Operation Manual MA 27 / MA 27e
6.3 Screening
A hearing screening utilizes a “pass” or “refer” result and is used
to determine if further testing is required as a hearing problem
may exist. Patients are typically screened at a level of 20 dB HL
at 500, 1000, 2000, and 4000 HZ in each ear. If a patient hears
all the tones in each ear, the result would be considered a “pass.”
Failure to hear any of the tones in either ear would result in a
“refer”. This is an example of one screening protocol. Each
state may have their own screening protocol. Please contact
your state health department for guidelines in your area.
NOTE: Background noise can produce false test results,
especially at lower frequencies. It is important to test in a quiet
environment.
For hygienic reasons, it is important to clean the headphone ear
cushions after testing.
6.4 Auto Threshold (Hughson-Westlake)
In addition to traditional manual testing, the MA 27e incorporates
a Hughson-Westlake patient controlled automatic threshold test
complying with ISO 8253. When the test is completed the results
are easily recalled from the internal memory of the MA 27e.
Hughson-Westlake is a procedure used to determine pure tone
thresholds. The MA 27e utilizes this procedure to perform an
automatic pure tone test procedure. Threshold is defined as 2
out of 3 (or 3 out of 5) correct responses obtained at a certain
level in a 10 dB decrease and 5 dB increase procedure.
The patient should sit at a distance of at least 1 m from the
device.