip Kit Instructions for Use
Test procedure
Precautions
• National Safety Regulations must be strictly followed.
• The PrioCHECK™ Porcine HEV Ab Strip Kit must be performed in
laboratories suited for this purpose.
• Samples should be considered as potentially infectious and all items
which contact the samples as potentially contaminated.
Notes
To achieve optimal results with the PrioCHECK™ Porcine HEV Ab Strip
Kit, the following aspects must be considered:
• The test procedure must be strictly followed.
• Pipette tips have to be changed for every pipetting step.
• Separate solution reservoirs must be used for each reagent.
• Kit components must not be used after their expiry date or if changes
in their appearance are observed.
• Kit components of different kit lot numbers must not be used together.
• Demineralized water or equal must be used for the test.
Sample preparation
• Serum can be obtained using standard methods.
• If meat juice is tested, a piece of muscle tissue e.g. 10 g is either freeze-
thawed in a dedicated device or alternatively the piece of meat can be
squeezed to obtain meat juice.
• Heat inactivated samples may lead to an increased background and
are therefore not recommended to be used.
Sample dilution for serum samples
Serum samples are diluted 1:100, controls 1:100.
1. Use a Dummy Plate or equivalent for first sample dilution.
2. Add 10 µL of Positive Control to wells A1 and B1 of the Dummy Plate.
3. Add 10 µL of Cut-off Control to wells C1 and D1 of the Dummy Plate.
4. Add 10 µL of Negative Control to wells E1 and F1 of the Dummy Plate.
5. Add 10 µL of serum samples to the remaining wells of the Dummy
Plate.
6. Add 90 µL of Dilution Buffer to each well of the Dummy Plate and mix
by pipetting up and down 5 times.
7. Add 90 µL of Dilution Buffer to each well of the Test Plate.
8. Transfer 10 µL of the diluted samples and controls from the Dummy
Plate to the Test Plate and mix by pipetting up and down 5 times.
Sample dilution for meat juice samples
Meat juice samples are diluted 1:10, controls 1:100.
1. Use a Dummy Plate or equivalent for first sample dilution.
2. Add 10 µL of Positive Control to wells A1 and B1 of the Dummy Plate.
3. Add 10 µL of Cut-Off Control to wells C1 and D1 of the Dummy Plate.
4. Add 10 µL of Negative Control to wells E1 and F1 of the Dummy Plate.
5. Add 90 µL of Dilution Buffer to wells containing controls A1 to F1 of
the Dummy Plate.
6. Add 100 µL of meat juice samples to the remaining wells of the
Dummy Plate.
7. Provide 90 µL of Dilution Buffer to all wells of the Test Plate.
8. Transfer 10 µL of meat juice samples and diluted controls from the
Dummy Plate to the Test Plate and mix by pipetting up and down
5 times.
Sample incubation
1. Carefully tape over the Test Plate with an unused sealing film.
2. Incubate the samples on the Test Plate for 60±1 minutes at 37±3°C.
3. Wash the Test Plate four times with 300 µL Wash Fluid working
solution (see Appendix A).
4. Tap the plate over a paper towel to remove any residual liquid from
the wells.
Conjugate incubation
Preparatory steps
Dilute the needed amount of Conjugate (100x) 100 fold in Dilution Buffer
(e.g. add 100 µL Conjugate to 9.9 mL Dilution Buffer for one full plate; see
Appendix A).
Conjugate incubation
1. Add 100 µL of the diluted Conjugate to each well of the Test Plate.
2. Carefully tape over the Test Plate with an unused sealing film.
3. Incubate the Test Plate for 30±1 minutes at 37±3°C.
4. Wash the Test Plate four times with 300 µL Wash Fluid working
solution (see Appendix A).
5. Tap the plate over a paper towel to remove any residual liquid from
the wells.
Detection
Substrate reaction
1. Add 100 µL of the Chromogen (TMB) Substrate to each well on the
Test Plate.
2. Incubate the Test Plate for 30±1 minutes at 22±3°C.
3. Add 100 µL of the Stop Solution to each well of the Test Plate.
Remark: Add the Stop Solution in the same order and time as the
Chromogen (TMB) Substrate solution was dispensed.
Note: The color of the Positive Controls will change from blue to yellow.
Detection
1. Shake the Test Plate shortly (5–10 s) either on an orbital shaker
(~300 rpm) or manually on the working bench.
2. Read the Test Plate in the ELISA reader at 450 nm within 60 minutes.
Recommendation: Use a reference filter at 620 nm.
Result interpretation
Validation criteria
1. The mean OD450 of the Positive Controls minus the OD450 of the Cut-off
Control must be ≥0.3.
2. The mean OD450 of the Cut-off Control minus the OD450 of the negative
control must be ≥0.05.
3. The mean OD450 of the Negative Controls must be <0.15.
If these criteria are not met, the results are invalid and the plate has to be
retested.
Cut-off calculation
The Cut-off is calculated as mean OD450 of the Cut-off Control multiplied
with 1.2.
mean OD450 of Cut-off Control × 1.2 = cut-off
Interpretation of results
• Results obtained above or equal to the cut-off are considered positive.
• Results between the OD450 of the Cut-off Control and the cut-off (see
calculation above) are doubtful and it is recommended to retest those
samples. If they remain in the doubtful are, a second sample taken at a
later time point should be taken and tested.
• Results obtained below the cut-off are negative.