MENU/DEFINITIONS StepFour
FAST/SLOW:SettingforechoaveragingaswellasthereactiontimeforSAF_settingintheLVU‐201A:
FAST:Is the typical and default setting for the
LVU‐201A to operate.The LVU‐201A will
average 2 signal responses per second and
updateeverysecond.In FASTmode, the sensor
willcontinue to search fora valid readingfor 30
secondsbeforegoingintotheLOSTstate.
SLOW:Is designed to help dampen out effects
causedbysevereturbulence.TheLVU‐201Awill
average signal returns over the preceding 10
seconds.In SLOW mode, the sensor will
continue to search for a valid reading for 2.5
minutesbeforegoingintotheLOSTstate.
TochangetheFAST/SLOWsetting,followthestepsbelow:
1. Hold[MENU]keyuntilFASTorSLOWappearsinthedisplay.
2. Release[MENU]keyandhold[SET]keytotogglebetweenFASTandSLOW.
3. Whendesiredsettingisreached,release[SET]key.Thelastdisplaysettingwillbelockedintomemory.
ALIN:Usethismodeasanaidtolevelingthesensorinstalledinaself‐aligningorrotationalbulkheadfitting.
DisplaywillshowthereturnsignalstrengthindB’s.Foroptimumalignment,firstenergizetheunitandreceive
a valid return signal.Then select the ALIN mode and adjust the LVU‐201A until the display is maximized.
Typicalreadingsrangebetween2and60dB’s.
ON/OFF:ThisistheactualactivationsettingforALINmode.Defaultis[OFF].Setto[ON]toviewtheALIN
readings.TheALINmodemustbeturned[OFF]whenalignmentiscompleted.
ToALINasensor,followthestepsbelow:
1. Hold[MENU]keyuntilOFFappearsinthedisplay.
a. OFFwillappearrightafterALIN
2. Release[MENU]keyandimmediatelyhold[SET]keytotogglefromOFFtoON.
3. Release[SET]key.TheLVU‐201AisnowinALINmode.
4. Adjustthesensorsothedisplayshowsthelargestvalue.
a. Whenthehighestvalueisshown,lockthesensorinplace.
5. ToexitALINmode,repeatsteps1‐4changingfromONtoOFF.
MAXR:The MAXR se
ts the maximum tank height and will filter out all returns greater that this value.
Regardless of how the EC4 and EC20 are set, set MAXR to the Empty tank
distanceorgreater.
(VALUE):The number after MAXR shows the setting for MAXR.This value is
the actual MAXR setting that can be changed for the application.The
maximumdistanceis216.0inches.
MINR:TheMINRvalueisthedeadbandclosesttothetransducerfacewhere
nosignalwillbegenerated.TheMINRsetstheminimumdistancebetweenthe
liquid and the transducer.Regardless of how the EC4 and EC20 are set, set
MINRtotheFULLtankdis
tance.
(Value):ThenumberafterMINRshowsthesettingforMINR.Thisvalueisthe
actualMINRsettingthatcanbechangedfortheapplication.Theminimumdistanceis6.0inches.