  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo LabPac-M For PM balances Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Operating instructions
METTLER TOLEDO
LabPac-M
2
The METTLER TOLEDO LabPac-M
The METTLER TOLEDO LabPac-M has been developed especially for use in the chemistry lab.
It comprises a terminal (input keypad) and a plug-in program cassette. Connected to METTLER
TOLEDO balances employing M technology, the LabPac-M provides an extremely easy way of
performing the three most commonly required tasks in the lab.
- Net total weighing: Weighing in of any number of components from zero with the possibility
of calling up the net total at any time. With target weight setting,
weighing-in aid (DeltaTrac) is activated.
- % Formulation: With or without preset component weights as a percentage of the
formula weight. With target weight setting, weighing-in aid (DeltaTrac)
is activated.
- Differential weighing: Determination of weight change - due to a physical or chemical process
- in weight units or percent.
When METTLER TOLEDO balances of the SM series are used, there is a possibility to work
with the SM terminal instead of the LabPac-M terminal. Further details are given in the Section “SM
keypad” on pages 18 and 19.
To connect the terminal and other units
Two connection sockets at the rear of the balance are provided for the following units:
Socket “DATA I/ O”: Units with RS232C or CL interface, e.g. GA44 Printer (for GA44 in the
configuration register, section I-Face; select standard setting: S.Stb,
b 2400 and interval 1 s)
Socket “GM”: LabPac-M terminal and GM units with adapter plug
The coding pins in the sockets prevent improper connection.
3
To insert the program cassette
Insertion of the program cassette is described in the Operating Instructions of the balance used.
Notes: - Before changing the program cassette, disconnect power cord of the balance (with SM
balances, press key [OFF]).
- Replace standard program cassette by LabPac-M cassette.
- Please keep the standard cassette; it may be useful in subsequent use without the
LabPac-M.
Configuration register with LabPac-M
The LabPac-M can be adapted to units specific to the application. In the section -Unit- of the
configuration register (see Operating Instructions of the balance, “Configuration”), the following
can be selected whatever the standard software of the balance:
Language: -E -Yes- = English (standard setting)
-F -Oui- = French
-d -Ja- = German
-E -Si- = Spanish
-I -Si- = Italian
Date: EU date = day, month, year (standard setting)
US date = month, day, year
Printout: Au-Prot = printout automatic, depending on application (standard setting)
SEL.Prot = printout only with [PRINT] key
Notes: - If the configuration register cannot be selected, check whether the jumper of the
program cassette is in position “unsecured” (cf. Operating Instructions of the
balance, “Configuration”).
- Reset does not reset language and date.
4
The METTLER DeltaTrac
With METTLER balances employing M technology, the DeltaTrac can be used as a weighing-in aid or as a dynamic graphic display. Depending on
the application, the DeltaTrac displays "target weight” when the weight on the balance corresponds to the preset (keyed in) weight value or the piece
number. The display has two accuracy ranges for coarse and fine weighing.
Operation of the DeltaTrac is illustrated here with a weighing to target weight in the application net total.
Dynamic graphic display
A container is placed on the balance and tared with the control bar.
The display continues to function as though the LabPac-M were not connected. It does not switch
on until entries are made via the keypad during an application.
The segments of the dynamic graphic display show the container weight; the digital readout
remains at zero (container tared).
Weighing-in aid
The keys [NET TOTAL] and [COMP ] are now pressed, the component weight inputted and the
entry closed with [ENTER]. This alters the way in which the dynamic graphic display functions:
The radial bars indicating the container weight are blanked out. A horizontal bar appears at the
“9 o’clock” position. This is the coarse display and the 9 o’clock position is the zero position (no
material has been weighed yet).
Think of the graphic display, which is like a clock face, as being divided into a lower and upper half.
The lower half shows the coarse reading, the upper half is the fine readout.
If the bars are in the left-hand half, the weight is below the target value. However, if the bars are
in the right-hand half, the weight is greater than the target value.
minus zone
fine display
plus zone
coarse display
5
Weighing-in, coarse (material to be weighed added at a brisk rate)
The bar previously pointing in a horizontal direction now rotates downwards. With 1/3 of the weight,
it shows “8 o’clock”, with 2/3 “7 o’clock”. The target weight is reached when the bar points vertically
downwards (“6 o’clock”).
If more material is added (i.e. overfilling) the bar moves from the 6 o’clock position towards
3 o’clock (shown dashed in the illustration). The 3 o’clock position denotes 100% overfill. The bar
remains in this position even if more material is added.
Weighing-in, fine
When the coarse pointer is just before the 6 o’clock position, the fine pointer appears. It also begins
at the 9 o’clock position and moves towards the 12 o’clock position. Each step in the pointer’s
progress corresponds to 0.5% of the target weight. The target weight is reached at the 12 o’clock
position (accuracy ±0.25% of target weight). If still more material is added, i.e. overfilling, the fine
pointer moves on towards the 3 o’clock position: shown dashed in the illustration. The 3 o’clock
position signifies 7.5% overfilling. The fine pointer remains in this position even if filling continues.
At the same time, the coarse pointer moves from 6 o’clock in the direction 3 o’clock.
Weight OK
When the exact target weight (±0.25%) is reached, the fine and coarse pointers form a vertical line
between 12 and 6 o’clock. Tolerance markers are provided to the left and right of the 12 o’clock
position. They indicate a tolerance range of ±2.5% of the target weight. If an accuracy of ±0.25%
is not sufficient, material can be introduced at a brisk rate with the weighing-in aid until the fine
pointer is positioned just before 12 o’clock; the digital readout then provides further guidance.
The pointers jump back to 9 o’clock when the component weight is stored by pressing the [RE-
ZERO] key or [ENTER] after a new entry with [COMP ].
fine display
coarse display
target weight
6
Functions of general system keys
If a printer is connected, the date, time, identification number and code, if inputted, are printed out
at the top of the printout. On power failure, the printout inscription is cleared.
[ID NR] Begin entry of identification number with max. 7 digits and 6 decimal points (e.g. user
or workstation).
[DATE] Begin date entry (always 6 digits). The calendar runs until a power failure occurs.
Cleared by overwriting with “0”.
[TIME] Begin time entry (00...24 h; 6 digits, hours, minutes, seconds). The clock runs until
a power failure occurs. Cleared by overwriting with “0”.
[CODE] Begin entry of the code number with max. 7 digits and 6 decimal points, e.g. to identify
each weighing task of a series. Printed out immediately after entry and then cleared
automatically.
[RESET] Clears all entries except for date and time.
[RESET] must always be pressed before a new weighing task.
[CLEAR] Clears the entry of the last numeric key pressed (numeric keypad), then the next to
last, etc. provided [ENTER] has not yet been pressed.
[PRINT] Manual print command to print out entries or results which are not printed out
automatically (by an application).
[ENTER] Concludes the numerical input on the numeric keypad.
[SAVE] Saves maximum 3 stored values in addition to the application and the weighing
modes (statistics only) to prevent loss on power failure.
Use [SAVE] only if power failure likely. For details, see applications.
7
To enter values
Three steps are needed to enter values with the numeric keypad ([0] ... [9], [
.
], [-]):
1. Select desired function The selected function determines what type of value follows, e.g. a code number or a component
weight. The balance is now in the value entry mode.
The keys [ID NR] [DATE] [TIME] and [CODE] have two functions: When pressed once, the
appropriate function is selected. As pressing the keys activates the numeric keypad, the keys
become numeric keys of this keypad.
2. Input the value on the numeric keypad
3. Press [ENTER] key The value entry mode is terminated and the value assigned to the appropriate function.
Corrections Wrong entries which have already been closed with [ENTER] must be corrected by reinputting
them.
Entries not yet closed with [ENTER] can be cleared with [CLEAR].
Examples The keys mentioned should be pressed in the specified sequence:
To enter date e.g. 24.7.87 [DATE] [2] [4] [0] [7] [8] [7] [ENTER]
(for EU date)
To enter time e.g. 9.45.38 [TIME] [0] [9] [4] [5] [3] [8] [ENTER]
To enter a component e.g. 83.5 g [NET TOTAL] [COMP ] [8] [3] [
.
] [5] [ENTER]
weight in grams
To enter a component e.g. 27.8% [% FORMULA] [COMP IN %] [2] [7] [
.
] [8] [ENTER]
weight in %
8
Net total (information)
Typical tasks and possibilities
- Weighing in each component from zero (when weight entered with weighing-in aid).
- Net total (sum of component weights) can be called up at any time.
- Addition of top-up component (e.g. liquid) from net total to weight of formulation (add-weighing).
Notes
- Minimum weight entry = 1 digit
- Procedure when working without input of component weight (Delta-
Trac not activated):
[RESET] - [NET TOTAL] - load weight - [RE-ZERO] - load next weight,
etc.
- Switching off the DeltaTrac during a net total weighing with [COMP]
[0]. Input component weight to reactivate DeltaTrac.
- The component weight can be protected against loss on power failure
with [SAVE].
- A protected value can be cleared with [RESET] [SAVE].
Specific key functions
Activates the application net total and the 3 related keys.
Cleared by [RESET] or selection of another application.
Begin entry of component weight. Switches weighing-in
aid to coarse and fine readout. Target weight is displayed
when preset weight reached.
Adds effective component weight to net total. Sets digital
readout and weighing-in aid to zero (ready for next compo-
nent).
Switches between current component weight and net total
so far. (If key [RE-ZERO] has already been pressed,
current component weight is zero.) The weighing-in aid
shows the total weight.
9
Net total (example)
Task: 1st component 1000 g; 2nd component 125 g; fill with liquid to 2500 g (balance type PM4600 DeltaRange)
Procedure Press keys Display Printout
Place container on balance,
select application net total [NET TOTAL] 409.86 g
Tare Balance control bar 0.00 g
Key in weight of 1st component
(1000 g) [COMP ] [1] [0] [0] [0] [ENTER] 0.00 g
Weigh in 1st component
(slightly too much in this case) 1003.60g
Store true weight of
1st component [RE-ZERO] 0.00 g
Key in weight of 2nd component
(125 g) [COMP ] [1] [2] [5] [ENTER] 0.00 g
Weigh in second component
(slightly too little in this case) 122.96 g
Store true weight of
2nd component [RE-ZERO] 0.00 g
Call up net total so far [TOTAL/ LAST] [PRINT] 1126.60g
Fill to 2500 g [PRINT] 2501.10g
--------------------
NET TOTAL
--------------------
15.12.87 09:42
1*Comp 1000.00 g
Weight 1003.6 g
2*Comp 125.00 g
Weight 122.96 g
Total 1126.6 g
--------------------
Total 2501.1 g
--------------------
10
% Formula (information)
Typical tasks and possibilities
- Setting target weight (= 100%) of formula in grams and of components in percent of target weight.
- Mixing solutions, starting with the weight of the 1st component on the weighing pan (input of 1st component in %). The target weight (100%) is
calculated by the balance.
- Weighing in each component from zero with weighing-in aid.
- Net total (sum of components) can be called up any time in % and weight.
Specific key functions
Activates the % FORMULA application and the 3 related keys. Cleared by [RESET] or selection of another application.
Begin entering the formula target weight in weight units (e.g. gram). This target weight represents 100% in relation to the components.
Without entry of the target weight, the weight on the balance pan can be entered with [COMP IN %]. The target weight is then calculated.
Begin entering the components in % of target weight. Weighing-in aid with coarse and fine pointers indicates “target weight” when
specified percentage reached. After weighing in, [COMP IN %] must be pressed to add effective percentage to net total. The display
is then set to zero automatically.
To call up the net total reached so far: Press key once to give net total in % of target weight, press twice for net total in weight units
(e.g. gram). Clears previously shown, current component percentage. Thus press this key only when weighing of a component is
complete. Return to % display by inputting the next component percentage.
11
Notes
- The target value entry can be greater than the weighing range of the balance. “Err 2” flashes as a warning, but work can be continued.
- Minimum value entry = 0.01%
%- values which, based on the target weight, are less than 1 digit (smallest display unit of the balance), are not supported by the DeltaTrac +/ display.
- Formula target weight and first value of the component entry can be protected against loss on power failure with [SAVE].
- Protected values can be cleared with [RESET] [SAVE].
Application examples for %- formula, see following two pages.
12
% Formula (example 1)
Task: Formula target weight 2500 g (= 100%); 1st component 20% (= 500 g); 2nd component 7.5 % (= 187.5 g); fill with 3rd component to
target weight (2500 g) (balance type PM4600 DeltaRange)
Procedure Press keys Display Printout
Place container on balance,
select application % formula [% FORMULA] 409.86 g
Tare Balance control bar 0.00 g
Key in target weight (2500 g) [TARGET ] [2] [5] [0] [0] [ENTER] 0.00 g
Key in 1st component (20%) [COMP IN %] [2] [0] [ENTER] -500.00 g
Weigh in 1st component
(slightly too little here) -2.16 g
Store 1st component [COMP IN %] 0.00 g
Key in 2nd component (7.5%) [COMP IN %] [7] [
.
] [5] [ENTER] -187.50 g
Weigh in 2nd component
(slightly too much in this case) 1.44 g
Store second component [COMP IN %] 0.00 g
Call up net total in % [TOTAL %/ TOTAL ] [PRINT] 27.47 %
Fill with 3rd component
to 100% (= 2500 g) [PRINT] 100.05 %
--------------------
% FORMULA
--------------------
15.12.87 10:44
Target: 2500.00 g
*******
1*Comp 20.00 %
Weight 19.91 %
Weight 497.84 g
2*Comp 7.50 %
Weight 7.56 %
Weight 188.94 g
Total 686.8 g
Total 27.47 %
--------------------
Total 2501.1 g
Total 100.05 %
--------------------
13
% Formula (example 2)
Task: Prepare solution with 3 components. 1st component 700 g = 18%; 2nd component 10%; fill to 100% with 3rd component.
Target weight calculated automatically with aid of 1st component weight. (balance type PM4600 DeltaRange)
Procedure Press keys Display Printout
Place container on balance,
select application % formula [% FORMULA] 181.95 g
Tare Balance control bar 0.00 g
Weigh in 1st component
(target weight 700 g) 699.80 g
Key in 1st component (18%) [COMP IN %] [1] [8] [ENTER] 699.80 g
Store 1st component [COMP IN %] 0.00 g
Key in 2nd component (10%) [COMP IN %] [1] [0] [ENTER] -388.76 g
Weigh in 2nd component 0.11 g
Store 2nd component [COMP IN %] 0.00 g
Call up net total in % [TOTAL %/ TOTAL ] [PRINT] 28.00 %
Fill with 3rd component
to 100% [PRINT] 100.03 %
--------------------
% FORMULA
--------------------
15.12.87 11:03
1*Comp 18.00 %
Weight 699.8 g
2*Comp 10.00 %
Weight 10.00 %
Weight 388.87 g
Total 1088.6 g
Total 28.00 %
--------------------
Total 3888.7 g
Total 100.03 %
--------------------
14
Differential weighing (information)
Typical tasks and possibilities
- Weighed in samples are reweighed after treatment. Weights can be read in % or g.
- Balance can be used for other weighings while samples undergo treatment, e.g. net total or % formula.
- Storage of up to 20 tare weights and initial weighings possible. The results can be displayed in 4 different ways.
Specific key functions
Activates the differential weighing application and the 3 related keys. Cleared by [RESET] or selection of another application. Key must
also be pressed again before the actual differential weighing.
Stores the container weight. Sets digital readout to zero, but not the weighing-in aid. Occupied tare memory shown by LED.
Stores the sample weight weighed. Occupied weight memory is shown by LED. The container weight is subtracted automatically.
Ends the weighing process (single weighing or a series of weighings). Only the LED [END © RESULTS] remains on. The key must
be pressed to store the weighing and to start the differential weighing. After pressing the key [END © RESULTS], work can also be
carried out in another application.
15
Notes
- Calculation method for weight difference or differential weighing
After loading the sample for differential weighing and pressing the key [END © RESULTS], the result is displayed in the calculation mode last
used (CLEAR 1, 2 or 3) and recorded. As long as the sample remains on the balance, it is possible to switch to a different calculation method and
print out the corresponding value with [PRINT].
The following calculation methods are possible:
[CLEAR] weight difference in % of initial weighing (default, if nothing else has been selected)
[1] weight difference in weight units
[2] differential weighing in weight units
[3] differential weighing in % of initial weighing
- The following two types of weighings are possible:
a) Store all tare weights (1...20) in succession, then the weights (1...20) in the same order © example 1
b) First the tare and then the initial weight are stored, sample by sample (1...20) © example 2
Between the individual weighings (container or sample) of a series, the balance can be tared with the control bar if necessary.
Stored values are indicated with LEDs.
- Repetition of the differential weighing:
As long as [END © RESULTS] [RESET] is not pressed during the differential weighing, the values remain stored. The differential weighing can
thus be performed again.
- Clearing of an unusable weighing series:
Stored weight values of an unusable sample series which has been closed with [END © RESULTS] can be cleared with [DIFF. WEIGHING]
[END © RESULTS ] [RESET].
Application examples for differential weighing, see following two pages.
16
Differential weighing (example 1)
Task: Drying process with 2 samples. 1st sample 500 g, 2nd sample 800 g. Determine residual weight in % of initial weight (1st and 2nd sample),
as well as weight difference in % of initial weight (2nd sample) (balance type PM4600 DeltaRange)
Procedure Press key Display Printout
Select diff. weighing application [DIFF. WEIGHING] 0.00 g
Place 1st container on balance, 151.75 g
then store tare weight [STORE TARE] - - - 1 - t -
Weigh in 1st sample, store, 501.50 g
then remove [STORE WEIGHT] - - - 1 - - -
Place 2nd container on balance, 206.53 g
then store tare weight [STORE TARE] - - - 2 - t -
Weigh in 2nd sample, store, 801.60 g
then remove [STORE WEIGHT] - - - 2 - - -
End weighing, perform
drying procedures [END © RESULTS] 0.00 g
Start diff. weighing [DIFF. WEIGHING] 0.00 g
Place 1st dry sample on balance, - - - 1 - - -
then remove sample [END © RESULTS] A69.60%
- - - 2 - - -
Place 2nd dry sample on balance [END © RESULTS] A43.14%
Display and printout of weight
difference in % of initial weighing [CLEAR] [PRINT] - 56.86 %
--------------------
DIFF WEIGHING
SAMPLE PREP
--------------------
1 StoT 151.75 g
1 StoW 501.5 g
2 StoT 206.53 g
2 StoW 801.6 g
--------------------
--------------------
DIFF WEIGHING
RESULTS
--------------------
1 Abs 69.60 %
2 Abs 43.14 %
2 Dif -56.86 %
17
Differential weighing (example 2)
Task: Drying process with 2 samples. 1st sample 250 g, 2nd sample 150 g. Determine weight difference in % of initial weighing (1st and 2nd
samples) (balance type PM4600 DeltaRange)
Procedure Press key Display Printout
Select diff. weighing application
(with difference display in %) [DIFF. WEIGHING] 0.00 g
Place 1st container on balance, 121.39 g
then store tare weight [STORE TARE] - - - 1 - t -
Place 2nd container on balance, 142.96 g
then store tare weight [STORE TARE] - - - 2 - t -
Weigh in 1st sample, store, 371.35 g
then remove [STORE WEIGHT] - - - 1 - - -
Weigh in 2nd sample, store, 294.70 g
then remove [STORE WEIGHT] - - - 2 - - -
End weighing, perform drying
procedures [END © RESULTS] 0.00 g
Start diff. weighing [DIFF. WEIGHING] 0.00 g
Place 1st dry sample on balance, - - - 1 - - -
then remove sample [END © RESULTS] -47.65 %
- - - 2 - - -
Place 2nd dry sample on balance [END © RESULTS] -44.45 %
Note: After storage of the second tare weight, the differential weighing application can be interrupted with [END © RESULTS]. All values are stored.
Work can now be started with another application. Return to unfinished differential weighing application with [DIFF. WEIGHING]. The
weighing of the samples can now be continued.
--------------------
DIFF WEIGHING
SAMPLE PREP
--------------------
15.12.87 08:16
1 StoT 121.39 g
2 StoT 142.96 g
1 StoW 249.96 g
2 StoW 151.74 g
--------------------
--------------------
DIFF WEIGHING
RESULTS
--------------------
15.12.87 08:19
1 Dif -47.65 %
2 Dif -44.45 %
--------------------
18
Operation of METTLER TOLEDO SM balances with the SM terminal
METTLER TOLEDO balances of the SM type can be operated with the SM terminal in the net total
application. Inputs can be made only via the LabPac-M terminal, however. After entry of the
component weight, the LabPac-M terminal can thus be removed until a change in the setting is
desired.
Notes
- Values which are not saved with [SAVE] are lost when the balance is switched off or on power
failure.
- Saved values are cleared with the keys [RESET] [SAVE] of the LabPac-M terminal.
- In the % formula and differential weighing applications, the keys on the SM terminal have no
function.
19
SM function keys in the net total application
Switching between current component weight and net total so far.
Corresponds to the key [TOTAL/ LAST] of the LabPac-M keypad.
Corresponds to the [PRINT] key of the LabPac-M keypad.
Corresponds to the [RE-ZERO] key of the LabPac-M keypad.
20
Error messages
Err 2 Input value too high - Component weight is greater than balance capacity.
Err 3 Input value too small - Percentage input too small (display Err 2 or Err 3), i.e. 100% would be greater than weighing
range.
After a timeout, the balance returns to the weighing mode.
Err 5 Date or time not complete Format for both entries: XX.XX.XX (6 digit)
or implausible EU date: DD.MM.YY
US date: MM.DD.YY
Err 8 Sample number too large Sample number entry greater than 255.
All these error messages are displayed for approx. 5 seconds. The balance then displays the weight again. The cause of the error display is ignored.
Other error messages do not originate with the LabPac-M and are explained in the Operating Instructions of the balance in question.
/