Swann AMI Solicon4 User manual

Category
Measuring, testing & control
Type
User manual
AMI Solicon4
Version 6.00 and higher
A-96.250.321 / 191015
Operator’
s Manual
© 2015, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
Document Status
Title:
Monitor AMI Solicon4 Operator’s Manual
ID:
A-96.250.321
Revision Issue
00 April 2005 First Edition
01 April 2013 Update to FW-Release 5.30
02 October 2015 Update to FW-Release 6.00
AMI Solicon4
A-96.250.321 / 191015 1
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4. Single Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4.1 AMI Solicon4 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4.2 Flow Cell M-Flow PG and M-Flow G1 . . . . . . . . . . . . . . . . . . . . . 14
2.4.3 Swansensor Shurecon P . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4.4 Swansensor Shurecon S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.4.5 Swansensor DeltaT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.1. Installation Checklist Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3. Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 19
3.4. Install Swansensor deltaT (Option) . . . . . . . . . . . . . . . . . . . . . . . 20
3.4.1 Mount the deltaT Sensor and Connect the Sample Inlet . . . . . . . 21
3.4.2 Connect the Sensor Cable to the Transmitter . . . . . . . . . . . . . . . 22
3.4.3 Change Firmware Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.4.4 Adjust the deltaT sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.1 Connection Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.5.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.6. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.2 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.3 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.7.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . . . 32
3.8. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.8.1 Signal Output 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.8.2 Profibus, Modbus Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.8.3 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
2 A-96.250.321 / 191015
AMI Solicon4
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.2. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 39
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.1. Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 40
6.3. Maintenance of the Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.1 Remove the Sensor form the Flow Cell . . . . . . . . . . . . . . . . . . . . 41
6.3.2 Install the Sensor into the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . 41
6.4. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.5. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.6. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9. Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . . 53
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
11. Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
12. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
AMI Solicon4
Safety Instructions
A-96.250.321 / 191015 3
AMI Solicon4 - Operators Manual
This document describes the main steps for instrument setup, oper-
ation and maintenance.
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
Operator: Qualified person who uses the equipment for its intended
purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
4 A-96.250.321 / 191015
AMI Solicon4
Safety Instructions
1.1. Warning Notices
The symbols used for safety-related notices have the following sig-
nificance:
DANGER
Your life or physical wellbeing are in serious danger if such
warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor-
rect process can be the consequence if such warnings are ig-
nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
AMI Solicon4
Safety Instructions
A-96.250.321 / 191015 5
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
6 A-96.250.321 / 191015
AMI Solicon4
Safety Instructions
1.2. General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation
of the instrument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi-
bility for any claim resulting from unauthorized modification or alter-
ation.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
relay 1,
relay 2,
alarm relay
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.
AMI Solicon4
Product Description
A-96.250.321 / 191015 7
2. Product Description
2.1. Description of the System
This instrument is applicable for the measurement of the specific
conductivity in surface water, potable water and cooling water. The
complete system consists of the Transmitter AMI Solicon4, the con-
ductivity sensor Swansensor Shurecon P and the flow cell
M-Flow PG.
Transmitter, flow cell and sensors are also available as single com-
ponents.
Application
Range
Conductivity is a parameter for the total quantity of ions present in
the solution. It can be used for the controlling of:
the condition of waters
water purification
water hardness
completeness of ion analysis
Signal
Outputs
Two signal outputs programmable for measured values (freely scal-
able, linear, bilinear, log) or as continuous control output (control
parameters programmable).
Current loop: 0/420 mA
Maximal burden: 510 Ohm
Third signal output with the same specifications as option.
(Only possible if no communication interface is used.)
Relays Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A/250 VAC
Alarm Relay One potential free contact.
Alternatively:
Open during normal operation, closed on error or loss of
power.
Closed during normal operation, open on error or loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.
8 A-96.250.321 / 191015
AMI Solicon4
Product Description
Input For potential-free contact to freeze the signal outputs or to interrupt
control in automated installations (hold function or remote-off).
Communica-
tion interface
(optional)
USB interface to store logger data.
RS485 interface with fieldbus protocol Modbus, Profibus DP or
Webserver connection via Modbus.
Safety
Features
No data loss after power failure. All data is saved in non-volatile
memory.
Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs.
The analyzer is factory tested and ready for installation and opera-
tion.
Temperature
compensation
The mobility of ions in water increase with higher temperature
which enlarges the conductivity. Therefore, the temperature is mea-
sured simultaneous by an integrated Pt1000 temperature sensor
and the conductivity is compensated to 25 °C.
Measuring
principle
When a voltage is set between two electrodes in an electrolyte so-
lution, the result is an electric field which exerts force on the
charged ions: the positively charged cations move towards the neg-
ative electrode (cathode) and the negatively charged anions to-
wards the positive electrode (anode). The ions, by way of capture
or release of electrons at the electrodes, are discharged and so a
current I flows through this cycle and the Ohms law V = I×R ap-
plies. From the total resistance R of the current loop, only the resis-
tance of the electrolyte solution, respectively its conductivity
1
/
R
, is
of interest.
The cell constant of the sensor is determined by the manufacturer
and is printed on the sensor label. If the cell constant has been pro-
grammed in the transmitter, the instrument measures correctly. No
calibration must be done, the sensor is factory calibrated.
AMI Solicon4
Product Description
A-96.250.321 / 191015 9
Fluidics The flow cell M-Flow PG consists of the flow cell block [B] and the
calibration vessel [D].
The conductivity sensor [A] is fixed in the flow cell block [B] with a
threaded sleeve.
The sample enters at the sample inlet [F]. It flows through the del-
taT flow sensor [C] (if installed) and then through the flow cell block
into the calibration vessel [D], were the conductivity is measured.
The sample leaves the calibration vessel via flow cell block through
the sample outlet [E] and flows into the drain [G].
A
B
C
D
Conductivity sensor
Flow cell block
deltaT flow sensor (option)
Calibration vessel
E
F
G
Sample outlet
Sample inlet
Drain
A
C
B
E
F
D
G
10 A-96.250.321 / 191015
AMI Solicon4
Product Description
2.2. Instrument Specification
Power Supply
Voltage:
Power consumption:
100240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
Electronics
Aluminium with a protection degree of IP 66 / NEMA 4X
housing
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
1090% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
4 to 15 l/h
up to 50 °C
up to 1 bar
pressure free
On-site
The analyzer site must permit connections to:
requirements
Sample inlet:
Sample outlet:
Max. Altitude:
Flow celll with elbow hose nozzle 1/4”
diam.10 mm
or
deltaT sensor with hose nozzle
diam. 10 mm
1/2” hose nozzle for flexible tube
diam. 20 x 15 mm
2000 m above sea level
Measuring
Range
Range
0.10 to 9.99 S/cm
10.0 to 99.9 S/cm
100 to 999 S/cm
1.00 to 9.99 mS/cm
10.0 to 29.9 mS/cm
30 to 100 mS/cm
Resolution
0.01 S/cm
0.1 S/cm
1 S/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
Accuracy
± 0.5% of measured value
AMI Solicon4
Product Description
A-96.250.321 / 191015 11
Dimensions
Panel:
Dimensions:
Screws:
Weight:
white PVC
280x850x150 mm
5 mm or 6 mm diameter
9.0 kg
AMI Solicon4
850 mm / 33½”
13 mm / ½”
6 x dia. 6.5 mm / ¼”
254 mm/ 10”
280 mm/ 11”
412 mm /16 ¼”
824 mm / 32
7
/
16
12 A-96.250.321 / 191015
AMI Solicon4
Product Description
2.3. Instrument Overview
A
B
C
D
E
Transmitter
Panel
Conductivity sensor
Flow cell block
Calibration vessel
F
G
H
I
Sample outlet
DeltaT flow sensor (option)
Drain
Sample inlet
A
B
C
D
E
F
G
H
I
AMI Solicon4
Product Description
A-96.250.321 / 191015 13
2.4. Single Components
2.4.1 AMI Solicon4 Transmitter
Electronic transmitter and controller for conductivity measurement.
140
4.5
180
165
103
18.5
7.5
Dimensions
Width:
Height:
Depth:
Weight:
140 mm
180 mm
70 mm
1.5 kg
Specifications
Electronics case:
Protection degree:
Display:
Electrical connectors:
Cast aluminum
IP 66 / NEMA 4X
backlit LCD, 75 x 45 mm
screw clamps
14 A-96.250.321 / 191015
AMI Solicon4
Product Description
2.4.2 Flow Cell M-Flow PG and M-Flow G1
Flow cell M-Flow PG, made of PVC and acrylic glass with one
13.5 mm sensor opening for Swansensor Shurecon P.
or
Flow cell M-Flow G1, made of PVC and acrylic glass with one
G 1” sensor opening for Swansensor Shurecon S.
60
102
23 23
161
10
33 38
90
142
9
Sample
connections
Inlet:
Outlet:
G 1/4” thread
G 1/4” thread
Each equipped with an elbow hose nozzle for 10 mm tube
Sample
Only valid for the flow cell without sensor.
conditions
Flow rate:
Temperature:
Inlet pressure:
Sample outlet:
Particle size:
4 to 15 l/h
up to 50 °C
up to 1 bar @ 25 °C
Pressure-free outlet
below 0.5 mm
No strong acids and bases.
No organic solvents.
Dimensions
Width:
Front-to-back:
Height:
Panel mounting:
90 to 142 mm
105 mm
161 mm
3 screws M5
AMI Solicon4
Product Description
A-96.250.321 / 191015 15
2.4.3 Swansensor Shurecon P
Sensor with integrated cable for the measurement of the specific
conductivity. Four electrode design with platinum electrodes and
built-in Pt1000 temperature sensor.
16.5
12
46
43
Sensor adapter
16 mm / PG 13.5
PG 13.5
120 ± 2
Specifications
Measuring range
Accuracy:
Temperature sensor:
Operating conditions:
Electrical connection:
Process connection:
0.1S/cm to 100mS/cm
± 1.5 % or ± 0.2 S/cm whichever is
greater
Pt 1000
> 50°C
Max. temperature short-time 90 °C
Max.pressure 10 bar at 25 °C.
Sensor with integrated cable
PG 13.5 mm
16 A-96.250.321 / 191015
AMI Solicon4
Product Description
2.4.4 Swansensor Shurecon S
Sensor for the measurement of the specific conductivity. Four elec-
trode design with integrated Pt 1000 temperature sensor.
Screw connector with cable must be ordered separately.
94
25
164
258
G 1”
thread
108
Specifications
Measuring range
Accuracy:
Temperature sensor:
Operating conditions:
Electrical connection:
Process connection:
Space around sensor
tip:
0.1S/cm to 100mS/cm
± 1 % or ± 0.2 S/cm whichever is
greater
Pt 1000
Max. temperature: 120°C at 6.5 bar
Max. pressure: 12 bar at 20°C
Sterilizable at: 120°C / 5 bar / 30 min.
Plug M16 male (IP 67)
G 1" thread
20 mm
AMI Solicon4
Product Description
A-96.250.321 / 191015 17
2.4.5 Swansensor DeltaT
Calorimetric flow meter based on heat dissipation. For applications
in potable water, surface water treatment and effluent.
250
80
70
12
4.5
175
163
Specifications
Measuring range/Flow rate
Accuracy:
Response time t
90
:
Sample temperature:
Sample inlet and outlet:
Max. cable length:
040 l/h
± 20%
ca. 1 min
535 °C
for tubing diam. 1011 mm
1 m
18 A-96.250.321 / 191015
AMI Solicon4
Installation
3. Installation
3.1. Installation Checklist Monitors
Check Instrument’s specification must conform to your AC power rat-
ings. Do not turn on power until instructed to do so.
On site require-
ments
100240 VAC (± 10%), 50/60 Hz (± 5%) or 24 VDC, isolated
(±10%) power outlet with ground connection and 30 VA.
Sample line with sufficient sample flow and pressure (see Instru-
ment Specification, p. 10
Installation Mount the instrument in vertical position. Display should be at
eye level.
Connect sample and waste.
Electrical Wir-
ing
Connect all external devices like limit switches, current loops
and pumps.
Connect power cord; do not switch on power yet!
Power-up Turn on the sample flow and wait until the flow cell is completely
filled.
Switch on power.
Instrument
set-up
Adjust sample flow.
Program all parameters for sensor and external devices (inter-
face, recorders, etc.).
Program all parameters for instrument operation (limits, alarms).
Run-in period Let the instrument run continuously for 1 h.
Calibration Perform a calibration if necessary, see Calibration, p. 42.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80

Swann AMI Solicon4 User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI