CEM iPrep 12 User manual

  • Hello! I am an AI chatbot trained to assist you with the CEM iPrep 12 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
iPrep®
High Performance Vessels
Manual
THIS PAGE INTENTIONALLY LEFT BLANK
Contents
Safety Notations
Safety Notations 1
Introduction
About the iPrep Vessels 2
Important Safeguards
Operating Parameters 3
Sample Preparation Safeguards 3
Vessel Safeguards 4
Microwave Safeguards 4
Compounds Unsuitable for Closed Vessel Microwave
Digestion 5
Vessel and Accessories Overview
Vessel Components 6
Accessories 7
New Liner and Plug Conditioning
Nitric Acid Liner and Plug Conditioning 8
Non-Nitric Acid Liner and Plug Conditioning 9
Perform a Digestion
Vessel Assembly 10
Turntable Setup 13
Load and perform a Method 14
Vessel Removal 14
Vessel Cleaning and Inspection
Liners and Plugs 17
Liner and plug Bake-out Procedure 17
Turntable, Frame, Load Disc, Screw & Racks 18
Sleeves 18
Vessel Distribution
12 Place Turntable 19
16 Place Turntable 20
Methods
Create a One Touch Method 21
Create a Classic Method 21
Routine Maintenance
After Every Use 22
Monthly 22
Annually 22
Bake-out Procedure 22
Troubleshooting
Problems and Possible Solutions 23
THIS PAGE INTENTIONALLY LEFT BLANK
1Safety Notations
Safety Notations
Safety Notations
This manual uses three safety alert words at points in the documentation where the user should be aware of
potential hazards. The safety alerts are shown in color-coded boxes. The three words–NOTE, CAUTION, and
WARNING–indicate differing levels of observation or action as described below:
NOTE
A NOTE is intended to provide emphasis of procedures that may be misinterpreted or overlooked, or to
otherwise clarify confusing situations.
CAUTION
A CAUTION is intended to provide essential information and to emphasize procedures which, if not strictly
followed, may result in improper instrument operation.
WARNING
A WARNING is intended to emphasize dangerous or hazardous conditions which may result in personal
injury to the user and damage the instrument.
2 Introduction
Introduction
About the iPrep Vessels
The iPrep vessel is designed for microwave digestion in a MARS 6TM with iWave®temperature control. This vessel
is well suited for digestion of difcult matrices such as waste oils, APIs with multiple aromatic rings, polymers, and
geological samples. This vessel is also a great option when digesting 1 g or larger samples due to ppt detection
limit needs. Preinstalled methods take the guesswork out of method creation and sample preparation.
The MARS 6 requires the iWave temperature control with software version 1.43 or higher to operate iPrep vessels.
This Manual should be used in conjunction with the MARS 6 iWave Manual, PN 600832. Read and fully
understand all documentation before operating the instrument.
3Important Safeguards
Important Safeguards
Microwave sample preparation imposes a unique set of safety considerations beyond the basics of good laboratory
practice. Additional Safeguards for the MARS 6 instrument are included in the manual, PN 600832. General
guidelines for safe operation of the iPrep are presented below and all specic safety messages are located
throughout the manual.
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should be worn
as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s safety data sheet.
Refer to these guidelines for proper handling and disposal of the reagents.
Operating Parameters
iPrep 12 iPrep 16
iPrep Parameters Minimum Volume 5 mL 5 mL
Maximum Volume 30 mL 30 mL
Maximum Pressure 1500 psi 1500 psi
Maximum Hold Time 45 min 45 min
Maximum Temperature Maximum Programmable Temperature 300 ºC 300 ºC
Nitric Acid 250 ºC 250 ºC
Hydrochloric Acid* 250 ºC 250 ºC
Hydrouoric Acid* 250 ºC 250 ºC
Sulfuric Acid 300 ºC 300 ºC
Phosphoric Acid 300 ºC 300 ºC
Aqua Regia* 250 ºC 230 ºC
Water 180 ºC 180 ºC
Maximum Sample Weight Food, Plant, and Animal Tissue 1.5 g 1.5 g
Oil and Plastic 0.5 g 0.5 g
Pharmaceutical 1.5 g 1.5 g
Inorganic 1.5 g 1.5 g
Metallurgical 1.5 g 1.5 g
Clinical 1.5 g 1.5 g
* The maximum achievable temperature is 210 °C if running more than 6 vessels.
Sample Preparation Safeguards
• Do not use water to dilute the acid(s) greater than 25% when running over 220 ºC or to add volume before
digestion. This can cause uneven heating of vessels
• Run similar sample types with the same acid(s) and acid volume in each vessel
• Optimal Sample Volume is 10mL and will result in more even heating
• Minimum acid volume is 5 mL
• Maximum volume is 30 mL
• Maximum sample weight varies per sample type. See "Operating Parameters" for weights
• Acid blanks cannot account for more than 25% of turntable when mixed with samples
• Acid blanks cannot be run with organic samples > 230 °C
• Acid blanks cannot be run with samples when acid mixtures are > 50% hydrouoric acid. Acid blanks must
be run separately
4 Important Safeguards
Vessel Safeguards
• Only use vessels and vessel components purchased directly from CEM or through its authorized dealer
network.
• All vessel components must be dry prior to use
• Always condition new liners and plugs
• Do not batch old liners with new liners unless new liners have been conditioned with at least 3 runs using
the customer’s sample
• Ensure vessels are assembled properly
• Align tab on load disc in center notch of liner
• Look for vent hole through the frame
• Align liner properly on insulator disc at bottom – liner ush with tab
• Use torque block
• Invert sleeves after every run to extend lifetime and promote better sealing & even heating
• Slide liner into the end of the sleeve with most resistance felt if unsure which side was previously
used
• CEM recommends using a high temperature pen to mark sleeves for a visual reference.
• Try to group liners with similar sample history and usage
• Segregate liners for HNO3 and non-HNO3 applications
• Do not use a new liner with a set of liners that has been run consistently until the new liner has
been conditioned with that sample type
• Pull ones that run 20 ºC hotter than average and run them together
• Bake out liners and plugs to remove residual HNO3 & NOx gases
• Use Bake Out procedure for iPrep vessels (210 °C for 6 hrs – overnight)
• Bake out liners if TempGuardTM is triggered
Microwave Safeguards
• Always use the TempGuard feature in the method
• Set the TempGuard limit at 30 °C > control temperature if control temperature is < 250 °C.
• Set the TempGuard limit at 20 °C > control temperature if control temperature is > 250 °C
• Limit TempGuard to 270 °C for digestion of organics with HNO3 (primarily)
• Limit TempGuard to 300 °C for H3PO4/H2SO4 (primarily) applications
• Limit control temperature of organics with HNO3 to 250 °C
• Choose Sample Type in method with software version 1.45 or higher as follows:
• Standard = Any acid mixture other than HCl or HF dominant
• HCl = aqua regia or acid mixtures that are ≥ 75% HCl
• Always use the Max Power Monitor and set the alarm time for 10 minutes
• Maximum programmable temperature is 300 °C
• Maximum temperature varies per acid. See "Operating Parameters" for temperatures
• Maximum programmable Hold Time is 45 minutes
• If a run is stopped or paused for more than 30 seconds during the digestion, do not restart the method
with hot vessels
• Allow vessels to cool < 80 °C
• Vent vessels and reseal
• Restart method from the beginning
• Allow vessels to cool below 80 ºC before opening the microwave door.
5Important Safeguards
Compounds Unsuitable for Closed Vessel Microwave Digestion
WARNING
CEM will not be responsible for damage to equipment and facilities or personal injuries resulting from
microwave digestion of samples that are deemed unsuitable for closed vessel microwave digestion.
Acid decomposition of certain chemical compounds or types of samples constitutes unreasonable, hazardous
misuse of CEM microwave digestion systems. The classes of compounds listed below are unsuitable for closed
vessel microwave digestion because they are highly reactive with oxidizing acids and/or may become nitrated and
potentially explosive. Absence of a particular chemical compound from this list does not imply microwave acid
decomposition of such a sample is safe under all conditions.
• Explosives (TNT, Nitrocellulose, etc.)
• Propellants (Hydrazine, Ammonium Perchlorate, etc.)
• Pyrophoric chemicals
• Hypergolic mixtures (Nitric Acid and Phenol, Nitric Acid and Triethylamine, Nitric Acid and Acetone,
etc.)
• Animal Fats (Esters of glycerol capable of nitration and the formation of nitroglycerin or other nitrated
organic compounds)
• Aviation Fuels (JP-1, etc.)
• Acetylides
• Glycols (Ethylene Glycol, Propylene Glycol, etc.)
• Perchlorates (Ammonium, Potassium, etc.)
• Ethers (Cellosolve - Ethylene Glycol Phenyl ether, etc.)
• Lacquers
• Alkanes (Butane, Hexane, etc.)
• Ketones (Acetone, Methyl Ethyl Ketone, etc.) and alcohols (methanol, etc.)
6 Vessel and Accessories Overview
Vessel and Accessories Overview
For a list of parts and part numbers, visit our website at www.cem.com. A list of parts can be found on the
“Products” page under “MARS 6”.
Vessel Components
WARNING
Only use vessels and vessel components purchased directly from CEM or through its authorized dealer
network. The use of vessels and vessel components other than original CEM parts will put operators and
equipment at risk; therefore, CEM will not assume any liability for use of these parts and all warranties,
either written or implied, will immediately become void.
1
5
4
3
2
6
Item Description
1 Frame Module Supports sleeve, load disc, vessel liner, plug and screw
2 Sleeve Supports vessel liner
3 Load Disc Evenly distributes load from plug
4 Vessel Liner Contains acid and sample. 110 mL Liner (5 -30 mL working volume)
5 Plug Seals liner to contain acid and sample
6 Frame Screw Uniformly provides pressure seal among vessels when torqued
7Vessel and Accessories Overview
Accessories
1
2
3
4
5
Item Description
1Torque Tool Provides appropriate vessel seal. Includes plug removal tool
2 Vessel Liner Holder Holds up to a set of 16 vessel liners
312 Place Turntable Holds assembled vessels inside MARS 6 cavity and provides rotation
416 Place Turntable Holds assembled vessels inside MARS 6 cavity and provides rotation
5 Frame Carrier Holds up to 4 assembled vessels
8 New Liner and Plug Conditioning
New Liner and Plug Conditioning
All new liners and plugs should be conditioned as a set prior to being put into service. The break-in procedure is
used to condition the Teon® to ensure a proper seal. CEM recommends conditioning the plugs and liners based
on the acid type that will be used: HNO3, HCl, H2SO4, H3PO4, HF. Aqua Regia, or H2O2.
Nitric Acid Liner and Plug Conditioning
1. Use hot soapy water and foam brush to clean the liners and plugs. Rinse thoroughly with DI water and allow to
dry completely before use.
NOTE
Micro 90 Cleaner® liquid detergent works well on Teon. Avoid detergents that contain ingredients that
could cause potential contamination such as Na and P. Avoid abrasive cleansers and stiff brushes which
can scratch the Teon surface.
2. Place 0.25 g ± 0.05 g of tea leaves or appropriate sample into each liner. Never add sample or liquid to liner
with sleeve in place.
WARNING
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should
be worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
safety data sheet. Refer to these guidelines for proper handling and disposal of the reagents.
3. Add 10 mL of HNO3 to each liner. Use the acid to rinse the inside of the liner to ensure that the sample is in
the base of the liner.
4. Wait 15 minutes to allow pre-digestion.
5. Assemble and torque each vessel. See “Vessel Assembly” for detailed instructions.
6. Place the turntable into the MARS 6 cavity and symmetrically distribute vessels. See “Turntable Setup” and
"Vessel Distribution" for detailed instructions.
7. Select and load the “Plant Material” method from the One Touch directory.
8. Close the instrument door and run the preprogrammed “Plant Material” method.
9. After completion of the method, permit the vessels to cool in the automated sequence until the vessels cool to
below 80 ºC.
WARNING
Allow vessels to cool below 80 ºC before opening the microwave door.
To prevent the possibility of severe burns, ensure that gloves and protective gear as outlined in your safety
program are worn.
10. Remove the vessel assemblies from the turntable. See "Vessel Removal" for detailed instructions.
11. If the normal method temperatures are greater than 230 °C AND new liners will be mixed with old liners, the
new liners will need to be conditioned three times using the method that exceeds 230 °C before mixing in.
12. Rinse the liners and plugs with deionized water.
9New Liner and Plug Conditioning
Non-Nitric Acid Liner and Plug Conditioning
1. Use hot soapy water and foam brush to clean the liners and plugs. Rinse thoroughly with DI water and allow to
dry completely before use.
NOTE
Micro 90 Cleaner® liquid detergent works well on Teon. Avoid detergents that contain ingredients that
could cause potential contamination such as Na and P. Avoid abrasive cleansers and stiff brushes which
can scratch the Teon surface.
2. Place 0.25 g ± 0.05 g of your sample into each liner. Never add sample or liquid to liner with sleeve in place.
WARNING
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should
be worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
safety data sheet. Refer to these guidelines for proper handling and disposal of the reagents.
3. Add 10 mL of the digestion acid to each liner. Use the acid to rinse the inside of the liner to ensure that the
sample is in the base of the liner.
4. Wait 15 minutes to allow pre-digestion.
5. Assemble and torque each vessel. See “Vessel Assembly” for detailed instructions.
6. Place the turntable into the MARS 6 cavity and symmetrically distribute vessels. See “Turntable Setup” and
"Vessel Distribution" for detailed instructions.
7. Select and load the “Easy Prep Clean” method from the One Touch directory.
8. Close the instrument door and run the preprogrammed “Plant Material” method.
9. After completion of the method, permit the vessels to cool in the automated sequence until the vessels cool to
below 80 ºC.
WARNING
Allow vessels to cool below 80 ºC before opening the microwave door.
To prevent the possibility of severe burns, ensure that gloves and protective gear as outlined in your safety
program are worn.
10. Remove the vessel assemblies from the turntable. See "Vessel Removal" for detailed instructions.
11. Rinse the liners and plugs with deionized water.
NOTE
When using HCl, if a dark ring appears in liner, use hot soapy water and foam brush to clean. Rinse
thoroughly with DI water and allow to dry completely before use.
10 Perform a Digestion
Perform a Digestion
Vessel Assembly
CAUTION
Never add sample or liquid to liner with sleeve in place. Use the vessel liner holder.
1. Place liners in vessel liner holder. All vessel components must be dry prior to use.
2. Add sample to the liner(s).
WARNING
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should
be worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
safety data sheet. Refer to these guidelines for proper handling and disposal of the reagents.
2.1. Weigh and place the sample into the liner. The maximum sample weight varies per sample
type. See "Operating Parameters" for specic weights.
2.2. Add the acid (5 mL - 30 mL working volume). Use the acid to rinse the inside of the liner to
ensure that the sample is in the base of the liner. If running more than one vessel, you must
run similar samples and the same acids and acid volume in each vessel. See "Sample Prepa-
ration Safeguards" for information regarding acid blanks along with additional safeguards.
2.3. Ensure that the acid covers the sample and the base of the liner (5 mL minimum).
CAUTION
Do not use water to dilute the acid(s) greater than 25% when running over 220 ºC or to add volume before
digestion. This can cause uneven heating of vessels.
Do NOT batch old liners with new liners unless new liners have been conditioned with at least 3 runs.
3. Wait 15 minutes to allow pre-digestion.
4. Install the plug into the liner. Slight resistance should be felt when placing the plug into the liner. If resistance
is not felt, try using another plug.
11Perform a Digestion
CAUTION
If multiple plugs fall freely into liner, discard liner. Invert sleeves between tests to prevent liners from
stretching.
5. Carefully slide the liner into the sleeve.
NOTE
Invert the vessel sleeves after every test to help maintain uniformity of the liner and sleeves.
6. Place the load disc on the plug with the tab on the disc facing downward so that it ts into the center notch of
the liner.
7. Position the sleeve into the frame. Verify vertical placement by ensuring the sleeve is aligned on the disc at the
bottom of the frame.
12 Perform a Digestion
8. Tighten the screw onto the top of the load disc until it is nger tight.
9. Use the provided torque block and tighten the screw until a single click is heard or felt.
13Perform a Digestion
WARNING
Vessel torquing is critical to optimize controlled venting of the system and for the integrity of the vessel.
Over or under torquing may lead to sealing issues. If any issue is noted, contact CEM Corporation
10. Verify the vessel alignment by looking through the viewing port located on the vessel frame. A tiny vent hole
aligned with the window should be visible.
Turntable Setup
1. Verify the oor mat positioning, ensuring that the mat is not covering the iWave sensors.
2. Load the empty turntable into the instrument by aligning position “1” with the front of the instrument.
3. Turntable must be seated down on drive lug.
NOTE
CEM suggests that the iPrep vessel set be loaded into the turntable inside the instrument due to the
weight and height of the vessels.
4. Load the vessel assemblies into the turntable one by one, ensuring that the notch in the bottom rear of the
frame engages the locking ring on the turntable. For assistance in loading the vessels, use the Turntable Icon
in the upper right corner of the MARS 6 screen to rotate the turntable. See “Vessel Distribution” section
14 Perform a Digestion
of this manual for optimal vessel distribution.
5. With the door open, use the Turntable Icon to rotate the turntable. Observe the positioning and balance of
the turntable. Make adjustments by reseating the turntable if necessary.
Load and perform a Method
1. Verify Max Power Monitor is enabled and the alarm time is programmed for 10 minutes (Settings > Run > Max
Power Monitor).
2. Select/load the method from the directory, One Touch or Classic. See "Methods" section or contact CEM Cor-
poration if assistance is needed creating a new method.
3. Close the instrument door.
4. Select “Start” to run the method. The instrument will begin the method according to the selected parameters
– initializing, ramping, holding, cooling, etc.
NOTE
If a run is stopped during the digestion, do not restart the method with hot vessels. Allow vessels to cool
< 80°C, vent vessels and reseal, and restart method from the beginning.
CAUTION
Observe the entire digestion sequence when establishing a procedure for the rst time.
5. Verify that vessel count, type and position are correct.
6. After completion of the method, permit the vessels to cool in the microwave cavity below 80 ºC.
Vessel Removal
1. Allow the vessels to cool below 80 ºC before opening the microwave door.
WARNING
To prevent the possibility of severe burns, ensure that gloves and protective gear as outlined in your safety
program are worn.
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should
be worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
safety data sheet. Refer to these guidelines for proper handling and disposal of the reagents.
2. Remove the vessel assemblies from the turntable one by one, keeping the vent hole pointed away from the
user, and load them into the vessel carrier rack. Keep the vent holes facing away from the user during trans-
portation of the rack.
15Perform a Digestion
3. Place the vessel carrier rack into a fume hood, keeping the vent hole pointed away from the user.
4. Use the torque block to carefully loosen the vessel screws, allowing them to vent.
5. Remove the vessels from the frame carrier rack.
6. Place the liner and plug into the liner holder rack.
7. Use the plug removal tool to remove the plug from each liner.
16 Perform a Digestion
8. Transfer the digested sample from liner and proceed with work-up.
CAUTION
Never pour sample or liquid from liner with sleeve in place.
9. Clean vessels as outlined in “Vessel Cleaning ” Section.
/