Differential weighing (example 2)
Task: Drying process with 2 samples. 1st sample ≈ 250 g, 2nd sample ≈ 150 g. Determine weight difference in % of initial weighing (1st and 2nd
samples) (balance type PM4600 DeltaRange)
Procedure Press key Display Printout
Select diff. weighing application
(with difference display in %) [DIFF. WEIGHING] 0.00 g
Place 1st container on balance, 121.39 g
then store tare weight [STORE TARE] - - - 1 - t -
Place 2nd container on balance, 142.96 g
then store tare weight [STORE TARE] - - - 2 - t -
Weigh in 1st sample, store, 371.35 g
then remove [STORE WEIGHT] - - - 1 - - -
Weigh in 2nd sample, store, 294.70 g
then remove [STORE WEIGHT] - - - 2 - - -
End weighing, perform drying
procedures [END © RESULTS] 0.00 g
Start diff. weighing [DIFF. WEIGHING] 0.00 g
Place 1st dry sample on balance, - - - 1 - - -
then remove sample [END © RESULTS] -47.65 %
- - - 2 - - -
Place 2nd dry sample on balance [END © RESULTS] -44.45 %
Note: After storage of the second tare weight, the differential weighing application can be interrupted with [END © RESULTS]. All values are stored.
Work can now be started with another application. Return to unfinished differential weighing application with [DIFF. WEIGHING]. The
weighing of the samples can now be continued.
--------------------
DIFF WEIGHING
SAMPLE PREP
--------------------
15.12.87 08:16
1 StoT 121.39 g
2 StoT 142.96 g
1 StoW 249.96 g
2 StoW 151.74 g
--------------------
--------------------
DIFF WEIGHING
RESULTS
--------------------
15.12.87 08:19
1 Dif -47.65 %
2 Dif -44.45 %
--------------------