1-8 Synergy Safety Guidelines 03/2002
Applied Biosystems
Laboratory Ventilation Requirements
The information presented here reflects U.S. regulations and practices for venting
all Applied Biosystems, Inc. instruments to a fume hood or to a duct.
WARNING
POTENTIAL FOR EXPOSURE TO HAZARDOUS WASTES
Do
not operate a vented instrument unless it is connected in
accordance with all the ventilation requirements presented
here. The wastes produced by certain chemicals in Applied
Biosystems instruments are hazardous and can cause injury,
illness, or death. Always wear protective clothing when
handling hazardous waste. Mix and prepare hazardous
materials beneath a fume hood. Read the Waste Profile in
Appendix 2.
Fume Hood
Always mix and prepare hazardous materials beneath a fume hood. The vent line to
the fume hood should be no longer than 15 feet (5 meters). Be certain that the line
is not punctured or otherwise damaged.
Following are important points about the fume hood:
• It should operate at all times, including nights and weekends, when vented
waste bottle contents can escape to surroundings.
• It is to be located away from air currents generated by air conditioning ducts,
fans, windows, doors, and moving equipment and persons.
• It should exhibit a sign or label indicating where doors and windows must be
positioned to give an average flow of 100 ft/min (linear) face velocity. The
minimum velocity of flow at any point in the hood must be 80 ft/min (linear);
the maximum flow velocity must not exceed 125 ft/min.
• It must meet local, state, and federal health and safety requirements. In
addition, see current fume hood standards established by the American Society
of Heating, Refrigeration, Air Conditioning Engineers (ASHRAE); American
Conference of Governmental Industrial Hygienists (ACGIH); and
Occupational Safety and Health Agency (OSHA).
• It must be constructed of materials that are compatible with the waste
materials/chemicals being generated/exhausted.
Figure 1-3 shows a fume hood configuration with movement of gaseous waste
through the hood into a duct (the duct is described in the following section).