ALTAIR® 4XR
Multigas Detector
Order No.: 10175896/08
Print Spec: 10000005389 (EO)
CR: 800000057847
MSAsafety.com
US
© MSA 2022, 2023. All rights reserved
WARNING!
Read this manual carefully before using or maintaining the device. The device will perform as designed only if it is used and
maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform as designed, and persons
who rely on this device could sustain serious injury or death.
The warranties made by MSA with respect to the product are voided if the product is not installed and used in accordance with
the instructions in this manual. Please protect yourself and your employees by following the instructions.
Please read and observe the WARNINGS and CAUTIONS inside. For additional information relative to use or repair, call 1-
800-MSA-2222 during regular working hours.
For countries of Russian Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector will be delivered with a
passport document that includes valid approval information. On the CD with manual instruction attached to the gas detector the
user will find the documents "Type Description" and "Test Method" - appendixes to Pattern Approval Certificate of Measuring
instrument, valid in the countries of use.
MSA is a registered trademark of MSA Technology, LLC in the US, Europe and other Countries. For all other trademarks visit
https://us.msasafety.com/Trademarks.
This product incorporates Bluetooth® wireless technology. The Bluetooth word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc., and any use of such marks by MSA is under license. Other trademarks and trade names are
those of their respective owners. Versions of this product manufactured July 2022 and later may not contain Bluetooth wireless
technology. This will be indicated by the front case of the detector not utilizing a Bluetooth logo. All references in this manual to
Bluetooth will not pertain to these versions of the device.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may
not cause harmful interference, and (2) this device must accept any interference received, including interference that may
cause undesired operation.
You are cautioned that changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.
English:
This device complies with RSS-210 of the Industry Canada Rules. Operation is subject to the following two conditions: (1) This
device may not cause harmful interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
French:
Ce dispositif est conforme à la norme CNR-210 d'Industrie Canada applicable aux appareils radio exempts de licence. Son
fonctionnement est sujet aux deux conditions suivantes: (1) le dispositif ne doit pas produire de brouillage préjudiciable, et (2)
ce dispositif doit accepter tout brouillage reçu, y compris un brouillage susceptible de provoquer un fonctionnement
indésirable.
1000 Cranberry Woods Drive
Cranberry Township, PA 16066
USA
Phone 1-800-MSA-2222
For your local MSA contacts, please go to our website www.MSAsafety.com
US ALTAIR 4XR 3
Contents
1 Calibration Certification and Conformance 5
1.1 Calibration Certification 5
1.2 Conformance Statement 5
2 Safety Regulations 5
2.1 Correct Use 5
2.2 Liability Information 6
2.3 Safety and Precautionary Measures to be Adopted 6
2.4 Warranty 8
3 Description 10
3.1 Overview 10
3.2 Device Hardware Interfaces 10
3.3 On-Screen Indicators 12
3.4 Battery Care 13
3.5 Viewing Additional Pages 15
3.6 Sensor Missing Alarm 17
3.7 Sensor End of Life Warning 17
3.8 Sensor End of Life Indicator 17
3.9 Monitoring Toxic Gases 17
3.10 Monitoring Oxygen Concentration 18
3.11 Monitoring Combustible Gases 19
4 Operation 19
4.1 Environmental Factors 19
4.2 Turning ON and Fresh Air Setup 20
4.2.1 Fresh Air Setup (FAS) 23
4.3 Measurement Mode [Normal Operation] 24
4.4 Device Setup 24
4.5 Bluetooth Operation 28
4.6 Data Logging 30
4.7 Function Tests 30
4.8 Bump Test 30
4.9 Bump LED 31
4.10 Calibration 31
4.11 Time of Day Testing 34
5 Maintenance 34
5.1 Troubleshooting 35
5.2 Live Maintenance Procedure - Replacing and Adding a Sensor 35
5.3 Cleaning 37
5.4 Storage 37
5.5 Scope of Delivery 37
6 Technical Data 38
6.1 Factory-set Alarm Thresholds and Setpoints 39
6.2 Performance Specification 40
6.3 XCell Sensor Patents 43
7 Ordering Information 45
8 Appendix 47
8.1 Start Up Sequence (Power on) 47
8.2 Fresh Air Setup (FAS) 48
4 ALTAIR 4XR US
8.3 Reset Screen Controls 49
8.4 Bump Test 51
8.5 Options Setup 52
8.6 Sensor Setup 53
8.7 Calibrations 54
8.8 Alarm Setup 55
8.9 Time and Date Setup 56
8.10 Bluetooth Setup 56
1 Calibration Certification and Conformance
US ALTAIR 4XR 5
1 Calibration Certification and Conformance
1.1 Calibration Certification
All applicable inspections, testing, and calibrations were performed using NIST traceable equipment, where available, in
accordance with MSA's ISO 9001 Certified Quality System. Each material, component, and/or instrument must be
installed, operated and maintained in strict accordance with its labels, cautions, warnings, instructions, and within the
limitations stated in the supplied instruction manual. Routine calibration checks, equipment inspections, and applicable
preventative maintenance measures must be performed to verify that the materials, components, and/or instruments are
operating properly. Failure to perform these tasks on a routine basis, or suggested intervals, with specified equipment or
methods, may result in inaccurate readings.
1.2 Conformance Statement
MSA certifies that the materials, components, and/or instruments delivered in this shipment conform to all applicable
specifications. The items delivered have been processed through the appropriate approved document controlled
procedures for Receiving, Manufacturing and Inspection. The materials, components, and/or instruments were inspected,
tested, and calibrated, as applicable, per the associated drawings, standards requirements, and/or specifications, and
were deemed acceptable by appropriate authorized personnel.
2 Safety Regulations
2.1 Correct Use
The ALTAIR 4XR Multigas Detector is for use by trained and qualified personnel. It is designed to be used when performing
a hazard assessment to:
Assess potential worker exposure to combustible and toxic gases and vapors as well as low level of oxygen.
Determine the appropriate gas and vapor monitoring needed for a workplace.
The ALTAIR 4XR Multigas Detector can be equipped to detect:
Combustible gases and certain combustible vapors
Oxygen-deficient or oxygen-rich atmospheres
Oxygen for monitoring inertization applications. The device is suitable and certified for the measurement of the
oxygen concentration in gas mixtures for inertization according to EN50104 but without alarm function.
Specific toxic gases for which a sensor is installed.
NOTE: While the device can detect up to 30% oxygen in ambient air, it is approved for use only up to 21% oxygen.
The ATEX directive is only valid until 25% Vol O2.
It is imperative that this operating manual be read and observed when using the device. In particular, the safety
instructions, as well as the information for the use and operation of the device, must be carefully read and observed.
Furthermore, the national regulations applicable in the user's country must be taken into account for a safe use.
WARNING!
This product is supporting life and health. Inappropriate use, maintenance or servicing may affect the function of the
device and thereby seriously compromise the user's life.
Before use the product operability must be verified. The product must not be used if the function test is unsuccessful,
it is damaged, a competent servicing/maintenance has not been made, genuine MSA spare parts have not been
used.
Alternative use, or use outside this specification will be considered as non-compliance. This also applies especially
to unauthorized alterations to the product and to commissioning work that has not been carried out by MSA or
authorized persons.
Failure to follow these warnings can result in serious personal injury or death.
6 ALTAIR 4XR US
2 Safety Regulations
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
WARNING!
This is a class A product in accordance with CISPR 22. In a domestic environment, this product may cause radio
interference, in which case the user may be required to take adequate measures.
Failure to follow this warning can result in serious personal injury or death.
This Class A digital apparatus complies with Canadian ICES-003.
2.2 Liability Information
MSA accepts no liability in cases where the product has been used inappropriately or not as intended.
The selection and use of this product must be under the direction of a qualified safety professional who has carefully
evaluated the specific hazards of the jobsite where it will be used and who is completely familiar with the product and its
limitations. The selection and use of this product and its incorporation into the safety scheme of the jobsite is the exclusive
responsibility of the employer.
Changes and modifications not expressly approved by the manufacturer will void the user’s authority to operate the
equipment.
2.3 Safety and Precautionary Measures to be Adopted
WARNING!
Carefully review the following safety limitations and precautions before placing this device in service. Do not alter or modify
the device.
Failure to follow this warning can result in serious personal injury or death.
WARNING!
All device readings and information must be interpreted by someone trained and qualified in interpreting device readings in
relation to the specific environment, industrial practice and exposure limitations.
Failure to follow this warning can result in serious personal injury or death.
Check Function
Before each day’s use, check the function of the device (4.7 Function Tests). MSArecommends carrying out a routine
inspection prior to each day’s use.
Bluetooth Operation
If your device contains Bluetooth functionality, Bluetooth operation is dependent upon signal availability of the wireless
service(s) necessary to maintain the communication link. Loss of wireless signal will prevent communication of alarms and
other information to linked devices. Take appropriate precautions in the event a loss of wireless signal occurs.
Perform a Bump Test
2 Safety Regulations
US ALTAIR 4XR 7
Bump test frequency is often stipulated by national or corporate regulations; however, bump testing before each day's use
is generally the accepted best safety practice and is therefore MSA’s recommendation. The device must pass the bump
test. If it fails the test, perform a calibration before using the device.
Perform a bump test (see 4.8 Bump Test) more frequently if the device is subjected to physical shock or high levels of
contaminants. Also, perform a bump test more frequently if the tested atmosphere contains the following materials, which
may desensitize the combustible gas sensor and reduce its readings:
Organic silicones
• Silicates
Lead-containing compounds
Hydrogen sulphide exposures over 200ppm or exposures over 50ppm for oneminute.
Check minimum concentration of a combustible gas
The minimum concentration of a combustible gas in air that can ignite is defined as the Lower Explosive Limit (LEL). A
combustible gas reading of "XXX” indicates the atmosphere is above 100%LEL or 5.00% vol CH4, and an explosion
hazard exists. Move away from hazardous area immediately.
Observe atmosphere
Do not use the device to test for combustible or toxic gases in the following atmospheres as this may result in erroneous
readings:
Oxygen-deficient or oxygen-rich atmospheres
Reducing atmospheres
Furnace stacks
Inert environments
Atmospheres containing combustible airborne mists/dusts.
Use the device only to detect gases/vapors for which a sensor is installed.
The device is suitable and certified for the measurement of the oxygen concentration in gas mixtures for inertization
according to EN50104 but without alarm function.
Make sure >10% oxygen is present for accurate combustible readings with the catalytic sensor.
Not to be used for gases having a flash point in excess of 38°C (100°F)
Do not use the device to test for combustible gases in atmospheres containing vapors from liquids with a high flash point
(above 38°C, 100°F), as this may result in erroneously low readings.
Physical shock
Recheck calibration if device is subjected to physical shock.
Sensor maintenance
Do not block sensor openings as this may cause inaccurate readings. Do not press on the face of the sensors, as this may
damage them and cause erroneous readings. Do not use compressed air to clean the sensor holes, as the pressure may
damage the sensors.
Observe Proper Time for Display Stabilizing
Allow sufficient time for device to display accurate reading. Response times vary based on the type of sensor being
utilized.
Observe proper battery maintenance
Use only battery chargers made available by MSA for use with this device; other chargers may damage the battery pack
and the device. Dispose of in accordance with local health and safety regulations.
8 ALTAIR 4XR US
2 Safety Regulations
Use of the MSA GALAXY®GX2 Automated Test System is an alternate approved method for charging ALTAIR 4XR
devices.
As the battery pack ages, there will be a reduction in usable device runtime.
Risk of explosion: Do not recharge device in hazardous area.
Be Aware of Environmental Conditions
A number of environmental factors may affect the sensor readings, including changes in pressure, humidity and
temperature. Pressure and humidity changes also affect the amount of oxygen actually present in the atmosphere.
Be Aware of the Procedures for Handling Electrostatically Sensitive Electronics
The device contains electrostatically sensitive components. Do not open or repair the device without using appropriate
electrostatic discharge (ESD) protection. The warranty does not cover damage caused by electrostatic discharges.
Be Aware of the Product Regulations
Follow all relevant national regulations applicable in the country of use.
Be Aware of the Warranty Regulations
The warranties made by MSA The Safety Company with respect to the product are voided if the product is not used and
maintained in accordance with the instructions in this manual. Please protect yourself and others by following them. We
encourage our customers to write or call regarding this equipment prior to use or for any additional information relative to
use or service.
Special Conditions for Safe Usage
In the event of combustible sensor overrange, the device will enter a Lock Alarm state, which must be reset in a fresh
air environment. To reset this alarm, cycle power off and on, in fresh air. Keep the device in the fresh air environment
until LEL or CH4 readings have stabilized and then follow the Fresh Air Setup and Zero Calibration instructions
contained in this manual.
The RF radiation power used to activate the RFID tag antenna shall not exceed 6 Watts for EPL Group I applications
or 2 Watts for EPL Group IIC applications.
2.4 Warranty
ITEM WARRANTY PERIOD*
Chassis and electronics Four years
All sensors unless otherwise specified Four years
XCell EX-H Sensor One year
In-Box Accessories including
Replacements
Two years
*Period begins date of received shipment.
Specific battery run-time over temperature is not warranted.
This warranty does not cover filters, fuses, etc. As the battery pack ages, there will be a reduction in usable device run time.
Certain other accessories not specifically listed here may have different warranty periods. This warranty is valid only if the
product is maintained and used in accordance with Seller's instructions and/or recommendations.
The Seller shall be released from all obligations under this warranty in the event repairs or modifications are made by
persons other than its own or authorized service personnel or if the warranty claim results from physical abuse or misuse of
the product. No agent, employee or representative of the Seller has any authority to bind the Seller to any affirmation,
representation or warranty concerning this product. Seller makes no warranty concerning components or accessories not
manufactured by the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components.
2 Safety Regulations
US ALTAIR 4XR 9
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND IS
STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for breach of the above warranty, for any tortuous
conduct of Seller, or for any other cause of action, shall be the replacement at Seller's option, of any equipment or parts
thereof, which after examination by Seller is proven to be defective.
Replacement equipment and/or parts will be provided at no cost to Purchaser, F.O.B. Seller's Plant. Failure of Seller to
successfully replace any nonconforming equipment or parts shall not cause the remedy established hereby to fail of its
essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will seller be liable to purchaser for economic,
special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of
anticipated profits and any other loss caused by reason of non-operation of the goods. This exclusion is applicable to
claims for breach of warranty, tortuous conduct or any other cause of action against seller.
10 ALTAIR 4XR US
3 Description
3 Description
3.1 Overview
Figure 1 Device overview
1Datalink Communication
port 8 Display
2Bump LED (green/red) and
Fault LED (-yellow) 9 Alarm LEDs (4)
3 Sensor Inlets 10 Belt Clip
4 Horn 11 Charging Connection
5
p
Button 12 Screws (4)
6
q
Button 13 Charge LED
(red/green/orange)
7 Button 14 Bluetooth Status LED
The device monitors gases in ambient air and in the workplace.
It is available with a maximum of three sensors, which can display readings for four separate gases (one Two-Tox Sensor
can detect two toxic gases with a single sensor).
While the device can detect up to 30% oxygen in ambient air, it is approved for use only up to 21% oxygen.
The alarm levels for the individual gases are factory-set and can be changed through the Setup Menu. These changes can
also be made through the GALAXY GX2 Test Stand or MSA Link software. Ensure that the latest version of the GALAXY
GX2 firmware or MSA Link software has been downloaded from MSA’s website www.msasafety.com.
If gas is present during Fresh Air Setup, the device will fail and enters Normal Operation mode.
3.2 Device Hardware Interfaces
Device operation is dialog driven from the display with the aid of the three function buttons (3.2 Device Hardware
Interfaces).
The device has three buttons for user operation. Each button can function as a "soft key", as defined on the display directly
above the button.
3 Description
US ALTAIR 4XR 11
Button Definitions
Button Description
The button is used to turn device on or off and to confirm user action selections. When the
pbutton and the button are pressed simultaneously for device start-up, the Options Setup Mode
displays.
qThe qbutton is used to move forward through data screens in measuring mode, or as page back
and to decrease the values in set-up mode. Holding this button for 3seconds while in Normal
Operation mode activates the InstantAlert alarm.
pThe pbutton is used to reset peak, STEL TWA and acknowledge alarms (where possible) or
access calibration in measuring mode. It is also used as page up or to increase the values in set-up
mode.
LED Definitions
LED Description
GREEN/RED(Bump LED) After a successful Bump Test, the green LED will flash every 15 seconds for a 24-hour
period.
When the device fails a bump test or when the 24-hour period expires, the red LED
flashes every 15seconds.
This option can be turned off through the MSA GALAXY GX2 Test Stand or MSA Link
software.
RED(Alarm LED) The red alarm LEDs are visual indications of an alarm condition or any type of error in the
device.
YELLOW
(Fault LED)
The yellow fault LED is a visual indication of an device fault condition. This LED turns on
with the following conditions:
Device memory error
Sensor Missing
Sensor Error
RED/GREEN/ORANGE
(Charge LED)
The charge LED is a visual indication of charge status.
RED: device is charging
GREEN: charge is complete
ORANGE: problem detected during charging
BLUE (BluetoothStatus) If your instrument supports Bluetooth, the blue LED is a visual indication of the Bluetooth
connection status.
Off = Bluetooth OFF or Undiscoverable
Fast Flash = Discoverable Mode
Slow Flash = Connected
Vibrating Alarm
The device is equipped with a vibrating alarm.
Backlight
The backlight automatically activates when any front panel button is pressed. The backlight remains on for the duration of
the user-selected timeout. This duration can be set through MSA GALAXY GX2 Test Stand or MSA Link software.
Horn
The horn provides an audible alarm.
12 ALTAIR 4XR US
3 Description
Operating Beep
This operating beep activates every 30 seconds by momentarily beeping the horn and flashing the alarm LEDs under the
following conditions:
Operating beep is enabled
Device is in Normal Operation mode
Device is not in battery warning
Device is not in gas alarm
Audible and visual options are enabled
3.3 On-Screen Indicators
Figure 2 Display
1 Graphic Symbols 3 Gas Concentration
2 Gas Type
Alarm Symbol – Indicates alarm state
Motion Alert – Indicates Motion Alert is active
Bump Check Symbol – Indicates successful bump or cal
Indicates required interaction
Battery Condition – Indicates the battery charge level
Sensor Labels
Calibration Setup
Cal Gas Cylinder – Indicates cal gas must be applied
3 Description
US ALTAIR 4XR 13
No Gas Cylinder – Indicates cal gas should not be applied and device must be exposed to fresh air.
Hourglass – Indicates user should wait
Minimum – Indicates a minimum value or low alarm
PEAK Symbol – Indicates a PEAKreading or high alarm
STEL Symbol – Indicates a STELalarm
TWA Symbol – Indicates a TWAalarm
Sensor life Symbol – Indicates the end of sensor life
Bluetooth Symbol - Indicates enabled Bluetooth operation (if applicable)
3.4 Battery Care
Battery Life Indicator
The battery condition icon is continuously displayed in the upper right-hand corner of the display. As the battery is
depleted, battery icon segments blank until only the battery icon outline remains.
Each indicator segment represents approximately 25% of the total battery capacity.
Battery Warning
WARNING!
If battery warning alarm activates while using the device, leave the area immediately as the end of battery life is
approaching.
Failure to follow this warning can result in serious personal injury or death.
The nominal run-time of the device at room temperature is 24hours. The runtime with an active Bluetooth connection is
approximately 22 hours. The runtime of the device at -20°C/- 4°F will be approximately 14hours.
Actual run-time will vary depending on ambient temperature and alarm conditions.
The alarm levels for the individual gases are factory-set and can be changed through the Setup Menu.
A Low Battery Warning indicates that a nominal 30minutes of operation remain before the battery’s charge is depleted.
The duration of remaining device operation during a Low Battery Warning depends on ambient
temperatures.
When the device goes into Low Battery Warning, the:
battery life indicator flashes
alarm sounds
14 ALTAIR 4XR US
3 Description
alarm LEDs flash
display shows "LOWBATT" and
device repeats this warning every 60 seconds and continues to operate until it is turned off or battery shutdown
occurs.
Battery Shut Down
WARNING!
If the battery shutdown alarm activates, stop using the device as it no longer has enough power to indicate potential
hazards, and persons relying on this device for their safety could sustain serious personal injury or death.
The device goes into battery shutdown mode 60seconds before final shutdown (when the batteries can no longer operate
the device):
On the display flashes "BATT ALARM" and
Alarm sounds and lights flash; alarm cannot be silenced,
No other pages can be viewed; after approximately one minute, the device -automatically turns off.
When battery shutdown condition occurs:
1. Leave the area immediately.
2. Recharge the battery.
Battery Charging
WARNING!
Risk of explosion: Do not recharge device in hazardous area.
Failure to follow this warning can result in serious personal injury or death.
WARNING!
Use of any charger, other than the charger supplied with the device, may damage or improperly charge the batteries.
The charger is capable of charging a completely depleted pack in less than four hours in normal, room-temperature
environments.
Failure to follow these warnings can result in serious personal injury or death.
Allow very hot or cold devices to stabilize for one hour at room temperature before attempting to charge.
To Charge the Device
Ensure the charger connector is firmly seated in the charge port on the back of the device.
The battery symbol will scroll through a progressively increasing number of segments and the charge LED will be red
until 90% of full charge has been obtained. Then the battery symbol will remain fully illuminated and the charge LED
will be green while the battery is trickle charged to its full capacity.
If a problem is detected during charging, the battery symbol will flash and the charge LED will be orange. Disconnect
the device from the power module and reconnect it to the power module to reset the charge cycle.
During periods of non-use, the charger may remain connected to the device/battery pack.
Minimum and maximum ambient temperature to charge the device is 10°C (50°F) and 35°C (95°F), respectively.
For best results, charge the device at room temperature 23°C (73 °F).
3 Description
US ALTAIR 4XR 15
3.5 Viewing Additional Pages
The Main Screen appears at device turn-on.
Optional displays can be viewed by pressing the qbutton to move to:
Bump Mode
1. To select the Bump mode, press the button.
2. To move forward, press the qbutton.
3. To move backward to the Main page, press the pbutton.
Peak Readings (PEAK page)
The peak icon (3.3 On-Screen Indicators) shows the highest levels of gas recorded by the device since turn-on or since
peak readings were reset.
To reset the peak readings:
1. Access the PEAK page.
2. Press the pbutton.
Minimum Readings (MIN page)
This page shows the lowest level of oxygen recorded by the device since turn-on or since the MIN reading was reset.
The minimum icon (3.3 On-Screen Indicators) appears on the display.
To reset the MIN reading:
1. Access the MIN page.
2. Press the pbutton.
Short Term Exposure Limits (STEL page)
WARNING!
If the STEL alarm activates, leave the contaminated area immediately; the ambient gas concentration has reached the
preset STEL alarm level. Failure to follow this warning will cause over-exposure to toxic gases and persons relying on this
product for their safety could sustain serious personal injury or death.
Failure to follow this warning can result in serious personal injury or death.
The STEL icon (3.3 On-Screen Indicators) appears on the display to show the average exposure over a 15minute
period.
When the amount of gas detected by the device is greater than the STEL limit:
Alarm sounds
Alarm LEDs flash
STEL icon flashes.
To reset the STEL:
1. Access the STEL page.
2. Press the pbutton.
The STEL alarm is calculated over a 15minute exposure.
STEL calculation examples:
Assume the device has been running for at least 15 minutes:
16 ALTAIR 4XR US
3 Description
15minute exposure of 35ppm:
(15minutes x 35ppm) = 35ppm
15minutes
10minute exposure of 35ppm and 5minutes exposure of 5ppm:
(10 minutes x 35 ppm) + (5 minutes x 5 ppm) = 25ppm
15minutes
This page can be de-activated through MSA GALAXY GX2 Test Stand or MSA Link.
Time Weighted Average (TWA page)
WARNING!
If the TWA alarm activates, leave the contaminated area immediately; the ambient gas concentration has reached the
preset TWA alarm level. Failure to follow this warning will cause over-exposure to toxic gases and persons relying on this
product for their safety could sustain serious personal injury or death.
The TWA icon (3.3 On-Screen Indicators) appears on the display to show the average exposure since the device was
turned on or since the TWA reading was reset. When the amount of gas detected is greater than the eight-hour TWA limit:
Alarm sounds
Alarm LEDs flash
TWA icon flashes.
To reset the TWA:
1. Access the TWA page.
2. Press the pbutton.
The TWA alarm is calculated over an eight-hour exposure.
TWA calculation examples:
1hour exposure of 50ppm:
(1 hour x 50 ppm) + (7 hours x 0 ppm) = 6.25ppm
8hours
4hour exposure of 50ppm and 4hour exposure of 100ppm:
(4 hours x 50 ppm) + (4 hours x 100
ppm)
= 75ppm
8hours
12hour exposure of 100ppm:
(12 hours x 100 ppm) = 150ppm
8hours
This page can be deactivated through the MSA GALAXY GX2 Test Stand or MSA Link software.
Time Display
Current time appears on the display in a 12-hour format by default.
A 24-hour format can be selected using MSA GALAXY GX2 Test Stand or MSA Link.
Date Display
3 Description
US ALTAIR 4XR 17
Current date appears on the display in the format: MMM-DD-YYYY.
Discoverable Mode page
Allows the user to put the device into Bluetooth discoverable mode in order to pair with another device. This page can be
deactivated through the BT SETUP page.
Motion Alert Activation
To activate or deactivate the Motion Alert feature, press the pbutton while the MotionAlert Activation page is displayed.
When the Motion Alert feature is active, the MotionAlert symbol (3.3 On-Screen Indicators) will flash every 3seconds.
The device will enter pre-alarm when no motion is detected for 20seconds. This condition can be cleared by moving the
device.
After 30seconds of inactivity, the full MotionAlert alarm is triggered. This alarm can only be cleared by pressing the
pbutton.
3.6 Sensor Missing Alarm
The device enters the SensorMissing alarm if the device detects that the sensor is not properly installed in the device or is
not functional.
If a sensor is detected as missing, the following occurs:
• "SENSOR ERROR" displays.
The flag above the sensor detected as missing flashes on the display.
Alarm sounds and lights flash.
Yellow Fault LED is on solid.
If there is a sensor error at startup, the device shuts off in 60seconds.
3.7 Sensor End of Life Warning
If a sensor is nearing its end of life, the device will warn the user following a sensor calibration. The sensor is still fully
functional at this point, but the warning gives the user time to plan for a replacement sensor and minimize downtime. The
symbol will be continuously displayed. For more details, see 4.10 Calibration.
3.8 Sensor End of Life Indicator
If the device cannot calibrate one or more sensors, the device will display "SPAN ERR" followed by the Alarm symbol and
symbol to indicate end of sensor life. For more details, see 4.10 Calibration.
3.9 Monitoring Toxic Gases
The device can monitor the concentration of the following toxic gases in ambient air:
Carbon Monoxide (CO)
Hydrogen Sulfide (H2S)
Sulfur Dioxide (SO2)
Nitrogen Dioxide (NO2)
The device displays the gas concentration in parts per million (PPM) or mg/m3on the Normal Operations page until
another page is selected or the device is turned off.
WARNING!
If an alarm activates while using the device, leave the area immediately.Remaining in the area under such circumstances
can cause serious personal injury or death.
The device has four gas alarms for each toxic gas:
HIGH Alarm
18 ALTAIR 4XR US
3 Description
LOW Alarm
STEL Alarm
TWA Alarm
If the gas concentration reaches or exceeds the alarm set point, the device:
backlight turns on
triggers a vibrating alarm
displays and flashes the Alarm symbol and either the Minimum icon (LOWalarm) or the Maximum (PEAK) icon
(HIGHalarm)
enters an alarm state.
When a gas reading exceeds the full scale range of the sensor, the device displays "+ + +" in place of the
actual reading.
When a gas reading falls below a defined under-range threshold, the device displays " - - -" in place of the
actual reading.
3.10 Monitoring Oxygen Concentration
The device monitors the oxygen concentration in ambient air. The alarm set points can be set to activate on two different
conditions:
Enriched - oxygen concentration > 20.8% or
Deficient - oxygen concentration < 19.5%.
While the device can detect up to 30% oxygen in the ambient air, it is approved for use only up to 21% oxygen-content.
WARNING!
If an alarm activates while using the device, leave the area immediately.Remaining in the area under such circumstances
can cause serious personal injury or death.
When the alarm set point is reached for either of the above conditions:
an alarm sounds
alarm LEDs flash
a vibrating alarm triggers
device displays and flashes the Alarm icon and either the Minimum icon (Enriched alarm) or the Maximum icon
(Deficient alarm) (3.3 On-Screen Indicators) along with the corresponding oxygen concentration.
The LOW alarm (oxygen deficient) is latching and will not automatically reset when the O2concentration
rises above the LOW set point. To reset the alarm press the pbutton. If the alarm condition still exists, the
pbutton only silences the alarm for five seconds.
False oxygen alarms can occur due to changes in barometric pressure (altitude) or extreme changes in
ambient temperature.
It is recommended that an oxygen calibration be performed at the temperature and pressure of use. Be
sure that the device is in known fresh air before performing a calibration.
When a gas reading exceeds the full scale range of the sensor, the device displays "+ + +" in place of the
actual reading.
When a gas reading falls below a defined under-range threshold, the device displays " - - -" in place of the
actual reading.
4 Operation
US ALTAIR 4XR 19
3.11 Monitoring Combustible Gases
The device can monitor these concentrations in ambient air:
• Methane
Combustible gases
The device displays the combustible gas concentration in % LEL or % CH4on the Measuringpage until another page is
selected or the device is turned off.
WARNING!
If an alarm activates while using the device, leave the area immediately.Remaining in the area under such circumstances
can cause serious personal injury or death.
The device has two alarm set points:
HIGH Alarm
LOW Alarm
If the gas concentration reaches or exceeds the alarm set point, the device:
backlight turns on
a vibrating alarm triggers
displays and flashes the Alarm symbol and either the Minimum icon (LOWalarm) or the Maximum (PEAK) icon
(HIGHalarm)
enters an alarm state.
When a gas reading reaches 100% LEL or 5.00% CH4, the device enters a LockAlarm state: the
combustible sensor shuts down and displays “XXX” in place of the actual reading. This state can only be
reset by turning the device off and on in a fresh air environment.
When a gas reading falls below a defined under-range threshold, the device displays " - - -" in place of the
actual reading.
WARNING!
A combustible gas reading of “100” or “5” indicates the atmosphere is above 100%LEL or 5.00% vol CH4
respectively, and an explosion hazard exists. Move away from contaminated area immediately.
In such cases, the device enters a Lock Alarm state.
Failure to follow this warning can result in serious personal injury or death.
Check your national standard values for 100%LEL (EN60079-20-1).
4 Operation
Device operation is dialog driven from the display with the aid of the three Function buttons (see Figure 1 Device overview).
4.1 Environmental Factors
A number of environmental factors may affect the gas sensor readings, including changes in pressure, humidity and
temperature. Pressure and humidity changes affect the amount of oxygen actually present in the atmosphere.
Pressure Changes
If pressure changes rapidly (e.g., stepping through airlock), the oxygen sensor reading may temporarily shift and possibly
cause the device to go into alarm. While the percentage of oxygen may remain at or near 20.8Vol%, the total amount of
20 ALTAIR 4XR US
4 Operation
oxygen present in the atmosphere available for respiration may become a hazard if the overall pressure is reduced by a
significant degree.
Humidity Changes
If humidity changes by any significant degree (e.g., going from a dry, air conditioned environment to outdoor, moisture
laden air), oxygen readings can be reduced by up to 0.5%, due to water vapor in the air displacing oxygen.
The oxygen sensor has a special filter to reduce the effects of humidity changes on oxygen readings. This effect will not be
noticed immediately, but slowly impacts oxygenreadings over several hours.
Temperature Changes
The sensors have built-in temperature compensation. However, if temperature shifts dramatically, the sensor reading may
shift. Zero the device at the work site temperature for the least effect.
4.2 Turning ON and Fresh Air Setup
Device operation is dialog driven from the display with the aid of the three Function buttons (see Figure 1 Device overview).
For more information, see the flow charts in 8 Appendix.
1. Turn the device on with the button.
The device performs a self test and then goes to Fresh Air Setup:
all display segments are activated
audible alarm sounds
alarm LEDs light
vibrating alarm is activated.
During the self test, the device checks for missing sensors. In the case of a missing sensor, the device displays the Sensor
Missing screen and alarms until it is turned off. Otherwise, the turn-on sequence continues.
The device displays:
Alarm & display self test
Manufacturer name
Device name
Software version
FCC ID
IC ID
Bluetooth software version (if activated)
Sensor discovery
Combustible gas type
Toxic gas units
Alarm set points (PEAK, MIN) (STEL, TWA)
Calibration values
Date and time display
Last cal date (if activated)
CAL due date (if activated)
Fresh Air Setup option.
Screen Displays during Startup
During the startup sequence, all automatic page display timeouts are preset to a range from two to four
seconds.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56

MSA Altair User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI