I
NSTALLATION
2-2
September 2000 Rosemount Analytical 748262-LModel 400A Hydrocarbon Analyzer
WARNING: INSTALLATION RESTRICTIONS
This instrument was shipped from the factory configured to operate on 115 VAC, 50/60
Hz electric power. Verify that the power source conforms to the requirements of the
individual instrument, as noted on the name-rating plate.
Note:
230 VAC operation requires an accessory transformer mounted external to the
instrument. If external 230 VAC power is provided, do not change the switch
setting on the Temperature Control Board, which is factory-selected for 115
VAC. A 230 VAC switch setting will cause the case temperature controller to
malfunction.
2.4 FUEL AND AIR REQUIREMENTS
Fuel, air and calibration gas cylinders are under pressure. Mishandling of gas
cylinders could result in death, injury or property damage. Handle and store
cylinders with extreme caution and in accordance with manufacturer's
instructions. Refer to General Precautions for Handling and Storing Gas
Cylinders, which follows Section 7 of this manual.
During normal operation, the analyzer uses fuel and air to maintain the burner flame.
Criteria for selection of these gases are given in Sections 2.4.1 and 2.4.2. In addition,
the analyzer requires suitable standard gas(es) for calibration. Refer to Section 3.1.1.
Each gas used should be supplied from a tank or cylinder equipped with a clean,
hydrocarbon-free, two-stage regulator. In addition, a shutoff valve is desirable. Install
the gas cylinders in an area of relatively constant ambient temperature.
2.4.1 F
UEL
G
AS
The standard analyzer is equipped to use only mixed fuel, i.e. 40% hydrogen/60%
nitrogen or helium. Such blends are supplied by many gas vendors specifically for this
use, with a guaranteed maximum total hydrocarbon content of 0.5 ppm, measured as
methane (THC < 0.5 ppm). This specification should be used when buying such
mixtures.
When the analyzer is equipped with the optional hydrogen fuel kit, P/N 622576, 100%
hydrogen fuel is to be used. This is also supplied by many gas vendors specifically for
this use, with the same guaranteed total hydrocarbon content (THC < 0.5 ppm) which,
again, should be specified when buying the gas.
Note:
Always assure the sample flow is present when using the 100% hydrogen fuel
option. Absence of sample flow can result in burning of detector tip when using
100% hydrogen.