Optos 200tx User manual

Type
User manual
Copyright 2007-2010, Optos plc. All rights reserved.
The information contained within this document is subject to change without notice. The latest version of
this information can be found at www.optos.com.
This document should be used in conjunction with the help files supplied in each application and the
Technical Data Specifications supplied with the device. Instructions and service information can be
obtained by contacting the Optos Customer Service Department through the contact details given in this
handbook.
The device is a prescription device.
Caution
Federal law restricts this device to sale by or on the order of a physician or
practitioner. See CFR 801.109(b)(1) for more information.
Optos does not provide, and the optomap® plus guidelines do not constitute advice on making
reimbursement claims. Diagnostic tests should be ordered by the treating physician and this physician is
responsible for appropriate usage, adequate documentation and proper coding. It is the responsibility of
the physician to comply with Medicare regulations, and check with the local insurance carrier for
reimbursement information and instructions. Optos does not accept any liability for reimbursement claims
made while using optomap® plus.
Copyright
This document may only be used for your personal, non-commercial use. This document and all its
contents (including without limitation, all text, formats, graphics and logos) are the property of Optos plc
and are protected from unauthorized copying and dissemination by the Copyright, Designs and Patents
Act 1988 (as amended), by various intellectual property laws and by international conventions.
Trademarks
Optos, optos, optomap, V2, ResMax, 3D Wrap and P200 are ® Marks. All rights reserved.
WINDOWS is a Trademark of Microsoft Corporation.
ADOBE and ACROBATREADER are Trademarks of Adobe Systems Incorporated.
All other trademarks are the property of their respective owners.
Part Number: G-00085/1
Print Date: February 2011
Table of Contents
1 What's New 1
2 Introduction 3
About the device
3
Indications for use
3
About images
3
Image types
5
3 Getting to know the device 7
Understanding the equipment
7
About the head rest
8
Attaching and detaching the Patient Arm Support
9
Attaching and detaching the Head Rest
9
About the detector settings
9
How to use the hand control
11
4 About the software applications 13
Getting Help
13
Admin application features
13
Capture application features
14
Review application features
14
Storage application features
16
Scheduler application features
16
About software updates
17
5 About capturing an image 19
Working with patient records
19
Cleaning before each patient
20
Patient instructions
21
Capturing images
21
Capture Single Eye Image
23
Single Image Preliminary checks
24
Capturing Stereo Images
24
Multiple Image Preliminary Checks
26
How to capture Autofluorescence images
27
Multi-Frame Settings
27
Capturing Multiple Images (Angiography)
28
How to do eye steering
29
View captured image
30
Checking the quality of captured images
30
Patient positioning
30
6 About the Patient History view 35
Using the Patient History window
36
Opening images
36
Selecting images from the Image Series view
38
Opening images using alternative methods
39
Discarding unwanted images
40
7 Using the viewing options 43
Color Channels
43
Review Tools
43
Recommended reading protocol
44
Using the Automated view
45
Using the Magnifier
45
Using the 3D Wrap view
46
Using the Comparison Overlay
47
Simulating a Binocular Indirect Ophthalmoscope (BIO) view
47
Using False Color views
48
Viewing stereo images
48
Comparing with sample disease images
49
Using the Adjustment and Information panels
49
8 Documenting your review 51
About annotations, retinal drawings and measurements
52
Using annotations
52
Using retinal drawings
53
Using measurements
54
About review notes, diagnostic (pathology) codes and procedure codes
55
Using procedure codes
56
Using review notes
57
Using diagnostic codes
58
9 Distributing images 61
Emailing images
61
Exporting images
62
Importing an image to a patient’s record
63
Printing images
64
Changing eye images from Right to Left (and Left to Right)
66
Assigning an image to another patient’s record
66
10 Using the Image Library 67
11 About archiving 69
When to Archive
69
Changing the Archive Task Settings
69
Archiving to disk
71
Archiving to a network
74
Scheduling storage tasks
74
12 Getting help 75
Accessing help from the device
75
Accessing help from the applications
75
Optos on the Web
76
13 Where to find features 77
Capture application user reference
77
Review application user reference
81
Storage application user reference
91
Scheduler application user reference
92
14 Contact us 93
Global Headquarters
93
Regional Offices
93
15 Training Checklist 95
Chapter 1 - What's New
1 What's New
This release supports the display of optomap® af images.
Note:
optomap® af images are captured using a green laser light for illumination. This contrasts with the
traditional use of blue illumination or white light for autofluorescence imaging. Green light
autofluorescence differs, and allows a different visualization of macular pathologies since macular
pigment does not interfere. Also the Optic Nerve Head is clearly visible when green laser light is used
for autofluorescence imaging.
Reference Materials
About the device 3
Checking the quality of captured images 30
How to capture Autofluorescence images 27
Selecting images from the Image Series view 38
Using the Automated view 45
Using the Magnifier 45
Using the 3D Wrap view 46
Using the Comparison Overlay 47
Using False Color views 48
Comparing with sample disease images 49
Using measurements 54
Using the Image Library 67
Scheduling storage tasks 74
Capture application user reference 77
Review application user reference 81
Scheduler application user reference 92
Storage application user reference 91
1 of 100
2 of 100
Chapter 1
Chapter 2 - Introduction
2 Introduction
Your device will be installed by Optos trained personnel. Do not operate the device until they have
completed the installation and training.
Warning
Please read the Safety Guidelines in the Introductory Handbook before using the
device.
About the device
This scanning laser ophthalmoscope is a widefield digital imaging device capable of capturing from the far
periphery of the retina. The retinal images are captured in an automated, patient friendly way without
scleral depression or contact with the cornea.
The 200Tx uses red, green and blue lasers to produce a digital, high-resolution image, which is displayed
on a PC monitor screen. This device can be used to capture angiography images. Red and green lasers
are used for digital color imaging. These laser wavelengths penetrate the retinal structures to different
depths, each wavelength providing information for interpretation and diagnosis.In autofluorescence mode,
the device uses the green laser to illuminate the eye. This allows an image of the natural fluorescence of
the eye to be captured. optomap® af images are captured using a green laser light for illumination. This
contrasts with the traditional use of blue illumination or white light for autofluorescence imaging. Green
light autofluorescence differs, and allows a different visualization of macular pathologies since macular
pigment does not interfere. Also the Optic Nerve Head is clearly visible when green laser light is used for
autofluorescence imaging. No fluorescent dye has to be introduced into the patient.The blue laser is used
when capturing angiography images. A series of images is captured as the fluorescein flows through the
retinal vessels.
Optos’ technology is designed to operate through a minimum pupil diameter of 2mm. Although pupil
dilation is not required, the decision to dilate is a medical decision to be made by the eyecare professional.
Images are captured on the scan head and then viewed, magnified, annotated and separated into their
color components in the Review application.
Indications for use
The 200Tx scanning laser ophthalmoscope is intended to be used as a wide field and retinal fluorescence
imaging ophthalmoscope to aid in the diagnosis and monitoring of diseases or disorders that manifest in
the retina. Depending on the enabled options, some devices are capable of autofluorescence imaging.
Note:
Depending on the enabled options, the device may have a subset of the above indications.
About images
The device can be used in many clinical areas and can assist in disease detection, monitoring and
treatment.
The images can be separated to present distinct retinal sub-structures and can be used to assist in the
assessment and recognition of different retinal pathologies.
Traditional ophthalmoscopes produce a reddish picture derived from a white light source. The Optos
devices use red and green lasers to compose the retinal images.
3 of 100
Chapter 2 - Introduction
Sample image from 200Tx device.
The composite image supports analysis of the retinal image by allowing individual review of the green and
red channel information.
l
The Green Channel image contains information from the sensory retina through the pigment
epithelium layers of the retina.
l
The Red Channel image contains information from the deeper structures of the retina, from the
pigment epithelium through the choroid.
Compare the green and red channels derived from the composite image shown above.
Sample image from 200Tx device.
Green Channel view Red Channel view
The diagram below summarizes the retinal layers and structures reached by the lasers. These laser
wavelengths penetrate the retinal structures to different depths, each wavelength providing information for
interpretation and diagnosis.
4 of 100
Chapter 2 - Introduction
Image types
There are different types of image for each type of procedure:
l
optomap® – captures a retinal image. The standard optomap® procedure is a wellness exam. This
procedure captures a standard optomap® retinal image.
l
optomap® plus – captures a medical retinal image. The optomap® plus procedure is a medical
retinal exam. Following this procedure will allow the use of the enhanced features in the Review
application.
l
optomap® fa – some systems are capable of capturing fluorescein images. The optomap® fa
medical procedure allows the capture of fluorescein optomap® fa images. Following this
procedure will allow the use of the enhanced features in the Review application. These images are
captured at a higher resolution than standard optomap® images. The image capture rate for the
early-phase images can be set in the Capture application settings. You must ensure that the
necessary resources are available to administer the fluorescein.
Note
l
Some systems are capable of capturing autofluorescence images, How to capture
Autofluorescence images on page 27.optomap® af images are captured using a green laser
light for illumination. This contrasts with the traditional use of blue illumination or white light for
autofluorescence imaging. Green light autofluorescence differs, and allows a different
visualization of macular pathologies since macular pigment does not interfere. Also the Optic
Nerve Head is clearly visible when green laser light is used for autofluorescence imaging.
l
Some image types offer additional options, for example, eye steering, ResMaxâ„¢, images
optimized for the periphery or central pole. These options vary depending on the device being
used and the image type selected. All available options are displayed in the Procedure Selection
dialog box.
5 of 100
6 of 100
Chapter 2
Chapter 3 - Getting to know the device
3 Getting to know the device
The device comprises the scan head module, Image Server PC, and Viewing PC. For more information,
please refer to the Technical Data Specification supplied with the device.
Images are captured on the scan head. The scan head runs the Capture application. The Image Server PC
runs the Storage application. Admin application, and Scheduler application. The Review application is run on the
Viewing PCs. In a typical installation, the images and database will be stored on the Image Server PC. You
can review captured images at any Viewing PC that is connected to the Image Server PC across the
network.
Understanding the equipment
The device is part of a system of networked PCs.
Scan Head Module
The Scan Head module runs the Capture application. This application lets you select and perform the
required patient imaging procedure.
Caution
The scan head monitor should not be used to review images for diagnostic
purposes.
l
Scan Head – Comprises the lasers and electronics used to capture patient images.
l
Scan Head table – Supports the adjustment of the scan head height. The table can be raised or
lowered using the hand control.
l
Face Pad – The face pad supports the patient’s face when the patient is being imaged. The face
pad can be removed for cleaning.
l
Scan Head monitor arm – Links the monitor to the scan head.
l
Scan Head monitor – Displays the Capture application. The touch screen monitor is used to interact
with the Capture application. It displays alignment feedback and captured images.
l
Hand Control – Comprises buttons used to raise and lower the scan head table, align the patient,
and capture images.
l
Head Rest - Comprises the head support and adjustable chin cup. The head rest can be removed
for cleaning.
7 of 100
Chapter 3 - Getting to know the device
l
Patient arm support and table wings (if supplied) – Devices used to capture optomap® fa images
may be fitted with a patient arm support and table wings. They are used to support the patient when
the injection is administered. The patient arm support can be removed for cleaning.
Viewing PCs
Viewing PCs run the Review application. The Review application lets you analyze patient images. You can
review images in a variety of ways. You can add annotations to highlight areas of interest, add diagnostic
codes, add notes, and email images to third parties.
Image Server
The Image Server PC runs the management applications:
l
The Admin application lets you configure the network and data management environment. You can
define security levels, create and modify users, and set a variety of system configuration options.
l
The Storage application lets you archive and protect database and patient image files. It is essential
to archive regularly. For further information refer to the Storage application help file.
Depending on your network configuration you may also run the Review application and Scheduler application
from the Image Server PC. When the Review application has been installed on an Image Server PC, the
Image Server PC may be used as an additional Viewing PC.
The Scheduler application lets you schedule patient appointments and manage patient details. The Scheduler
application may be installed on any PC on the same network as the device.
About the head rest
The head rest supports the patient's head while images are being captured. It is made up of a forehead
strap and a chin cup.
The head rest can be removed for cleaning and should be cleaned between patients.The full instructions
can be found in the Introductory Handbook.
Moving the chin cup
l
The chin cup height can be altered by turning the knobs on the side of the head rest.
l
The chin cup should be slid into position for the eye being imaged. When imaging the left eye slide
the chin cup to the first position on the right. When imaging the right eye slide the chin cup to the
first position on the left. There is an additional position on each side if the patient's features require
the chin cup to move further.
l
Release the chin cup by pulling the lever and slide the chin cup into position.
l
The chin cup should be turned to ±15°.
8 of 100
Chapter 3 - Getting to know the device
Attaching and detaching the Patient Arm Support
The patient arm support is fitted with a clamp and can be attached and removed from the table wings.
Warning
Some devices are supplied with a patient arm support. The patient arm support
supplied with the device is designed to support the weight of a resting arm. Do not
exceed the weight indicated on the patient arm support. When in use, ensure the
patient arm support is securely clamped to the table wing, and that it is covered
with a piece of absorbent paper.
Attaching and detaching the Head Rest
To remove the head rest for cleaning, pull the head rest bolt down and turn one-quarter turn to lock it open.
You can then remove the head rest by sliding it toward you.
When attaching the head rest, push the head rest back in place and then turn the head rest bolt a quarter
turn to release the lock.
About the detector settings
There are a several factors to be considered when selecting the appropriate detector settings:
l
The patient
l
The system response
l
The operator
The patient
When meeting a patient at the start of the imaging session, it is important to assess the amount of
pigmentation in the patient's iris.
Less laser light is reflected by the retina in patients who have a greater amount of pigmentation. This can
result in dark images being captured because less light has been returned from the eye. You should select
the appropriate iris category setting to suit the patient's pigmentation levels.
Small pupil size can also restrict the amount of laser light returned from the retina. Since less laser light is
returned through the smaller pupil, the captured image may seem dark. To minimize this, you should
capture images in a dimly lit room.
Working in Low Light Conditions
9 of 100
Chapter 3 - Getting to know the device
Warning
The device requires low levels of room light to operate efficiently. Take care to
avoid accidents when working under low light conditions.
Media opacities also absorb or scatter laser light and can block retinal structures.
System response
The device has a laser power of 4mW at the patient's eye and includes a laser safety system that would
prevent the system being used if the lasers were operating at unsafe levels. If the power drops, perhaps
due to a build up of dust, the amount of light returned from the eye will be reduced and will impair image
quality. The accumulation of dust and debris in the return path will also impair image quality.
Detector gains
As the laser power at the eye is a fixed value, the returned retinal information will vary depending on the
amount of light reflected by the pigmentation in the patient's retina. To ensure the greatest amount of
information is detected and that the image quality is optimum, the red and green detectors are set to a
specific gain level. This gain level is initially set when the patient's iris category is entered into the patient
details.
The iris category sets the detectors to expect a certain amount of light back from the patient's retina. You
should assess the patient's hair color, eye color and skin tone to determine the most appropriate iris
category. You can create your own iris categories if you need to.
l
In patients who have high pigmentation levels the "dark" iris category is selected. This sets the
detectors to expect less light back from the patient's retina.
l
In patients who have medium pigmentation levels the "medium" iris category is selected.
l
In patients who have low pigmentation levels the "light" iris category is selected. This sets the
detectors to expect a greater amount of light back from the patient's retina.
The detector gains can be adjusted and used when the next image is captured.
Note
l
Clicking the Optimize button will optimize the current image and sets the detector gains to give a
balanced image of the central pole and periphery in the next image.
l
Review application optimization - maximizes the range of pixel levels displayed on the screen
and adjusts the gamma setting, so that the average intensity of the display matches an ideal
value.
10 of 100
Chapter 3 - Getting to know the device
How to use the hand control
200Tx Hand Control Hand Control table height buttons
Use these buttons to adjust the table height.
Hand Control eye position buttons
Use these buttons to move the pupil circle to
the center of the patient's eye.
The patient's eye should be displayed inside
the target shown on the External Eye
Camera view. Use the up, down, left and right
buttons on the hand control to align the pupil
circle within the patient's pupil. When the eye
is close to the correct position the system will
indicate which fine alignment button needs to
be pressed to set the distance.
Hint: You can select the eye to be captured
using the Hand Control. Press the left button
to select Right Eye (OD) and the right button
to select Left Eye (OS). Once the eye has
been selected this button reverts back to the
alignment functionality.
Hand Control in and out fine alignment
buttons
Fine adjustment feedback (either - or +) will
be shown on the External Eye Camera view.
These fine adjustment symbols correspond to
the in and out buttons on the hand control.
Press the button indicated to align the limbus
circle on the patient's limbus. When the
patient is close enough the symbol will
change to a filled in circle. The rings change
color, from red to green, when the patient is
correctly aligned.
Hand Control Capture button
Once the patient's eye has been correctly
aligned press the capture button to capture
the image.
11 of 100
12 of 100
Chapter 3
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108

Optos 200tx User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI