Nuaire NU-5510E Operation and Maintenance Manual

Category
Carbon monoxide (CO) detectors
Type
Operation and Maintenance Manual
OM0135 1
Air-Jacketed Automatic CO2 Incubator
Models
NU-5510
NU-5510E
Operation and Maintenance Manual
October, 2010
Revision 12
(Series 6 & Higher)
Manufactured By:
NuAire, Inc.
2100 Fernbrook Lane
Plymouth, MN 55447
Toll-Free: 1-800-328-3352
In Minnesota: (763)-553-1270
Fax: (763)-553-0459
OM0135 2
Air-Jacketed DHD Autoflow Automatic CO2 Incubator
Operation & Maintenance Manual
NU-5510/E
TABLE OF CONTENTS
Section No. 1.....................................................................General Description
Section No. 2.....................................................................Performance Parameters and Features
Section No. 3.....................................................................Models & Features
Section No. 4.....................................................................Test Performance & Procedures
Section No. 5.....................................................................Warranty
Section No. 6.....................................................................Shipments
Section No. 7.....................................................................Installation
Section No. 8.....................................................................DHD Autoflow Operation
8.1 .............................................................................Sterility
8.2 .............................................................................Humidity
8.3 .............................................................................System Introduction
8.4 .............................................................................Front Control Panel
8.5 .............................................................................DHD Autoflow Rear Panel
8.6 .............................................................................Run Mode Operator Interactions
8.7 .............................................................................Setup Mode Operator Interactions
8.8 .............................................................................Diagnostic Interactions
8.8.1 .......................................................................Diagnostic Mode Test
Section No. 9.....................................................................Calibration
9.1 .............................................................................Chamber Temperature Calibration
9.2 .............................................................................Door Temperature Calibration
9.3 .............................................................................CO2 Calibration
Section No. 10...................................................................Maintaining Your DHD Autoflow
10.1...........................................................................Heated Decon Cycles
10.2...........................................................................Chemical Decontamination
Section No. 11...................................................................Error Indicators & Troubleshooting
Section No. 12...................................................................Remote Alarm Contacts
Section No. 13...................................................................Electrical/Environmental Requirements
Insert..................................................................................Replacement Parts List
MANUAL DRAWINGS
BCD-09807.............................................NU-5510/E Direct Heat CO2 Incubator
BCD-09409.............................................Plenum/Shelf Installation
BCD-09709.............................................Front Control Panel
BCD-09809.............................................Rear Panel
ASSEMBLY DRAWINGS
BCD-09810.............................................Front Panel Assembly
BCD-09811.............................................Chamber Assembly
BCD-09812.............................................Component Layout
BCD-06782.............................................Door Heater & Switch Assembly
BCD-09813.............................................Rear Panel Assembly
BCD-09834.............................................NU-5510/E Tubing Configuration
ELECTRICAL SCHEMATICS
BCD-09710.............................................Electrical Schematic, NU-5510/E
BCD-09444.............................................Main Control Board Electrical Schematic
OM0135 3
DHD Autoflow Air-Jacketed
Automatic CO2 Incubator
Models: NU-5510/E
Operation and Maintenance Manual
1.0 General Description
The NuAire DHD Autoflow Automatic CO2 Air-Jacketed Incubator has been designed to provide a reliable
controlled in-vitro environment for optimum tissue cell culture growth. The chamber also provides an environment
for the storage and preservation of embryos, gametes and animal tissue cell cultures at on near body temperature.
There are five parameters that contribute to optimum growth conditions. These are:
1. Humidity
2. Precise temperature control
3. Precise CO2 control
4. Sterility
5. Reliability
Like all NuAire equipment, this Incubator has been designed to provide the highest quality standards of
performance with matching computer technology, precise temperature control, and CO2 gas control system
combining state-of-the-art technology with years of design, quality, and manufacturing experience.
In order to accomplish the foregoing objectives, this Incubator features the following:
1.1 Incubator Chamber
The design and size of the DHD Autoflow inner chamber provides a large capacity, and ease of use. The
chamber walls are directly heated by physically attached foil heating elements on the sides, bottom, top,
and back of the chamber, providing a temperature uniformity of +0.3C. A space-age high-density
insulation that has a high "R" rating covers the complete outer surfaces of the Incubator inner chamber.
1.2 Incubator Blower Fan & HEPA Filter
A continuous operating fan motor drives a blower wheel within the upper air plenum and sidewall duct
system. Air is constantly re-circulating within the chamber keeping every cubic inch of volume at a
uniform temperature. This airflow is distributed uniformly and at very low velocity, so as not to influence
culture growth. A large replaceable HEPA filter cartridge continually filters the air that circulates in the
chamber.
1.3 Incubator Control Electronics
The NuAire Incubator Control Electronics is a state-of-the-art microcomputer based control system
specifically designed to service the precise control requirements of the chambers environment, providing
optimum programmable conditions for culture growth. The microcomputer is “user-friendly” with status
indicators, LED display of control parameters and five touch control keypads to permit efficient operator
entry of data.
The microcomputer is supported with Read Only Memory (ROM) containing executable software,
Random Access Memory (RAM) for temporary storage, and Electronically Erasable Programmable Read
Only Memory (EEPROM) for control set points and parameters. The EEPROM provides for indefinite
storage of these values during periods of power off or power interruption (power fault tolerant). The
microcomputer includes a complete internal diagnostic software package that permits fault isolation
detection down to the failed component.
OM0135 4
1.4 Incubator CO2 Control
The NuAire direct heat Incubator incorporates a microprocessor-based, non-dispersive infrared CO2
induction sensor. The amount of energy received at the detector is an approximate logarithmic function of
the CO2 concentration in the gas between source and detector. The wavelengths used are absorbed only by
CO2 making the measurement insensitive to other components, such as water vapor. Detector linearization
is performed with 32 bit digital accuracy. Advanced design provides a very stable output minimizing drift
and requiring less frequent calibration. The output is digital, alleviating tolerances brought about by
analog signals. The sensor is within the chamber air plenum so very accurate CO2 control is achieved.
The CO2 sensor consists of a control board and a detector assembly connected by a cable. Calibration of
this control is accomplished through the front of the unit where there is a CO2 test port.
1.5 Incubator Construction
The outer shell of the air-jacketed Incubator is cold-rolled steel with a powder coat paint finish. The front
frame surface of the outer shell is heated with a foil type heater directly attached behind the front perimeter
opening.
The front frame perimeter heater, as well as the outer front door heater is duty cycle controlled (manually
adjusted for specific ambient conditions) to balance the heat that reaches the chamber and thus reduces the
possibility of condensation forming on the inner glass door and the inner chamber walls.
The inner chamber is 16 gauge, type 304, polished stainless steel using crevice-free construction, which
provides an easily cleanable inert surface that does not in itself promote biological growth. In addition, all
shelves, shelf supports, guide rails and the air plenum are easily removable and can be autoclaved to
remove contamination.
Remember: The chamber is not selective. The growth environment is applied equally to all
microorganisms within the chamber.
1.6 Incubator Humidity
A relative humidity level of up to 95% is achieved in the Incubator by the use of a stainless steel pan filled
with distilled water no purer than 1 mega ohm, and placed on the bottom of the chamber. It is necessary to
set the duty cycle (0-100%) of the door and front perimeter heater in proper proportions to reduce the
possibility of condensation forming on the glass inner door and the chamber walls. It is also important to
thoroughly wipe the walls and the glass door clean before adding the humidity water pan. Condensation
will occur more readily at contamination points. There is no electronic sensing and thus no automatic
control of the humidity level in the air jacketed Incubator. An air pump injects fresh air into the chamber
at a preset duty cycle to reduce condensation forming on the chamber walls and front glass door. The air
injections are settable by the user to help control condensation. (See section 8.8 for this option.)
1.7 Cabinet Ventilator Fan
An axial fan is mounted to the bottom cover panel of the Incubator and runs continuously when the unit is
switched on. This fan pulls air into the cabinet shell thru the top panel louver openings and discharges it
out of the bottom of the unit. The operation of this fan is necessary to assure accurate chamber
temperature control over the 60F to 85F ambient temperature range that the equipment may see.
OM0135 5
2.0 Performance Parameters and Features
Both the interior and exterior of the DHD Autoflow are constructed of 16-gauge material. The interior
is highly polished type 304 stainless steel, using crevice-free construction. All exposed edges are
deburred to insure no sharp edges. The exterior is cold rolled steel finished in a powder coated
polyurethane finish, which is resistant to chemicals and easily cleaned using mild household
detergents.
The DHD Autoflow's microcomputer temperature control system has two temperature sensors located
in the chamber. The temperature sensors compare the values to a set point and execute a differential
control algorithm that energizes a solid-state switch, supplying power to the heaters.
Up to 17 shelves can be placed inside the chamber (4 supplied).
Space-age high temperature material is used to insulate the inner chamber walls.
Foil heaters are directly attached to the chamber walls and the top and bottom surfaces of the inner
chamber.
Incubators can be stacked for space saving, and can still be serviced from the front.
Easily removable inner chamber plenum shelves and rails for sterilization.
Most electronics, motors, pumps, and valves are fully accessible from the front of the unit.
A CO2 sample port is provided on the front panel to check the concentration of CO2 in the chamber.
The CO2 percentage is controlled by a solid-state gas infrared sensor, which provides accurate
monitoring of CO2 affected minimally by changes in temperature or humidity levels in the chamber.
Automatic recovery of the CO2 level after a 1minute door opening to 5.0 + 0.2% is within a 4-1/2
minute period.
The outer door includes an internal radiant heater in order to minimize condensation on the inner glass
door. A magnetic outer door gasket helps to insure a tight seal against the cabinet.
The inner glass door is 3/16” tempered with smooth-ground edges and seals are tight against a U-
grooved silicone rubber gasket. The door latch is cam action. An all solid-state magnetic switch
monitors door motion.
All control electronics are protected with a circuit breaker that may trip at 110% of loading rating but
will trip at 145% of load rating in less than 2 seconds. Should the circuit breaker open (pop-out button
will appear), merely depress to reset.
The Incubator has factory-installed adjustable leveling legs to compensate for uneven laboratory
surfaces.
OM0135 6
3.0 Models & Features
NuAire offers a basic Model NU-5510/E Air-Jacketed DHD Autoflow CO2 Incubator.
3.1 Dimensions (see also Specification Drawing BCD-09807)
Overall Dimensions - inches (mm): Model NU-5510/E
Height: Exterior: 39.5 Inches (1003.3mm)
Width: 25.5 Inches (647.7mm)
Depth: 26.5 Inches (673.1mm)
Height: Interior: 25.5 Inches (647.7mm)
Width: 21.5 Inches (546.1mm)
Depth: 21.0 Inches (533.4mm)
Shelf Capacity:
Size: 19.25 Inches (489mm) x 19.25 Inches (489mm)
Supplied: 4 Shelves
Max. Capacity: 17 Shelves
Max. Weight Capacity: 30 lbs (Do not try to remove shelf with more than 20 lbs on it.)
Water Pan:
Dimensions: Mean Length 16.25” (413mm)
Mean Width 11.04” (280mm)
Depth 2.375” (60mm)
Capacity: Maximum 6.75 Liters
Recommended Fill 5.5 Liters
3.2 Performance Parameters
Temperature Range: 18C to 55C (5C above ambient to 30C ambient max.)
Temperature Uniformity: +0.3°C @ 37°C.
Temperature Accuracy: +0.1°C.
Minimum Temperature for 145 DEG Cycle 135°C
Decontamination: 95 DEG Cycle 87°C
CO2 Range: 0.1 to 20%.
CO2 Accuracy: +0.1%
CO2 Recovery: Up to 5% +0.2% in 4 minutes.
Temperature Recovery: 0.3°C/min. (Average)
Temperature Display Resolution: 0.1°C
CO2 Display Resolution: 0.1%
Door and Perimeter Heater Control Logic: Proportional 0-100% (adjustable)
Door & Perimeter Heater Duty cycles are processor controlled.
Temperature Sensor Type: Precision Integrated Circuit
CO2 Control Logic: Fixed Algorithm/Manual Environmental Adaptable.
CO2 Sensor Type: Infrared
Connectors on rear panel are provided for chart recorder, remote alarm, and printer connection.
NOTE: RATINGS APPLY ONLY AT DEFAULT CONTROL AND OPTION SETTINGS
OM0135 7
3.3 Standard Items Packed With Unit
Four (4) stainless steel shelves
Eight (8) stainless steel shelf brackets
Two (2) sidewall plenums (installed)
One (1) blower plenum (installed)
One (1) water pan
Gas tube and filter
Air intake tube and filter
Access port plug with breather holes (installed)
Outer shell plastic hole ring (installed)
2 meters long electrical cord
Operation and Maintenance Manual
Operating Instructions
3.4 Accessories (Ordered Separately)
Model NU-1550 Automatic Tank Switch (External) (115 VAC)
Model NU-1550E Automatic Tank Switch (External) (230 VAC)
Model NU-1552 CO2 Tank Alarm (115 VAC)
Model NU-1551E CO2 Tank Alarm (230 VAC)
Model NU-1557 Additional Shelves
Model NU-1559 CO2 Analyzer Fyrite Kit (Dry) 0-20% (replacement fluid required)
Model NU-1561 Replacement Fluid for CO2 Analyzer (two bottles/carton)**
Model NU-2568 Surge Protector (115 VAC)
Model NU-1564 CO2 Regulator (2 Stage)
**Fyrite Replacement Fluid may only be ordered when shipment is possible by UPS Ground Service.
OM0135 8
OM0135 9
4.0 Test Performance & Procedures
All equipment is thoroughly inspected at the NuAire Factory at the time of shipment. Quality control is
maintained by constant surveillance over the product, beginning at the receipt of purchased material and
concluding with a final inspection, which certifies the Incubator performance to the specifications. In all
instances where product quality cannot be easily assessed on the end item, the product is inspected during
sub-assembly fabrication. The following test procedures are conducted on each cabinet and a copy of the
test report is included with each unit.
4.1 Visual Inspection
4.1.1 Each Incubator is visually inspected to insure that the interior is clean and free from scratches,
nicks, and burrs and that all welds, both interior and exterior are ground and polished smooth.
4.1.2 Painted surfaces are inspected to be free of scratches, nicks, insufficient covering, and runs.
4.1.3 The doors open and close freely without binding of the hinges. The gasket seals the inner glass
door tightly. The glass door is free of scratches.
4.2 Electrical Tests
4.2.1 Electrical Leakage Test
All Incubators are required to have primary-circuit filtering to meet EMC (electromagnetic
compatibility) regulations. Electrical leakage may exceed 0.5 milliampere provided the leakage
current does not exceed 1.0 milliamperes.
4.2.2 Dielectric Voltage - Withstand
1770 volts (VDC ONLY) is applied between dead metal parts and the hot/neutral power source
lead with no electrical breakdown using an Associated Research Model 4045AI or 3560D for 115
VAC units. 2100 volts (VDC ONLY) is applied for 230 VAC units.
4.2.3 Grounding Continuity
The resistance between the green bonding conductor of the supply cord and any dead metal part of
the cabinet shall not exceed 0.10 ohms.
4.3 Functional Tests
These functional tests are performed on every unit after a 48-hour burn-in period.
4.3.1 Control Systems
All diagnostic functions are exercised to insure proper operation of control systems, components
and alarms.
4.3.2 CO2 Control
Each unit is calibrated to function at a 5% CO2 level. The concentration is checked with an
independent instrument. Each unit is monitored during the 48-hour burn-in period and only
accepted with zero failures.
4.3.3 CO2 Recovery
Each unit is exercised for CO2 recovery time at the end of the 48-hour burn in period. The door is
opened for 1 minute to deplete the CO2. After the door is closed, the unit shall recover to 5%
+0.2% within a 4-minute period.
4.3.4 Temperature
Each unit is monitored for stable temperature control over the 48-hour burn-in period (37°C).
4.3.5 Humidity
Each unit is subjected to the highest achievable RH during the 48-hour burn-in period and any
condensation problems corrected. Door and perimeter duty cycles are set at 18% and 32%
respectively for this test. These are the default settings from the factory.
OM0135 10
5.0 Warranty
NuAire, Inc. warrants that it will repair F.O.B. its factory or furnish without charge F.O.B. its factory a
similar part to replace any material in its equipment within 24 months after the date of sale if proven to the
satisfaction of the company to have been defective at the time it was sold provided that all parts claimed
defective shall be returned, properly identified to the company at its factory, charges prepaid. Factory
installed equipment or accessories are warranted only to the extent guaranteed by the original manufacturer
and this warranty shall not apply to any portion of the equipment modified by the user. Claims under this
warranty should be directed to NuAire, Inc. setting forth in detail the nature of the defect, the date of the
initial installation and the serial and model number of the equipment.
This warranty shall not apply to any NuAire product or part thereof which has been subject to misuse,
abuse, accident, shipping damage, improper installation or service, or damage by fire, flood or acts of God.
If the serial number of this product is altered, removed, or defaced as to be illegible, the Warranty shall be
null and void in its entirety.
The warranty is for the sole benefit of the original purchaser and is not assignable or transferable. Prior to
returning any item, for any reason, contact NuAire for a Return Authorization Number. This number must
accompany all returns. Any product shipped to NuAire without this number will be returned, refused
shipment or collect freight.
6.0 Shipments
NuAire, Inc. takes every reasonable precaution to assure that your Incubator arrives without damage.
Motor carriers are carefully selected and shipping cartons have been specifically designed to insure your
purchase. However, damage can occur in any shipment and the following outlines are steps you should
take on receipt of a NuAire Incubator to be sure that if damage has occurred, the proper claims and actions
are taken immediately.
6.1 Damaged Shipments
6.1.1 Terms are F.O.B. factory, unless stated otherwise. Therefore, it is important to check each
shipment before acceptance.
6.1.2 If there is visible damage, the material can be accepted after the driver makes a notation on the
consignee's copy of the freight bill. Then an inspection must be made to verify the claim against
the carrier. This inspection is the basis of your filing the claim against the carrier.
6.1.3 If concealed damage is found, it is absolutely necessary to NOTIFY THE FREIGHT AGENT AT
ONCE, and request an inspection. Without this inspection, the transportation company may not
accept a claim for loss or damage. If the carrier will not perform the inspection, an affidavit must
be prepared stating that he was contacted on a certain date and that he failed to comply with the
request. This, along with other papers in the customer's possession will support the claim.
OM0135 11
7.0 Installation
The Incubator is fastened to the base skid and it is usually the best procedure to leave the skid attached
until the Incubator is located in its approximate position to facilitate ease in handling. The base skid can
then be removed by removing the four bolts holding the cabinet to the skid. Examine the Incubator
carefully. INSPECT both the exterior and the interior for any transit damage before discarding the
shipping crate.
7.1 Location
In locating the Incubator, consider all possible conditions that might affect its performance as well as
laboratory procedures for its intended purpose. Do not locate near heating or cooling ducts, or next to
equipment that generates heat (steam radiators, stoves, ovens, autoclaves, etc). Avoid direct sunrays
and rapidly moving air currents. These conditions adversely affect the even heat dissipation required
from the exterior surfaces of the Incubator and will cause the temperature variation in the chamber to
exceed specifications as stated in section 3.2. Since the Incubator needs even heat dissipation on all
surfaces in order to maintain an internal temperature variation of less than 0.2 degrees C, a minimum of 2
inches (50mm) must be allowed between the rear and sides of the Incubator and any walls, partitions or
obstructions to facilitate adequate convection of air around the unit. Confirm clearance with a tape
measure if needed. Adjust the Incubator location accordingly. In addition, since the Incubator is cord
connected, the cord must be readily accessible for disconnection if necessary. For maintenance/service
purposes, the control center electronics should remain accessible.
Place Incubators 10” apart when placed side-by-side to enable running the heated Decon Cycle while
surrounding Incubators are in normal mode.
7.2 Leveling
Prior to use the incubator should be leveled using a bubble level on a middle shelf in the chamber. The
Incubator should have all 4 feet firmly on the bench or floor. Level from side to side and front to back.
Leveling feet are provided for this purpose, factory installed into the base of the Incubator. By turning the
adjustable leveling feet counter-clockwise, raises the Incubator. The leveling feet height should be a
minimum of 1/4 inch (6mm) below the base. IT IS IMPORTANT THAT THE OUTER CABINET BE
OFF THE FLOOR BY AT LEAST 1/4-INCH (6MM) TO ASSURE CABINET AIR FLOW
RELIEF. Confirm with a tape measure if needed.
OM0135 12
OM0135 13
7.3 Shelf Plenums & Water Pan Installation
Before installation of the shelves, and water pan, NuAire recommends to decontaminate all surfaces within
the interior chamber, glass door, and outer door with gasket. They can be wiped down with a disinfectant
of 70 percent alcohol or similar non-corrosive antimicrobial agent. Use only disinfectants that are
compatible with the vinyl gasket and the powder coat paint on the outer door. If the user desires, the top
and side plenums can be removed to disinfect the chamber surfaces that they cover. Absolutely no
chlorinated or halogen materials are to be used in the chamber.
Provided with each Incubator, are four shelves. The shelves are easily installed by attaching the shelf
guide supports to the sidewall plenums. If the shelf does not bind or disengage from the horizontal bracket
it is installed correctly. Additional shelves and shelf guides are available. The water pan shall be placed
directly on the bottom of the chamber.
7.4 Electrical
The electrical supply circuit to the Incubator must conform to all national and local electrical codes.
Consult the serial-data plate, located at the front of the right side of the Incubator, for voltage, cycle, phase,
and ampere requirements before making connection. Plug the power cord securely into a grounded power
source. VOLTAGE SHOULD NOT VARY MORE THAN 5% FROM SERIAL PLATE RATINGS.
Have a qualified technician check with the power source with a properly rated volt meter if needed. A
separate branch circuit is recommended to prevent possible loss of product due to overloading or failure of
other equipment on the same circuit. A SURGE PROTECTOR IS STRONGLY RECOMMENDED to
avoid power-related faults.
OM0135 14
7.5 Air Inlet Connection
An air inlet tubing kit consists of one-foot clear vinyl tube and (1) 50mm polypropylene .3-micron HEPA
filter. Locate the air inlet port on the back panel of the unit. Remove the cover cap; connect one end of
tubing to the air inlet port, and the other end of tubing to the air filter.
CAUTION: This is a free air supply. DO NOT connect to pressurized source.
7.6 CO2 Connection
High concentrations of CO2 gas can cause asphyxiation! Install Incubator in well ventilated area.
This Incubator is designed to be operated with CO2 gas only. Connecting a flammable or toxic gas can
result in a hazardous condition. Gases other than CO2 should not be connected to this equipment. CO2 gas
cylinders have a UN1013 label on the cylinder and are equipped with a CGA 320 outlet valve. Check the
gas cylinder for the proper identification labels.
Do not use CO2 gas cylinders equipped with siphon tubes. A siphon tube is used to extract liquid CO2 from
the cylinder which can damage the pressure regulator. Consult with your gas supplier to ensure that the
CO2 cylinder does no contain a siphon tube.
7.6.1 CO2 Tube Connection
Included with the Incubator are a tubing kit consisting of (1) six foot (2m) vinyl tube and
(1) 50mm polypropylene 0.3 micron HEPA filter.
7.6.2 CO2 Supply
1. CO2 of medical grade is recommended.
2. A two-stage pressure regulator, Linde# 19590, or equal, is recommended.
3. DO NOT USE a single stage regulator. It will not give a stable output at 20 psi and
exposes the Incubator to the gas cylinder pressure.
OM0135 15
7.6.3 CO2 Regulators
The regulator’s high-pressure stage direct from the supply cylinder must have a range of from 0 to
2000 PSI or 0 to 140 BAR. This gauge indicates actual tank pressure. The low-pressure stage
should have a range of 0 to 30 PSI or 0 to 2 BAR (100 PSI or 6 BARS maximum). This gauge
will indicate the actual CO2 pressure to the Incubator. Some single stage CO2 pressure regulators
have two gauges. USE A TWO-STAGE REGULATOR. All NuAire Incubators use CO2 in such
small quantities that precise metering of CO2 input pressure is important for maximum
performance.
To connect the regulator: First open the CO2 cylinder slightly, for an instant (this is termed
“cracking the valve.”) This will blow out dust or dirt that may have collected in the valve outlet.
BE SURE to keep your face away from the valve outlet to protect your eyes from dust or dirt.
Second, MAKE SURE the regulator pressure-adjusting screw is released by turning it
counterclockwise until it turns freely. Third, attach the regulator to the cylinder valve and tighten
the connection nut with a wrench. BE SURE DISC SEAL IS IN PLACE BEFORE MAKING
CONNECTION.
7.6.4 CO2 Connection
Connect the CO2 supply from the low-stage of the two-stage regulator, to the inlet fitting located
on the Incubator back panel labeled “CO2 tank 1”. The filter should be inserted downstream of the
low-stage regulator before the inlet fitting on the Incubator as shown in the figure. Secure each
connection with the clamps supplied. Observe proper flow orientation of the filter (look for “in” or
a green dot on the filter). The tubing is easily cut with a sharp knife.
7.6.5 CO2 Supply Adjustment
With the regulator OFF (i.e. fully counterclockwise), open the cylinder valve slowly usually 1 to 2
turns is sufficient.
NEVER STAND IN FRONT OR BEHIND THE REGULATOR WHEN OPENING THE
VALVE. ALWAYS STAND TO ONE SIDE.
The cylinder tank pressure should read 700 to 800 PSI or 48 to 55 BAR, more or less, depending
on the temperature of the cylinder. Next, turn the regulator’s pressure adjusting screw clockwise,
until the low-pressure gauge reads 20 PSI or 1.4 BAR. The CO2 connection is now complete.
NOTE: OSHA requires the CO2 tanks to be physically restrained (i.e. via chained to wall) to
prevent accidental damage to cylinder.
If optional feature Model Number I01, CO2 Automatic Tank Switch (Internal) is purchased,
separate installation instructions are provided.
7.6.6 Checking the Connections
The connections can be checked for leakage by brushing a small amount of soapy water on each
one. Observe to see if any bubbles are blown. If not the connection is secured properly. If it is,
the cause of the leaking needs to be determined, (i.e. loose clamp or damaged hose) and corrected.
OM0135 16
7.7 Air Pump Assembly Shipping Foam: One or more blocks of foam is used to secure the air pump
during shipping.
IMPORTANT: FOAM BLOCK MUST BE REMOVED BEFORE OPERATION
To remove foam block:
1. Remove control center cover by unscrewing the (4) screws holding it to the top of the
Incubator. Lift the cover straight up and set it aside.
2. Remove the foam block from under the air pump allowing the pump to hang freely in the
mounting bracket. It is located to the right of the sensor shroud on the right side of the
Incubator.
3. Reinstall the control center cover on top of the Incubator and secure it with the screws
removed in Step 1.
NOTE: Foam block can be removed through the front control panel. Remove the 2 locking screws
(one each) either side of the control center lid. Open the outer door and the control panel will swing
down and open.
7.8 Water Pan
Place water pan in the center on the bottom of the chamber and fill with Single distilled water no purer
than 1 Mega Ohm. It is recommended to fill the pan to about 1/2 inch below the top rim. See Sections 3.1
and 8.2 for more specifications and operational details.
7.9 Stacking Units: Refer to BCD-09812 in the ASSEMBLY Section
Align the feet of the top Incubator with the indentations in the lid of the bottom Incubator. Ensure that
there is a minimum of 1/4 inch air gap between the Incubators as explained in section 7.2 to avoid
obstructing the cabinet air flow. Secure the stacking straps that are attached to the bottom of the top
Incubator to the lid of the bottom Incubator using the 4 screws that secure the lid. Only stack units 2 high.
7.10 Correct Installation
When the Incubator is installed correctly connected to the power source it is rated for gas connections
made properly, the water pan filled, the shelves in place, and the unit is leveled. Read section 8 and follow
all instructions for setting set points at the desired value. Then calibrate the Incubator control systems as
explained in section 9. If the Incubator is installed and calibrated correctly it will meet the performance
specs listed in section 3.2.
OM0135 17
8.0 DHD Autoflow Operation
CAUTION: A qualified technician who is familiar with the proper maintenance procedures required
for this equipment, as well as repair must perform all maintenance actions on this equipment.
The Incubator is designed to provide a sterile, constant temperature, constant CO2 level and naturally
humidified atmosphere for optimum growth of tissue cell cultures and other organisms requiring this
precise environment. To operate the Incubator properly, the following parameters must be reviewed,
carefully set, and/or prepared.
8.1 Sterility
The environment provided by this Incubator is not selective. As a result, any contamination within the
chamber is subjected to the same environment as the specimens. Therefore, before placing any cultures in
the Incubator, the shelves and sidewall top plenums should be sterilized. The interior sidewalls, top,
bottom, door, as well as the gasket should be wiped clean with a 70% solution of isopropyl alcohol or other
disinfectant compatible with the chamber construction, to remove any contamination. Use mild detergent
to clean the exterior of the Incubator. This Incubator gives the lab professional a choice of 2 heated
decontamination cycles. (See section 10.1 for further information.)
8.2 Humidity
Humidification of the Incubator chamber is achieved through the process of water evaporation (vapor
water pressure) from a stainless steel water pan placed on the bottom of the chamber. Materials of
different thermal resistance (i.e. glass, plastic) do not offer sufficient thermal recovery and are not
recommended for use. Although some metals offer better thermal coefficients than stainless steel,
dissimilar metals cause electrolysis in the acid atmosphere (carbonic acid) and should never be used, or
placed within the chamber.
Use only distilled or mineral-free water, no purer than 1 mega ohm, in the stainless steel pan. The water
should be changed at least once a week, preferably more often. FLOODING THE BOTTOM OF THE
INCUBATOR IS NOT RECOMMENDED since it is difficult to change the water weekly and almost
always necessitates the use of chemicals, which are not recommended and may damage the stainless steel.
Also, it promotes condensation on the inner walls because it steals the natural convection, heat flow
through the inner chamber and condensation points occur. ABSOLUTELY NO CHLORINATED OR
HALOGEN MATERIALS ARE TO BE USED IN THE CHAMBER.
Humidity recovery to 90% of original level within 20-40 minutes after a 15-second door opening with a
water reservoir area of 210 square inches. Contamination in the water pan may be avoided by adding a
small amount of copper sulfate to the water pan after each decontamination of the chamber.
564
191
ATTENTION ACCOMPANY'S
INFORMATION OR IMPORTANT
SYMBOL
POTENTIAL ELECTRICAL
HAZARD ONLY QUALIFIED
PERSON TO ACCESS
188
HOT SURFACE
BURN POTENTIAL
!
OM0135 18
Condensation on the glass door, walls, top, or bottom of the chamber indicates an incorrect balance of door
and front perimeter heat. Both the door and front perimeter heaters operate on a duty cycle. A good
starting point for these duty cycles is the default setting for the door & perimeter heaters in a room ambient
temp of 22ºC (72ºF) at a temperature set point of 37.0ºC. The fresh airflow that the air pump delivers to
the chamber has been preset at the factory. If condensation starts forming on the sides or backwall in the
chamber, the number and length of air injections into the chamber can be increased. (See section 8.8.1.2
for instructions.) To alleviate the condensation, increase the airflow. To increase RH in the chamber
reduce the airflow. If airflow is decreased, some condensation may be unavoidable. The port plug with
breather hole must be used on the back port. DO NOT SEAL THE OPENING IN THE PORT PLUG
EXCEPT WHERE INSTRUCTED TO DO SO. THIS IS A PRESSURE RELIEF FOR THE
DECON CYCLES.
8.3 Control System
The NuAire Incubator Control Electronics system is designed to serve the control requirements of the
Incubator chamber. Temperature and CO2 levels are controlled by preset values to provide the optimum
conditions for culture growth within a chamber. Operator input is coordinated through the control panel
keypad and status displays. Figure 1 shows the various inputs and outputs of the system.
Figure 1: Direct Heat Control Inputs & Outputs
OM0135 19
The NuAire Incubator Control Electronics is a state-of-the-art microcomputer based system that provides:
8.3.1 Chamber control in a single electronic package.
8.3.2 Enhanced information presentation
A chamber temperature (set point and actual)
A CO2 level (set point and actual)
An alarm LED
A mode switch status LED
A temperature status digital LED
A CO2 status digital LED
A heat, CO2 inject, door ajar, air class, and two CO2 tank status LEDs.
A heated decon. Cycle LED
8.3.3 Simplified operator controls
The control panel is operated using 6 keys. The keypad switches and their functions:
Run/Setup: Switches between “Run” mode and the “Setup” mode when pressed for 3 seconds.
Run mode: The green LED above the switch pad is lit continuously and Incubator is fully
functional. System calibrations are performed in this mode.
Setup mode: The green LED is blinking with the Temperature & CO2 displays flashing
alternating between the set point values and “SEt” “UP”. Changes in the set point
values are performed in this mode by pressing the “SEL” key to choose the
system indicated by lighting the green LED next to the display and using the
up/down arrows to change the value in the display. The Incubator is in a
“standby” condition and heat or gas functions are not active.
NUAIRE Logo: When pressed for 3 seconds it gives access to the “tSt”, “Opt” &
master reset menus.
Up and Down Arrows: Increase or decrease the value in the display during a calibration in run mode
changing the system set points in setup mode.
They are also used for advancing to the “tSt”, “Opt” and master reset menus after
the logo key is pressed. These keys are also used to change
state of systems in the “tSt” menus and adjust values of the systems in
the “Opt” menu.
SELect key: Advances between and selects the active system displays for calibration functions
in run mode and changing set points in setup mode. The lit green LED indicates
selection. This key also advances between systems in the “tSt” & “Opt” menus.
95/145 cycle: Starts the heated Decon Cycle when pressed for 3 seconds and it also
advances through the phases of the Decon Cycle.
OM0135 20
8.3.4 Automatic notification of abnormal situations
The red alarm LED on the control panel indicators will light to indicate a fault within the system.
An audible alarm is also heard.
System intermittent
Temperature Control Fault
Temperature is over set point by more than 1.0C.
Temperature does not reach set point within 4 hours within 0.2º.
Sensors disagree by 4º
CO2 Control Fault
CO2 percent is over set point by more than 1%.
CO2 percent does not reach set point within 30 minutes within 0.2%.
Tank Switch
NOTE: See Section 11 for all alarm conditions.
8.3.5 Password Protection
If desired, the NuAire Incubator Control Electronics may be configured to prevent users who are
unauthorized to change the set points (CO2%, Temp), running conditions of the Incubator or
initiate a Decon Cycle. The password protection may be initiated by accessing the option
configuration parameters in diagnostic mode (see Section 8.8). Once the password protection is
initiated, the user must use the correct sequence of the UP, DOWN, and SELECT keys to access
the setup mode. To operate, first press the SETUP key. The displays will then indicate "Pas rEq",
or “password required”, then enter the correct sequence of the UP, DOWN, and SELECT keys
followed by pressing and holding the SETUP key for three seconds to access the setup mode.
8.3.6 Provision for add-on expansion capability
Chart recorder output (option)
Automatic tank switch (option)
RS-232 output
8.3.7 Diagnostic and calibration assists.
Individual analog inputs may be displayed to assist calibration.
Individual outputs may be forced to an on or off condition.
Individual digital inputs may be displayed.
Front panel lamps may be tested.
Memory and internal processor diagnostics may be selected.
All options may be individually tested.
Packaging is designed such that all accessories are easily removed and replaced
8.4 Front Control Panel Descriptions
The system front control panel contains the following functions (BCD-09709).
8.4.1 Heat Jacket Status LED
The green LED indicates when the chamber heater is turned on. A blinking LED indicates the
chamber heater is being cycled to maintain chamber set point temperature.
8.4.2 Inject CO2 Status LED
The green LED indicates when the CO2 control valve is open and CO2 is flowing into the chamber.
When it blinks on/off briefly, the system is injecting CO2 to maintain set point.
8.4.3 Door Ajar Status LED
The door ajar yellow LED indicates when the inner glass door is not closed and latched. The LED
acts upon a magnetic switch located along the lower right corner of the inner glass door. The LED
blinks after door is closed, indicating door delay time (40 second default)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64