Thermo Fisher Scientific QuantStudio 3 and 5 Calibration and Performance Check Record Owner's manual

Type
Owner's manual
QuantStudio 3 and 5 Real-Time PCR Systems
Pub. No. MAN0019221 Rev. A.0
The QuantStudio 3 and 5 Real-Time PCR Systems Calibration and Performance Check Record is a document designed to allow end
users to perform and record the results of dye calibration and an instrument performance check. This document does not take the place
of a qualication service performed by a certied Thermo Fisher Scientic eld service engineer.
IMPORTANT! This document does not represent qualication of the QuantStudio 3 or 5 Real-Time PCR System.
Customer and service information
Customer organization name:
Customer contact name:
Customer contact title:
Customer organization location:
Date performed:
Performed by:
Instrument information
Instrument model: QuantStudio 3 Real-Time PCR Instrument (96Well 0.1 mL Block)
QuantStudio 3 Real-Time PCR Instrument (96Well 0.2 mL Block)
QuantStudio 5 Real-Time PCR Instrument (96Well 0.1 mL Block)
QuantStudio 5 Real-Time PCR Instrument (96Well 0.2 mL Block)
QuantStudio 5 Real-Time PCR Instrument (384Well 0.02 mL Block)
Instrument serial number:
Software version:
Responsibilities
Customers should ensure that the QuantStudio 3 and 5 Real-Time PCR Systems Installation, Use, and Maintenance Guide (Pub. No.
MAN0010407) is available, and that the appropriate consumables (listed in each section) have been obtained.
CALIBRATION AND PERFORMANCE
CHECK RECORD
For Research Use Only. Not for use in diagnostic procedures.
Optical calibrations
3Task
Obtain the appropriate dye calibration plates listed below.
96-well (0.1mL) 96-well (0.2mL) 384-well (0.02mL)
ROI and background kit 4432426 4432364 4432320
10-dye spectral calibration kit A26342 A26343 A26341
Follow the instructions in the user guide to complete the ROI, background, and uniformity calibrations.
ROI: Passed
Background: Passed
Uniformity: Passed
Note: If the background calibration fails, clean the block according to the instructions in the user guide.
For 96Well 0.1mL and 96Well 0.2mL instruments only. If not applicable, select "N/A" below, then skip to the next section in this
table.
N/A
Follow the instructions in the user guide to complete the pure dye calibrations.
Plate Dyes Result
Plate 1 FAM, VIC, ROX, SYBR GREENPassed
Plate 2 ABY, JUN, MUSTANG PURPLEPassed
Plate 3 NED, TAMRA, Cy5Passed
For 384Well 0.02mL instruments only. If not applicable, select "N/A" below, then skip to the next section in this document.
N/A
Follow the instructions in the user guide to complete the pure dye calibrations.
Plate Dyes Result
Plate 1 FAM, VIC, ROX, TAMRA, SYBR GREENPassed
Plate 2 ABY, JUN, MUSTANG PURPLE, NED, Cy5Passed
Comments
If you have no comments, enter "Not applicable" below.
2QuantStudio 3 and 5 Real-Time PCR Systems Calibration and Performance Check Record
Performance check
3Task
Obtain the appropriate RNase P plate listed below.
96-well (0.1mL) 96-well (0.2mL) 384-well (0.02mL)
RNase P Instrument Verification Plate 4351979 4432382 4455280
Follow the instructions in the user guide to run the RNase P plate to complete the performance check. The instrument passes performance
specifications if the following inequality is true and the instrument successfully distinguishes between unknown populations A (5000 copies) and
B (10000 copies) with a statistical confidence level of 99.7%. Performance is verified by the analysis equation specified below:
[(CtA) – 3(σCtA)] > [(CtB) + 3(σCtB)]
where:
• CtA = Average Ct of unknown population A
σCtA = Standard deviation of unknown population A
• CtB = Average Ct of unknown population B
σCtB = Standard deviation of unknown population B
When the run is complete, perform a primary analysis of the data and enter the values:
CtA:σCtA:
CtB:σCtB:
Performance check: Passed
Comments
If you have no comments, enter "Not applicable" below.
QuantStudio 3 and 5 Real-Time PCR Systems Calibration and Performance Check Record 3
Approval
Customer Name:
Customer Signature:
Date:
Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT,
PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. MAN0019221
Revision Date Description
A.0 03 April 2020 New document.
©2020 Thermo Fisher Scientic Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Cy5 is a trademark
of GE Healthcare.
thermofisher.com/support | thermofisher.com/askaquestion
thermofisher.com
3 April 2020
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Thermo Fisher Scientific QuantStudio 3 and 5 Calibration and Performance Check Record Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI