ABB SM500F User manual

Type
User manual
ABB MEASUREMENT & ANALYTICS | INSTRUCTION | IN/RANDC/004-EN REV. A
SM500F
Field mountable paperless recorder
Color processor board replacement
This instruction is applicable only
to instruments manufactured after
October 2016
Measurement made easy
1 Introduction
This publication details replacing an SM500F
paperless recorder’s color processor board. The
procedure must be carried out by a trained
technician. The contents of the replacement kit are
shown in Table 1.
Description Part number Qty
Processor board assembly SM500/0215 1
Battery B12454 1
Screw B12406 6
Windows licence B12423 1
This publication IN/RandC/004-EN 1
Table 1 Color processor board assembly replacement
kit SM500/0736
Tools required
No. 2 Pozidriv screwdriver
T8 Torx driver
2 For more information
Further information is available from:
www.abb.com/recorders
or by scanning these codes:
Sales Service
SM500F
field mountable
paperless recorder
2 IN/RandC/004-EN Rev. A | SM500F | Field mountable paperless recorder
3 Processor board replacement
1. Disconnect the recorder from the power supply.
2. Referring to Fig. 3.1, unlock the door with the key A,
press release catch B and open the door.
3. Referring to Fig. 3.2, remove 6 Torx screws C securing
the door assembly rear cover and remove the rear cover.
Referring to Fig. 3.3:
4. Disconnect processor board flexi-cable D, display
flexi-cable E and keypad membrane cable F from the
processor board and lift out and discard the processor
board.
5. Position the new processor board in the door housing and
reconnect keypad membrane cable F, display flexi-cable
E and processor board flexi-cable D.
6. Referring to Fig. 3.2, refit the door assembly rear cover and
secure using 6 Torx screws C.
7. Close and lock the door.
8. Restore the recorder's power supply.
NOTICE – Property damage
The recorder is vulnerable to electrostatic damage.
Wear an anti-static strap or place the instrument on
an anti-static workbench.
Fig. 3.1 Opening the door
Fig. 3.2 Removing the rear cover
A
B
C
Fig. 3.3 Disconnecting the processor board
D
E
F
SM500F | Field mountable paperless recorder | IN/RandC/004-EN Rev. A 3
4 Entering the recorder's serial number
1. Locate the serial number label on the outside of the
recorder and note the serial number.
2. Access the configuration level as described in Section 6 of
the User Guide (IM/SM500F).
3. Access the 'Common' configuration level and select the
Setup tab. Use the and keys to select the 'Options
enabled' parameter and press .
4. Select any option from the list box and press .
5. Use the and keys to select the button next to
the 'Serial number' field. Press the key and using the
displayed keypad, enter the recorder's serial number noted
at step 1.
6. Repeat steps 4 and 5 to enter software option security
codes in the security code field for each option as required.
7. Exit the configuration level as described in Section 6 of the
User Guide (IM/SM500F), saving the configuration as the
'Current Configuration'.
8. Reset archiving as described in Section 5.2 of the User
Guide (IM/SM500F).
IMPORTANT (NOTE) The format of the serial
number is either */*****/*/* or 3K21**********.
IN/RandC/004-EN Rev. A 12.2018
We reserve the right to make technical changes or modify the contents of this document
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in
whole or in parts – is forbidden without prior written consent of ABB.
© ABB 2018
ABB Limited
Measurement & Analytics
Howard Road, St. Neots
Cambridgeshire, PE19 8EU
UK
Tel: +44 (0)0870 600 6122
Fax: +44 (0)1480 213 339
Email: instrumentat[email protected]
ABB Inc.
Measurement & Analytics
125 E. County Line Road
Warminster
PA 18974
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183
abb.com/measurement