Toxalert ToxCO/1/M/C, ToxCO/1/M/B Calibration Procedures

  • Hello, I've reviewed the user manual for the Toxalert TOX-CO/1/M/C CO sensor. This document details the calibration process, emphasizing the need for a 72-hour power-on period before calibration. It also outlines the specific tools and procedures, including the use of 100ppm CO gas and a milliamp meter, required for accurate calibration. I'm ready to assist with any questions you have about this device and its calibration.
  • What tools are required for calibration?
    How long should the sensor be powered before calibration?
    What milliamp reading should be expected when using 100ppm CO gas?
TOX-CO/1/M/C CO SENSOR CALIBRATION SET-UP
TOOLS REQUIRED:
Tank of calibration gas (usually 100ppm CO)
0.2 LPM regulator
• Wetter bottle
Calibration cover and tubing
• Milliamp meter
PROCEDURE:
1. Fill water bottle half full of tap water.
NOTE: MAKE SURE THAT WATER BOTTLE IS BELOW SENSOR SO THAT
WATER IS NOT FORCED INTO CO SENSOR.
2. Screw the regulator on to the tank of calibration gas.
3. Press the sensor calibration cover inside the sensor shield for a snug fit.
CALIBRATION CHECKING:
NOTE: SENSOR MUST BE CONTINUOUSLY POWERED FOR 72 HOURS PRIOR
TO CALIBRATION.
1. Connect a milliamp meter in series with the output signal that is capable of reading 4 to 20
milliamps. (A controller or other instrument may already be attached, and may be used
instead of the meter).
2. Start the gas flow by pushing in and turning the control knob on the regulator.
3. The sensors electronics reads the CO concentration every 2.5 minutes, therefore allow the
gas to flow for at least 10 minutes to allow the ambient air to be flushed out and an
accurate reading can be taken.
4. Using 100 ppm CO test gas, the milliamp reading should be apprx.12mA after 10 minutes
of continuous gas flow.
5. Upon completion of the test, ensure that the tank valve is off and remove the calibration
cover from the sensor.
TOXALERT TOXCOCALKIT
WATER BOTTLE
1/8" I.D. TUBING
SETUP
TEST GAS
CAP
0.2 LPM REG
CONTROL BUTTON-
PUSH IN AND TURN
TO START GAS FLOW
DO NOT ADJUST
MIXING VALVE-
TOX-CO/1/M/C CALIBRATION
100ppm CO TOX-CO/1/M/C
Toxalert International, Inc.; P.O. ox 159; Mound, MN. 55364 Phone: 952-472-4541 Fax: 952-472-4960
/