SICK MERCEM300Z Operating instructions

Category
Measuring & layout tools
Type
Operating instructions

This manual is also suitable for

MMM
I
MMMOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS
MERCEM300Z
Mercury Analyzer
Installation, Operation, Maintenance
Title Page
2 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Document Information
Document Identification
Title: Operating Instructions MERCEM300Z
Part No.: 8013255
Version: 2.0
Release: 2014-09
Described Product
Product names: MERCEM300Z
MERCEM300Z Indoor
Manufacturer
SICK AG
Erwin-Sick-Str. 1 · D-79183 Waldkirch · Germany
Phone: +49 7641 469-0
Fax: +49 7641 469-11 49
E-mail: info.pa@sick.de
Original documents
The English edition 8013255 of this document is an original
document of SICK AG.
SICK AG assumes no liability for the correctness of an
unauthorized translation.
In case of doubt, contact SICK AG or your local representative.
Legal information
Subject to change without notice.
© SICK AG. All rights reserved.
Glossary
CAN-Bus: Control Area Network. A field bus.
Ethernet: Computer networking technology. Basis for network
protocols, such as TCP/IP.
Firewall: Safety concept of software and hardware components to
restrict access to computer networks.
Modbus®: Field bus communication protocol
PROFIBUS®: Field bus communication protocol
OLE: Object Linking and Embedding. Standardized data interface
(Microsoft Corporation)
OPC: Openness, Productivity, Collaboration. Standardized data
interface (OPC Foundation
TM
).
Span gas: Test gas with a concentration of approx. 75% of the
upper measurement range limit
SOPAS (SICK Open Portal for Applications and Systems): SICK
Parameter Setting and Data Calculation Software.
SOPAS ET: SOPAS PC Engineering Tool. Configuration protocol.
TCP/IP: Network protocol.
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 3
Warning Symbols
Warning Levels/Signal Words
DANGER
Risk or hazardous situation which will result in severe personal
injury or death.
WARNING
Risk or hazardous situation which could result in severe personal
injury or death.
CAUTION
Hazard which could result in less severe or minor injuries.
NOTICE
Hazard which could result in property damage.
Information Symbols
Hazard (general)
Hazard by voltage
Hazard by explosive substances/mixtures
Hazard by corrosive substances
Hazard by unhealthy substances
Hazard by laser radiation
Important technical information for this device
Important information on electric or electronic func-
tions
Supplementary information
Link to information at another place
Nice to know
Contents
4 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
1 Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.1 Purpose of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 Product identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3.1 Installation location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.4 Responsibility of user. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4.1 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4.2 Correct use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4.3 Special local conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4.4 Retention of documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.5 Additional documentation/information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2 Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1 Device characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2 Gas flow plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2.1 Method of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2.2 Test gas generator CALSIC300 (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2.3 Instrument air conditioning (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.3.1 Operation via SOPAS ET (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5 Remote control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5.1 Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5.2 Modbus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5.3 OPC (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.6 SOPAS ET (PC program) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.6.1 Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3 Preparations for Initial Start-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1 Overview of work to be performed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3 Preparing the installation location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.4 Transport and installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5 Laying the gas lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5.1 Laying the sample gas lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.2 Connecting the gas outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.3 Test gas feeding on gas sampling system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6 Electrical installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.6.1 Laying the tube bundle cable of the gas sampling system. . . . . . . . . . . . . . . . . . . . . 28
3.6.2 Connecting the potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.6.3 Connecting the signal lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.6.4 Preparation of mains supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.7 Ethernet interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Contents
Contents
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 5
4Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1 Putting the device into operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.3 Controls and indicators/displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.3.1 Assignment of buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.4 Status and classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.4.1 Status (operating state). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.4.2 Classification, LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.5 Measuring screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.5.1 “List” Measuring screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.5.2 “Bar graph” Measuring screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.5.3 “Line graph” Measuring screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.6 Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.1 Menu tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.2 Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
5.3.1 Maintenance/maintenance signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.3.2 Maintenance/operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.4 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.4.1 Adjustment/test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.4.2 Adjustment/manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.4.3 Adjustment/automatic. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.4.4 Adjustment/Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.5 Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.5.1 Diagnosis/check values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.5.2 Diagnosis/temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.5.3 Diagnosis/System info. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.5.4 Diagnosis/error messages and Diag button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.6 Parameter setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.6.1 Parameter settings/display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.6.2 Parameter/reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6Shutdown. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.1 Switching off (for a period up to approx. 2 weeks) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.2 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.3 Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Contents
6 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
7 Scheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
7.1 Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.1.1 Maintenance by operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.1.2 Maintenance by Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.2 Expendable and wearing parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.3 Maintenance work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.3.1 Before commencing maintenance work: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.3.2 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.3.3 Replacing the test solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8 Clearing Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
8.1 If the device does not function at all … . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.2 When measured values are obviously incorrect … . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.3 Malfunction messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.4 Stand-by . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.5 External UPS (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.6 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.7 Error messages and possible causes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
9 Technical Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.1 Compliances and approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.1.1 Electrical protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.2.1 MERCEM300Z. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.2.2 MERCEM300Z Indoor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
9.3 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Important Information
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 7
Subject to change without notice
MERCEM300Z
1 Important Information
Main instructions for operation
Intended use
Own responsibility
8 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Important Information
Subject to change without notice
1.1 Important Information
Potentially dangerous gas in device
1.2 Intended use
1.2.1 Purpose of the device
MERCEM300Z and MERCEM300Z Indoor (referred to as "device" in the following) are
extractive gas analysis systems and serve continuous measurement of mercury and
mercury compounds in the exhaust gases of industrial combustion plants (emission
measuring system).
The sample gas is taken at a sampling point and fed through the gas analysis system
(extractive measurement).
1.3 Product identification
Type plates
Number of type plates: 2
Position: Top right, inside and outside.
NOTICE: During measurement: Keep cabinet door closed
An open cabinet door results in measurement errors.
Keep the cabinet door closed during operation.
WARNING: Risk to health through contact with toxic gases
The modules and devices contain enclosed potentially dangerous gases that
can escape due to a defect or leak. Refer to the Table below for these gases.
If a leak occurs, the concentrations inside the enclosed device can increase up
to a certain concentration. These concentrations are also shown in the Table.
Regularly check the state of the device/module seals.
Only open the device when well ventilated, especially when a leak of a
device component is suspected.
Device / Module Gas Max. total volume
(ml)
Max. concentration inside the device
with a leak (defect)
MERCEM300Z
MERCEM300Z Indoor
Hg 150 μg 80 ppb
Hydrocarbons 6.5 ml
Calibration gas gener-
ator
HgCL
2
solution 10 l 30 ppb
Product names:
MERCEM300Z Indoor
MERCEM300Z
Manufacturer:
SICK AG
Erwin-Sick-Str. 1 · D-79183 Waldkirch · Germany
It is possible that the device has a different configuration to that described in
this Manual.
Refer to the System Documentation delivered with your device for the
individual configuration of your system.
Important Information
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 9
Subject to change without notice
1.3.1 Installation location
MERCEM300Z Indoor
Installation location: In air-conditioned rooms (IP43).
MERCEM300Z
Installation location: Indoors or outdoors (IP55).
10 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Important Information
Subject to change without notice
1.4 Responsibility of user
1.4.1 Target group
This Manual is intended for persons installing, operating and maintaining the
MERCEM300Z and MERCEM300Z Indoor.
Operation
The device may only be operated by qualified persons who, based on their device-specific
training and knowledge as well as knowledge of the relevant regulations, can assess the
tasks given and recognize the hazards involved.
Installation and maintenance
Skilled persons may be required at certain times during installation and maintenance.
Please observe the information at the beginning of the respective Sections.
1.4.2 Correct use
Use the device only as described in these Operating Instructions.
The manufacturer bears no responsibility for any other use.
Perform the specified maintenance work.
Do not remove, add or modify any components to or on the device unless described and
specified in the official manufacturer information. Otherwise:
The device could become dangerous.
Any warranty by the manufacturer becomes void.
1.4.3 Special local conditions
In addition to these Operating Instructions, follow all local laws, technical rules and
company-internal operating directives applicable at the respective installation location
of the device.
1.4.4 Retention of documents
These Operating Instructions and the System Documentation:
Must be available for reference.
Must be passed on to new owners.
NOTICE: Responsibility for system safety
The person setting up the system in which the device is to be integrated is
responsible for the safety of the system.
Important Information
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 11
Subject to change without notice
1.5 Additional documentation/information
The following documents are applicable in addition to these Operating Instructions:
Instructions delivered with the System Documentation
Operating Instructions SCU
Operating Instructions Cooling Unit
Operating Instructions Gas Sampling System
Operating Instructions Heated Sample Gas Line
Operating Instructions Instrument Air Conditioning (Option)
Operating Instructions Modular I/O System
Documentation on individual settings
Installation Plan
12 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Important Information
Subject to change without notice
Product Description
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 13
Subject to change without notice
MERCEM300Z
2 Product Description
Device characteristics
Functional principle
Device description
14 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Product Description
Subject to change without notice
2.1 Device characteristics
Fig. 1 MERCEM300Z front side view
Fig. 2 MERCEM300Z Indoor front side view
1 Operator panel (
p. 39, §5.2)
2 Air conditioner
1 Operator panel (
p. 39, §5.2)
Product Description
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 15
Subject to change without notice
The internal layout of the MERCEM300Z and MERCEM300Z Indoor is identical.
Fig. 3 Inside view (MERCEM300Z example)
1 Thermal converter and cell
2 Optical unit
3 Electronics with data interfaces and heater control
4 Power supply
Fuses (
p. 69, §8.6)
5Ejector
6 Sample gas line (inlet)
(Gas outlet at device base)
7 Vaporizer for test gas generator CALSIC300
(option)
8 Pressure control module
9 Valve block
10 Instrument air conditioning
11 Test gas generator CALSIC300 with storage
container (option)
16 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Product Description
Subject to change without notice
2.2 Gas flow plan
Fig. 4 Gas flow plan
1 Gas sampling system
2 Instrument air conditioning
3 Thermal converter and cell
4 Test gas generator CALSIC300 (option)
5 Vaporizer (of test gas generator CALSIC300)
Product Description
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 17
Subject to change without notice
2.2.1 Method of operation
The device operates independently.
The operating states are signaled by status signals and entered in a logbook (SOPAS ET
p. 19, §2.6).
Feeding of the sample gas to the analyzer in a heated sample gas line.
Conversion of mercury compounds into elementary mercury in a thermal converter.
Measuring principle: Photometric with Zeeman effect.
Standardized (pressure, temperature) output of measured values.
Output, relative to humid flue gas.
Test gas feed (adjustment)
Integrated adjustment cell for automatic adjustment without test gases (
p. 44, §5.4.3).
Operation with external UPS (uninterruptible power supply) (
p. 69, §8.5) possible.
2.2.2 Test gas generator CALSIC300 (option)
Option: Integrated test gas generator CALSIC300 (position
p. 15, Fig. 3).
The initial solution, an HgCl
2
solution, is in a storage container (position
p. 15, Fig. 3).
Using a vaporizer, the test gas generator generates test gas from the HgCl
2
solution.
The test gas is transported to the cell via the gas sampling system (“test gas feeding on
gas sampling system”) (
p. 16, Fig. 4 and
p. 26, §3.5.3).
The test gas concentrations of the test gas generator are set using a menu.
Test gas feeding is started cyclically or manually (
p. 41, §5.4).
2.2.3 Instrument air conditioning (option)
Option: Integrated instrument air conditioning.
Menu item Test gas feed ... Reference
Test ... manually with internal test gas generator CALSIC300
(option)
p. 41, §5.4.1
manual ... manually with external test gas feed on gas sampling
system
p. 26, §3.5.3 and
p. 42, §5.4.2
automatic ... cyclically with internal test gas generator CALSIC300
(option)
p. 44, §5.4.3
For more information on instrument air conditioning Operating Instructions
for instrument air conditioning.
18 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Product Description
Subject to change without notice
2.3 Operation
The unit is operated via an operator panel in the front door of the device.
The device signals an uncertain operating state with status indicators (
p. 33, §4.3).
The device remains in Measuring mode.
Should a malfunction occur, the device switches automatically to “Stand-By” (
p. 68,
§8.4).
The sample gas line and the sample gas path in the device are automatically purged
with instrument air in this mode.
If a test gas generator CALSIC300 is used: Test gas generator and vaporizer are also
purged.
The analog outputs remain at the last valid measured value.
2.3.1 Operation via SOPAS ET (option)
Operator menus and measured value displays are also available on an external PC via the
Ethernet connection (with the SOPAS ET engineering tool
p. 19, §2.6).
2.4 Interfaces
2 x analog outputs
4 x digital inputs
4 x digital outputs
1 x Ethernet interface
More interfaces depending on configuration (refer to “Documentation of individual set-
tings”).
2.5 Remote control
2.5.1 Ethernet
Standard: Ethernet (Modbus TCP/IP).
2.5.2 Modbus
Modbus
®
is a communication standard for digital controls to connect a »Master« device
with several »Slave« devices. The Modbus protocol defines the communication commands
only but not their electronic transfer; therefore it can be used with different digital inter-
faces (Ethernet).
2.5.3 OPC (option)
OPC is a standardized software interface that allows to exchange data between the appli-
cations of different manufacturers.
The SICK OPC server is required.
System bus: Ethernet.
Product Description
MERCEM300Z · Operating Instructions · 8013255 V 2.0 · © SICK AG 19
Subject to change without notice
2.6 SOPAS ET (PC program)
SOPAS ET can also be used to set the device parameters and provides access to the log-
book.
SOPAS ET runs on an external PC connected to the device via the Ethernet interface
(
p. 29, §3.7).
2.6.1 Language
Language setting exclusively in SOPAS ET:
Menu: Diagnosis/System info (User level: Authorized operator).
Then change once to any menu on the operator panel.
More information on SOPAS ET:
SOPAS ET Help menu
20 MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
Product Description
Subject to change without notice
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82

SICK MERCEM300Z Operating instructions

Category
Measuring & layout tools
Type
Operating instructions
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI