Buhler PKE 42 Installation And Operation Instruction Manual

Type
Installation And Operation Instruction Manual
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
2
BX440009, 10/2008 Art. Nr. 90 31 100
Inhaltsverzeichnis Seite
1 Einleitung........................................................................................................................................ 3
2 Wichtige Hinweise ......................................................................................................................... 3
2.1 Allgemeine Gefahrenhinweise............................................................................................... 4
3 Aufbauen und Anschließen .......................................................................................................... 5
3.1 Montage................................................................................................................................. 5
3.2 Elektrische Anschlüsse..........................................................................................................5
4 Betrieb und Wartung ..................................................................................................................... 6
4.1 Warnhinweise........................................................................................................................ 6
4.2 Betrieb ................................................................................................................................... 7
4.3 Wartung................................................................................................................................. 7
5 Instandsetzung, Entsorgung ........................................................................................................ 7
5.1 Fehlerbehebung..................................................................................................................... 7
5.2 Entsorgen .............................................................................................................................. 7
6 Anhang............................................................................................................................................ 8
6.1 Fehlersuche und Beseitigung................................................................................................ 8
6.2 Auswechseln der Feinsicherung ........................................................................................... 8
6.3 Austausch des Schlauches der peristaltischen Pumpe (wenn Pumpe vorhanden).............. 8
6.4 Demontage und Reinigung des Wärmeaustauschers........................................................... 9
6.5 Ersatzteile und Zusatzteile ....................................................................................................9
6.6 Beiliegende Unterlagen......................................................................................................... 9
Contents Page
1 Introduction.................................................................................................................................. 10
2 Important Advice.......................................................................................................................... 10
2.1 General indication of risk..................................................................................................... 11
3 Installation and Connection........................................................................................................ 12
3.1 Mounting.............................................................................................................................. 12
3.2 Electrical connection............................................................................................................ 12
4 Operation and Maintenance........................................................................................................ 13
4.1 Indication of risk................................................................................................................... 13
4.2 Operation............................................................................................................................. 14
4.3 Maintenance........................................................................................................................ 14
5 Repair and Disposal .................................................................................................................... 14
5.1 Repair ..................................................................................................................................14
5.2 Disposal............................................................................................................................... 14
6 Appendices................................................................................................................................... 15
6.1 Trouble shooting.................................................................................................................. 15
6.2 Replacing the electrical fuse................................................................................................ 15
6.3 Replacement of the peristaltic pump’s hose (just in case pump is fitted) ........................... 15
6.4 Cleaning of the heat exchanger........................................................................................... 16
6.5 Spare parts.......................................................................................................................... 16
6.6 Attached documents............................................................................................................ 16
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
1 Introduction
The sample gas coolers of model series PKE42 are designed for installation in gas analysis systems. Check
technical data according to the datasheets attached with regard to the specific application, used material com-
binations, as well as pressure- and temperature limits.
2 Important Advice
Operation of the device is only valid if
the product is used under the conditions described in the installation- and operation instruction, the inten-
ded application according to the type plate and the intended use,
the performance limits given in the datasheets and in the installation- and operation instruction are obey-
ed,
monitoring devices and safety devices are installed properly,
service and repair is carried out by Bühler Technologies GmbH,
only original spare parts are used.
This manual is part of the equipment. The manufacturer keeps the right to modify specifications without ad-
vanced notice. Keep this manual for later use.
Definitions for warnings:
NOTE
Signal word for important information to the product.
CAUTION
Signal word for a hazardous situation with low risk, resulting in damaged to the device or
the property or minor or medium injuries if not avoided.
WARNING
Signal word for a hazardous situation with medium risk, possibly resulting in severe injuries
or death if not avoided.
DANGER
Signal word for an imminent danger with high risk, resulting in severe injuries or death if
not avoided.
Warning against hazardous
situation
Warning against possible
explosive atmospheres
disconnect from
mains
Warning against electrical
voltage
Warning against hot surface
wear respirator
Warning against respiration
of toxic gases
wear face pro-
tection
Warning against acid and
corrosive substances
wear gloves
BX440009, 10/2008 Art. Nr. 90 31 100
11
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
2.1 General indication of risk
Check all relevant safety regulations and technical indications for the specific installation place. Prevent fail-
ures and protect persons against injuries and the device against damage.
The person responsible for the system must secure that:
safety and operation instructions are accessible and followed,
local safety regulations and standards are obeyed,
performance data and installation specifications are regarded,
safety devices are installed and recommended maintenance is performed,
national regulations for disposal of electrical equipment are obeyed.
Maintenance and repair
Repairs on the device must be carried out by Bühler authorized persons only.
Only perform modifications, maintenance or mounting described in this manual.
Only use original spare parts.
During maintenance regard all safety regulations and internal operation instructions.
DANGER
Electrical voltage
Electrocution hazard.
Disconnect the device from power supply. Make sure that the equipment
cannot be reconnected to mains unintentionally.
The device must be opened by trained staff only.
DANGER
Toxic and corrosive gases
Sample gas can be hazardous.
Take care that the gas is exhausted in a place where no persons are in danger.
Protect yourself during maintenance against toxic / corrosive gases. Use
gloves, respirator and face protector under certain circumstances.
DANGER
Explosion hazard if used in hazardous areas
The device is
not suitable for operation in hazardous areas with potentially
explosive atmospheres.
Do not expose the device to combustible or explosive gas mixtures.
12
BX440009, 10/2008 Art. Nr. 90 31 100
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
3 Installation and Connection
The PKE-42 sample cooler is to be attached to vertical panels. The heat exchangers either made from
stainless steel, glass or PVDF are inserted from the top. Make sure, that there is enough space left above and
under the cooler to get the pipes/or hoses connected to the unit.
The place of installation must be weather shielded and air shall circulate freely around the cooler. The ambient
temperature shall not exceed the range from +5 to +50°C (+41 to +122°F) for the PKE-42.
Free air circulation must be provided. On both sides a gap of at least 10 cm (4 inches) must be kept clear.
Free air circulation must also be provided if the cooler is installed inside a cabinet. In some cases a fan is nec-
essary to establish sufficient circulation within the cabinet.
3.1 Mounting
Depending on the application parameters the connecting pipes or hoses must be of adequate material and
fastened tight. Connect the hoses to the heat exchangers made of duran glass with care to avoid breaking the
glass.
Make sure that all sample gas lines leading to the cooler are installed with downward slope to enable conden-
sate flow into the heat exchanger by gravity.
In some applications with very high condensate content separators upstream the cooler could become neces-
sary (see catalogue for appropriate types).
The gas entrance is marked with red.
If the sample gas pump is located upstream of the cooler, the condensate can be drained off by automatic
condensate drains. If the pump is located downstream peristaltic pumps or condensate vessels must be used
for removal (see our catalogue for appropriate equipment).
Important: The PVDF heat exchanger type DTV cannot be used with automatic drainers but with peristaltic
pumps only.
The condensate drains can be attached directly to such coolers with stainless steel heat exchangers. In case
of glass heat exchangers the condensate drains must be connected with flexible lines fixed with by brackets
separately.
The condensate lines must be installed with considerable slope and should not have less than 8 mm (0,3 inch)
inner diameter.
3.2 Electrical connection
WARNING
The device must be installed by trained staff only.
CAUTION
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
BX440009, 10/2008 Art. Nr. 90 31 100
13
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
WARNING
Damage to the device in case of insulation testing
Do not proceed insulation tests with high voltage to the device as a whole.
The device is equipped with extensive EMC protection. If insulation tests are carried out the electronic filter
devices will be damaged. All necessary tests have been carried out for all concerned groups of components at
the factory (test voltage 1 kV or 1.5 kV respectively, depending on the device).
The device is equipped with connectors according to DIN 43650. After correct wiring the connectors cannot be
interchanged and must be wired according to the following diagram (numbering can be found on the connec-
tor). Make sure the connector is correctly refitted after wiring.
Mains voltage is 230 VAC 50/60 Hz or 115 VAC 50/60 Hz (regard type plate).
If the cooler is equipped with peristaltic pumps, they must be connected to a separate power supply.
pin numbering
1
1
1
L1
N
2
2
2
3
3
3
PE
PE
PE
function ok
alarm
power supply
status output
4 Operation and Maintenance
4.1 Indication of risk
All repairs must be carried out by Bühler authorised personnel only.
Only perform modifications, servicing or mounting described in this manual.
Only use original spare parts.
Regard all relevant safety regulations and internal operating instructions during maintenance.
DANGER
Electrical voltage
Electrocution hazard.
Before any manipulation on the device, disconnect the electrical equipment
from mains power supply. Make sure that the electrical equipment cannot be
reconnected during repair or maintenance.
The wiring must be done by trained staff only. Regard the correct mains
voltage.
14
BX440009, 10/2008 Art. Nr. 90 31 100
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
DANGER
Toxic, corrosive gases
Sample gas may be harmful.
Please exhaust sample gas to a safe place.
Protect yourself against toxic / corrosive gas during maintenance. Use gloves,
respirator and face protector under certain circumstances.
4.2 Operation
After turning on the power supply the display will show the present temperature of the cooling block. The
status-LED is blinking until the unit has reached the operational temperature within the ±3K (± 5°F) range.
Then the LED will turn of and the output will switch to o.k.
If the status-LED starts blinking during the operation see “6.1 Troubleshooting“. For technical data see data
sheet.
4.3 Maintenance
The unit runs free of maintenance.
If the cooler has already mounted peristaltic pumps (optional), their tubes have to be checked regularly. Re-
placement is described in chapter 6.3.
5 Repair and Disposal
5.1 Repair
If the device shows irregularities see chapter 6 for troubleshooting
If you need help or more information
call +49(0)2102-498955 or your local agent.
If the device doesn’t work correctly after elimination of failures and turning power on, the device must be
checked by the manufacturer. Please ship the device with suitable packing to
Bühler Technologies GmbH
- Service -
Harkortstraße 29
40880 Ratingen
Germany
5.2 Disposal
The device contains no coolants. Regard the local regulations for disposal of electronic equipment.
BX440009, 10/2008 Art. Nr. 90 31 100
15
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
16
BX440009, 10/2008 Art. Nr. 90 31 100
6 Appendices
6.1 Trouble shooting
Problem / Failure Possible cause Solution
no power check power supply
No display
fuse blown Check fuse and change it if nec-
essary
Cooler not running
high temperature at the compressor
casing
wait until cooled off and care for
enough ventilation
Status-LED blinks with
Cooling capacity to low, even
though cooler is running
Make sure that air can circulate
free and that ventilation louvres
are not obstructed
gas flow / dew point / gas tempe-
rature to high
check application parameters,
install pre-separator
High Temperature
fan broken check fan, replace if necessary
Low temperature controller defect send cooler for inspection
condensate vessel full drain vessel
stuck valve in automatic conden-
sate drain
flush both directions
Condensate in gas outlet
cooler overloaded check limiting parameters
clogged gas path check / flush heat exchanger
Reduced gas flow
condensate outlet clogged by ice send cooler for inspection
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
6.2 Safety Advice
DANGER
Electrical voltage
Electrocution hazard.
Disconnect the device from power supply. Make sure that the equipment
cannot be reconnected to mains unintentionally.
The device must be opened by trained staff only.
DANGER
Toxic and corrosive gases / condensate
Sample gas / condensate can be hazardous.
Shut off gas supply prior to maintenance or repairs.
Protect yourself during maintenance against toxic / corrosive gases /
condensate. Use gloves, respirator and face protector under certain
circumstances.
6.3 Replacing the Fuse of the Cooler
¾ Disconnect cooler from power supply
¾ Remove the black cap on the front side of the cooler holding the fuse. Use a screwdriver big enough
for pressing the cap and turn it counter-clockwise about 90 degrees.
¾ Replace the fuse recommended for the voltage of the cooler (See
6.7).
¾ Refit the cap by pressing it and turning it clockwise.
¾ Reconnect power supply.
6.4 Replacing the Fuse of the peristaltic Pump
¾ Disconnect the pump from the mains.
¾ Remove the insulation cover from the fuse holder at the pump’s mounting bracket. For this, push the
cover using a screw driver and quarter-turn it to the left.
¾ Replace the fuse and refit the insulation cover by quarter-turning it to the right.
¾ Reconnect the cooler to the mains.
6.5 Replacing the Hoses of the peristaltic Pump
¾ Shut off the gas supply and drain residual condensate from the hoses.
¾ Disconnect the pump from the mains.
¾ Remove the supplying and draining hoses from the pump.
¾ Loosen but don’t remove the centre knurled thumb screw. Push screw downwards.
¾ Remove cover
¾ Push terminals sidewards.
¾ Replace the hose and remount the pump in reverse order.
BX440009, 10/2008 Art. Nr. 90 31 100
17
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Meßgaskühler / Sample Gas Coolers PKE-42
18
BX440009, 10/2008 Art. Nr. 90 31 100
6.6 Cleaning of the heat exchanger
The heat exchanger must be replaced only in case of damage or when clogged. In the latter case we recom-
mend to check if the use of a filter will avoid such clogging in the future.
¾ Disconnect the cooler from the mains and shut off gas supply.
¾ Loosen gas fittings and condensate drain fitting. Collect condensate.
¾ Pull out the heat exchanger upwards out of the cooler.
¾ Clean the cooling nest (hole in cooling block).
¾ Clean the heat exchanger until all impurities are disposed.
¾ Grease the heat exchanger on the cooling surfaces with silicone grease.
¾ Introduce the heat exchanger with turning movement into cooling nest.
¾ Re-establish fittings for gas tubes and condensate drain.
¾ Reconnect the cooler to the mains.
6.7 Spare parts
Please indicate with spare part requirements type of cooler and serial no.
For accessories and enhancement see data sheets and/or catalogue.
The following parts are recommended for stocking:
Spare part Part no.
Replacement hose for peristaltic pump 0.3 l/h
(only if your cooler is equipped with pump)
91 24 03 00 27
230V 44 10 030 Fan
115V 44 00 03 0
230V 91 00 01 00 81 Electronic board
115V 91 00 01 00 99
Temperature display 44 00 003
230 V 5x20 mm, 1 A slow blow 91 100 00 011 Fuse cooler
115 V 5x20 mm, 4 A slow blow 91 100 00 051
Fuse peristaltic pump 230 V / 115 V 5x20 mm, 1 A fast-acting 91 100 00 061
6.8 Attached documents
Data sheet DE 44 0011
Declaration of conformity KX 44 0001
Sample Gas Cooler
PKE 42
DE 44 0011
09/2009
Page 1/4
The PKE 42 cooling system consists of semiconductor Peltier
§ Quick installation
cooling elements with an aluminum cooling block. Fitted into
the block is a removable, high-efficiency heat exchanger
made of stainless steel, DURAN-glass or PVDF with 1 or 2
§ No maintenance required
gas paths.
§ Low noise
The unit maintains a constant outlet dew point of 40 °F with
an electronic controller. The temperature of the cooling block
§ Efficient heat exchangers made of
is shown on an LED-display. The status is indicated by a
flashing LED which shows high or low temperature alarms
stainless steel, Duran-glass or PVDF
and operates together with relay to halt the flow of sample gas
in fail-safe mode.
§ Nominal cooling capacity 140 kJ/h
The relay maybe used to control the sample gas pump when
the cooler reaches the desired temperature range.
§ Dew point stability 0.1°C
Condensate is removed either by peristaltic pumps, by
§ Status display and output
automatic condensate drains, or condensate vessels.
§ Cooling temperature display
§ Heat exchanger with 1 or 2 gas paths
§ Compact design
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 49 89-0 Fax: + 49 (0) 2102 / 49 89-20
Internet: www.buehler-technologies.com
10
9
8
1
2
3
4
5
6
7
A000037X
ALARM-
AUSGANG
ALARM-
OUTPUT
NETZ
POWER
A000130D
320
337
148
017
91
2
,
5
2 ,
5
1 0
1
9
2
ø7
Alarm
C
05°
Technical Data
Ready for operation after max. 10 minutes
Cooling capacity (at 25°C) 140 kJ/h
Ambient temperature +5 °C to 50 °C
Factory set dew point 5 °C
Dew point noise static 0.1 K
Drift over full specified range ± 1.5 K
Power supply 115 V or 230 V ,50/60 Hz, plug acc. to DIN 43650
Power consumption max. 350 VA
Status output Electrical spec. max. 230V AC, 150 VDC
2A, 50 VA
free of potential
Plug acc. to DIN 43650
Protection class IP 20
Housing stainless steel
Weight approx. 11 kg
Typical Installation Diagram:
Dimensions (mm)
1 Sample probe
2 Sample tube
3 3 way valve
4 Sample gas pump
5 Sample gas coole EGK-1/2
6 Automatic condensate drain or perist. pump
7 Moisture detector
8 Fine filter
9 Flow meter
10 Analyser
For models and specs of components see individual data
sheets.
calibration gas
air
out
condensate
out
sample gas
in / out
air
in
DE 44 0011
09/2009
Page 2/4
0
20
40
60
80
100
120
140
160
kJ/h
5 10 15 20 25 30 35 40
°C
A000131D
180
200
45 50
ambient temperature
coolong capacity
PKE 42
with DTS
Limits for
TV-SS
3)
Heat Exchanger TS TG TV-SS DTS DTG DTV
1)
Flow rate v 500 l/h 400 l/h 235 l/h 2 x 250 l/h 2 x 200 l/h 2 x 160 l/hI
max
1)
Inlet dewpoint t 80 °C 80 °C 65 °C 80 °C 65 °C 65 °C
e,max
1)
Gas inlet temperature. J 180 °C 140 °C 140 °C 180 °C 140 °C 140°C
G,max
Max. cooling capacity Q 450 kJ/h 230 kJ/h 120 kJ/h 450 kJ/h 230 kJ/h 185 kJ/h
max
Gas pressure p 160 bar 3 bar 3 bar 25 bar 3 bar 2 bar
max
Pressure drop Dp (v=150 l/h) 8 mbar 8 mbar 8 mbar each 5 mbar each 5 mbar each 15 mbar
Dead volume V 69 ml 48 ml 129 ml 28 / 25 ml 28 / 25 ml 21 / 21 ml
tot
2)
Sample gas connections G 1/4“ i GL 14 DN 4/6 tube 6 mm GL 14 DN 4/6
2)
Condensate out connections G 3/8“ i GL 25 G 3/8" i tube 10 mm GL 18 DN 4/6
1)
max. cooling capacity of the cooler must be considered
2)
NPT-threads upon request
3)
Con only be used with peristaltic pumps
Heat Exchanger
The energy content of the sample gas and, as a result, the required cooling capacity of the gas cooler is determined by 3
parameters: gas temperature J , dewpoint t (moisture content) and flow v. The outlet dew point rises with increasing energy
G e
content (heat) of the gas. The required cooling capacity is determined by the maximum acceptable level of the outlet dew point.
The following table shows cooler performance assuming the following conditions: t =50°C and J =70°C. Indicated is the v in
e G max
Nl/h cooled air (i.e. after the moisture has condensated). If the actual values stay below the parameters t and J , v can be
e G max
increased. For example (TG), instead of t = 50°C, J = 70°C and v = 345 l/h the values t =40°C, J =70°C and a maximum flow
e G e G
rate of v=425 l/h could be achieved.
Please contact one of Buhlers application specialists for assistance and further information.
Performance Data
DE 44 0011
09/2009
Page 3/4
4 4 7 0 0 0
0
1
2
0 0 0
1 1 0
1 2 0
1 3 0
2 6 0
2 7 0
2 8 0
0
1
PKE 42
Power Supply
115 V
230 V
Gas Path/ Material/ Version
without heat exchanger
single path heat exchanger TS, stainless steel
single path heat exchanger TG, glass
single path heat exchanger TV SS, PVDF
dual path heat exchanger DTS, stainless steel
dual path heat exchanger DTG, glass
1)
dual path heat exchanger DTV , PVDF
1)
Condesate Discharge
without condensate discharge
3)
peristaltic pump(s)
Automatic condensate drain AK 5.2
Condensate vessel GL1, 0,4l
Peristaltic pump 0.3 l/h, 230V, for separate mounting
Peristaltic pump 0.3 l/h, 115V, for separate mounting
1)
Condensate outlet only suitable for connecting peristaltic pumps
2)
Peristaltic pumps for separate mounting available
3)
Each gas path is equipped with a peristaltic pump with the same mains supply requirements as the cooler.
Please indicate with order
Please extract the part number for the cooler fulfilling your requirements from the type code below.
Please note: Each gas path should be equipped with a peristaltic pump or an automatic condensate drain.
Part No.
DE 44 0011
09/2009
Page 4/4
We reserve the right to amend specification.
Accessories
451 00 08
441 00 05
912 40 30 104
912 40 30 105
Sample Gas Cooler
PKE 42
The PKE 42 cooling system consists of semiconductor Peltier
cooling elements with an aluminum cooling block. Fitted into
the block is a removable, high-efficiency heat exchanger
made of stainless steel, DURAN-glass or PVDF with 1 or 2
gas paths.
The unit maintains a constant outlet dew point of 40 °F with
an electronic controller. The temperature of the cooling block
is shown on an LED-display. The status is indicated by a
flashing LED which shows high or low temperature alarms
and operates together with relay to halt the flow of sample gas
in fail-safe mode.
The relay maybe used to control the sample gas pump when
the cooler reaches the desired temperature range.
Condensate is removed either by peristaltic pumps, by
automatic condensate drains, or condensate vessels.
§ Compact design
§ Quick installation
§ No maintenance required
§ Low noise
§ Efficient heat exchangers made of
stainless steel, Duran-glass or PVDF
§ Nominal cooling capacity 133 Btu/hr
§ Dew point stability 0.2°F
§ Status display and output
§ Cooling temperature display
§ Heat exchanger with 1 or 2 gas paths
DA 440011
09/2009
Page 1/4
Buhler Technologies LLC
1030 West Hamlin Road, Rochester Hills, MI 48309
Phone: 248.652.1546 Fax: 248.652.1598
Internet: www.buhlertech.com
10
9
8
1
2
3
4
5
6
7
A000037X
ALARM-
AUSGANG
ALARM-
OUTPUT
NETZ
POWER
A000130E
12.6”
13.3”
5.8”
6
7”.
7
6
.
7
4.
7 6
.
ø0.3”
Technical Data
Ready for operation After max. 10 minutes
Cooling capacity (at 75°F) 133 Btu/hr
Ambient temperature 40 °F to 120 °F
Factory set dew point 5 °C (40 °F)
Dew point noise static 0.1 K
Drift over full specified range ± 1.5 K
Power supply 115 V or 230 V, 50/60Hz, plug DIN 43650
Power consumption max. 350 VA
Status output Electrical spec. max. 230 V AC, 150 V DC
2 A, 50 VA
dry contacts
Plug DIN 43650
Protection class IP 20
Housing Stainless Steel
Weight approx. 24 lb.
1 Sample probe
2 Sample tube
3 3 way valve
4 Sample gas pump
5 Sample gas cooler EGK-1/2
6 Automatic condensate drain or perist. pump
7 Moisture detector
8 Fine filter
9 Flowmeter
10 Analyzer
For models and specs of components see individual data
sheets.
Typical Installation Diagram:
Dimensions (inch)
Sample gas/
Calibration gas
air
out
condensate
out
sample gas
in / out
air
in
DA 440011
09/2009
Page 2/4
0
20
40
60
80
100
120
140
160
180
btu/h
40 50 60 70 80 90 100 110 120
°F
A00
PKE 42
fitted with
DTS
Limit for
TV-SS
Cooling capacity
Ambient temperature
3)
Heat Exchanger TS TG TV-SS DTS DTG DTV
1)
Flow rate v 8.3 lpm 6.7 lpm 3.9 lpm 2 x 4.2 lpm 2 x 3.3 lpm 2 x 2.7 lpm
max
1)
Inlet dewpoint t 175 °F 175 °F 150 °F 175 °F 150 °F 150 °F
e,max
1)
Gas inlet temperature. J 355 °F 285 °F 285 °F 355 °F 285 °F 285°F
G,max
Max. cooling capacity Q 427 Btu/hr 218 Btu/hr 114 Btu/hr 427 Btu/hr 218 Btu/hr 175 Btu/hr
max
Gas pressure p 2321 psig 44 psig 44 psig 363 psig 44 psig 29 psig
max
Pressure drop Dp (v=2.5 lpm) 0.12 psig 0.12 psig 0.12 psig each 0.07 psig each 0.07 psig each 0.22 psig
Dead volume V 4.2 cu. in. 2.9 cu. in. 7.9 cu. in. 1.7 / 1.5 cu. in. 1.7 / 1.5 cu. in. 1.3 / 1.3 cu. in.
tot
2)
Sample gas connections G 1/4“ i GL 14 DN 4/6 tube 6 mm GL 14 DN 4/6
2)
Condensate outlet connections G 3/8“ i GL 25 G 3/8" i tube 10 mm GL 18 DN 4/6
1) 2) 3)
Max. cooling capacity of the cooler must be considered NPT-threads upon request Can only be used with peristaltic pumps
Heat Exchanger
The energy content of the sample gas and, as a result, the required cooling capacity of the gas cooler is determined by 3 parameters: gas
temperature J , dewpoint t (moisture content) and flow v. The outlet dew point rises with increasing energy content (heat) of the gas. The
G e
required cooling capacity is determined by the maximum acceptable level of the outlet dew point.
The following table shows cooler performance assuming the following conditions: t =120°F and J =160°F. Indicated is the v in lpm cooled
e G max
air (i.e. after the moisture has condensed).If the actual values stay below the parameters t and J , v can be increased. For example (TG),
e G max
instead of t = 120°F, J = 160°F and v = 5.7 lpm the values t =105°F, J =160°F a maximum flow rate of v=7.1 lpm could be achieved.
e G e G
Please contact one of Buhlers application specialists for assistance and further information.
DA 440011
09/2009
Page 3/4
Performance Data
4 4 7 0 0 0
0
1
2
0 0 0
1 1 0
1 2 0
1 3 0
2 6 0
2 7 0
2 8 0
0
1
PKE 42
Power Supply
115 V
230 V
Gas Path/ Material/ Version
without heat exchanger
single path heat exchanger TS, stainless steel
single path heat exchanger TG, glass
single path heat exchanger TV SS, PVDF
dual path heat exchanger DTS, stainless steel
dual path heat exchanger DTG, glass
1)
dual path heat exchanger DTV , PVDF
1)
Condesate Discharge
without condensate discharge
3)
peristaltic pump(s)
1)
Condensate outlet only suitable for connecting peristaltic pumps
2)
Peristaltic pumps for separate mounting available
3)
Each gas path is equipped with a peristaltic pump with the same mains supply requirements as the cooler.
Please indicate with order
Please extract the part number for the cooler fulfilling your requirements from the type code below.
Please note: Each gas path should be equipped with a peristaltic pump or an automatic condensate drain.
Part No.
DA 440011
09/2009
Page 4/4
We reserve the right to amend specifications.
Automatic condensate drain AK 5.2
Condensate vessel GL1, 0,4l
Peristaltic pump 0.3 l/h, 230V, for separate mounting
Peristaltic pump 0.3 l/h, 115V, for separate mounting
Accessories
451 00 08
441 00 05
912 40 30 104
912 40 30 105
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31

Buhler PKE 42 Installation And Operation Instruction Manual

Type
Installation And Operation Instruction Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI

in other languages