CEM EDGE Nitrogen Option User guide

Type
User guide
EDGE Nitrogen Accessory Page 1 of 2
2021-01
www.cem.com © 2020 CEM Corporation
All rights reserved. This may not be reproduced or published without written permission from CEM.
Nitrogen Accessory
The Nitrogen accessory for EDGE systems allows nitrogen gas to be connected and regulated into the EDGE
to replace ambient air during the extraction, purge and agitation. This accessory can be purchased with an
EDGE or added onto any existing EDGE. For a preexisting EDGE, an outside cover with a nitrogen fitting must
be present as shown in the image below. If this fitting is not present, please contact your local sales
representative, and this fitting can be installed on your current EDGE as well as the Nitrogen accessory.
The Nitrogen accessory will include:
Nitrogen accessory
18tubing to connect Nitrogen accessory to EDGE
Items not included:
0.25OD polypropylene tubing to connect accessory to gas source
EDGE Nitrogen Accessory Page 2 of 2
2021-01
© 2020 CEM Corporation
Installation
1. The EDGE should be unplugged without a solvent or waste bottle attached.
2. Place the EDGE so you can access the nitrogen fitting on the clear cover.
3. Connect the assembly to the EDGE using the supplied 0.25tubing by pressing it into the push-to-
connect fitting as shown in the image below. Ensure the tubing is pushed in fully by pulling on the
tubing once installed. If it does not move, it is locked in place.
4. Connect the other side of the tubing to the Nitrogen accessory.
5. Connect the Nitrogen accessory to the nitrogen source using polypropylene tubing with an OD of 0.25”.
Note: 0.25polypropylene tubing is required to connect the nitrogen accessory to the nitrogen
source.
6. Turn the gas on when ready to run samples.
Note: The gas must be turned on in order for the EDGE to function.
7. Run your EDGE as normal, and turn the gas off at the source to prevent gas from escaping when
EDGE is idle.
Caution
A regulator that restricts the pressure/flow to no more than 100 psi entering the box is required. Anything
above this flow will cause damage to the EDGE and Nitrogen accessory.