1154563 A V60/V60 Plus User Manual Addendum: Resistive Circuit Component Warning 3
V60/V60 Plus User Manual Addendum:
Resistive Circuit Component Warning
Make these changes to the V60/V60 Plus user manual.
Replace this warning (in the “Warnings, cautions, and notes,” “Setting up the
ventilator for use,” and “Parts and accessories” chapters):
WARNING: Avoid adding resistive circuit components on the patient side of the
proximal pressure line. Such components may defeat the disconnect
alarm.
with this warning:
WARNING: Avoid adding resistive circuit components anywhere within the patient
circuit. Such components may defeat the disconnect alarm.
Replace this note (in the “Setting up the ventilator for use” and “Parts and
accessories” chapters):
NOTE: Resistive components can include but are not limited to HMEs,
proximal flow sensors, a filter at the patient connection, or a narrow
diameter circuit attached to a mask.
with this note:
NOTE: Resistive components can include but are not limited to HMEs,
unapproved humidifiers, proximal flow sensors, a filter at the patient
connection, or a narrow diameter circuit attached to a mask.