Service Guide
II 34450A Service Guide
Notices
© Agilent Technologies, Inc., 2012–2013
No part of this manual may be reproduced in
any form or by any means (including elec-
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ment and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
34450-90029
Edition
Third Edition, August 30, 2013
Agilent Technologies, Inc.
5301, Stevens Creek Blvd.
Santa Clara, CA 95051 USA
Warranty
The material contained in this document is
provided “as is,” and is subject to change,
without notice, in future editions. Further,
to the maximum extent permitted by the
applicable law, Agilent disclaims all war-
ranties, either express or implied, with
regard to this manual and any information
contained herein, including but not limited
to the implied warranties of merchantabil-
ity and fitness for a particular purpose.
Agilent shall not be liable for errors or for
incidental or consequential damages in
connection with the furnishing, use, or
performance of this document or of any
information contained herein. Should Agi-
lent and the user have a separate written
agreement with warranty terms covering
the material in this document that conflict
with these terms, the warranty terms in
the separate agreement shall control.
Technology Licenses
The hardware and or software described in
this document are furnished under a license
and may be used or copied only in accor-
dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft-
ware and technical data rights granted to
the federal government include only those
rights customarily provided to end user cus-
tomers. Agilent provides this customary
commercial license in Software and techni-
cal data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Com-
puter Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a haz-
ard. It calls attention to an operat-
ing procedure, practice, or the likes
of that, if not correctly performed
or adhered to, could result in dam-
age to the product or loss of impor-
tant data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the likes of that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARN-
ING notice until the indicated
conditions are fully understood
and met.
34450A Service Guide III
Safety Symbols
The following symbols on the instrument and in the documentation
indicate precautions that must be taken to maintain safe operation of the
instrument.
Earth (ground) terminal
Caution, risk of electric shock
Frame or chassis terminal
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information)
IEC Measurement Category II. Inputs
may be connected to mains (up to
300 VAC) under Category II overvoltage
conditions.
CAT II
300 V
IV 34450A Service Guide
Regulatory Markings
The CE mark is a registered trademark
of the European Community. This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
The C-tick mark is a registered
trademark of the Spectrum
Management Agency of Australia. This
signifies compliance with
the Australia EMC Framework
regulations under the terms of the
Radio Communication Act of 1992.
ICES/NMB-001 indicates that this ISM
device complies with the Canadian
ICES-001.
Cet appareil ISM est confomre a la
norme NMB-001 du Canada.
This instrument complies with the
WEEE Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard
this electrical/electronic product in
domestic household waste.
The CSA mark is a registered
trademark of the Canadian Standards
Association.
This symbol indicates the time period
during which no hazardous or toxic
substance elements are expected to
leak or deteriorate during normal use.
Forty years is the expected useful life
of the product.
34450A Service Guide V
General Safety Information
The following general safety precautions must be observed during all
phases of operation, service, and repair of this instrument. Failure to
comply with these precautions or with specific warnings elsewhere in this
manual violates safety standards of design, manufacture, and intended
use of the instrument. Agilent Technologies assumes no liability for the
customer’s failure to comply with these requirements.
VI 34450A Service Guide
Protection Limits
The Agilent 34450A 5½ Digital Multimeter
provides protection circuitry to prevent
damage to the instrument and to protect
against the danger of electric shock, pro-
vided that the Protection Limits are not
exceeded. To ensure safe operation of the
instrument, do not exceed the Protection
Limits shown on the front panel, as defined
below:
Note: The front-panel terminals and current
protection fuse are shown above.
Input Terminal Protection
Limits
Protection Limits are defined for the input
terminals:
Main Input (HI and LO) Terminals. The HI
and LO input terminals are used for voltage,
resistance, capacitance, and diode test
measurements. Two Protection Limits are
defined for these terminals:
HI to LO Protection Limit. The Protection
Limit from HI to LO ("A" in the figure
above) is 1000 VDC or 750 VAC, which is
also the maximum voltage measurement.
This limit can also be expressed as
1000 Vpk maximum.
LO to Ground Protection Limit. The LO
input terminal can safely "float" a maxi-
mum of 500 Vpk relative to ground. This is
Protection Limit "B" in the figure.
Although not shown on the figure, the Pro-
tection Limit for the HI terminal is a maxi-
mum of 1000 Vpk relative to the ground.
Therefore, the sum of the “float” voltage
and the measured voltage must not exceed
1000 Vpk
Current Input Terminal. The current input
("I") terminal has a Protection Limit of
100 mA (rms) maximum current flowing
from the LO input terminal. This is Protec-
tion Limit "C" in the figure. Note that the
current input terminal will be at approxi-
mately the same voltage as the LO terminal.
Note: The current-protection circuitry
includes a fuse on the back panel. To main-
tain protection, replace this fuse only with a
fuse of the specified type and rating.
10 A Current Input Terminal. The 10 A cur-
rent input terminal has a Protection Limit of
10 A (rms) maximum current flowing from
the LO input terminal. This is Protection
Limit "D" in the figure. Note that the current
input terminal will be at approximately the
same voltage as the LO terminal.
Note: The current-protection circuitry
includes an internal fuse. To maintain pro-
tection, service-trained personnel should
replace this fuse only with a fuse of the
specified type and rating.
Sense Terminal Protection
Limits
The HI and LO sense terminals are used
only for four-wire resistance measurements
(" Ω 4W"). The Protection Limit is 200 Vpk
for all of the terminal pairings ("E" in the fig-
ure):
LO sense to LO input.
HI sense to LO input.
HI sense to LO sense.
Note: The 200 Vpk limit on the sense termi-
nals is the Protection Limit. Operational
voltages in resistance measurements are-
much lower - less than 5 V in normal opera-
tion.
IEC Measurement Category II
Overvoltage Protection
To protect against the danger of electric
shock, the Agilent 34450A 5½ Digital Multi-
meter provides overvoltage protection for
line-voltage mains connections meeting
both of the following conditions:
The HI and LO input terminals are con-
nected to the mains under Measurement
Category II conditions, defined below, and
The mains are limited to a maximum line
voltage of 300 VAC.
IEC Measurement Category II includes elec-
trical devices connected to mains at an out-
let on a branch circuit. Such devices include
most small appliances, test equipment, and
other devices that plug into a branch outlet
or socket. The 34450A may be used to make
measurements with the HI and LO inputs
connected to mains in such devices, or to
the branch outlet itself (up to 300 VAC).
However, the 34450A may not be used with
its HI and LO inputs connected to mains in
permanently installed electrical devices
such as the main circuit-breaker panel,
sub-panel disconnect boxes, or permanently
wired motors. Such devices and circuits are
subject to overvoltages that may exceed the
protection limits of the 34450A.
Note: Voltages above 300 VAC may be mea-
sured only in circuits that are isolated from
mains. However, transient overvoltages are
also present on circuits that are isolated
from mains. The 34450A is designed to
safely withstand occasional transient over-
voltages up to 2500 Vpk. Do not use this
multimeter to measure circuits where tran-
sient overvoltages could exceed this level.
A
C
D
B
E
34450A Service Guide VII
WARNING
Do not defeat the power cord safety ground feature. Plug in to a grounded
(earthed) outlet.
Do not use the instrument in any manner that is not specified by the
manufacturer.
To avoid electric shock or injury, do not operate the multimeter without panels or
case in place.
Do not substitute parts or modify the instrument to avoid the danger of
introducing additional hazards. Return the instrument to Agilent Technologies
Sales and Service Office for service and repair to ensure the safety features are
maintained.
Main Power and Test Input Disconnect: Unplug the instrument from the wall
outlet, remove the power cord, and remove all probes from all terminals before
servicing. Only qualified, service-trained personnel should remove the cover from
the instrument.
Line and Current Protection Fuses: For continued protection against fire, replace
the line fuse and the current-protection fuse only with fuses of the specified type
and rating.
IEC Measurement Category II. The HI and LO input terminals may be connected to
mains in IEC Category II installations for line voltages up to 300 VAC. To avoid the
danger of electric shock, do not connect the inputs to mains for line voltages
above 300 VAC. See “IEC Measurement Category II Overvoltage Protection” on
the following page for further information.
Protection limits: To avoid instrument damage and the risk of electric shock, do
not exceed any of the Protection Limits defined in the following section.
If the Test Lead Set is used in a manner not specified by Agilent Technologies, the
protection provided by the Test Lead Set may be impaired. Also, do not use a
damaged or worn Test Lead Set. Instrument damage or personal injury may
result.
VIII 34450A Service Guide
Environmental Conditions
This instrument is designed for indoor use and in an area with low
condensation. The table below shows the general environmental
requirements for the instrument.
Environmental Condition Requirement
Operating temperature Full accuracy at 0 °C to 55 °C
Operating humidity Full accuracy up to 80% RH at 30 °C
(non-condensing)
Storage temperature 40 °C to 70 °C
Altitude Operating up to 3,000 meters
Pollution degree Pollution Degree 2
NOTE
The Agilent 34450A 5½ Digit Multimeter complies with the following EMC and
safety requirements:
IEC 61010-1:2001 / EN 61010-1:2001 (2nd Edition)
IEC 61326-2-1:2005 / EN61326-2-1:2006
CISPR 11:2003 / EN 55011:2007 Group 1 Class A
Canada: ICES/NMB-001:Issue 4, June 2006
Australia/New Zealand: AS/NZS CISPR 11:2004
CAUTION
Degradation of some product specifications can occur in the presence of ambient
electromagnetic (EM) fields and noise that are coupled to the power line or I/O
cables of the instrument. The instrument will self-recover and operate to all
specifications when the source of ambient EM field and noise are removed or when
the instrument is protected from the ambient EM field or when the instrument
cabling is shielded from the ambient EM noise.
34450A Service Guide IX
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard
this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is shown as below:
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent office, or
visit www.agilent.com/environment/product for more information.
X34450A Service Guide
Additional Notices
The Agilent 34450A is provided with an Agilent 34138A Test Lead Set,
described below.
Test Lead Ratings
Test Leads - 1000 V, 15 A
Fine Tip Probe Attachments - 300 V, 3 A
Mini Grabber Attachment - 300 V, 3 A
SMT Grabber Attachments - 300 V, 3 A
Operation
The Fine Tip, Mini Grabber, and SMT Grabber attachments plug onto the
probe end of the Test Leads.
Maintenance
If any portion of the Test Lead Set is worn or damaged, do not use. Replace
with a new Agilent 34138A Test Lead Set.
WARNING
If the Test Lead Set is used in a manner not specified by Agilent Technologies, the
protection provided by the Test Lead Set may be impaired. Also, do not use a
damaged or worn Test Lead Set. Instrument damage or personal injury may result.
34450A Service Guide XI
Declaration of Conformity (DoC)
The Declaration of Conformity (DoC) for this instrument is available on the
Web site. You can search the DoC by its product model or description.
http://regulations.corporate.agilent.com/DoC/search.htm
NOTE
If you are unable to search for the respective DoC, please contact your
local Agilent representative.
XII 34450A Service Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
34450A Service Guide XIII
Table of Contents
1 Performance Tests and Calibration
Calibration Overview 2
Closed-case calibration 2
Agilent Technologies calibration services 2
Calibration interval 3
Time required for calibration 3
Automating calibration procedures 3
Recommended Test Equipment 4
Test Considerations 5
Input connections 6
Performance Verification Tests Overview 7
Self test 7
Quick performance check 8
Performance Verification Tests 9
Zero offset verification 9
Gain verification 11
DC current gain verification test 12
Ohms gain verification test 14
Frequency gain verification test 16
AC voltage verification test 17
AC current verification test 19
Capacitance performance verification test 22
Calibration Security 23
Unsecuring the instrument for calibration 24
Calibration Process 26
Using the front panel for adjustments 27
Adjustments 29
XIV 34450A Service Guide
Zero adjustment 29
Gain adjustments 29
DC voltage gain adjustment procedure 30
DC current gain adjustments procedure 32
AC voltage gain adjustment procedure 33
AC current gain adjustment procedure 35
Ohms gain adjustment procedure 36
Capacitance gain adjustment procedure 38
Frequency gain adjustment procedure 39
Finishing the adjustments 39
Calibration Message 40
To read the calibration count 40
Calibration Errors 41
2 Disassembly and Repair
Operating Checklist 44
Types of Services Available 45
Repackaging for Shipment 46
Cleaning 46
To Replace the Power Line Fuse 47
To Replace a Current Input Fuse 48
To Verify your Device License 49
Self Test Errors 50
Electrostatic Discharge (ESD) Precautions 51
Mechanical Disassembly 52
Replaceable Parts 58
Rack Mounting 60
34450A Service Guide XV
List of Tables
Tab l e 1- 1 Recommended test equipment 4
Tab l e 1- 2 Zero offset verification test 10
Tab l e 1- 3 DC voltage gain verification test 12
Tab l e 1- 4 DC current gain verification test 13
Tab l e 1- 5 Ohms gain verification test 15
Tab l e 1- 6 Frequency gain verification test 16
Tab l e 1- 7 AC volts verification test 17
Tab l e 1- 8 AC current verification test 20
Tab l e 1- 9 Capacitance verification test 22
Tab l e 1- 1 0 Valid gain and frequency compensation input values 30
Tab l e 1- 1 1 DC voltage gain adjustment 31
Tab l e 1- 1 2 DC current gain adjustment 32
Tab l e 1- 1 3 AC voltage gain adjustment 34
Tab l e 1- 1 4 AC current gain adjustment 35
Tab l e 1- 1 5 Ohms gain adjustment 37
Tab l e 1- 1 6 Capacitance gain adjustment 38
Tab l e 1- 1 7 Frequency gain adjustment 39
Tab l e 2- 1 Self test error numbers 50
Tab l e 2- 2 Replaceable parts list 58
XVI 34450A Service Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
1
Agilent 34450A 5½ Digit Multimeter
Service Guide
Agilent Technologies
1
Performance Tests and Calibration
Calibration Overview 2
Recommended Test Equipment 4
Test Considerations 5
Performance Verification Te sts O v e r v i e w 7
Performance Verification Tests 9
Calibration Security 23
Calibration Process 26
Adjustments 29
Calibration Message 40
Calibration Errors 41
This chapter contains performance test procedures and
calibration procedures. The performance test procedures
allow you to verify that the multimeter is operating within
its published specifications.
WARNING
Only service–trained personnel who are aware of the hazards
involved should perform the procedures in this chapter. To avoid
electrical shock and personal injury, make sure you have read and
follow all test equipment safety instructions. Use only completely
electric insulated test lead sets with connectors which are out of
contact with test voltages.
1 Performance Tests and Calibration
Calibration Overview
234450A Service Guide
Calibration Overview
.
Closed-case calibration
The instrument features closed- case electronic calibration. No internal
mechanical adjustments are required. The instrument calculates
correction factors based upon the input reference value you set. The new
correction factors are stored in non- volatile memory until the next
calibration adjustment is performed. Non- volatile EEPROM calibration
memory does not change when power has been off or after a remote
interface reset.
Agilent Technologies calibration services
When your instrument is due for calibration, contact your local Agilent
Service Center for a low- cost re- calibration. The 34450A is supported on
automated calibration systems, which allow Agilent to provide this
service at competitive prices.
NOTE
Make sure you have read “Test Considerations” on page 5 before calibrating the
instrument.
Performance Tests and Calibration 1
Calibration Overview
34450A Service Guide 3
Calibration interval
A one year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at regular
calibration intervals. Accuracy specifications are not warranted beyond
the one year calibration interval. Agilent does not recommend extending
calibration intervals beyond two years for any application.
Time required for calibration
The 34450A can be automatically calibrated under computer control.
With computer control you can perform the complete calibration
procedure and performance verification tests in less than 60 minutes
once the instrument is warmed- up (see“Test Considerations” on page 5).
Refer to the 34450A Programmer’s Reference online help for more
information.
Automating calibration procedures
The adjustment procedures provided in this Service Guide demonstrate
front panel adjustment. You can automate the complete verification and
adjustment procedures outlined in this manual. You can program the
instrument configurations specified for each test over the remote
interface. You can then enter read back verification data into a test
program and compare the results to the appropriate test limit values.
The instrument calibration must be unsecured to perform a calibration.
1 Performance Tests and Calibration
Recommended Test Equipment
434450A Service Guide
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed in Table 1- 1 below. If the exact
instrument is not available, substitute calibration standards of equivalent
accuracy.
A suggested alternate method would be to use the Agilent 3458A 8½
Digit Digital Multimeter to measure less accurate yet stable sources. The
output value measured from the source can be entered into the
instrument as the target calibration value.
Table 1- 1 Recommended test equipment
Application Recommended Equipment Recommended Accuracy Requirements
Zero calibration Fluke 5520A or shorting Plug-Dual banana plug with
copper wire short between the two terminals
DC voltage Fluke 5520A <1/5 instrument 1 year spec
DC current Fluke 5520A <1/5 instrument 1 year spec
Resistance Fluke 5520A <1/5 instrument 1 year spec
AC voltage Fluke 5520A <1/5 instrument 1 year spec
AC current Fluke 5520A <1/5 instrument 1 year spec
Frequency Fluke 5520A <1/5 instrument 1 year spec
Capacitance Fluke 5520A <1/5 instrument 1 year spec
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77