4RH25‐en‐GB_v1.15/15
Selecting
o
C/
o
Funitsofmeasure
WiththemeterOFF,pressandholdthePOWERandMODEbuttonssimultaneouslyuntiltheunit
of measure appears on the lower right corner of the LCD. Use the MODE button to select the
desiredunit of measure(
o
Cor
o
F).PressPOWER twicetostep pastthealarm settings untilthe
normalmeasurementmodeisdisplayed.
HighAlarmforHeatIndexandWetBulbGlobetemperature
1. WiththemeterOFF,pressandholdthePOWERandMODEbuttonssimultaneouslyuntilthe
temperatureunit‘C’ or‘F’appears on thelowerrightcorneroftheLCD (ifdesired,usethe
MODEbuttontoselectthedesiredunitofmeasure).
2. PressthePOWERbuttontostepto
theHeatIndex(HI)AlarmON/OFFdisplay.
3. UsetheMODEbuttontoselectON(A‐on)orOFF(A‐oF);seedisplayexamplesbelow:
4. PressPOWERtosteptotheHeatIndextemperatureAlarmlimitvalue.Seeexamplebelow:
5. PresstheMODEbuttonmomentarily tochangethevalueinsmallsteps.Pressandholdthe
MODEbuttontoscrollquickly.
6. When the desired Heat Temperature alarm limit is displayed, press the POWER button to
continuetotheWetBulbGlobetemperatureAlarmprogrammingscreens.
7. Repeat steps
3‐6 to program the Wet Bulb Globe temperature ON/OFF and Alarm limits.
UsethePOWERbuttonwhendonetostepbacktothenormaloperatingmode.
8. Thedefaultsettings are 27.8
o
C(82.0
o
F)forHeat Indextemperature alarmlimits and 28.0
o
C
(82.4
o
F)forWetBulbGlobetemperaturealarmlimits
9. When either alarm limit is exceeded, the meter continuously beeps. To determine which
alarm is sounding, check the HI (Heat Index) and WB/HI (Wet Bulb Globe) temperature
displaystoseewhichunitofmeasureisblinking,thatwillindicatethetrippedalarm.
10. TosilenceanalarmtheusermusteitherswitchthealarmOFForchangeitsalarmlimitvalue.
ErrorMessageDisplays
E‐2:HumiditySensorfailure.Replaceunit.
E‐3:TemperatureSensorfailure.Replaceunit.
E‐4:Temperatureover‐range;E‐5:Temperatureunder‐range
E‐6:Hardwarefailure.Replaceunit