Steris System 1 Endo Liquid Chemical Sterilant Processing System Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Steris System 1 Endo Liquid Chemical Sterilant Processing System Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
OPERATOR MANUAL
SYSTEM 1® endo LIQUID CHEMICAL
STERILANT PROCESSING SYSTEM
MODEL P6900
S40® STERILANT CONCENTRATE
CATALOG NO. S4000
T6900 Rev A
WARNING — COPYING PROHIBITED
This manual is protected by Federal Copyright Law, which provides for damages of up to USD $20000,
as well as criminal fines and imprisonment for unauthorized copying.
Introduction Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
i
A WORD FROM STERIS
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended
for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical
heat-sensitive medical devices and their accessories in healthcare facilities. The
SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate
to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the
load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load
with 0.2 micron filtered water. The SYSTEM 1 endo Processor uses only S40
Sterilant Concentrate to liquid chemically sterilize medical devices.
After completion of a cycle in the SYSTEM 1 endo Processor (which includes
liquid chemical sterilization and rinse phases), the processed load should
be used immediately or prepared for storage according to the endoscope
manufacturer’s written instructions for drying and storage following a liquid
chemical sterilization or high-level disinfection process.
SYSTEM 1 endo Processor is to be used only by trained personnel who are fully
familiar with, and strictly comply with, the operating instructions and conditions
detailed in the Operator Manual, as well as any applicable Quick Connect
processing instructions. The warranty pertaining to equipment or products sold
by STERIS is contained in STERIS’s Terms and Conditions of Sale.
Important: Liquid chemical sterilization (LCS) of a medical device by the
SYSTEM 1 endo Processor requires that the liquid sterilant makes contact with
all exterior and interior surfaces of the device. In the case of a device having
internal channels or other interior surfaces, the device manufacturer’s instructions
should be reviewed to determine the manufacturer’s recommendations for the
preparation of the device for liquid chemical sterilization or high-level disinfection.
STERIS is not responsible for the adequacy of the device manufacturer’s
instructions or the design of a device that fails to permit liquid to contact the
surfaces of all internal channels and components of the devices. The device
will not be liquid chemically sterilized if the sterilant does not contact all interior
and exterior surfaces of the device.
Overview
T6900 Rev A Operator Manual Introduction
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
ii
©2019, STERIS Corporation. All rights reserved. This publication is protected by copyright.
Copying, disclosure to others, or other use of this publication is prohibited without the express
written consent of STERIS Corporation. STERIS Corporation reserves the right to make
changes in specifications shown herein without notice or obligation. Contact your STERIS
representative or STERIS Customer Service for the latest information.
STERIS, SYSTEM 1, S40, VERIFY and MaxLife are registered trademarks of STERIS
Corporation.
U.S. Patents
7,135,142
7,569,182
Printed in Canada.
Manufactured exclusively by
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060-1834 USA
440-354-2600 n 800-548-4873
www.steris.com
Customer Service
440-354-2600
800-548-4873
Certified to ISO 9001 and 13485
Introduction Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
iii
A WORD FROM STERIS ...............................................................................................i
Overview .............................................................................................................................................. i
1. LISTING OF WARNINGS AND CAUTIONS ................................................................. 1-1
2. SITE SPECIFICATIONS ........................................................................................... 2-1
3. S40 STERILANT CONCENTRATE ............................................................................ 3-1
Part 1: Overview ............................................................................................................................3-2
Part 2: Safety Information for Peracetic Acid ..................................................................................3-3
Part 3: Disposal .............................................................................................................................3-4
4. COMPONENT IDENTIFICATION............................................................................... 4-1
Part 1: Illustrations of Processor and Components .........................................................................4-1
Part 2: SYSTEM 1 endo Processor ................................................................................................4-4
5. GENERAL PROCESSING INSTRUCTIONS ................................................................ 5-1
Part 1: Processor with Open-Style Processing Tray ........................................................................5-1
Part 2: Processor with Processing Tray and Container ...................................................................5-2
Part 3: Preparation of Devices for Liquid Chemical Sterilization ......................................................5-3
6. SPECIFIC DEVICE PROCESSING INSTRUCTIONS .................................................... 6-1
Part 1: Quick Connect Information .................................................................................................6-1
7. OPERATION OF THE PROCESSOR .......................................................................... 7-1
Part 1: LCS Cycle ..........................................................................................................................7-1
Part 2: Diagnostic Cycle ...............................................................................................................7-10
Part 3: Aborting Cycles ................................................................................................................7-12
Part 4: Troubleshooting ................................................................................................................7-14
8. QUALITY ASSURANCE ........................................................................................... 8-1
Part 1: Documentation ...................................................................................................................8-1
9. MAINTENANCE ...................................................................................................... 9-1
Part 1: Daily Cleaning and Checks .................................................................................................9-1
Part 2: Changing 0.2 Micron MaxLife Water Filter ...........................................................................9-5
Part 3: Changing Sterile Air Filter ..................................................................................................9-11
Part 4: Changing the Printer Paper for the Optional Printer ..........................................................9-13
Part 5: Changing Pre-filters ..........................................................................................................9-15
Part 6: Supervisor Mode ..............................................................................................................9-17
Part 7: Replacement Parts ...........................................................................................................9-23
Part 8: Routine Maintenance Parts ...............................................................................................9-24
TABLE OF CONTENTS
Section Title Page
T6900 Rev A Operator Manual Introduction
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
iv
10. PROCESS MONITORING ....................................................................................... 10-1
Part 1: Overview ..........................................................................................................................10-1
Part 2: Process Assurance Monitoring .........................................................................................10-2
11. INSERVICE ........................................................................................................... 11-1
Listing of Warnings and Cautions Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
1-1
LISTING OF WARNINGS AND CAUTIONS 1
WARNINGS:
Do not use the SYSTEM 1 endo Processor until it has been properly installed and its electro/mechanical
performance is verified.
The SYSTEM 1 endo Processor may be used only in hospital-grade rooms when it is installed according to the
applicable specification (VDE 0107, Class 1 Group C NFPA Rating).
The SYSTEM 1 endo Processor is not intended for operation in explosion-endangered areas.
The SYSTEM 1 endo Processor is not protected against water spillage or spray. Take precautions in the use of
liquid during cleaning or insertion of devices.
Connect the SYSTEM 1 endo Processor to hospital grade receptacle only following national and local codes
(GFCI, RCBO, etc.).
Do not attempt to access non-user serviceable components of the SYSTEM 1 endo Processor.
Failure to thoroughly clean devices, endoscopes, etc., may result in ineffective liquid chemical sterilization.
Failure to properly position devices so that all surfaces will be exposed to the liquid sterilant or overloading the
processing container may result in ineffective liquid chemical sterilization and/or damage the devices.
Always verify that the sterilant container is empty at the completion of the cycle.
Devices are not liquid chemically sterilized and/or adequately rinsed when an LCS Cycle is cancelled.
Danger: Risk of explosion if used in the presence of flammable substances.
Repairs by unauthorized individuals should not be attempted and may result in damage or malfunction. Non-user
serviceable components should be serviced by an authorized STERIS Service Representative only.
Disconnect electrical power before cleaning.
The performance of the SYSTEM 1 endo Processor has been validated and optimized with components defined by
STERIS in the Operator Manual for the SYSTEM 1 endo Processor. Use of pre-filters that do not meet STERIS's
specifications and have not been validated by STERIS can lead to premature failure of the MaxLife water filter
inside the processor. STERIS’s warranty will not apply to damage or early life failure of the MaxLife water filter
resulting from the use of pre-filters that do not meet STERIS’s specifications. DO NOT USE components that do
not meet STERIS’s specifications and have not been validated by STERIS.
Do not position equipment so that it is difficult to disconnect the plug from the duplex receptacle.
Do not substitute any other product for S40 Sterilant Concentrate.
Contact with or inhalation of the inert powdered ingredients in the outer cup of S40 Sterilant Concentrate may
cause irritation to eye, skin or respiratory system.
Throughout this manual, WARNING statements have been included to alert the Operator to important information
regarding safety and liquid chemical sterilization efficacy.
CAUTION statements advise the Operator of important information regarding processor and device safety, and proper
operating technique.
Please review and be familiar with these statements before using the SYSTEM 1 endo Processor and S40 Sterilant
Concentrate.
IMPORTANT:
These instructions are to be placed in an easily seen location.
T6900 Rev A Operator Manual Listing of Warnings and Cautions
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
1-2
S40 Sterilant Concentrate contains 35% peracetic acid in the inner cup of the container. Eye contact with the sterilant
concentrate can cause irreversible eye damage, including blindness.
Contact with 35% peracetic acid can cause destruction of skin tissue. Avoid contact with the skin.
Peracetic acid is a flammable material. Do not smoke when handling, storing, transporting, or disposing of S40
Sterilant Concentrate.
To avoid accumulated concentration of peracetic acid vapor, use S40 Sterilant Concentrate in a well ventilated room
or area.
High concentrations of peracetic acid vapor will cause tearing of the eyes and irritate the nose, throat, and lungs.
Persons who are asthmatics may be more sensitive to the effects of inhaled peracetic acid vapors.
Although the use dilution of the S40 Sterilant Concentrate is not corrosive to skin or the eyes, it may cause eye
irritation.
Adequate cleaning of the medical device is required. Residual patient soil or other material or cleaning agents may
decrease effectiveness of S40 Sterilant Concentrate.
Do not use S40 Sterilant Concentrate past the expiration date printed on the carton.
Do not use partially filled, leaking, or damaged containers of S40 Sterilant Concentrate.
When disposing of S40 Sterilant Concentrate, thoroughly rinse the carton and container before attempting to
manually open the sterilant container.
Extreme care should be taken if an attempt is made to manually open a sealed container of S40 Sterilant Concentrate.
Use extreme care when cutting through the lid and the inner cup when disposing of a submerged container of S40
Sterilant Concentrate to avoid personal injury.
The SYSTEM 1 endo Processor must be installed by qualified personnel and according to correct national and local
electrical safety codes. Failure to do so can result in shock hazard that could endanger the health of the Operator.
The 0.2 micron MaxLife water filter must be changed every 180 days.
Use of a blocked or damaged aspirator may result in ineffective liquid chemical sterilization.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel.
Maintenance performed by unqualified personnel or installation of unauthorized parts could cause personal injury,
result in improper equipment performance, invalidate the warranty, or result in costly damage. Contact STERIS
regarding service options.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment. Contact
STERIS Customer Service to schedule preventive maintenance.
WARNINGS (continued):
Listing of Warnings and Cautions Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
1-3
CAUTIONS:
Do not push the container into the sterilant compartment with excessive force. NEVER slam the container into the
sterilant compartment. Damage to the container may occur.
Do not force lid to close.
If fluid remains in the chamber at the end of a cycle, call Customer Service.
Never use sterilant for the Diagnostic Cycle.
Water spills on the microcomputer housing and/or the barcode reader may result in damage to the Processor.
Do not touch the membrane of the MaxLife water filter while inserting into the filter housing.
Appropriate Personal Protective Equipment (PPE) is required when handling containers of S40 Sterilant Concentrate.
Minimally, PPE should consist of chemical-resistant gloves, apron, goggles or face shield, and any other protection
required by facility procedures.
If the carton of S40 Sterilant Concentrate is damaged, do not use the sterilant container inside the damaged carton.
T6900 Rev A Operator Manual Listing of Warnings and Cautions
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
1-4
Site Specifications Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
2-1
SITE SPECIFICATIONS 2
Water Requirements*:
Pressure: Minimum 40 psig (276 kPa), Preferred 50 - 60 psig (345 - 414 kPa), maximum
90 psig (620 kPa)
Flow Rate: Minimum 2.5 gpm (9.5 Lpm)
Temperature: 109 -140°F (43 - 60°C)
Recommended: For optimum cycle time, water temperature should be between 115 - 118°F
(46 - 48°C)
Supply Line Size: 1/2” (1.27 cm) I.D. minimum / 3/4” (1.9 cm) I.D. optimum
Connection: 3/4” (1.9 cm) male hose connector
Quality: Potable/Tap water, 140 ppm hardness as CaCO3
Drain: 1 1/4” (3.18 cm) I.D. (minimum) non-back pressuring drain
Water usage: Approximately 8.7 gal. (33 L) per processing cycle
Space and Weight Requirements:
SYSTEM 1 endo Processor
Width: 40” (102 cm.)
Depth: 24” (61 cm.)
Height: 38” (97 cm.) (includes overhead clearance)
Weight: 136 lbs. (62 kg.)
Operating Weight: 162 lbs. (74 kg.)
Electrical Requirements:
Specifications: 115 VAC, 60 Hz, 15 A
Service: 115 VAC, 20 A, 60 Hz dedicated circuit
Hospital grade GFCI duplex receptacle
WARNING: The SYSTEM 1 endo Processor must be installed by qualified personnel and according to
correct national and local electrical safety codes. Failure to do so can result in shock hazard that could
endanger the health of the Operator.
NOTE: There are no special requirements for venting the drain.
WARNING: Do not position equipment so that it is difficult to disconnect the plug from the duplex
receptacle.
WARNING: Connect the SYSTEM 1 endo Processor to hospital grade receptacle only following national and
local codes (GFCI, RCBO, etc.).
NOTE: Back flow (back siphonage) prevention provided by user (check local codes).
* Potable water should meet specifications identified in AAMI TIR 34: Water for the Reprocessing of Medical Devices.
Failure to meet these specifications may result in cancelled cycles and require the installation of additional water
treatment solutions by the health care facility.
T6900 Rev A Operator Manual Site Specifications
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
2-2
Pre-filter Specifications:
Pre-filter A: Filter Rating - 2.5 micron absolute
Flow Rate – 14.5 gpm/psid
Pre-filter B: Filter Rating - 0.1 micron nominal
Flow Rate - >1.0 gpm/psid
Water Pressure: Pressure Setting - regulator pre-set to 50 psig (345 kPa) static
WARNING: The performance of the SYSTEM 1 endo Processor has been validated and optimized with
components defined by STERIS in the Operator Manual for the SYSTEM 1 endo Processor. Use of pre-filters
that do not meet STERIS's specifications and have not been validated by STERIS can lead to premature
failure of the MaxLife water filter inside the Processor. STERIS’s warranty will not apply to damage or
early life failure of the MaxLife water filter resulting from the use of pre-filters that do not meet STERIS’s
specifications. DO NOT USE components that do not meet STERIS’s specifications and have not been
validated by STERIS.
WARNING: Do not use the SYSTEM 1 endo Processor until it has been properly installed and its electro/
mechanical performance is verified.
WARNING: To avoid accumulated concentration of peracetic acid vapor, use S40 Sterilant Concentrate
in a well ventilated room or area.
Environmental Requirements:
Room Temperature: 60 - 90°F (16 - 32°C)
Humidity: 10 - 90% relative, non-condensing
Ventilation: Recommend conformance to AAMI ST58: Chemical Sterilization and High-Level
Disinfection in Healthcare Facilities
Classification:
Protection against electric shock: Class 1
S40 Sterilant Concentrate Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
3-1
ASPIRATOR PROBE ASSEMBLY
(PROBE UNIT AND TUBING)
LID
INNER CUP
OUTER CUP
STERILANT COMPARTMENT
CUP PUNCH
DRAIN SCREEN
S40 STERILANT CONCENTRATE 3
S40 STERILANT CONCENTRATE
AND
STERILANT COMPARTMENT OF THE PROCESSOR
"POP-OUT" PLUG
T6900 Rev A Operator Manual S40 Sterilant Concentrate
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
3-2
S40 Sterilant Concentrate is a single-use chemistry labeled for use in the
SYSTEM 1 endo Processor. The sterilant concentrate container holds both the
liquid active ingredient (35% peracetic acid) and the powdered inert ingredients.
Review the information provided in this section of the Operator Manual and the
package insert for S40 Sterilant Concentrate prior to use. Contact STERIS with
any questions regarding the safe and proper use of S40 Sterilant Concentrate.
Always read the package insert for S40 Sterilant Concentrate for detailed
information, including Dangers and Precautionary statements, directions
for use, storage conditions and expiration date, emergency and safety
information, and sterilant container disposal instructions.
Part 1:
Overview
S40 Sterilant Concentrate Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
3-3
The 35% peracetic acid used in this product is a corrosive liquid due to its low
pH. The container is designed to minimize user contact with the contents. Care
should be taken to minimize exposure to vapors, contact with the skin, and
contact with the eyes. The following safety precautions should be taken when
handling the S40 Sterilant Concentrate:
The sterilant concentrate should be used in a well ventilated room.
Conformance to AAMI ST58 is recommended.
Chemical-resistant gloves should be worn to protect hands from potential
contact with the liquid portion of the sterilant concentrate.
It is recommended that appropriate eye protection be worn to protect the
eyes from potential contact in the event of a splash. Note: This should not
occur if the integrity of the sterilant container has not been compromised
and the Operator Manual instructions for handling and inserting the
container into the Processor are followed.
The peracetic acid concentrate solution is considered to have effects on the
eyes, skin, nose, throat and lungs based on acute exposure. These same sites
show signs of irritation upon repeated exposure.
Orally administered peracetic acid, in the concentrations found in the S40
Sterilant Concentrate, has been found to cause gastrointestinal tract irritation,
respiratory irritation and possible necrosis of renal tubules, in rats. Note: Orally
administered peracetic acid is classified as a GHS Category 3 material.
Inhalation of the sterilant concentrate has been shown to cause irritation of the
respiratory tract. Persons who are asthmatics may be more sensitive to the
effects of inhaled peracetic acid vapors. Note: Peracetic acid vapor is classified
as a GHS Category 1 material.
Application to the skin of a 17% peracetic acid solution is characterized by mild
skin irritation with reversible effects in dermal tissue and has caused severe
erythema, blanching of the skin, eschar formation, and exfoliation in rabbits. It
is severely irritating and corrosive to the skin. It will cause whiteness and can
cause blistering. Note: Application of a concentrated solution of peracetic acid
classifies it as a GHS Category 3 material.
Application of a 17% peracetic acid solution is severely irritating and corrosive
to the eye of rabbits. Note: It is classified as a Corrosive material.
The SDS for S40 Sterilant Concentrate is available upon request from STERIS
or at www.steris.com.
WARNING: High
concentrations of peracetic
acid vapor will cause tearing
of the eyes and irritate the
nose, throat, and lungs.
WARNING: Peracetic acid
is a flammable material. Do
not smoke when handling,
storing, transporting, or
disposing of S40 Sterilant
Concentrate.
WARNING: To avoid
accumulated concentration
of peracetic acid vapor, use
S40 Sterilant Concentrate
in a well ventilated room or
area.
WARNING: Persons who are
asthmatics may be more
sensitive to the effects
of inhaled peracetic acid
vapors.
WARNING: S40 Sterilant
Concentrate contains 35%
peracetic acid in the inner
cup of the container. Eye
contact with the sterilant
concentrate can cause
irreversible eye damage,
including blindness.
WARNING: Contact with
or inhalation of the inert
powdered ingredients in the
outer cup of S40 Sterilant
Concentrate may cause
irritation to eye, skin or
respiratory system.
WARNING: Contact with
35% peracetic acid can
cause destruction of skin
tissue. Avoid contact with
the skin.
Part 2:
Safety Information for
Peracetic Acid
T6900 Rev A Operator Manual S40 Sterilant Concentrate
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
3-4
Part 3:
Disposal
CAUTION: Appropriate
Personal Protective
Equipment (PPE) is
required when handling
containers of S40 Sterilant
Concentrate. Minimally,
PPE should consist of
chemical-resistant gloves,
apron, goggles or face
shield, and any other
protection required by
facility procedures.
WARNING: When disposing
of S40 Sterilant Concentrate,
thoroughly rinse the carton
and container before
attempting to manually
open the sterilant container.
WARNING: Extreme care
should be taken if an attempt
is made to manually open a
sealed container of S40
Sterilant Concentrate.
WARNING: Use extreme
care when cutting through
the lid and the inner cup
when disposing of a
submerged container of
S40 Sterilant Concentrate
to avoid personal injury.
WARNING: Do not use
partially filled, leaking, or
damaged containers of S40
Sterilant Concentrate.
CAUTION: If the carton of
S40 Sterilant Concentrate
is damaged, do not use the
sterilant container inside
the damaged carton.
Following a successful LCS Cycle, dispose of empty sterilant containers in the
trash (no special precautions are required).
To dispose of partially filled, leaking, damaged, or expired sterilant containers,
put on appropriate personal protective equipment (chemical-resistant gloves,
apron, goggles or face shield, and any other protection required by facility
procedures). Wear protective attire for the entire procedure. Perform one of the
following three disposal procedures or follow facility guidelines for disposal of
hazardous waste. To contain the powders and/or liquid until time of disposal,
place the partially filled, leaking, damaged, or expired sterilant container in a bag,
box or other suitable receptacle. Use either the automated or manual disposal
method as described below. Dispose of each sterilant container one at a time.
1. Automated Disposal (preferred method)
IMPORTANT: Automated disposal method should not be used if there are no
dry powders in the container.
a. Place the partially filled, leaking, damaged, or expired sterilant container
in the sterilant compartment in the SYSTEM 1 endo Processor.
b. Do not place a medical device in the Processor.
c. Press down until the sterilant container is firmly seated.
d. Line up the tip of the aspirator probe over the center of the sterilant
container lid. Insert the probe in a downward motion until the top of the
aspirator is resting on the lid of the container. Make sure the tubing is
not kinked.
e. Start the LCS Cycle according to the Operator Manual.
f. Dispose of the empty sterilant container in the trash.
If the SYSTEM 1 endo Processor is unavailable, isolate the partially filled,
leaking, damaged, or expired sterilant container until the Processor is available
or perform a manual disposal as described below.
2. Manual Disposal - Check facility procedures, local and state regulations
for waste disposal requirements.
a. Increase room ventilation to remove any strong, tear-producing
vapors.
b. Carefully handle the partially filled, leaking, damaged, or expired sterilant
container and place upside down on a flat surface or in a sink. Note: If
the sterilant container is inside a carton, place the entire carton under
running water and remove the sterilant container from the carton. Treat
the sterilant container as instructed above. Rinse the cardboard carton
thoroughly before disposing of it in the trash.
c. In a deep sink, place a bucket filled with at least 2 gallons (8 L) of tap
water.
d. With the sterilant container placed upside down on a flat surface, grasp
S40 Sterilant Concentrate Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
3-5
the sterilant container firmly and, using a blunt object (brush handle,
marking pen, hemostats, etc.), push the sterilant container’s "pop out"
plug into the sterilant container. (See Figure 1)
IMPORTANT: A small puff of powder may escape from bottom of the sterilant
container.
Figure 1
Figure 2
e. Gently shake the contents (dry powders) into the bucket of water. Avoid
inhaling or contacting contents.
f. Using a long object suitable for stirring liquids (a metal/glass/plastic
rod) stir until the contents are dissolved.
g. Submerge the sterilant container into the bucket containing the dissolved
powders.
h. Using a pair of scissors, carefully cut through the lid at the center and
then into the inner cup. Allow the acid to mix with the water and dissolved
powders in the bucket. (See Figure 2)
i. Avoid standing directly over the bucket or irritation may occur
from the peracetic acid vapors. Avoid inhaling or contacting
contents.
j. With the cold water running, carefully pour the solution down the sink
drain.
k. Flush the bucket and sink with a large quantity of cold tap water.
l. Thoroughly rinse the sterilant container and inner cup and then dispose
of them in the trash.
3. Disposal When Water is Not Available
a. Increase ventilation to the area - open truck or warehouse doors, use
fans to remove any strong tear-producing vapors.
b. Shut off ignition sources and DO NOT SMOKE.
c. Seal boxes with sterilant containers into plastic bag(s) and take to a
water source.
d. When water is available, follow the procedure for manual disposal.
CAUTION: Appropriate
Personal Protective
Equipment (PPE) is
required when handling
containers of S40 Sterilant
Concentrate. Minimally,
PPE should consist of
chemical-resistant gloves,
apron, goggles or face
shield, and any other
protection required by
facility procedures.
T6900 Rev A Operator Manual S40 Sterilant Concentrate
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
3-6
Component Identification Operator Manual T6900 Rev A
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
4-1
PRE-FILTER
ASSEMBLY (A1563)
REGULATOR
HOUSING BODY
PRE-FILTER B
PRE-FILTER A
COMPONENT IDENTIFICATION 4
BARCODE READER
AND BRACKET
WINDOW
HMI CONTROL DISPLAY
(TOUCHSCREEN)
RELEASE LATCH/HANDLE
PROCESSOR CHAMBER
(INSIDE)
SYSTEM 1 endo PROCESSOR
(FRONT VIEW)
Part 1:
Illustrations of
Processor and
Components
(COVER NOT SHOWN)
USB PORT
T6900 Rev A Operator Manual Component Identification
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
4-2
INFLATABLE SEAL
MICROCOMPUTER
HOUSING
HINGED PANEL (SWING ARM)
INTERNAL PROCESSOR COMPONENTS
STERILE AIR FILTER
SEAL
MAXLIFE
WATER FILTER/HOUSING
CONNECTOR
POSTS
CHAMBER ETHERNET PORT
(STERIS-TRAINED TECHNICIAN
USE ONLY)
/