Thermo Fisher Scientific Plus/Minus Assay: Applied Biosystems 7900HT Fast Real-Time PCR Systems Owner's manual

  • Hello! I am an AI chatbot trained to assist you with the Thermo Fisher Scientific Plus/Minus Assay: Applied Biosystems 7900HT Fast Real-Time PCR Systems Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Introduction
Designing a
Plus/Minus Assay
Experiment
Preparing the
Samples and
Reaction Plate
Performing a
Plus/Minus Assay
Pre-Read Run
Performing an
Amplification
Run
Performing and
Analyzing a
Plus/Minus Assay
Post-Read Run
Applied Biosystems 7900HT Fast
Real-Time PCR System
Plus/Minus Assay
Getting Started Guide
© Copyright 2007, 2010 Applied Biosystems. All rights reserved.
For Research Use Only. Not for use in diagnostic procedures.
Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that may appear in this
document. This document is believed to be complete and accurate at the time of publication. In no event shall Applied Biosystems be liable for incidental,
special, multiple, or consequential damages in connection with or arising from the use of this document.
Notice to Purchaser
The Applied Biosystems 7900HT Fast Real-Time PCR System is a real-time thermal cycler covered by US patents and corresponding claims in their non-US
counterparts, owned by Applied Biosystems. No right is conveyed expressly, by implication or by estoppel under any other patent claim, such as claims to
apparatus, reagents, kits, or methods such as 5’ nuclease methods. Further information on purchasing licenses may be obtained by contacting the Director of
Licensing, Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.
TRADEMARKS:
ABI PRISM, Applied Biosystems, MicroAmp, NucPrep, PrepMan, Primer Express and VIC are registered trademarks and AB (Design), Applera, BloodPrep,
Celera Genomics, FAM, ROX and TAMRA are trademarks of Applied Biosystems or its subsidiaries in the U.S. and/or certain other countries.
TaqMan is a registered trademark of Roche Molecular Systems, Inc.
All other trademarks are the sole property of their respective owners.
Part Number 4364017 Rev. D
06/2010
Contents
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System iii
Preface v
How to Use This Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
How to Obtain More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
How to Obtain Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Chapter 1 Introduction 1
About the 7900HT Fast System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
About Plus/Minus Assays Using an IPC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
About Plus/Minus Assay Experiments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Before You Begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 2 Designing a Plus/Minus Assay Experiment 9
Using TaqMan Probe-Based Chemistry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Selecting the Probes and Primers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Chapter 3 Preparing the Samples and Reaction Plate 13
Preparing the DNA Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Preparing the Reaction Mix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Preparing the Reaction Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Chapter 4 Performing a Plus/Minus Assay Pre-Read Run 19
About Plus/Minus Assay Plate Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Creating a Plus/Minus Assay Plate Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Performing the Pre-Read Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Chapter 5 Performing an Amplification Run 33
About Standard Curve (AQ) Plate Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Creating a Standard Curve (AQ) Plate Document . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Performing the Amplification Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
iv Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Chapter 6 Performing and Analyzing a Plus/Minus Assay Post-Read
Run 47
Performing the Post-Read Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Analyzing the Run and Evaluating the Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Post-Analysis Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Appendix A Sample Plus/Minus Assay Experiment 69
About the Sample Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Plus/Minus Assay Experiment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Appendix B SDS Automation Controller Software 75
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Using the Automation Controller Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Appendix C Analyzing and Viewing Amplification Data 81
Terms Used in Quantitation Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Analyzing the Amplification Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Viewing the Amplification Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
References 93
Index 95
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System v
Preface
How to Use This Guide
Purpose of This
Guide
This guide provides procedures for conducting plus/minus assays using the Applied
Biosystems 7900HT Fast Real Time PCR System (7900HT Fast System).
Audience This guide is intended for principal investigators and laboratory staff who conduct plus/
minus assays using the 7900HT Fast System.
Assumptions This guide assumes that you have:
Familiarity with the Microsoft® Windows® XP operating system.
Knowledge of general techniques for handling DNA samples and preparing them
for PCR.
A general understanding of hard drives and data storage, file transfers, and copying
and pasting.
Networking experience, if you want to integrate the 7900HT Fast System into your
existing laboratory data flow.
Text Conventions This guide uses the following conventions:
Bold indicates user action. For example:
Type 0, then press Enter for each of the remaining fields.
Italic text indicates new or important words and is also used for emphasis. For
example:
Before analyzing, always prepare fresh matrix.
A right arrow bracket (>) separates successive commands you select from a drop-
down or shortcut menu. For example:
Select File > Open.
User Attention
Words
The following user attention words appear in Applied Biosystems user documentation.
Each word implies a particular level of observation or action as described below:
Note – Provides information that may be of interest or help but is not critical to the use
of the product.
Preface
How to Use This Guide
vi Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
IMPORTANT! – Provides information that is necessary for proper instrument operation,
accurate chemistry kit use, or safe use of a chemical.
Indicates a potentially hazardous situation that, if not avoided, may
result in minor or moderate injury. It may also be used to alert against unsafe practices.
Indicates a potentially hazardous situation that, if not avoided, could
result in death or serious injury.
Safety Follow specific safety practices when using this instrument.
For safety guidelines, refer to the "Safety and EMC Compliance" section in the
Applied Biosystems 7900HT Fast Real-Time PCR System Site Preparation Guide
(PN 4351923) and the Sequence Detection Systems Software version 2.3 Online Help
(SDS Online Help).
1. You can obtain from Applied Biosystems the MSDS for any chemical supplied by
Applied Biosystems. This service is free and available 24 hours a day. To obtain
MSDSs:
2. Go to www.appliedbiosystems.com, click Support, then click MSDS Search.
3. In the Keyword Search field, enter the chemical name, product name, MSDS part
number, or other information that appears in the MSDS of interest. Select the
language of your choice, then click Search.
4. Find the document of interest, right-click the document title, then select any of the
following:
Open – To view the document
Print Target – To print the document
Save Target AsTo download a PDF version of the document to a destination
that you choose
Note: For the MSDSs of chemicals not distributed by Applied Biosystems, contact the
chemical manufacturer.
5. To have a copy of a document sent by fax or e-mail, select Fax or Email to the left
of the document title in the Search Results page, then click RETRIEVE
DOCUMENTS at the end of the document list.
6. After you enter the required information, click View/Deliver Selected Documents
Now.
For chemicals not manufactured or distributed by Applied Biosystems, contact the
chemical manufacturer.
Preface
How to Obtain More Information
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System vii
How to Obtain More Information
Related
Documentation
For more information about using the 7900HT Fast System and performing assays, refer
to:
Applied Biosystems 7900HT Fast Real-Time PCR System Plus/Minus Assay Getting
Started Guide (PN 4364017)
Applied Biosystems 7900HT Fast Real-Time PCR System Absolute Quantitation
using Standard Curve Getting Started Guide (PN 4364014)
Applied Biosystems 7900HT Fast Real-Time PCR System Allelic Discrimination
Getting Started Guide (PN 4364015)
Applied Biosystems 7900HT Fast Real-Time PCR System Relative Quantitation
using Comparative CT Getting Started Guide (PN 4364016)
Sequence Detection Systems Software version 2.3 Online Help (SDS Online Help)
Applied Biosystems 7900HT Fast Real-Time PCR System Maintenance and
Troubleshooting Guide (PN 4365542)
Applied Biosystems 7900HT Fast Real-Time PCR System Site Preparation Guide
(PN 4351923)
ABI PRISM® 6100 Nucleic Acid PrepStation Users Guide (PN 4326242)
DNA Isolation from Fresh and Frozen Blood, Tissue Culture Cells, and Buccal
Swabs Protocol (using BloodPrep Chemistry, PN 4343586)
NucPrep® Chemistry Protocol: Isolation of Genomic DNA from Animal and Plant
Tissue (PN 4333959)
PrepMan® Ultra Sample Preparation Reagent Protocol (PN 4318925)
Primer Express® Software v3.0 Getting Started Guide (PN 4362460)
Real-Time PCR Systems Chemistry Guide (PN 4348358)
•TaqMan
® Exogenous Internal Positive Control Reagents Protocol (PN 4308335)
•TaqMan
® Low Density Array Getting Started Guide (PN 4319399)
•TaqMan
® Universal PCR Master Mix Protocol (PN 4304449)
TransPrep Chemistry Protocol: Purification of gDNA from Filtrates Obtained After
the Isolation of RNA from Homogenized Animal or Plant Tissue Samples
(PN 4326965)
Send Us Your
Comments
Applied Biosystems welcomes your comments and suggestions for improving its user
documents. You can e-mail your comments to:
How to Obtain Support
To contact Applied Biosystems Technical Support from North America by telephone,
call 1.800.899.5858.
For the latest services and support information for all locations, go to http://
www.appliedbiosystems.com, then click the link for Support.
Preface
How to Obtain Support
viii Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
At the Support page, you can:
Obtain worldwide telephone and fax numbers to contact Applied Biosystems
Technical Support and Sales facilities
Search through frequently asked questions (FAQs)
Submit a question directly to Technical Support
Order Applied Biosystems user documents, MSDSs, certificates of analysis, and
other related documents
Download PDF documents
Obtain information about customer training
Download software updates and patches
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System 1
Notes
1
Chapter 1
Introduction
About the
7900HT Fast System
About Plus/Minus
Assays Using an IPC
About
Plus/Minus Assay
Experiments
Before You Begin
Introduction
P
er
f
orm
i
n
g
an
d
Analyzing a
P
lus/Minus Assa
y
P
ost-
R
ea
d
R
un
D
esi
g
nin
g
a
P
lus
/
Minus Assay
E
x
p
erimen
t
P
reparin
g
the
S
am
p
les and
Reac
ti
o
n Pl
a
t
e
P
er
f
orm
i
ng t
h
e
P
lus/Minus Assa
y
re-
ea
u
P
erformin
g
an
A
mp
lifi
cat
i
on
R
u
n
See page 4
See page 7
See page 2
See page 3
Chapter 1 Introduction
About the 7900HT Fast System
2Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Notes
About the 7900HT Fast System
System
Description
The Applied Biosystems 7900HT Fast Real Time PCR System (7900HT Fast System)
uses fluorescent-based PCR chemistries to provide:
Quantitative detection of nucleic acid sequences using real-time analysis.
Qualitative detection of nucleic acid sequences using end-point and dissociation-
curve analysis.
For more information about the 7900HT Fast System, refer to the Sequence Detection
Systems Software version 2.3 Online Help (SDS Online Help) and the
Applied Biosystems 7900HT Fast Real-Time PCR System Maintenance and
Troubleshooting Guide (PN 4365542).
Note: To access the SDS Online Help, select Help > SDS Online Help from the SDS
software menu bar.
Supported Assay
Configurations
You can perform several assay types on the 7900HT Fast System using reaction plates in
the 96-well, 384-well and TaqMan® Low Density Array format.This guide describes
only the plus/minus assay type.
For information about the other supported assay configurations for the 7900HT Fast
System, refer to the following table.
Plus/Minus Assay
Configuration
Plus/minus assays are supported on the 7900HT Fast System using reaction plates in the
standard 96-well or standard 384-well format with standard reagents and protocols.
IMPORTANT! Plus/minus assays are not supported using Fast reaction plates, reagents
and protocols.
IMPORTANT! Use standard reaction plates on the 7900HT Fast System with a standard
block. Fast 96-well reaction plates do not fit into the standard 96-well block correctly
and standard 96-well reaction plates do not fit into the fast 96-well block.
Reaction Plate and System Block Options
Assay Type Fast
96-well
Standard
96-well
Standard
384-well
Ta q M a n ®
Low Density Array
Standard curve (AQ) Yes Yes Yes No
Comparative CT (RQ) Yes Yes Yes Yes
Allelic discrimination No Yes Yes No
Plus/minus No Yes Yes No
Chapter 1 Introduction
About Plus/Minus Assays Using an IPC
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System 3
Notes
1
About Plus/Minus Assays Using an IPC
Plus/Minus Assay
Definition
A plus/minus assay is an end-point assay that determines if a specific target sequence is
present (plus) or not present (minus) in a sample. In an end-point assay, data are
collected at the end of the PCR process.
What is an IPC? An internal positive control (IPC; see TaqMan® Exogenous Internal Positive Control
Reagents kit, PN 4308323) is used in plus/minus assays to monitor the PCR process and
to ensure that a negative result is not due to failed PCR in the sample (Kwok and
Higuchi, 1989). The IPC consists of a template, a primer set, and a VIC® dye-labeled
probe that are added to each well of a reaction plate (the IPC is part of the reaction mix;
see “Preparing the Reaction Mix” on page 15).
Plus/minus assays with an IPC use fluorogenic 5 nuclease chemistry (also known as
“TaqMan® probe-based chemistry”; see “Using TaqMan Probe-Based Chemistry” on
page 10). During amplification, the sample target and the IPC target generate reporter
fluorescence signals such that positive or negative calls can be made on unknown
samples (Saiki et al., 1985).
Note: Plus/minus assays can also be accomplished without the use of an IPC; however,
an IPC ensures that a failed PCR reaction is not mistaken for a negative target result.
Note: The SYBR® Green I dye-based chemistry and the TaqMan® Fast reagent-based
chemistry are not supported for plus/minus assays.
Terms Used in
Plus/Minus Assay
Analysis Term Definition
Internal positive control (IPC) A second TaqMan probe and primer set
added to the plate to identify well failure to
amplify. Provides a means of determining
amplification failure that might give a false
negative signal.
No amplification control (NAC) Wells that contain no target template and
blocked IPC (the IPC target template has
been blocked by a blocking agent).
No template control (NTC) A sample that does not contain template.
Shows background signal and is used as the
negative control. Provides a means of
measuring contamination that might give a
false positive signal.
Nucleic acid target (also called “target
template” or “target”)
Nucleotide sequence that you want to
identify as present or absent.
Unknown sample (U; also called “sample of
interest”)
The sample for which you want to determine
the presence or absence of a specific target.
Chapter 1 Introduction
About Plus/Minus Assay Experiments
4Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Notes
About Plus/Minus Assay Experiments
Plus/Minus Assay
Workflow
After you design a plus/minus assay experiment and prepare the DNA samples, you
perform three runs:
A pre-read run using a plus/minus assay plate document to determine the baseline
fluorescence associated with primers and probes before amplification.
An amplification run using a standard curve (AQ) plate document to generate real-
time PCR data, which can be used to analyze and troubleshoot the PCR data for the
plus/minus assay, if needed.
A post-read run using the original plus/minus assay plate document, which
automatically subtracts the baseline fluorescence determined during the pre-read
run from the post-read amplified data to calculate the result.
The following figure illustrates the complete process.
Sample Plus/
Minus Assay
Experiment
To illustrate how to design, perform, and analyze plus/minus assay experiments, a
sample experiment representing a typical plus/minus assay experiment is described in
Appendix A on page 69. References to parts of the sample experiment are made in the
subsequent chapters (2 to 6) of this guide where applicable. You can use the summarized
procedures of the sample plus/minus assay experiment in Appendix A on page 69 to
familiarize yourself with the entire plus/minus assay workflow.
Prepare
the Samples
Perform an
Amplification Run
Perform and Analyze
a Plus/Minus Post-Read Run
Perform a Plus/Minus
Pre-Read Run
Chapter 1 Introduction
About Plus/Minus Assay Experiments
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System 5
Notes
1
Required
User-Supplied
Materials
Chemistry/Reagents
Item Source
DNA isolation and purification chemistry systems:
•ABIP
RISM® 6100 Nucleic Acid PrepStation
• BloodPrep Chemistry (genomic DNA from
fresh or frozen blood or cells)
•NucPrep
® Chemistry (DNA from animal and
plant tissue)
•PrepMan
® Ultra Sample Preparation Reagent
Kit
•ABIP
RISM® TransPrep System (purification of
gDNA after the isolation of RNA from animal
and plant tissue)
Applied Biosystems (PN 6100-01)
Applied Biosystems (PN 4346860)
Applied Biosystems (PN 4330274)
Applied Biosystems (PN 4322547)
Applied Biosystems web site
Labeled primers and probes source:
Primer Express® Software (custom-designed
primers and probes)
Contact your Applied Biosystems sales
representative.
Ta q M a n ® reagents:
•TaqMan
® Exogenous Internal Positive Control
Reagents (VIC® dye-labeled probe)
•TaqMan
® Universal PCR Master Mix (2X)
Applied Biosystems (PN 4308323)
Applied Biosystems (PN 4304437)
Chapter 1 Introduction
About Plus/Minus Assay Experiments
6Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Notes
Reaction Plates and Covers
IMPORTANT! Do not use MicroAmp® Caps (domed) or Optical Tubes with the 7900HT Fast
System. You can use Optical Caps (PN 4323032) only with standard 96-well plates on the
7900HT Fast System.
Item Source
Standard 96-well reaction plates:
•MicroAmp
® 96-Well Optical Reaction Plate
with Barcode (code 128), 20 Plates/Pkg
•MicroAmp
® 96-Well Optical Reaction Plate
with Barcode (code 128), 500 Plates/Pkg
Applied Biosystems (PN 4306737)
Applied Biosystems (PN 4326659)
Standard 384-well reaction plates:
384-Well Clear Optical Reaction Plate with
Barcode (code 128), 50 Plates/Pkg
384-Well Clear Optical Reaction Plate with
Barcode (code 128), 500 Plates/Pkg
Applied Biosystems (PN 4309849)
Applied Biosystems (PN 4326270)
Optical adhesive covers:
•ABIP
RISM® Optical Adhesive Cover Starter Kit
Includes 20 ABI PRISM® Optical Adhesive
Covers, an Applicator, and an ABI PRISM®
Optical Cover Compression Pad
•ABIP
RISM® Optical Adhesive Covers,
100 Covers/Pkg
•ABIPRISM® Optical Adhesive Covers,
25 Covers/Pkg
Optical Caps,
8 caps/strip, 2400 caps/300 strips
Applied Biosystems (PN 4313663)
Applied Biosystems (PN 4311971)
Applied Biosystems (PN 4360954)
Applied Biosystems (PN 4323032)
Other Consumables and Equipment
Item Source
Centrifuge with adapter for 96-well and
384-well plates
Major Laboratory Supplier (MLS)
Gloves MLS
Microcentrifuge MLS
Microcentrifuge tubes, sterile 1.5-mL MLS
Pipette tips, with filter plugs MLS
Pipettors, positive-displacement MLS
Vortexer MLS
Nuclease-free water MLS
Tris-EDTA (TE) Buffer, pH 8.0 MLS
Chapter 1 Introduction
Before You Begin
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System 7
Notes
1
Before You Begin
Calibrating the
7900HT Fast
System
Check that background and pure-dye runs have been performed regularly for optimal
performance of the 7900HT Fast System. For information about calibrating the 7900HT
Fast System, refer to the Sequence Detection Systems Software version 2.3 Online Help
(SDS Online Help) and the Applied Biosystems 7900HT Fast Real-Time PCR System
Maintenance and Troubleshooting Guide (PN 4365542).
Accessing the
SDS Online Help
Some steps in this guide refer you to the SDS Online Help for more information. To
access the SDS Online Help, select Help > SDS Online Help from the SDS software
menu bar.
Automation
Options
The 7900HT Fast System can run prepared reaction plates individually or in groups
using the Automation Accessory with the Zymark® Twister Microplate Handler. If you
are not using the Automation Accessory, you must run reaction plates individually. For
clarity, this guide illustrates only running an individual reaction plate using the SDS
Software for the 7900HT Fast System.
For information on running multiple reaction plates using the Automation Controller
Software, refer to Appendix B on page 75. For information on automated operation of
the 7900HT Fast System using the Automation Accessory, see the SDS Online Help.
Chapter 1 Introduction
Before You Begin
8Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Notes
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System 9
Notes
Chapter 2
2
Designing a Plus/Minus Assay
Experiment
Using TaqMan®
Probe-Based Chemistry
Selecting the
Probes and Primers
Designing a
Plus/Minus Assay
Experiment
P
er
f
orm
i
ng an
d
Analyzing a
P
lus/Minus Assa
y
P
ost-
R
ea
d
R
un
P
erforming the
P
lus/Minus Assa
y
re-
ea
u
P
erformin
g
an
A
mp
lifi
cat
i
on
R
u
n
P
repar
i
n
g
t
h
e
S
am
p
les and
Reac
ti
o
n Pl
a
t
e
I
ntr
oduc
ti
on
See page 10
See page 11
Chapter 2 Designing a Plus/Minus Assay Experiment
Using TaqMan Probe-Based Chemistry
10 Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Notes
Using TaqMan Probe-Based Chemistry
About the
Chemistry
Plus/minus assays with an IPC use the fluorogenic 5 nuclease chemistry (also known as
TaqMan® probe-based chemistry).
Note: The SYBR® Green I dye-based chemistry and the TaqMan® Fast reagent-based
chemistry are not supported for plus/minus assays.
Two Ty p e s o f
TaqMan Probes
Applied Biosystems offers two types of TaqMan® probes:
TaqMan probes with TAMRA dye as quencher
TaqMan MGB (minor groove-binder) probes with non-fluorescent quencher (NFQ)
For more information about the TaqMan probe-based chemistry, refer to the Real-Time
PCR Systems Chemistry Guide (PN 4348358).
Description Process
Taq Man ® chemistry
TaqMan probe-based
chemistry uses a
fluorogenic probe to detect
a specific PCR product as it
accumulates during PCR
cycles (Mullis and Faloona,
1987).
This process, outlined in
the following figures, is
the basis for running
plus/minus assays on the
7900HT Fast System.
3
5
5
53
3
FORWARD
PRIMER PROBE
R = REPORTER
Q = QUENCHER
REVERSE
PRIMER
QR
3
5
5
53
3
5
Q
R
Step 1: Polymerization
A reporter (R) and a quencher (Q)
are attached to the 5' and 3' ends
of a TaqMan probe.
Step 2: Strand Displacement
When both dyes are attached
to the probe, reporter dye
emission is quenched.
3
5
5
53
3
5
Q
R
3
5
5
53
3
5
Q
R
Step 3: Cleavage
During each extension cycle, the
Applied Biosystems hot-start
DNA polymerase system cleaves
the reporter dye from the probe.
Step 4: Polymerization Completed
After being separated from the
quencher, the reporter dye emits
its characteristic fluorescence.
5
Chapter 2 Designing a Plus/Minus Assay Experiment
Selecting the Probes and Primers
Plus/Minus Assay Getting Started Guide for the 7900HT Fast System 11
Notes
2
Chemistry Kits for
Plus/Minus
Assays
The following reagents are available from Applied Biosystems for designing and
running plus/minus assays:
•TaqMan
® Exogenous Internal Positive Control Reagents with VIC®–TAMRA
dye-labeled probe (PN 4308323)
•TaqMan
® Universal PCR Master Mix (2X) (PN 4304437)
Note: The IPC DNA, primers, and probe supplied in these reagents can be used with all
target samples. Refer to the TaqMan® Universal PCR Master Mix Protocol
(PN 4304449) for instructions on optimizing amplification of your target.
Selecting the Probes and Primers
Design and select a probe and primer set for your target sequence(s) using
Applied Biosystems Primer Express® software. For more information about using this
software, refer to the Primer Express® Software v3.0 Getting Started Guide
(PN 4362460).
Sample Experiment
For the sample experiment, we extracted DNA from 84 batches of ground beef to test for the presence of E. coli using the
plus/minus assay on the 7900HT Fast System. We ran six no IPC/no target template controls, six IPC/no target template
controls, and 84 unknown samples.
The sample experiment used the TaqMan® Exogenous Internal Positive Control Reagents Kit, which is supplied with one
1-mL tube of 10 Exo IPC Mix. This mix contains the IPC primers and VIC® dye-labeled probe. We designed the primers/
probe set for E. coli, containing a FAM dye-labeled probe with TAMRA as the quencher, using Applied Biosystems
Primer Express® Software.
Chapter 2 Designing a Plus/Minus Assay Experiment
Selecting the Probes and Primers
12 Plus/Minus Assay Getting Started Guide for the 7900HT Fast System
Notes
/