ES-FLOWâ„¢ 1xxI
Ultrasonic Liquid Flow Meter/Controller
Doc. no.: 9.17.145 rev. I Date: 27-07-2022
Instruction Manual
ATTENTION
Please read this document carefully before installing and operating the product.
Not following the guidelines could result in personal injury and/or damage to the equipment.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I2
Copyright
© 2022 Bronkhorst High-Tech B.V.
All rights reserved. No part of this publication may be reproduced, distributed, or transmitted in any form or by any means,
without the prior written permission of the publisher.
Disclaimer
This document has been reviewed and is believed to be accurate. Bronkhorst High-Tech B.V. does not assume liability for
errors, inaccuracies or absence of information. The material in this document merely serves information and illustration
purposes; no rights can be derived from its contents.
Bronkhorst High-Tech B.V. reserves the right to modify or improve its products and documentation without prior notice.
Device specifications and the contents of the package may deviate from what is stated in this document.
Symbols in this document
Important information. Disregarding this information could increase the risk of damage to the equipment, or the risk of
personal injuries.
Tips, useful information, attention points. This will facilitate the use of the instrument and/or contribute to its optimal
performance.
Additional information available in the referenced documentation, on the indicated website(s) or from your Bronkhorst
representative.
Receipt of equipment
Check the outside packaging box for damage incurred during shipment. If the box is damaged, the local carrier must be
notified at once regarding his liability. At the same time a report should be submitted to your Bronkhorst representative.
Carefully remove the equipment from the box. Verify that the contents of the package was not damaged during shipment.
Should the equipment be damaged, the local carrier must be notified at once regarding his liability. At the same time a
report should be submitted to your Bronkhorst representative.
Check the packing list to ensure that you received all items included in the scope of delivery
Do not discard spare or replacement parts with the packaging material
Refer to Removal and return instructions about return shipment procedures.
Equipment storage
The equipment should be stored in its original package in a climate controlled storage location.
Care should be taken not to subject the equipment to excessive temperatures or humidity.
See technical specifications (data sheet) for information about required storage conditions.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 3
Warranty
Bronkhorst® products are warranted against defects in material and workmanship for a period of three years from the date
of shipment, provided they are used in accordance with the ordering specifications and not subject to abuse or physical
damage. Products that do not operate properly during this period may be repaired or replaced at no charge. Repairs are
normally warranted for one year or the balance of the original warranty, whichever is the longer.
See also section 9 (Guarantee) of the Conditions of sales:
www.bronkhorst.com/int/about/conditions-of-sales/
The warranty includes all initial and latent defects, random failures, and indeterminable internal causes. It excludes failures
and damage caused by the customer, such as contamination, improper electrical hook-up, physical shock etc.
Re-conditioning of products primarily returned for warranty service that is partly or wholly judged non-warranty may be
charged for.
Bronkhorst High-Tech B.V. or affiliated company prepays outgoing freight charges when any part of the service is
performed under warranty, unless otherwise agreed upon beforehand. The costs of unstamped returns are added to the
repair invoice. Import and/or export charges as well as costs of foreign shipping methods and/or carriers are paid by the
customer.
General safety precautions
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety
precautions required to prevent possible injury. Read the operating information carefully before using the product.
Before operating, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting
cables for cracks or breaks before each use.
The equipment and accessories must be used in accordance with their specifications and operating instructions, otherwise
the safety of the equipment may be impaired.
Opening the equipment is not allowed. There are no user serviceable parts inside. In case of a defect please return the
equipment to Bronkhorst High-Tech B.V.
One or more warning signs may be attached to the product. These signs have the following meaning:
General warning; consult the instruction manual for handling instructions
Surface may get hot during operation
Shock hazard; electrical parts inside
To maintain protection from electric shock and fire, replacement components must be obtained from Bronkhorst. Standard
fuses, with applicable national safety approvals, may be used if the rating and type are the same. Non-safety related
components may be obtained from other suppliers, as long as they are equivalent to the original component. Selected parts
should be obtained only through Bronkhorst, to maintain accuracy and functionality of the product. If you are unsure
about the suitability of a replacement component, contact your Bronkhorst representative for information.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I4
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 5
Table of contents
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introduction 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.1 Scope of this manual
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.2 Intended use
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.3 Product description
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81.4 Documentation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101.5 Model key
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Customized I/O options 1.5.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installation 2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132.1 Functional properties
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132.2 Mounting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Location in fluid system 2.2.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Orientation 2.2.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Hygienic applications 2.2.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142.3 Fluid connections
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152.4 Electrical connections
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152.5 Communication interface
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operation 3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.1 Powering up and powering down
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.2 First use
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.3 User interface
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Main screen functions 3.3.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Unlocking buttons 3.3.1.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Selecting display information 3.3.1.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Selecting screen area 3.3.1.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Editing setpoint 3.3.1.4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Resetting counter 3.3.1.5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Resetting alarm 3.3.1.6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Menu navigation 3.3.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Password protection 3.3.2.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Editing string 3.3.2.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Editing list selection 3.3.2.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Settings menu 3.3.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Readout 3.3.3.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Controller 3.3.3.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Counter/totalizer 3.3.3.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Alarm 3.3.3.4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Setup 3.3.3.5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Advanced 3.3.3.6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Security menu 3.3.4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Changing password 3.3.4.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Resetting password 3.3.4.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30LED indications 3.3.5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Interface status 3.3.5.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31DeviceNetâ„¢ 3.3.5.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Multifunctional switch 3.3.6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Normal operating functions 3.3.6.1
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Control mode - readout/change 3.3.6.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Network settings - readout/change 3.3.6.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Disabling multifunctional switch 3.3.6.4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353.4 Digital parameters
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36General 3.4.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Measurement and control 3.4.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Secondary outputs 3.4.2.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Advanced measurement and control 3.4.2.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Alarms 3.4.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Counter and totalizer 3.4.4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Network configuration 3.4.5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Fluid 3.4.6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Controller 3.4.7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Master/slave configuration (FLOW-BUS) 3.4.8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Device identification 3.4.9
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Special parameters 3.4.10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Default control mode 3.4.10.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 493.5 Adjusting zero point
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Autozero function 3.5.1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Manual procedure 3.5.2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Digital procedure 3.5.3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Maintenance 4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 514.1 Cleaning
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 514.2 Calibration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Troubleshooting and service 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 525.1 Common issues
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 545.2 Service
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Returns 6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 556.1 Removal and return instructions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 556.2 Disposal (end of lifetime)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Parameter index
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 7
1Introduction
1.1 Scope of this manual
This manual covers general product information, installation and operating instructions and troubleshooting tips for the
ES-FLOWâ„¢ 1xxI digital volume flow meter/controller for liquids.
1.2 Intended use
The ES-FLOWâ„¢ 1xxI has been designed to accurately measure and control low volume flows with high precision and a
limited pressure drop. A wide range of liquids can be measured independent of fluid density, temperature and viscosity.
When ordered as a 3-A compliant instrument, the ES-FLOWâ„¢ 1xxI is suited for hygienic (food processing) applications and
Clean-in-place (CIP). The end user is responsible for the selection and application of suitable (FDA-compliant) sealing
materials.
The wetted materials incorporated in the ES-FLOWâ„¢ 1xxI are compatible with media and conditions (e.g. pressure,
temperature) as specified at ordering time. If you are planning to use the product (including any third party components
supplied by Bronkhorst, such as pumps or valves) with other media and/or other conditions, always check the wetted
materials (including seals) for compatibility. See the technical specifications of the product and consult third party
documentation (if applicable) to check the incorporated materials.
Responsibility for the use of the equipment with regard to its intended use, suitability for the intended application, cleaning
and compatibility of process media with the applied materials lies solely with the user.
The user is responsible for taking the necessary safety measures to prevent damage and/or injury while working with the
equipment and process media (as described in the associated Material Safety Data Sheets).
Where appropriate, this document recommends or prescribes safety measures to be taken with respect to media usage or
working with the described equipment under the specified conditions. However, this does not relieve the user of
aforementioned responsibility, not even if such is not explicitly recommended or prescribed in this document.
Bronkhorst High-Tech B.V. cannot be held liable for any damage and/or injury resulting from unintended, improper or
unsafe use, or use with other media and/or under other process conditions than specified at ordering time.
1.3 Product description
The ES-FLOWâ„¢ 1xxI is a precise and compact volume flow meter with control function for liquids, based on a novel
ultrasonic technology.
Measuring principle
Measuring is done in a straight tube, without obstructions or dead spaces. Multiple transducers measure both the surface
acoustic wave and the transit time through the media. All up- and down-stream combinations are recorded and processed
in nanoseconds. The sound wave velocity and the surface area are recalculated to the volume flow value. This ultrasonic
measuring method is fast, accurate and inherently bi-directional.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I8
Application
The combination of a straight sensor tube with zero dead volume, self-drainability and hygienic connections, makes the ES-
FLOWâ„¢ especially suitable for hygienic applications. Examples of typical use are the measuring of food ingredients like
flavourings or syrups, or measuring concentrates in CIP units. For non-hygienic applications, the instrument can be
equipped with compression type fittings. Wetted parts are made of stainless steel, the housing has a high ingress
protection rating.
Operation
The ES-FLOWâ„¢ 1xxI can be operated with the integrated readout and control unit, as well as digitally, via RS-232 or fieldbus
(Modbus, FLOW-BUS, PROFIBUS DP, PROFINET, DeviceNetâ„¢ or CANopen), or in analog mode. The readout and control unit
has a capacitive touchscreen with a TFT display. The fluid temperature can be read out as a secondary output. An on-board
PID controller can be used to drive a control valve or pump, establishing a complete, compact control loop.
By default, the instrument is supplied with the readout and control unit above the measuring tube (assuming the flow
direction is to the right). If so required, the flow direction and display orientation can be customized to support a
convenient reading angle. The illustrations in this manual are based on the default orientation.
Multi-range
The ES-FLOWâ„¢ 1xxI offers multi-range functionality: factory calibrated ranges can be re-ranged to a different full scale
range. Because of the extremely high linearity of the sensor, this does not affect the original accuracy specifications.
The instrument comes with a calibration certificate for all supported full scale ranges. The actual full scale of the instrument
is set to a value as ordered.
1.4 Documentation
The documentation listed in the following table is available on the ES-FLOWâ„¢ product pages under
www.bronkhorst.com/products
Type
Document name
Document no.
Manuals
Instruction Manual ES-FLOWâ„¢ 1xxI (this document)
9.17.145
Hook-up diagrams
Hook-up diagram Analog/RS-232
9.16.252
Hook-up diagram CANopen
9.16.250
Hook-up diagram DeviceNetâ„¢
9.16.245
Hook-up diagram FLOW-BUS
9.16.247
Hook-up diagram Modbus
9.16.246
Hook-up diagram PROFIBUS DP
9.16.248
Hook-up diagram
EtherCAT® / EtherNet/IP / Modbus-TCP / POWERLINK / PROFINET
9.16.251
Hook-up diagram optional bus and I/O configurations
9.16.249
Dimensional drawings
Dimensional drawing ES-103I / ES-113I
Mk I: 7.15.194
Mk II: 7.15.217
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 9
The documentation listed in the following table can be downloaded from www.bronkhorst.com/downloads
Type
Document
Document no.
Manuals
Manual CANopen interface
9.17.131
Manual DeviceNetâ„¢ interface
9.17.026
Manual EtherCAT interface
9.17.063
Manual EtherNet/IP interface
9.17.132
Manual FLOW-BUS interface
9.17.024
Manual Modbus ASCII/RTU/TCP interface
9.17.035
Manual PROFIBUS DP interface
9.17.025
Manual POWERLINK interface
9.17.142
Manual PROFINET interface
9.17.095
Manual RS-232 interface
9.17.027
Certificates
EU Declaration of Conformity
9.06.021
Calibration certificates are part of the scope of delivery unless otherwise agreed.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I10
1.5 Model key
The model key on the serial number label contains information about the technical properties of the product as ordered.
The actual properties of your instrument can be retrieved from the diagram below.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 11
1.5.1 Customized I/O options
ES-FLOWâ„¢ 1xxI instruments offer various customized input/output functions through pin 5 of the power connector and
through pin 2 of the actuator output connector as an option. I/O options are factory installed as specified at ordering time,
and cannot be changed manually.
The last 3 groups of the model key on the serial number label indicate the installed I/O configuration. The possible
configurations are described in the table below. See the hook-up diagram for custom bus and I/O configurations for an
explanation of the codes.
Code
Description
000
Disabled, pin 5 is pulled down to 0 Vdc (default selection)
A1V*
0…10 Vdc sourcing output, controller
Analog signal for pump or external valve steering (control signal only)
When the controller output is used for pump or external valve steering (mass flow meters only), make sure
to set parameter Valve maximum to 0.3 [A]. For mass flow controllers, the controller output is limited to a
value below 10Vdc, due to the maximum valve current restriction.
B1V*
4…20 mA sourcing output, controller
Analog signal for pump or external valve steering (control signal only).
When the controller output is used for pump or external valve steering (mass flow meters only), make sure
to set parameter Valve maximum to 0.3 [A]. For mass flow controllers, the controller output is limited to a
value below 20 mA, due to the maximum valve current restriction.
B2V*
3.8…20.8 mA sourcing output, controller
Analog signal for Badger Meter valve with TEIP11 signal converter (control signal only)
C3A
Digital output, min/max alarm
During a min/max alarm, pin 5 is pulled down to 0 Vdc.
C4A
Digital output, counter alarm
During a counter alarm, pin 5 is pulled down to 0 Vdc.
C5S
Digital output, enabled by setpoint (for shut-off control)
Pin 5 is pulled down to 0 Vdc at a controller setpoint, e.g. for shut-off valve activation.
For factory selected analog control (…-A#-C5S):
If parameter Control mode is set for analog control by factory, the minimum setpoint at which the device
(shut-off valve) connected to pin 5 is activated is 1.9%. This prevents possible noise on the analog input
activating the device accidentally.
For factory selected digital control (…-D#-C5S):
If parameter Control mode is set for digital control by factory, the setpoint threshold for activating the
device connected to pin 5 is any value > 0.
Note: If the instrument is forced into Valve Safe State, the digital output is not affected, so a (n.c.) shut-off
valve connected to pin 5 will not close when the (n.c.) controller is in Valve Safe State’
Make sure to use 24 Vdc power supply corresponding to the shut-off valve specifications.
C0I
Digital output, high/low switch via remote parameter (e.g. for shut-off valve control)
Pin 5 is pulled down to 0 Vdc when writing value 1 to parameter IO switch status, this is undone by writing
value 0.
A device connected to pin5 (e.g. a shut-off valve) can be activated/deactivated by writing parameter IO
switch status.
Note: If the instrument is forced into Valve Safe State, the digital output is also affected, so a (n.c.) shut-off
valve connected to pin 5 will be closed when the (n.c.) controller is in ‘Valve Safe State’.
Make sure to use 24 Vdc power supply corresponding to the shut-off valve specifications.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I12
Code
Description
D9E
Digital frequency output, measure
Measurement value is translated to a frequency within given frequency range.
The default frequency range to represent 0…100% flow is 0…10000 Hz. Any other frequency range must be
specified on order.
F9B
Digital pulse output, batch counter
Pin 5 is pulled down to 0Vdc when a given batch size is reached (during a given pulse length).
By default, a pulse is given at each 1x the Counter unit batch value, with a pulse length of 1 second. For
instance, when Counter unit is set to 'ln', a pulse is given each time 1 ln has passed through the instrument.
An alternative pulse length must be specified on order.
Provide a pull-up resistor of 5…10 kOhm to create 15…24 Vdc at pin 5 (according to the applicable hook-up
diagram).
I3C
Digital input, controller mode valve close
Valve closes when pin 5 is connected to 0 Vdc.
This option switches between the default Control mode and mode ‘Valve Close’ (value 3). When the default
Control mode is digital, the default value is 0 (bus/RS-232), when the default Control mode is analog, the
default value is 1 (Analog input).
I8C
Digital input, controller mode valve purge
Valve is fully opened when pin 5 is connected to 0 Vdc.
This option switches between the default Control mode and mode ‘Valve Fully Open’ (value 8). When the
default Control mode is digital, the default value is 0 (bus/RS-232), when the default Control mode is analog,
the default value is 1 (Analog input).
I1R
Digital input, reset counter
The counter resets when pin 5 is connected to 0 Vdc.
I2R
Digital input, reset alarm
The alarm resets when pin 5 is connected to 0 Vdc.
*) Notes regarding controller options:
A controller output option (A1V, B1V or B2V) can only be installed in combination with controller mode A or D
(controller enabled, see model key).
If the controller is enabled, the M8 connector can always be used to control a Bronkhorst® valve (through pin 3).
Although it is theoretically possible to have 3 controller options installed, no more than 1 signal should be used at any
time.
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 13
2Installation
This chapter describes the steps to take in order to prepare the ES-FLOWâ„¢ 1xxI for first time use.
2.1 Functional properties
Before installing the ES-FLOWâ„¢ 1xxI, check if the functional properties match your requirements.
The ES-FLOWâ„¢ Brochure provides general technical specifications, specific technical data can be
found on the serial number label and in the purchase oder.
Flow/pressure rate(s)
Media to be used in the instrument
Upstream and downstream pressure(s)
Ambient temperature
Input and output signal
Bronkhorst® instruments are pressure tested to 1.5 times the requested pressure rating and outboard leak tested to at least
2 * 10-9 mbar l/s Helium. The tested pressure is specified on the serial number label. If the serial number label is missing or if
the specified pressure is insufficient, the instrument must not be used and should be returned to the factory.
Before installation, make sure that the tested pressure is in accordance with the safety factor of your application. The tested
pressure must always be higher than the maximum operating pressure.
2.2 Mounting
2.2.1 Location in fluid system
In general, the best location is a horizontal pipe segment or a segment where the fluid direction is upward.
Gas might accumulate in the horizontal segment if it is followed by a downward segment. Do not mount
the instrument in a location like this.
Mounting in a downward pipe segment with an open end is strongly dissuaded, especially if the pipe
diameter is 1/2" or more. Gravity might let the segment run empty; depending on the specific system
dimensions and the viscosity of the metered fluid, this effect might be stronger or weaker.
If the instrument is part of a closed fluid system, mounting the instrument in a downward pipe segment is
not preferable, but may be considered if other mounting locations are problematic.
To minimize the risk of gas inclusion by cavitation, the preferred location to install a control valve is downstream from the
instrument, the preferred location for a pump is upstream.
To prevent damage to the internal electronics, make sure the temperature inside the instrument housing does not get
above 60 °C. If necessary, take appropriate heat dissipation measures, especially if the instrument is operated inside an
enclosure (e.g. a control cabinet).
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I14
2.2.2 Orientation
In hygienic applications, the ES-FLOWâ„¢ 1xxI should be mounted in such a way that it allows self-draining (see further).
To prevent liquid from pooling on the display surface, the instrument should not be mounted with the display surface
facing up.
Other than that, the instrument has no preferred mounting orientation.
Note that the connection between the measuring tube and the electronics housing is fixed; the housing
cannot be rotated to get a better view of the display. Its orientation can only be altered by
repositioning the entire instrument.
On models starting from MkII, the flow direction and display orientation can be adjusted at the
factory, for which the instrument has to be returned. Contact your Bronkhorst representative for more
information and arrangements.
2.2.3 Hygienic applications
In order to comply with the 3-A Sanitary Standards, the ES-FLOWâ„¢ 1xxI should be mounted in a location where its
exterior cannot come into direct contact with food substances and/or the product contact surfaces of their packaging
material.
To allow self-draining, mount the ES-FLOW™ 1xxI vertically or with a minimum inclination of 10°, with the normal flow
going upward (note the FLOW arrow on the instrument).
2.3 Fluid connections
The FLOW arrow on the measuring tube indicates the normal flow direction. For normal use, install the ES-FLOWâ„¢ 1xxI in
the process line, in accordance with the direction of the FLOW arrow. For bi-directional measuring, install the instrument in
the direction in which the highest flow will be measured (if applicable). When deciding which direction to install the
instrument in, take into account that the measuring range in the reverse direction is approximately 73% of the full scale
range (whereas the instrument can measure 131% FS in the normal direction).
Tighten fittings according to the instructions of their manufacturer.
Check the fluid system for leak tightness after any modification and before applying full operating pressure, especially when
using hazardous media (e.g. toxic or flammable).
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 15
2.4 Electrical connections
The ES-FLOWâ„¢ 1xxI is equipped with one or more electrical connection ports. The image to
the right shows the locations of the different ports (the actual presence and appearance of
ports might be different, depending on the ordered fieldbus interface and instrument type):
1. Standard connection port, 8-pin M12 male
2. Connection port for actuator output, 4-pin M8 female
3. Fieldbus connection port, 5-pin M12 male/female (optional, fieldbus dependent)
4. Fieldbus connection port, 5-pin M12 male/female (optional, fieldbus dependent)
See sections Model key and Functional properties for configurable and installed options
respectively.
Upon delivery, all connection ports are capped. To maintain the original ingress protection rating, do not remove the caps
of unused connectors.
Electrical connections must be made with standard cables or according to the applicable hook-up diagram.
When using self-assembled cables, follow the guidelines provided by the connectors' manufacturer.
For use in a fieldbus system, follow the instructions of the cable supplier for the according fieldbus system.
Make sure that the power supply is suitable for the power ratings as indicated on the serial number label (see model key)
or in the technical specifications, and that double or reinforced insulation is used for the power supply.
Before powering up, make sure all required cabling is properly connected.
Before each use, inspect cabling and connectors for damage.
Never power the instrument simultaneously from two different power sources (e.g. fieldbus and Plug-in Power Supply).
Doing so will irreversibly damage the printed circuit board and the instrument will have to be repaired before it can be
used.
In order to be able to comply with all applicable guidelines and regulations, it is essential that electrical connections be
made by or under supervision of a qualified electrician.
The equipment described in this document contains electronic components that are susceptible to electrostatic
discharge.
When working on the electrical installation, take appropriate measures to prevent damage as a result of electrostatic
discharge.
The CE mark on the equipment indicates that it complies with requirements imposed by the European Union, including
electromagnetic compatibility (EMC).
EMC can only be guaranteed by applying appropriate cables and connectors or gland assemblies:
Cable wire diameters must be sufficient to carry the supply current and minimize voltage loss.
When connecting the product to other devices, ensure that the integrity of the shielding remains uncompromised; use
shielded cables and connectors where possible and/or required.
Preferably use the supplied cables (if applicable) to make electrical (signal) connections to and between the supplied
components. These cables are shielded, have the required wire diameter, and loose ends (if applicable) are marked to
facilitate correct connection.
If not all requirements for proper shielding can be met (for example, because a component is not equipped with shielded
connectors), take the following measures to ensure the best possible shielding:
Keep cable lengths at a minimum.
Route cables as closely as possible alongside metal structures or components.
Ensure all electrical components are grounded to earth.
When in doubt about the shielding of your cabling and/or electrical connections, contact your Bronkhorst representative.
2.5 Communication interface
Check if the default network settings match the configuration of the fieldbus system (if applicable). If necessary, the default
settings can be overruled by changing the appropriate parameters (see Network configuration).
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I16
3Operation
3.1 Powering up and powering down
To maintain control of the fluid system and ensure a safe situation, it is recommended to turn on power before applying
fluid pressure and to switch off power only after the fluid system is depressurized.
When pressurizing, prevent pressure shocks by gradually bringing the fluid system to the required operating pressure.
3.2 First use
Bronkhorst does the utmost to ensure that you receive a clean product.This does not, however, relieve the end user of the
responsibility to ensure that the instrument and the system in which it is incorporated meet (hygiene) requirements that
apply to the intended use of the instrument. The responsibility for cleaning the equipment to meet such requirements lies
exclusively with the end user.
Before starting measurement and control, make sure to remove gas from the fluid system by flushing all fluid lines with the
process fluid at a high flow rate.
For instruments with a control function, the highest possible flow rate is ensured by giving setpoint = 100% or using special
Control mode = 8 (valve fully open). Control mode 8 bypasses the PID controller, which is particularly useful when having
the ES-FLOWâ„¢ 1xxI set to a low capacity flow range.
3.3 User interface
The screen representations in this section are for illustration purposes only. They do not necessarily reflect the exact lay-out
and/or information displayed on the screen. Some representations might only be applicable to certain instrument types or
factory configurations.
The image to the right shows the screen display, immediately after powering up. The following screen areas can be
distinguished:
1. Top line
2. Measure readout
3. Custom readout 1 (only for instruments with control
function)
4. Custom readout 2
The navigational buttons have the following functionality:
Enter selected menu
Enter edit mode
Confirm selection/changes
Navigate up in menu
Change character or list item
Navigate down in menu
Change character or list item
Return to previous menu
Leave edit mode without making changes
Select display info
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 17
The communication status of the instrument can be monitored
with the three LEDs to the left of the screen. See LED indications
for a description of the possible indications.
A multifunctional switch ( ) to the right of the screen can be
used to start several functions without having to enter the
menu (see Multifunctional switch).
3.3.1 Main screen functions
3.3.1.1 Unlocking buttons
The switch lock function prevents accidental activation of the user interface (.e.g. by cleaning the instrument and/or its
environment with a hose or high-pressure cleaner). The lock is activated after a period of inactivity (lock time; 60 seconds by
default), after which the individual tip buttons are disabled. The buttons can be unlocked by pressing and holding and
for a couple of seconds (unlock time; default: 4 seconds).
The switch lock properties can be edited in the Settings menu.
3.3.1.2 Selecting display information
The Custom readout 2 area can show different parameters.
Depending on the instrument type, the following parameters
may be available (editable parameters are marked with an *):
setpoint*
percentage reading
actuator/valve percentage
counter*
alarm*
capacity
temperature
density
instrument*
To change the displayed parameter, press from the main
screen. Cycle through the available parameters by pressing
the button repeatedly.
3.3.1.3 Selecting screen area
Depending on the instrument type and the settings of the
readout and control unit, none, one or both of the custom
readout areas can display an editable parameter. If an area
contains an editable parameter, it can be selected and edited
in place. Selection is made visible by a red line above the
concerning area.
To cycle through the editable screen areas, press or
repeatedly
To enter edit mode for the selected area, press
When edit mode is active, press at any time to leave edit
mode without making changes
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I18
3.3.1.4 Editing setpoint
The setpoint of a controlling instrument can be changed if the parameter is displayed in the custom readout area.
Instruments without a control function have no editable setpoint and can only be monitored (Custom readout 1 not
available).
If the setpoint is configured to be entered as a character string, follow these steps to change its value:
1. Select a readout area
that displays the
setpoint:
2. Press to enter edit
mode (the first
character position is
highlighted):
3. Press or to select
the required character:
4. Press to advance to
the next character
position:
On confirmation of the last digit, the entered value is stored and edit mode is left (whereupon the character highlight is
removed).
If the setpoint is configured to be entered stepwise, follow these steps to change its value:
1. Select a custom
readout area that
displays a setpoint:
2. Press to enter edit
mode (the current
value is highlighted):
3. Press or to
change the parameter
value (hold the button
to scroll fast):
4. Press to store the
current value and leave
edit mode:
The setpoint input method (string or step-wise) can be set in the Settings menu (Settings > Setup > Customize > Setpoint)
3.3.1.5 Resetting counter
1. In the main screen,
select the Custom
readout 2 area:
2. Press to enter edit
mode:
3. Press or to
change the value to
'yes':
4. Press to confirm the
selected option and
reset the counter:
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI9.17.145I 19
3.3.1.6 Resetting alarm
Before resetting the alarm, be sure to eliminate the cause. Resetting the alarm without changing the conditions that
caused it will re-activate the alarm immediately.
When an alarm occurs, a message blinks in the top line of the main screen. If the alarm is
configured to be reset automatically, the blinking stops as soon as the alarm conditions no
longer apply. If the alarm is configured to be reset manually, follow these steps to reset it:
1. Custom readout 2 area is
selected automatically:
2. Press to enter edit
mode:
3. Press or to
change the value to
'yes':
4. Press to confirm the
selected option and
reset the alarm:
3.3.2 Menu navigation
The configurable parameters of the instrument and settings of the readout and control unit are organized in a menu
structure.
Items
Menus can contain items of 3 different types:
Sub menu
An arrow pointing to the right indicates a sub menu
Parameter with list selection
An arrow pointing down indicates a parameter that can be
changed by selecting a value from a list
Character string parameter
First line: unit (if applicable)
Second line: current parameter value
Not all parameters can be edited; some parameter values are protected, or display a value that is linked to (the value of)
another parameter:
Normal display
Parameter can be edited
Dimmed
Parameter is read-only
Bronkhorst®
Instruction Manual ES-FLOWâ„¢ 1xxI 9.17.145I20
Navigation
Inside menus and sub menus, the selected item is highlighted in red
Press or to navigate to the required menu item
Arrows pointing up and/or down in the top line indicate the menu contains more items than
can be displayed.
Press to enter the selected sub menu or to enter edit mode
Press to return to the previous screen or menu or to leave edit mode without making
changes
3.3.2.1 Password protection
1. By default, some items
are accessible only
after entering a
password:
2. Enter the password (if
the password contains
less than 8 characters,
fill the remaining
positions with spaces):
3. After entering 8
characters, the
password is validated.
If the password is
correct, the selected
function is displayed:
4. If the password is
incorrect, access is
denied:
The default password is 'abc' (without quotes) and is case sensitive.
To avoid unauthorized access, change the default password immediately after installation. See Changing password for
instructions.
3.3.2.2 Editing string
1. Select the parameter to
be edited:
2. Press to enter edit
mode (the first
character position is
highlighted):
3. Press or to select
the required
character/digit:
4. Press to advance to
the next character
position:
On confirmation of the last character/digit, the entered value is stored and edit mode is left (wherupon the character
highlight is removed).
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50