Munters Farm Silo Load Cell-10 Owner's manual

Type
Owner's manual
1
Manual for use and maintenance
Farm Silo Load Cell
Silo Weighing Scale
Ag/MIS/UmGB-2200-06/14 rev 1.3
P/N: 110306
Farm Silo
Load Cell
© Munters AB, 2014 2
Farm Silo Load Cell
Manual for use and maintenance
Revision
: N.1.3 of 06.2014
Product Software
: N/A
This manual for use and maintenance is an integral part of the apparatus together with the attached
technical documentation.
This document is destined for the user of the apparatus: it may not be reproduced in whole or in
part, committed to computer memory as a file or delivered to third parties without the prior
authorization of the assembler of the system.
Munters reserves the right to effect modifications to the apparatus in accordance with technical and
legal developments.
© Munters AB, 2014 3
Index
Chapter page
1
INTRODUCTION
----------------------------------------------------------------------------------------
4
1.1 Disclaimer ----------------------------------------------------------------------------------------------------------- 4
1.2 Introduction --------------------------------------------------------------------------------------------------------- 4
1.3 Notes ---------------------------------------------------------------------------------------------------------------- 4
2
INSTALLATION
------------------------------------------------------------------------------------------
5
2.1 Installing the Load Cell ------------------------------------------------------------------------------------------ 5
2.2 Wiring the Load Cells ------------------------------------------------------------------------------------------- 7
2.3 Load Cell Test Procedure ------------------------------------------------------------------------------------- 12
3
TROUBLESHOOTING
--------------------------------------------------------------------------------
13
4
WARRANTY
--------------------------------------------------------------------------------------------
18
© Munters AB, 2014 4
1 Introduction
1.1 Disclaimer
Munters reserves the right to make alterations to specifications, quantities, dimensions etc. for
production or other reasons, subsequent to publication. The information contained herein has been
prepared by qualified experts within Munters. While we believe the information is accurate and
complete, we make no warranty or representation for any particular purposes. The information is
offered in good faith and with the understanding that any use of the units or accessories in breach of
the directions and warnings in this document is at the sole discretion and risk of the user.
1.2 Introduction
Congratulations on your excellent choice of purchasing an Farm Silo!
In order to realize the full benefit from this product it is important that it is installed, commissioned and
operated correctly. Before installation or using the fan, this manual should be studied carefully. It is
also recommended that it is kept safely for future reference. The manual is intended as a reference
for installation, commissioning and day-to-day operation of the Munters Controllers.
1.3 Notes
Date of release: July 2010
Munters cannot guarantee to inform users about the changes or to distribute new manuals to them.
NOTE All rights reserved. No part of this manual may be reproduced in any manner whatsoever
without the expressed written permission of Munters. The contents of this manual are subject
to change without notice.
© Munters AB, 2014 5
2 Installation
The following sections detail how to install the:
Farm Silo -4: Four tons
Farm Silo -10: 10 tons
2.1 Installing the Load Cell
Figure 1 shows the unassembled silo components.
Figure 1: Unassembled Silo Parts
1. Take apart any assembled parts and spread them in front of you as shown in the picture.
The only missing component is the silo’s leg, which is shown in Figure 6.
2. Place the half ball on the load cell with the flat side facing down.
It is very important that the flat side faces down or the load cell will not function properly.
© Munters AB, 2014 6
Figure 2: Ball Placement on Load Cell
3. Place the base plate on the half ball.
4. Verify that the ground wires are connected shown in Figure 2.
5. Place the silo leg on the center of the base plate. A wooden board illustrates the silo leg.
Figure 3: Wooden Board
6. After the silo’s leg is on, place the metal ring first then place the spring above the ring, on
the leg.
7. The spring supplied measures 11 mm/0.43 inches. The bolt must be screwed to the point in
which the spring measures 10 mm/0.39 inches.
Figure 4: Measuring the Spring
CAUTION Do not over tighten the bolt! Use a reliable measurement instrument and follow the
instructions in full.
© Munters AB, 2014 7
8. To secure the bolt, insert the safety pin and split its ends. The silo leg can be so thick that
the end of the bolt will not be visible. In this case a safety pin can not be used.
Figure 5: Placing the Safety Pin
The load cell is now in place and ready to use.
Figure 6: Connection to the Ground Rod
CAUTION Connect one of the legs to ground rod with heavy-duty cable (Figure 6).
2.2 Wiring the Load Cells
The following section details how to wire the Load Cell to the FJB-4 Junction Box and FSW-2 silo
weight controllers.
1. Each load cell has six-wire shielded cables. Connect each load cell to the FJB-4 double
connector according to the colors inscribed on the FJB-4 PCB board .
Cable: Board
o Brown Brown
o White White
o Red Red
o Blue - Blue
o Black Black
o Green Green
2. Connect the cable that going from the FSW-2 to the fJB-4, to the “Controller” terminal block
within the FJB-4 junction box. The connection colors of this cable are direct connection.
© Munters AB, 2014 8
Cable: Board
o Brown Brown
o White White
o Red Red
o Blue - Blue
o BlackBlack
o Green Green
3. Connect the other side of the cable that going from the FJB-4 to the FSW-2 controller, to the
“Silo 1” terminal block within the FSW-2. The colors connections of this cable:
Cable: Board
o Brown Brown
o White Red
o Red White
o Blue - Blue
o Black Black
o Green Green
NOTE White and Red are swapped.
© Munters AB, 2014 9
Figure 7: FSW 2 Wiring Diagram *(up to Eight Load Cells) with an External Power Source
Key:
1: Place two jumpers on EXT.
2: For more than
six load cells
, connect an external power supply and move the
jumper to the
EXT
position.
© Munters AB, 2014 10
Figure 8: FSW 2 Wiring Diagram *(up to Eight Load Cells) with an Internal Power Source
Key:
1: Place two jumpers on INT.
© Munters AB, 2014 11
Figure 9: FJB Connectors
Key:
1: If there are six load cells, add the two remaining cells along with load cells 1 and
2.
© Munters AB, 2014 12
2.3 Load Cell Test Procedure
To perform a load cell test procedure, disconnect the load cell from any load and test following
points:
Table 1: Points to be Tested When Performing Test Procedure
Test resistance between
Limits
Green - Black 285 Ohm
Green - Red 285 Ohm
White - Black 285 Ohm
White - Red 285 Ohm
The Farm Silo Weight-2 metallic shield connects on the FJB-4 side and four wires according to RPLP-1
Unprotected colors.
Figure 10: FSW-2 Silo Weight Offset Adjustment
RPLP
-1
TR2 for Silo 2 Adjustment
© Munters AB, 2014 13
3 Specifications
FSLC-04 Specifications
FSLC-10 Specifications
Figure 11 details the lower base plate dimensions.
Figure 11: Lower Base Plate Dimensions (millimeters)
© Munters AB, 2014 14
3.1 FSLC-04 Specifications
Table 2: FSLC-04 Specifications
Parameter
V
alue
Unit
Rated capacityR.C. (E
max
) 4000 / 1800
lbs.
/ Kg.
NTEP/OIML accuracy class
NTEP
Maximum no. of intervals (n)
3000 single, 5000
multiple
Y = E
max
/V
min
6666 Maximum available
Rated output-R.O. 3.0
mV/V
Rated output tolerance 0.1 ±% of rated
output
Zero balance 2 ±% of rated
output
Zero return, 30 min. 0.0250 ±% of applied load
Total error (per OIML R60) 0.0200 ±% of rated
output
Temperature effect on
zer
o 0.0023 ±% of rated
output/°C
Temperature effect on output
0.0010 ±% of applied
load/°C
Temperature range 10 to +40 °C
Temperature range, safe 20 to +70 °C
Maximum safe central
overload
150 % of R.C.
Ultimate central overload 300 % of R.C.
Excitation, recommended 10 VDC or VAC RMS
Excitation, maximum 15 VDC or VAC RMS
Input impedance 385±10
Ω
Output impedance 351±5
Ω
Insulation resistance >2000
Cable length 6.0
m
Cable type 6-wire, braided, polyurethane, floating screen Standar
d
Construction Nickel-plated alloy steel and stainless steel
Environmental protection IP67
Recommended torque 136 N*m
© Munters AB, 2014 15
Figure 12 and Table 3 detail the FSLC-04 Load Cell dimensions.
Figure 12: FSLC-04 Load Cell Dimensions
Table 3: FSLC-04 Load Cell Dimensions Details
A
B
C
D
E
ØF
ØG
H
I
X
Y
30.5
130
76.2 25.4
16
Ø13.5
Ø13.5
34.0 30.5
57
4
3.2 FSLC-10 Specifications
Table 4: FSLC-10 Specifications
Parameter
V
alue
Unit
Rated capacity-R.C. (E
max
) 10,000 /4535
lbs.
/ Kg.
Rated output-R.O. 3.0 ± 0.10%
mV/V
Zero balance 3.0 ±% of rated
output
Zero return, 30 min. 0.050 ±% of applied load
Total error (per OIML R60) 0.0500 ±% of rated
output
Temperature effect on
zer
o 0.0070 ±% of rated
output/°C
Temperature effect on output
0.0040 ±% of applied
load/°C
Temperature range 30 to +80 °C
Temperature range, safe 10 to +40 °C
Maximum safe central
overload
150 % of R.C.
Ultimate central overload 300 % of R.C.
Excitation, recommended 10 VDC or VAC RMS
Excitation, maximum 15 VDC or VAC RMS
Input impedance 380 ± 10
Ω
© Munters AB, 2014 16
Parameter
V
alue
Unit
Output impedance 350 ± 3
Ω
Deflection at rated capacity < 0.4
mm
Insulation resistance >2000
Cable length 6.0
m
Cable type 6-wire, braided, polyurethane, dual floating
screen
Standar
d
Construction Electrolysis, nickel-plated steel
Environmental protection IP67
Figure 13 and Table 5 detail the FSLC-10 Load Cell dimensions.
Figure 13: FSLC-10 Load Cell Dimensions
Table 5: FSLC-10 Load Cell Dimensions Details
Letter
A B C D E ØF ØG H X Y
Dimension (mm)
37 171.5 95.4 38.1 19 Ø19.8 Ø19.8 40.9 76 7.5
© Munters AB, 2014 17
4 Troubleshooting
#
Problem Description
Troubleshooting
1
Installation and filling feed
has been done but silo
stops weighing.
If the display of Silo A does not show A/D '65536' counts on 16
bits, lower the offset (see Figure 7) until this number changes.
After receiving numbers shown on the screen, reducing the
numbers should be done according to:
[65536 (number of Kg missing in the silo * Silo 1 Scale
Factor)]
After new installation is done and the Silo is empty, the A/D
reading should be around 1000 counts (near zero), use the
offset feature to get it.
If the display shows '0', increase it by using offset feature.
2
The controller display
shows “disconnected” (or
Error) and the test menu
shows 65,536 or 0 rather
than a voltage
measurement that is
required in order to
identify the error.
Check if FJB-4 received voltage from controller (green and black
wires from controller should be approximately 10 DCV). If not
received, check wire connections on the controller's side.
If 10 DCV is received, load cells must be checked (see Table 1
and Figure 9 for further explanation):
Take White and Green wires from each load cell and check their
voltage using a digital multi meter (the amount is not important
but it must be identical throughout load cells with a difference of
up to 2 mV). If difference is more than 2mV there is a problem
with one of the load cells. The range of voltage should be
between 0 to 20 mV depending on Silo load.
In some cases, when the prior test is not effective and voltage
between Black (common) wire and White, and between Black
and Green, wires must be checked (amount of voltage must be
between 4-4.5V and identical throughout load cells).
© Munters AB, 2014 18
5 Warranty
Warranty and technical assistance
Munters products are designed and built to provide reliable and satisfactory performance but cannot be
guaranteed free of faults; although they are reliable products they can develop unforeseeable defects
and the user must take this into account and arrange adequate emergency or alarm systems if failure to
operate could cause damage to the articles for which the Munters plant was required: if this is not done,
the user is fully responsible for the damage which they could suffer.
Munters extends this limited warranty to the first purchaser and guarantees its products to be free from
defects originating in manufacture or materials for one year from the date of delivery, provided that
suitable transport, storage, installation and maintenance terms are complied with. The warranty does not
apply if the products have been repaired without express authorisation from Munters, or repaired in such
a way that, in Munters’ judgement, their performance and reliability have been impaired, or incorrectly
installed, or subjected to improper use. The user accepts total responsibility for incorrect use of the
products.
The warranty on products from outside suppliers fitted to Farm Silo Weighing Controller, (for example
cables, scales, power supplies, etc.) is limited to the conditions stated by the supplier: all claims must be
made in writing within eight days of the discovery of the defect and within 12 months of the delivery of
the defective product. Munters has thirty days from the date of receipt in which to take action, and has the
right to examine the product at the customer’s premises or at its own plant (carriage cost to be borne by
the customer).
Munters at its sole discretion has the option of replacing or repairing, free of charge, products which it
considers defective, and will arrange for their despatch back to the customer carriage paid. In the case of
faulty parts of small commercial value which are widely available (such as bolts, etc.) for urgent
despatch, where the cost of carriage would exceed the value of the parts, Munters may authorise the
customer exclusively to purchase the replacement parts locally; Munters will reimburse the value of the
product at its cost price.
Munters will not be liable for costs incurred in demounting the defective part, or the time required to travel
to site and the associated travel costs. No agent, employee or dealer is authorised to give any further
guarantees or to accept any other liability on Munters’ behalf in connection with other Munters products,
except in writing with the signature of one of the Company’s Managers.
WARNING:
In the interests of improving the quality of its products and services, Munters reserves the
right at any time and without prior notice to alter the specifications in this manual.
The liability of the manufacturer Munters ceases in the event of:
dismantling the safety devices;
use of unauthorised materials;
© Munters AB, 2014 19
inadequate maintenance;
use of non-original spare parts and accessories.
Barring specific contractual terms, the following are directly at the user’s expense:
preparing installation sites;
providing an electricity supply (including the protective equipotential bonding (PE) conductor, in
accordance with CEI EN 60204-1, paragraph 8.2), for correctly connecting the equipment to the
mains electricity supply;
providing ancillary services appropriate to the requirements of the plant on the basis of the
information supplied with regard to installation;
tools and consumables required for fitting and installation;
lubricants necessary for commissioning and maintenance.
It is mandatory to purchase and use only original spare parts or those recommended by the
manufacturer.
Dismantling and assembly must be performed by qualified technicians and according to the
manufacturer’s instructions.
The use of non-original spare parts or incorrect assembly exonerates the manufacturer from all liability.
Requests for technical assistance and spare parts can be made directly to the nearest Munters office. A
full list of contact details can be found on the back page of this manual.
© Munters AB, 2012
www.munters.com
Australia Munters Pty Limited, Phone +61 2 6025 6422, Brazil Munters Brasil Industria e Comercio Ltda, Phone +55 41 3317 5050, Canada Munters Corporation
Mason, Phone +1 517 676 7070, China Munters Air Treatment Equipment (Beijing) Co. Ltd, Phone +86 10 80 481 121, Denmark Munters A/S, Phone
+45 9862 3311, India Munters India, Phone +91 20 3052 2520, Indonesia Munters, Phone +62 818 739 235, Italy Munters Italy S.p.A., Chiusavecchia, Phone
+39 0183 52 11, Japan Munters K.K., Phone +81 3 5970 0021, Korea Munters Korea Co. Ltd., Phone +82 2 761 8701, Mexico Munters Mexico, Phone
+52 818 262 54 00, Russia Munters AB, Phone +7 812 448 5740, Singapore Munters Pte Ltd., Phone +65 744 6828, South Africa and Sub-Sahara Countries
Munters (Pty) Ltd., Phone +27 11 997 2000, Spain Munters Spain S.A., Phone +34 91 640 09 02, Sweden Munters AB, Phone +46 8 626 63 00, Thailand Munters
Co. Ltd., Phone +66 2 642 2670, Turkey Munters Form Endüstri Sistemleri A.Ş, Phone +90 322 231 1338, USA Munters Corporation Mason, Phone
+1 517 676 7070, Vietnam Munters Vietnam, Phone +84 8 3825 6838, Export & Other countries Munters Italy
S.p.A., Chiusavecchia Phone +39 0183 52 11
Ag/MIS/UmGB-2200-06/14 rev 1.3
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20

Munters Farm Silo Load Cell-10 Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI