OPERATION AND CYLINDER CHANGE OVER
1. START UP OF MANIFOLD
When the unit is completely installed and all shut o valves are closed.
Slowly open gas cylinder shut-o valves.
Slowly open shut-o valves V1, V2 and V3.
Open purge valve V5.
Slowly open main-shut-o valve V4 to ventilate the unwanted “air” from the collecting unit and
high pressure hoses and close it back.
Close purge valve V5. The purge valve V5 must be closed during operation.
You should keep at least one of the cylinders closed to be used as reserve one. Close shut-o
valve V1, V2 and/or V3 for concerned cylinder(s) and concerned gas cylinder shut-o valve.
Assign cylinder with indication panels (Fig. 4).
To check if there is enough gas in the cylinders, use inlet gauge M1.
Check gauge M2 if the outlet pressure is as required.
2. CYLINDER CHANGE OVER
When 3 gas cylinders are connected to the collecting unit C1 and the one which is in use indicates
low pressure,the following steps are to be taken:
SLOWLY open full cylinder and its corresponding shut o valve on the collecting unit.
Close empty cylinder and the corresponding shut-o valve.
The gas cylinder indication panels must be now adjusted so that it is clearly apparent which
cylinder is in use.
Exchange the empty cylinder(s) for a full one.
Close shut-o valves (V1 and/or V2 and/or V3 and/or corresponding to full gas cylinders or cylinders
you do not wish to exchange. The gas flow supply has now been stopped.
Open shut-o valve V4.
Close cylinder shut o valve of the empty cylinder(s) that you wish to exchange.
Open purge valve V5 to evacuate remaining gas from the collecting unit and high pressure hoses.
Close the purge valve V5 when all gas is evacuated.
Close the opened shut-o valves (V1 and/or V2 and/or V3 and V4).
Exchange empty cylinder(s) with full one.
Ventilate the unwanted gas as described in chapter 7.1 from the changed cylinders.
Then continue using the gas supply from the previously used cylinders. The indication panels
must always match the cylinder(s) in use to prevent unexpected running out of gas.
When 3 gas cylinder are connected to the collecting unit C1 and all 3 are empty, the following
steps are to be taken:
Open main shut o valve V4 .
Close cylinder shut o valves.
Open purge valve V5 to evacuate remaining gas from the collecting unit and high pressure hoses.
Close the purge valve V5 when all gas is evacuated.
Close the shut o valves V1, V2, V3 and V4.
Exchange empty cylinders for full ones.
Now follow steps from chapter 7.1 - start up the manifold.
3. SHUT DOWN
The manifold can be only shut down with agreement of a responsible person according to standard
EN ISO 7396-1.
For short term (hours) shutdown close shut-o valve V4.
For long term (days) shutdown close all opened shut-o valves on the unit (V1 - V4)and the cylinder
shut-o valves
CYLINDER SHUT-OFF VALVES
V1 V2 V3
V5V4M1
M2 C1
MEDICAL GAS
INDICATION PANELS
INDICATION OF GAS TYPE
KEEP THE PRODUCT AND ITS ASSOCIATED EQUIPMENT AWAY FROM:
heat sources (fire, cigarettes, ...)
flammable materials,
oil or grease (especially be careful if hand cream is used),
water,
dust.
The product and its associated equipment must be prevented from falling.
Always maintain oxygen cleanliness standards,
Use only the product and its associated equipment in a well ventilated area.
It’s very important to assure that the device can be only accessed by a qualified
person – possibility of misuse.
Doc. Nr. 735800000296; Date: 2018-07-16; Rev. 01; TI: A3, BAR, LAM
IN CASE OF EMERGENCY CALL: