Mettler Toledo Thornton 4000TOC Sensor THO_4000TOC_58 130 196_en_May 2014 Operating instructions

Category
Measuring, testing & control
Type
Operating instructions
Total Organic Carbon Measurement System
Operation Manual
4000TOC Sensor
M300 TOC Transmitter
4000TOC Sensor
M300 TOC Transmitter
58 130 196 Rev B
4000TOC Sensor 2
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4000TOC Sensor 3
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Operation Manual
4000TOC Sensor
M300 TOC Transmitter
4000TOC Sensor 4
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4000TOC Sensor 5
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Content
1 Introduction ______________________________________________________________________________________________ 9
2 Safety instructions________________________________________________________________________________________ 10
2.1 Denitionofequipmentanddocumentationsymbolsanddesignations __________________________________________ 10
2.2 Correctdisposaloftheunit ____________________________________________________________________________ 11
3 Unit overview____________________________________________________________________________________________ 12
3.1 Overview, 4000TOC Sensor ____________________________________________________________________________ 12
3.1.1 SensorDetails ________________________________________________________________________________ 12
3.1.2 HighPressureApplications ______________________________________________________________________ 13
3.1.3 PotentialCondensationApplications ______________________________________________________________ 13
3.1.4 HighTemperatureApplications ___________________________________________________________________ 13
3.2 Overview, M300 TOC _________________________________________________________________________________ 13
3.2.1 1/4DIN _____________________________________________________________________________________ 13
3.2.2 1/2DIN _____________________________________________________________________________________ 14
3.3 Control/NavigationKeys _______________________________________________________________________________ 14
3.3.1 MenuStructure _______________________________________________________________________________ 14
3.3.2 Navigationkeys _______________________________________________________________________________ 15
3.3.2.1 Navigatingthemenutree _______________________________________________________________ 15
3.3.2.2 Escape
_____________________________________________________________________________ 15
3.3.2.3 Enter _______________________________________________________________________________ 15
3.3.2.4 Menu
_______________________________________________________________________________ 15
3.3.2.5 Calibrationmode _____________________________________________________________________ 15
3.3.2.6 Infomode ___________________________________________________________________________ 15
3.3.3 Navigationofdataentryelds ____________________________________________________________________ 15
3.3.4 Entryofdatavalues,selectionofdataentryoptions ___________________________________________________ 15
3.3.5 NavigationwithuinDisplay _____________________________________________________________________ 16
3.3.6 ”Savechanges”dialog _________________________________________________________________________ 16
3.3.7 SecurityPasswords ____________________________________________________________________________ 16
3.4 Display
____________________________________________________________________________________________ 16
4 M300 TOC Installation instruction ___________________________________________________________________________ 17
4.1 UnpackingandinspectionoftheM300TOC _______________________________________________________________ 17
4.1.1 Panelcutoutdimensionalinformation–
1/4DINmodels _______________________________________________________________________________ 17
4.1.2 Installationprocedure–1/4DINmodels ____________________________________________________________ 18
4.1.3 Panelcutoutdimensionalinformation–
1/2DINmodels _______________________________________________________________________________ 19
4.1.4 Installationprocedure–1/2DINmodels ____________________________________________________________ 20
4.1.5 Assembly–1/2DINversion ______________________________________________________________________ 20
4.1.6 1/2DINversion–Dimensiondrawings _____________________________________________________________ 21
4.1.7 1/2DINversion–Pipemounting __________________________________________________________________ 21
4.1.8 1/4DINversion–Dimensiondrawings _____________________________________________________________ 22
4.2 Connectionofpowersupply ____________________________________________________________________________ 23
4.2.1 1/4DINhousing(panelmount) ___________________________________________________________________ 23
4.2.2 1/2DINhousing(wallmount) ____________________________________________________________________ 24
4.3 ConnectorPINdenition _______________________________________________________________________________ 24
4.3.1 TB1andTB2for1/2DINand1/4DINversions _______________________________________________________ 25
4.3.2 TB3for1/2DINand1/4DINversions ______________________________________________________________ 25
4.4 ConnectionofTOCSensor _____________________________________________________________________________ 25
5 4000TOC Installation Instructions ___________________________________________________________________________ 26
5.1 Unpackingandinspectionofthe4000TOC ________________________________________________________________ 26
5.2 Installationofthe4000TOC ____________________________________________________________________________ 26
5.3 SampleTubingConnections ____________________________________________________________________________ 27
5.4 SampleConditioningCoil ______________________________________________________________________________ 28
5.5 ACPowerConnection _________________________________________________________________________________ 29
5.6 SensorConnection ___________________________________________________________________________________ 30
6 Placing transmitter in, or out, of service _____________________________________________________________________ 31
6.1 Placingtransmitterinservice ___________________________________________________________________________ 31
6.2 Placingtransmitteroutofservice _______________________________________________________________________ 31
7 4000TOC Sensor Operation ________________________________________________________________________________ 32
7.1 InitialStartup _______________________________________________________________________________________ 32
7.2 SettingSampleFlowRate ______________________________________________________________________________ 32
7.3 StartTOCMeasurement _______________________________________________________________________________ 33
7.4 NormalOperation ____________________________________________________________________________________ 33
7.5 SystemSuitabilityTesting ______________________________________________________________________________ 34
4000TOC Sensor 6
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
7.6 Calibration _________________________________________________________________________________________ 34
7.6.1 TOCCalibrationMethods ________________________________________________________________________ 34
8 Quick Setup _____________________________________________________________________________________________ 36
9 Sensor Calibration _______________________________________________________________________________________ 37
9.1 EnterCalibrationMode ________________________________________________________________________________ 37
10 Conguration
____________________________________________________________________________________________ 38
10.1 EnterCongurationMode ______________________________________________________________________________ 38
10.2 Measurement
_______________________________________________________________________________________ 38
10.2.1 ChannelSetup ________________________________________________________________________________ 38
10.2.2 Conductivity/ResistivityTemperatureCompensation ___________________________________________________ 39
10.2.3 SetAveraging ________________________________________________________________________________ 39
10.3 TOCSetup __________________________________________________________________________________________ 40
10.3.1 TOCGeneralParameters ________________________________________________________________________ 40
10.3.2 LampControl _________________________________________________________________________________ 41
10.3.3 Autobalance
_________________________________________________________________________________ 41
10.4 AnalogOutputs ______________________________________________________________________________________ 42
10.5 ____________________________________________________________________________________________________
Setpoints __________________________________________________________________________________________ 43
10.6 Alarm
_____________________________________________________________________________________________ 46
10.7 Display
____________________________________________________________________________________________ 47
10.7.1 Measurement
_________________________________________________________________________________ 47
10.7.2 Resolution
___________________________________________________________________________________ 48
10.7.3 Backlight
____________________________________________________________________________________ 48
10.7.4 Name ______________________________________________________________________________________ 48
10.8 HoldAnalogOutputs _________________________________________________________________________________ 49
11 System _________________________________________________________________________________________________ 50
11.1 SetLanguage _______________________________________________________________________________________ 50
11.2 USB ______________________________________________________________________________________________ 50
11.3 Passwords
_________________________________________________________________________________________ 51
11.3.1 ChangingPasswords __________________________________________________________________________ 51
11.3.2 ConguringMenuAccessforOperator _____________________________________________________________ 51
11.4 Set/ClearLockout ____________________________________________________________________________________ 51
11.5 Reset _____________________________________________________________________________________________ 52
11.5.1 ResetSystem _________________________________________________________________________________ 52
11.5.2 ResetAnalogCalibration ________________________________________________________________________ 52
11.5.3 ResetSensorCalibrationDatatoFactorySettings _____________________________________________________ 53
11.5.4 ResetCalibrationDataofSensorElectronicstoFactorySettings __________________________________________ 53
12 4000TOC Sensor Service Requirements ______________________________________________________________________ 54
12.1 UVLampReplacement ________________________________________________________________________________ 54
12.2 HighCapacityInletFilterReplacement ____________________________________________________________________ 55
13 M300 TOC Service _______________________________________________________________________________________ 56
13.1 EnterServiceMenu ___________________________________________________________________________________ 56
13.2 Diagnostics
________________________________________________________________________________________ 56
13.2.1 Model/SoftwareRevision ________________________________________________________________________ 56
13.2.2 DigitalInput __________________________________________________________________________________ 57
13.2.3 Display
_____________________________________________________________________________________ 57
13.2.4 Keypad
_____________________________________________________________________________________ 57
13.2.5 Memory
_____________________________________________________________________________________ 57
13.2.6 SetRelay ____________________________________________________________________________________ 58
13.2.7 ReadRelays _________________________________________________________________________________ 58
13.2.8 SetAnalogOutputs ____________________________________________________________________________ 58
13.2.9 ReadAnalogOutputs ___________________________________________________________________________ 58
13.3 Calibrate
___________________________________________________________________________________________ 59
13.3.1 VerifyConductivityElectronics ____________________________________________________________________ 59
13.3.2 CalibrateAnalog ______________________________________________________________________________ 59
13.3.3 CalibrateUnlock ______________________________________________________________________________ 60
13.4 TechService ________________________________________________________________________________________ 60
14 Info ____________________________________________________________________________________________________ 61
14.1 InfoMenu __________________________________________________________________________________________ 61
14.2 Messages
__________________________________________________________________________________________ 61
14.3 CalibrationData _____________________________________________________________________________________ 61
14.4 Model/SoftwareRevision ______________________________________________________________________________ 62
14.5 TOCSensorInfo _____________________________________________________________________________________ 62
15 Maintenance ____________________________________________________________________________________________ 63
15.1 ForTechnicalSupport _________________________________________________________________________________ 63
15.2 FrontPanelCleaning _________________________________________________________________________________ 63
4000TOC Sensor 7
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
16 Troubleshooting the M300 TOC Transmitter ___________________________________________________________________ 64
16.1 ChangingtheFuse ___________________________________________________________________________________ 64
16.2 Warning-andAlarmindicationonthedisplay _____________________________________________________________ 65
16.2.1 Warningindication ____________________________________________________________________________ 65
16.2.2 Alarmindication ______________________________________________________________________________ 65
17 Troubleshooting the 4000TOC Sensor ________________________________________________________________________ 66
17.1 BasicTroublshooting _________________________________________________________________________________ 66
17.2 FaultandErrorMessages ______________________________________________________________________________ 66
18 Accessories and Spare Parts _______________________________________________________________________________ 68
18.1 M300 TOC Transmitter ________________________________________________________________________________ 68
18.2 4000TOC Sensor ____________________________________________________________________________________ 68
18.2.1 4000TOCSensorSpareParts ____________________________________________________________________ 68
18.2.2 4000TOCSensorConsumableItems ______________________________________________________________ 69
4000TOC Sensor 8
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4000TOC Sensor 9
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
1 Introduction
Statementofintendeduse–The4000TOCsensorisaTotalOrganicCarbonsensordesignedto
measuretheconcentrationoforganicsubstancesinpureandultrapurewaterapplications.The
M300TOCisasinglechanneltransmitterwithalargefourlinebacklitLiquidCrystalDisplay
whichconveysmeasuringandsetupinformationforthe4000TOCsensor.
Whenusedincombination,the4000TOCsensorandM300TOCtransmitter,formaTOCmea-
surementsystemwhichprovidesthedesignflexibilityforfullintegrationintowatersystems.The
separatecomponentsallowforclosepositioningofthesensortothesamplepointforfastestre-
sponse,whileallowingintegrationofthetransmitterintothecontrolsystem.
Thismanualappliestothe4000TOCsensor,andM300TOCtransmitter:
4000TOC
Designation Orderno.
4000TOCSensor,110VAC,50/60Hz 58036021
4000TOCSensor,220VAC,50/60Hz 58036022
M300 TOC
Designation Orderno.
M300 TOC 1/4DIN 58 006 302
M300 TOC 1/2DIN 58 006 312
Throughoutthismanual,theterms‘M300’and‘M300TOC’areused.Withinthecontextofthis
document,bothtermsshouldbetakentorefertotheM300TOCtransmitter.
Theprintscreenimagesinthismanualhaveageneralexplainingcharacterandcandifferfrom
therealdisplayinyourtransmitter.
Thisdescriptioncorrespondstothefirmwarerelease,version1.0oftheM300TOCtransmitter.
Changesaretakingplaceconstantly,withoutpriornotification
4000TOC Sensor 10
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
2 Safety instructions
Thismanualincludessafetyinformationwiththefollowingdesignationsandformats.
2.1 Denitionofequipmentanddocumentationsymbols
and designations
a
WARNING:POTENTIALFORPERSONALINJURY.
a
CAUTION:possibleinstrumentdamageormalfunction.
h
NOTE:Importantoperatinginformation.
a
On the transmitter or in this manual text indicates:Cautionand/orotherpossiblehazardin-
cludingriskofelectricshock(refertoaccompanyingdocuments).
Thefollowingisalistofgeneralsafetyinstructionsandwarnings.Failuretoadheretothesein-
structionscanresultindamagetotheequipmentand/orpersonalinjurytotheoperator.
TheM300Transmittershouldbeinstalledandoperatedonlybypersonnelfamiliarwith
thetransmitterandwhoarequalifiedforsuchwork.
TheM300Transmittermustonlybeoperatedunderthespecifiedoperatingconditions
(seesection19).
RepairoftheM300Transmittermustbeperformedbyauthorized,trainedpersonnelonly.
Withtheexceptionofroutinemaintenance,cleaningproceduresorfusereplacement,as
describedinthismanual,theM300Transmittermustnotbetamperedwithoralteredin
anymanner.
Mettler-Toledoacceptsnoresponsibilityfordamagecausedbyunauthorizedmodifica-
tions
tothetransmitter.
Followallwarnings,cautions,andinstructionsindicatedonandsuppliedwiththisprod-
uct.
Installequipmentasspecifiedinthisinstructionmanual.Followappropriatelocaland
nationalcodes.
Protectivecoversmustbeinplaceatalltimesduringnormaloperation.
Ifthisequipmentisusedinamannernotspecifiedbythemanufacturer,theprotection
providedbyitagainsthazardsmaybeimpaired.
WARNINGS:
Installationofcableconnectionsandservicingofthisproductrequireaccesstoshock
hazardvoltagelevels.
Mainpowerandrelaycontactswiredtoseparatepowersourcemustbedisconnectedbe-
foreservicing.
Switchorcircuitbreakershallbeincloseproximitytotheequipmentandwithineasy
reachoftheOPERATOR;itshallbemarkedasthedisconnectingdevicefortheequip-
ment.
Mainpowermustemployaswitchorcircuitbreakerasthedisconnectingdeviceforthe
equipment.
a
4000TOC Sensor 11
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
ElectricalinstallationmustbeinaccordancewiththeNationalElectricalCodeand/orany
otherapplicablenationalorlocalcodes.
Safetyandperformancerequirethatthisinstrumentbyconnectedandproperlygrounded
throughathree-wirepowersource.
h
NOTE! RELAY CONTROL ACTION:theM300transmitterrelayswillalwaysde-energizeonlossof
power,equivalenttonormalstate,regardlessofrelaystatesettingforpoweredoperation.Config-
ureanycontrolsystemusingtheserelayswithfail-safelogicaccordingly.
h
NOTE! PROCESS UPSETS:Becauseprocessandsafetyconditionsmaydependonconsistent
operationofthistransmitter,provideappropriatemeanstomaintainoperationduringsensor
cleaning,replacementorsensororinstrumentcalibration.
h
NOTE: Thisisa4-wire-productwithanactive4–20mAanalogoutput.
PleasedonotsupplytoPin1–Pin6ofTB2.
2.2 Correct disposal of the unit
Whenthetransmitterisfinallyremovedfromservice,observealllocalenvironmentalregulations
forproperdisposal.
4000TOC Sensor 12
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
3 Unit overview
M300modelsareavailableinbotha1/4DINand1/2DINcasesize.The1/4DINisapanel-
mountonlydesignandthe1/2DINmodelsprovideanintegralIP65housingforwall-,orpipe-
mount.
3.1 Overview, 4000TOC Sensor
3.1.1 Sensor Details
The4000TOCSensorisaTotalOrganicCarbonsensordesignedforusewiththeM300TOC
transmitter.TheM300TOCwillautomaticallyidentifythe4000TOCSensorwhenitisconnect-
ed,andallfactorycalibrationdataisautomaticallyreadandusedforoperationandindication.
The4000TOCSensorandM300TOCtransmitterareconfiguredwithfactorydefaultsettingsto
minimizemeasurementsetuptime.Thesedefaultsettingsallowforsensoroperationwithmini-
malkeystrokes.The4000TOCSensorhasmorefeaturesthanothersensorsandmayrequire
additionalconfigurationdependantupontheoperationalsetup.Oncetheseparametersareset,
theTOCsensorwillfunctionautomatically,andwillrequireuserinterfaceonlyforperiodicser-
viceandmaintenance.Measurementsavailablefromthe4000TOCSensorincludeTOC,Con-
ductivity/Resistivity(compensatedanduncompensated)andTemperature.
The4000TOCSensorisdesignedwithfourlocalLEDindicatorsandaUVLampcontrolkeyon
thefrontpanel,asshownintheillustrationbelow.TheLEDsaredesignedtoprovidelocalindi-
cationofsensorstatus.TheoperationoftheLEDsissynchronizedwiththeM300TOCdisplay
statusmessagesthatappearinthe6-digitmeasurementnamelocation.
Function Color Operation
Fault Red
FlashesduringFaultcondition,sensoroperationstopped
Error Amber IlluminatesduringErrorcondition,sensorremainsoperational
SensorStatus Green
IlluminateswheneverACPowerisonandtheM300TOCis
connectedtothesensor
UVLampOn Green IlluminateswhenevertheUVLampison
TheUVLampcontrolkeyisdesignedtoprovidealocalOn/OffcontrolfortheUVlamptoallow
forquicklampchange-outandassisttheoperatorduringtroubleshooting,ifnecessary.
NOTE:IftheUVLampKeyispressedtoturntheUVLampoff,theUVLampcannotbeturnedon
fromtheM300TOC,itcanonlybeturnedonbypressingthesensorkeyagain.
h
4000TOC Sensor 13
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Thefrontcoverofthesensorishingedontheleftside.Locatedatthetopandbottomrightcor-
nersofthesensorenclosurearetwotriangle-shapeddoorfasteners.Theinstallationkitprovid-
edwiththeSensorincludesthespecialtoolneededtoloosenthesefastenersandopenthefront
doortotheenclosure.Periodicaccesswillberequiredtoperformroutineserviceandmainte-
nance.AdditionalfrontcovertoolsareavailablefromMettler-ToledoThornton,Inc.See
Section18forpartnumbersanddescriptions.
3.1.2 High Pressure Applications
Forinstallationswithprocesspressureabove85psig(5.9bar),theHighPressureRegulator
(p/n58091552)isrequired(seesection5.3).Installationinstructionsaresuppliedwiththe
HighPressureRegulator.
3.1.3 Potential Condensation Applications
Certainambientandprocesstemperaturesmaycausecondensationtooccurinsidethe
4000TOCSensor.TheMettler-ToledoThorntonSampleConditioningCoilprovidedwiththe
sensorinstallationkitisdesignedtoreducecondensationconditionsbyallowingthesample
temperaturetopartiallyequalizewiththeambienttemperature.
3.1.4 High Temperature Applications
Forwatertemperaturesabove70°C(158°F),theSampleConditioningCoil(p/n58091511)
isrequired.Inhighpressureapplications,theHighPressureRegulator(p/n58091552)must
beinstalleddownstreamoftheSampleConditioningCoil.
3.2 Overview, M300 TOC
3.2.1 1/4DIN
4
5
6
7
8
4.01
[102]
4.01
[102]
EnterInfo
METTLER TOLEDO
ESC
Menu Cal
M300
1
2
3
1:HardPolycarbonatecase 5:USBInterfacePort
2:FiveTactile-FeedbackNavigationKeys 6:RelayOutputTerminals
3:Four-lineLCDDisplay 7:AnalogOutput/DigitalInputTerminals
4:PowerSupplyTerminals 8:SensorInputTerminals
4000TOC Sensor 14
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
3.2.2 1/2DIN
5.90
[150]
5.90
[150]
EnterInfo
CalMenu
M300
1
2
3
ESC
4
5
6
7
8
1:HardPolycarbonatecase 5:USBInterfacePort
2:FiveTactile-FeedbackNavigationKeys 6:RelayOutputTerminals
3:Four-lineLCDDisplay 7:AnalogOutput/DigitalInputTerminals
4:PowerSupplyTerminals 8:SensorInputTerminals
3.3 Control/Navigation Keys
3.3.1 Menu Structure
BelowisthestructureoftheM300menutree:
Measurement
Mode M300
CalMenu Info
Quick Setup Configure System Service Messages
TOC Sensor
Info
Measurement
Analog Outputs
Set Points
Alarm/Clean
Display
Set Language
USB
Passwords
Set/Clear
Lockout
Reset
Calibrate
Tech Service
Diagnostics
Hold Outputs
Calibration Data
Model/Software
Revision
Channel Setup
Analog Outputs
Set Points
Set Flow
Set Flow
UV Lamp
4000TOC Sensor 15
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
3.3.2 Navigation keys
Menu Cal
Info
Enter
ESC
3.3.2.1 Navigating the menu tree
EnterthedesiredmainMenubranchwiththe or keys.Usetheandkeystonavi-
gatethroughtheselectedMenubranch.
h
NOTE:Inordertobackuponemenupage,withoutescapingtothemeasurementmode,
movethecursorundertheUPArrowcharacter(u) atthebottomrightofthedisplayscreenand
press[Enter].
3.3.2.2 Escape
Presstheandkeysimultaneously(escape)toreturntotheMeasurementmode.
3.3.2.3 Enter
Usetheekeytoconfirmactionorselections.
3.3.2.4 Menu
PressthekeytoaccessthemainMenu.
3.3.2.5 Calibration mode
PressthekeytoenterCalibrationMode.
3.3.2.6 Info mode
PressthekeytoenterInfoMode
3.3.3 Navigationofdataentryelds
Usethekeytonavigateforwardorthe keytonavigatebackwardswithinthechangeable
dataentryfieldsofthedisplay.
3.3.4 Entry of data values, selection of data entry options
4000TOC Sensor 16
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Usethekeytoincreaseorthekeytodecreaseadigit.Usethesamekeystonavigatewith-
inaselectionofvaluesoroptionsofadataentryfield.
h
NOTE:Somescreensrequireconfiguringmultiplevaluesviathesamedatafield(ex:configuring
multiplesetpoints).Besuretousethe or keytoreturntotheprimaryfieldandthe or
keytotogglebetweenallconfigurationoptionsbeforeenteringtothenextdisplayscreen.
3.3.5 Navigation with u in Display
Ifauisdisplayedonthebottomrighthandcornerofthedisplay,youcanusetheorthe
keytonavigatetoit.Ifyouclick[ENTER]youwillnavigatebackwardsthroughthemenu
(gobackonescreen).Thiscanbeaveryusefuloptiontomovebackupthemenutreewithout
havingtoexitintothemeasuringmodeandre-enterthemenu.
3.3.6 ”Save changes” dialog
Threeoptionsarepossibleforthe”Savechanges”dialog:Yes&Exit(Savechangesandexitto
measuringmode),”Yes&u”(Savechangesandgobackonescreen)and”No&Exit”(Don’t
savechangesandexittomeasuringmode).The”Yes&u”optionisveryusefulifyouwantto
continueconfiguringwithouthavingtore-enterthemenu.
3.3.7 Security Passwords
TheM300transmitterallowsasecuritylock-outofvariousmenus.Ifthesecuritylock-out
featureofthetransmitterhasbeenenabled,asecuritypasswordmustbeenteredtoallowac-
cesstothemenu.Seesection11.3formoreinformation.
3.4 Display
h
NOTE:IntheeventofanalarmorothererrorconditiontheM300Transmitterwilldisplayaflash-
inga intheupperrightcornerofthedisplay.Thissymbolwillremainuntiltheconditionthat
causedithasbeencleared.
h
NOTE:Duringcalibrations,clean,DigitalInwithAnalogOutput/Relay/USBinHoldstate,
aflashingHwillappearintheupperleftcornerofthedisplay.Thissymbolwillremainfor
20secondsuntilafterthecalibrationorcleaniscompleted.Thissymbolwillalsodisappear
whenDigitalInisdeactivated.
4000TOC Sensor 17
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4 M300 TOC Installation instruction
4.1 Unpacking and inspection of the M300 TOC
Inspecttheshippingcontainer.Ifitisdamaged,contacttheshipperimmediatelyforinstructions.
Donotdiscardthebox.
Ifthereisnoapparentdamage,unpackthecontainer.Besureallitemsshownonthepacking
listarepresent.
Ifitemsaremissing,notifyMettler-Toledoimmediately.
4.1.1 Panel cutout dimensional information –
1/4DIN models
1/4DINModeltransmittersaredesignedforpanel-mountinstallationonly.Eachtransmitteris
suppliedwithmountinghardwaretoprovidefastandsimpleinstallationtoaflatpanelorflaten-
closuredoor.ToinsureagoodsealandmaintainIPintegrityofinstallation,thepanelordoor
mustbeflatandhaveasmoothfinish.Hardwareconsistsof:
2snap-onMountingbrackets
1mountinggasketseal
Transmitterdimensionsandmountingareshowninthefiguresbelow.
3.62
+0.02
–0
[92
+0.5
]
–0
3.62
+0.02
–0
[92
+0.5
]
–0
PANEL CUT-OUT
4000TOC Sensor 18
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4.1.2 Installation procedure – 1/4DIN models
Makecutoutinpanel(seedimensionscutoutdrawing).
Besuresurfacesurroundingcutoutisclean,smoothandfreeofburrs.
Slidefacegasket(suppliedwithtransmitter)aroundtransmitterfromthebackoftheunit.
Placetransmitterintocutouthole.Besuretherearenogapsbetweenthetransmitter
andpanelsurface.
Placethetwomountingbracketsoneithersideofthetransmitterasshown.
Whileholdingtransmitterfirmlyintothecutouthole,pushthemountingbrackets
towardthebacksideofpanel.
Oncesecure,useascrewdrivertotightenthebracketsagainstthepanel.Inordertopro-
videIP65environmentalenclosurerating,thetwoclampsprovidedshallbesecurely
tightenedtocreateanadequatesealbetweenthepanelenclosureandM300frontface.
Facegasketwillcompressbetweentransmitterandpanel.
a
CAUTION:Donotovertightenbrackets.
4.01
[102]
4.33
[110]
3.55
[90]
4.01
[102]
3.55
[90]
4.01
[102]
4.96
[126]
1.77
[45]
3.54
[90]
4.33
[110]
MAX. 0.118
[3]
FLAT GASKET
MOUNTING SCREW
2 PLACES
MOUNTING
BRACKET
2 PLACES
PowerFuse
USB
1234567
1234567
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
4.92
[125]
4.01
[102]
4.33
[110]
3.55
[90]
4.01
[102]
3.55
[90]
4.01
[102]
4.96
[126]
1.77
[45]
3.54
[90]
4.33
[110]
MAX. 0.118
[3]
FLAT GASKET
MOUNTING SCREW
2 PLACES
MOUNTING
BRACKET
2 PLACES
PowerFuse
USB
1234567
1234567
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
4.92
[125]
4000TOC Sensor 19
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4.1.3 Panel cutout dimensional information –
1/2DIN models
1/2DINModeltransmittersaredesignedwithanintegralrearcoverforstand-alonewallmount
installation.
Theunitmayalsobewallmountedusingtheintegralrearcover.Seeinstallationinstructionsin
Section4.1.4.
Belowarecut-outdimensionsrequiredbythe1/2DINmodelswhenmountedwithinaflatpanel
oronaflatenclosuredoor.Thissurfacemustbeflatandsmooth.Texturedorroughsurfacesare
notrecommendedandmaylimittheeffectivenessofthegasketsealprovided.
5.39
+0.02
–0.00
[137
+0.5
]
–0.0
5.39
+0.02
–0.00
[137
+0.5
]
–0.0
PANEL CUT-OUT
Optionalhardwareaccessoriesareavailablethatallowforpanel-orpipe-mount.
RefertoSection18“AccessoriesandSpareParts”fororderinginformation.
4000TOC Sensor 20
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4.1.4 Installation procedure – 1/2DIN models
General:
Orientthetransmittersothatthecablegripsfacedownward.
Wiringroutedthroughthecablegripsshallbesuitableforuseinwetlocations.
InordertoprovideIP65enclosureratings,allcableglandsmustbeinplace.Eachcable
glandmustbefilledusingacable,orsuitableCableGlandHoleSeal.
For Wall Mount:
Removerearcoverfromfronthousing.
Startbyunscrewingthefourscrewslocatedonthefaceofthetransmitter,ineachcorner.
Thisallowsthefrontcovertoswingawayfromtherearhousing.
Removethehinge-pinbysqueezingthepinfromeachend.
Thisallowsthefronthousingtoberemovedfromtherearhousing.
Mountrearhousingtowall.SecuremountingkittotheM300accordingtothesupplied
instructions.Attachtowallusingappropriatemountinghardwareforwallsurface.Be
sureitislevelandsecurelyfastenedandtheinstallationadherestoanyandallclearance
dimensionsrequiredfortransmitterserviceandmaintenance.Orientthetransmitterso
thatthecablegripsarefacingdownward.
Replacethefronthousingtotherearhousing.Securelytightentherear-coverscrewsto
ensurethatIP65enclosureenvironmentalratingisprovided.Theunitisreadytobe
wired.
For Pipe Mount:
Useonlymanufacturer-suppliedcomponentsforpipe-mountingtheM300transmitterand
installperthesuppliedinstructions.Seesection18fororderinginformation.
4.1.5 Assembly – 1/2DIN version
1
2
3
1:3Pg13.5cableglands
2: 2plasticplugs
3: 4screws
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78

Mettler Toledo Thornton 4000TOC Sensor THO_4000TOC_58 130 196_en_May 2014 Operating instructions

Category
Measuring, testing & control
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI