RBCBioscience MagCore Super Training manual

Type
Training manual
MagCore®
Super/HF16 Plus
Engineer In Training
manual
1
Handle the power cord carefully.
Hold the plug when unplugging the cord.
Keep the set free from moisture, water, and dust.
Unplug the power cord when not using the set for
long periods of time.
Do not obstruct the ventilation space.
Do not let foreign objects in the set.
Do not let insecticides, benzene, and thinner come in
contact with the set.
Never disassemble or modify the set in any way.
Avoid high temperatures.
Allow for sufficient heat dispersion when installed on a rack.
CAUTIONS
Risk Of Electric Shock Do Not Open
Safety Precautions
Before use
These WARNINGS and CAUTIONS are intended to protect you and other persons from injuries and damages. To ensure safe operation, please follow them carefully.
CAUTION:
To Reduce The Risk Of Electric Shock, Do Not Remove Cover (Or Back).
No Userserviceable Parts Inside. Refer Servicing To Qualified Service Personnel.
CAUTION:
1. Handle the power supply cord carefully
Do not damage or deform the power supply cord. If it is damaged or deformed, it
may cause electric shock or malfunction when used. When removing from wall
outlet, be sure to remove by holding the plug attachment and not by pulling the
cord.
2. Do not open the top cover
In order to prevent electric shock, do not open the top cover.
3. Do not place anything inside
Do not place metal objects or spill liquid inside the MagCore® System. Electric shock
or malfunction may result.
Note On Use:
1. Read Instructions – All the safety and operating instructions should be read before the
product is operated.
2. Retain Instructions – The safety and operating instructions should be retained for future
reference.
3. Heed Warnings – All warnings on the product and in the operating instructions should be
adhered to.
4. Follow Instructions – All operating and use instructions should be followed.
5. Cleaning – Unplug this product from the wall outlet before cleaning. Only allow to use 75% of
EtOH to clean the surface of instrument.
6. Attachments – Do not use attachments not recommended by the product manufacturer as
they may cause hazards.
7. Water and Moisture – Do not use this product near water – for example, near a bath tub,
wash bowl, kitchen sink, or laundry tub; in a wet basement; or near a swimming pool; and the
like.
8. AccessoriesDo not place this product on an unstable cart, stand, tripod, bracket, or
table. The product may fall, causing serious injury to a child or adult, and serious damage
to the product. Use only with a cart, stand, tripod, bracket, or table recommended by the
manufacturer, or sold with the product. Any mounting of the product should follow the
manufacturer’s instructions, and should use a mounting accessory recommended by the
manufacturer.
9. Ventilation – Slots and openings in the cabinet are provided for ventilation and to ensure
reliable operation of the product and to protect it from overheating, and these openings
must not be blocked or covered. The openings should never be blocked by placing the
product on a bed, sofa, rug, or other similar surface. This product should not be placed in a
built-in installation such as a bookcase or rack unless proper ventilation is provided or the
manufacturer’s instructions have been adhered to.
10. Power Sources – This product should be operated only from the type of power source
indicated on the marking label. If you are not sure of the type of power supply to your home,
consult your product dealer or local power company. For products intended to operate from
battery power, or other sources, refer to the operating instructions.
11. Grounding or Polarization – This product may be equipped with a polarized alternating-
current line plug (a plug having one blade wider than the other). This plug will fit into the
power outlet only one way. This is a safety feature. If you are unable to insert the plug fully
into the outlet, try reversing the plug. If the plug should still fail to fit, contact your electrician to
replace your obsolete outlet. Do not defeat the safety purpose of the polarized plug.
12. Power-Cord Protection – Power-supply cords should be routed so that they are not likely to
be walked on or pinched by items placed upon or against them, paying particular attention to
cords at plugs, convenience receptacles, and the point where they exit from the product.
13. Lightning – For added protection for this product during a lightning storm, or when it is left
unattended and unused for long periods of time, unplug it from the wall outlet and disconnect
the antenna or cable system. This will prevent damage to the product due to lightning and
power-line surges.
14. Overloading – Do not overload wall outlets, extension cords, or integral convenience
receptacles as this can result in a risk of fire or electric shock.
15. Object and Liquid Entry – Never push objects of any kind into this product through openings
as they may touch dangerous voltage points or short-out parts that could result in a fire or
electric shock. Never spill liquid of any kind on the product.
16. Servicing – Do not attempt to service this product yourself as opening or removing covers
may expose you to dangerous voltage or other hazards. Refer all servicing to qualified service
personnel.
17. Damage Requiring Service – Unplug this product from the wall outlet and refer servicing to
qualified service personnel under the following conditions:
a) When the power-supply cord or plug is damaged,
b) If liquid has been spilled, or objects have fallen into the product,
c) If the product has been exposed to rain or water,
d) If the product does not operate normally by following the operating instructions. Adjust
only those controls that are covered by the operating instructions as an improper adjustment
of other controls may result in damage and will often require extensive work by a qualified
technician to restore the product to its normal operation,
e) If the product has been dropped or damaged in any way, and
f) When the product exhibits a distinct change in performance
– this indicates a need for service.
18. Replacement Parts – When replacement parts are required, be sure the service technician
has used replacement parts specified by the manufacturer or have the same characteristics as
the original part. Unauthorized substitutions may result in fire, electric shock, or other hazards.
19. Safety Check – Upon completion of any service or repairs to this product, ask the service
technician to perform safety checks to determine that the product is in proper operating
condition.
20. Heat – The product should be situated away from heat sources such as radiators, heat
registers, stoves, or other products that produce heat.
The exclamation point within an equilateral triangle isintended
to alert the user to the presence of important operating
and maintenance (servicing) instructions in the literature
accompanying the appliance.
Be aware of contaimination by contagious specimens.
High temperature warning!!
DO NOT touch the heating block with this sticker attached, it may
cause serious burning injuries.
The lightning flash with arrowhead symbol, within an equilateral
triangle, is intended to alert the user to the presence of uninsulated
dangerous voltage” within the products enclosure that may
be of sufficient magnitude to constitute a risk of electric shock to
persons.
SAFETY INSTRUCTIONS
2
RBCBioscience
Model Super / HF16 Plus
Processing Capacity Max .16 samples per batch
System Components 1. Pipetting Unit: Dispensing, transferring, X -Y two axis movements.
2. Electric Control: PLC module and Arm-based main board embedded in.
3. UV Light: Power 8W, life duration 1,000 Hrs .
4. D2 Lamp life: 800 hours. ( MagCore Super Only )
5. Heating Block: RT~90° C .
6. OD detection range: ABS <6. ( MagCore Super Only )
7. Detection source: D2 lamp. ( MagCore Super Only )
8. Detection wavelength: 260nm, 280nm,320nm.
9. Display Screen: 7 inch colored touch panel .
10. Accessories: T - Rack, Cartridge Rack .
11. Barcode Scanner(MagCore HF16 Plus optional), embedded thermal printer. (MagCore Super only)
Power Supply Voltage: AC 100V~ 240V; Frequency: 50/60Hz
Dimension W760 x D710 x H770 (mm) / W27.95 x D29.92 x H30.31 (inches)
Net Weight 76.5kg / 168.68LB (MagCore Super) / 68.5kg / 151LB (MagCore HF16 Plus)
Specification
━ Operating Parameters
━ Operating Environment
━ Applications
Applications: Applications: Whole Blood DNA (200µl/400µl/1,200µl), Viral Nucleic Acid (DNA/RNA 200µl/400µl/1,200µl), FFPE Genomic
DNA, Forensic DNA, Plasma and Circulating Cell-Free DNA (1,200µl), Tissue Genomic DNA, Plant Genomic DNA, Bacterial DNA, Whole
Blood total RNA, Cultured Cells total RNA.
Processing Capacity 1~16 samples per batch
Processing Time 30-90 minutes (depends on sample type and method)
Sample Volume 200/400/1200 μl
Elution Volume 60/100/150/200 μl
Purity DNA : O.D A260/280 ratio 1.8 ± 0.1
RNA : O.D A260/280 ratio 2.0 ± 0.2
Temperatures allowed
during transportation
storage
packaging
15°C~35°C
Temperatures allowed
during
operation
18°C~30°C
Pollution Degree Indoor
3
INDEX
Safety Precautions…………………………………………………………………………………………………… 1
Specification…………………………………………………………………………………………………… 2
Accessories…………………………………………………………………………………………………… 5
Installation…………………………………………………………………………………………………… 6
Before Install…………………………………………………………………………………………………… 6
IQ/OQ/PQ Documents…………………………………………………………………………………………………… 7
IQ Documents…………………………………………………………………………………………………… 7
OQ Documents…………………………………………………………………………………………………… 8
PQ Documents…………………………………………………………………………………………………… 9
System Overview…………………………………………………………………………………………………… 10
Modulen Position Description…………………………………………………………………………………………………… 11
Unpacking/Prepare Unpacking Tools…………………………………………………………………………………………………… 12
Install MagCore® HF16 Plus/Super ……………………………………………………………………………………………………13
Installation settings and test…………………………………………………………………………………………………… 14
Touch Panel and User Interface…………………………………………………………………………………………………… 15
Install MagCore® Cuvette…………………………………………………………………………………………………… 16
Install reagent cartridge…………………………………………………………………………………………………… 16
Install Tube, Tip…………………………………………………………………………………………………… 16
Emergent Stop…………………………………………………………………………………………………… 17
Start Programs…………………………………………………………………………………………………… 18
If the test result of the optic
measurement is over detection
(MagCore® Super Only)…………………………………………………………………………………………………… 19
Stop…………………………………………………………………………………………………… 21
Enable the barcode function
(MagCore® HF16 Plus Only)…………………………………………………………………………………………………… 21
Barcode scanning and editing
(MagCore HF16 Optional)…………………………………………………………………………………………………… 22
Optic data
(MagCore® Super Only)…………………………………………………………………………………………………… 22
Easy maintenance…………………………………………………………………………………………………… 23
Replace O-Ring…………………………………………………………………………………………………… 23
Piston Test…………………………………………………………………………………………………… 23
Clean Piercing Needles…………………………………………………………………………………………………… 25
Reject Tip…………………………………………………………………………………………………… 25
Heater Test…………………………………………………………………………………………………… 26
Optic Test…………………………………………………………………………………………………… 26
UV sterilization…………………………………………………………………………………………………… 27
Alarm History…………………………………………………………………………………………………… 27
System Status…………………………………………………………………………………………………… 28
Set up the system time…………………………………………………………………………………………………… 28
Update Program…………………………………………………………………………………………………… 29
Clean the Waste Box…………………………………………………………………………………………………… 29
Replace thermal paper…………………………………………………………………………………………………… 30
Replace fuse…………………………………………………………………………………………………… 30
Advanced maintenance…………………………………………………………………………………………………… 31
Disassemble the back service lid…………………………………………………………………………………………………… 33
Disassemble the top service lid…………………………………………………………………………………………………… 34
Axis directions…………………………………………………………………………………………………… 35
Replace the UV lamp…………………………………………………………………………………………………… 36
Replace LED Light…………………………………………………………………………………………………… 37
4
RBCBioscience
Replace the optic lamp
(MagCore® Super Only)…………………………………………………………………………………………………… 38
Replace the pistons…………………………………………………………………………………………………… 39
Replace the ejector rod module…………………………………………………………………………………………………… 41
Replace the heater module…………………………………………………………………………………………………… 42
Replace the gas spring on front door…………………………………………………………………………………………………… 43
Replace Door Sensor…………………………………………………………………………………………………… 44
Replace the M-axis sensors…………………………………………………………………………………………………… 45
Replace the sensors of
C, V, X and Y axes…………………………………………………………………………………………………… 46
Replace Cooling Fans…………………………………………………………………………………………………… 47
Replace the X-axis stepper motor…………………………………………………………………………………………………… 48
Replace the Y-axis motor…………………………………………………………………………………………………… 49
Replace the Y-axis belt…………………………………………………………………………………………………… 50
Replace the V-axis motor…………………………………………………………………………………………………… 51
Replace the C-axis motor…………………………………………………………………………………………………… 53
Replace the C-axis timing belt…………………………………………………………………………………………………… 54
Replace the lift motor of the base…………………………………………………………………………………………………… 55
Replace the timing belt of the lift
motor of the base…………………………………………………………………………………………………… 56
Replace PCB…………………………………………………………………………………………………… 57
Replace PLC…………………………………………………………………………………………………… 58
Replace Power Supply…………………………………………………………………………………………………… 59
Replace Touch Panel…………………………………………………………………………………………………… 60
Replace Printer…………………………………………………………………………………………………… 61
Electrical control and engineering
software…………………………………………………………………………………………………… 62
System…………………………………………………………………………………………………… 63
Parts…………………………………………………………………………………………………… 64
Main board configuration…………………………………………………………………………………………………… 65
Copy the engineering software to the
desktop…………………………………………………………………………………………………… 66
Install the HMI updating software…………………………………………………………………………………………………… 66
Create an uploading folder…………………………………………………………………………………………………… 67
Install the driver…………………………………………………………………………………………………… 68
HMI updating…………………………………………………………………………………………………… 70
PLC updating…………………………………………………………………………………………………… 71
Position Teaching Tools…………………………………………………………………………………………………… 74
The pre-operation before
calibration…………………………………………………………………………………………………… 75
X-axis calibration…………………………………………………………………………………………………… 76
Y-axis calibration…………………………………………………………………………………………………… 78
C-axis calibration
( Magcore® Super Only)…………………………………………………………………………………………………… 79
Automatic Optical Measurement…………………………………………………………………………………………………… 81
Optical Calibration Flow Chart…………………………………………………………………………………………………… 82
Empty Cuvette Test of standard
cuvette…………………………………………………………………………………………………… 82
Standard Cuvette Vaule…………………………………………………………………………………………………… 83
Adjust integration time…………………………………………………………………………………………………… 84
Test Standard Sample…………………………………………………………………………………………………… 85
Trouble Shooting for Alarm Code…………………………………………………………………………………………………… 86
5
Accessories
1. T-Rack X 2
2. Cartridge Rack X 2
3. Power Cable X 2 (Europe standard and U.S.A standard)
4. Grease X 1
5. 3mm O-Rings X 16
6. 5mm O-Rings X 16
7. Empty Cartridge X 10
8. Sample Tube /Elution Tube X 20
9. Pipette Tip X 20
10. MagCore Cuvette Set(10pcs) X 1
11. Waste Box X 1
12. Barcode Scanner X 1 (MagCore Super Only)
1 2 345
7
6
8 9
10 11 12
You can purchase Cat. No.: MSC100 MagCore Cuvette Set (X100) from the
local distributor.
(MagCore Super Only) (MagCore HF16 Optional)
6
RBCBioscience
770 mm
Left Side (Keep space clear) Height (Keep space clear)
Cooling Fans
710mm
760mm
Depth
Width
━ Before Install
Installation
Please refer to the machine size shown in the above figure to prepare an appropriate position for the
machine and reserve space for thermal dissipation.
7
━ IQ Documents
IQ/OQ/PQ Documents
MagCore Super and HF16 Plus
Installation Qualification/Operation Qualification/Performance
Qualification
SerialNo.____________________Date____________
InstallationQualification:
IQchecklist PASSFAILNOTE
MachineInspection:
*Noobviousdefectoncasing.
*Checkhandle.
*NoobviousdefectonPMMALOGOsurface.
*CheckT‐rack.
*CheckCartridgerack.
*Checkdoorfunction.
FunctionTest:
*Checkoutletvoltage:110~220V.
*Turnonmachine.
*Check7inchtouchpanel.
*CheckLEDLOGOlight.
*Barcodescanfunction(NotforHF16Plus).
*UVlampTest.
*MotorTest(AllAxisTest).
*RejectTipTest(Actsmoothly).
*HeaterTest(RTto60Clessthan7mins).
*SystemTimeSetup(InSystemStatus).
Examination
Signature
 Verification
Signature
Page1of3
8
RBCBioscience
━ OQ Documents
OperationQualification:
OQchecklist PASSFAILNOTE
ProgramCode999:
(SampleVolume:400μL,ElutionVolume:200μL,
OpticalMeasurement:Disable)
*Initialcheck.
*Perforatemotioncheck.
*Tiptakingmotioncheck.
*Pistonaspirationcheck.
*Magneticseparationmotioncheck.
*STOPfunctioncheck(AbortandRejecttip)
OpticTest:(NotforHF16Plus)
(EmptyCuvette,PlaceEmptyCuvetteinSlotNo.1)
*Thecountvaluein260nmshould>7000.
Examination
Signature
 Verification
Signature
Page2of3
9
━ PQ Documents
PerformanceQualification:
ThePerformanceQualificationshouldbedesignedaccordingusersrequirement.
ThefollowingformatisanexampleofPerformanceQualification.
Sample KitSpec.ResultPASS/FAIL
400LWholeBlood
MagCore
GenomicDNA
WholeBloodKit
MGB400‐03
Cartridge102
Totalyield
>14g

Purity
OD260/280=
1.8~2.0

Examination
Signature
 Verification
Signature
Page3of3
10
RBCBioscience
System Overview
Front Door
Printer (MagCore® Super Only)
Touch Panel (HMI)
Top Service Lid
Back Service Lid
Cooling holes
Operation Area
Barcode Scanner
(HF16 Plus Optional)
Main Power Switch
and Power Outlet
Barcode
Scanner Port
USB Port
Slots
There are totally 16 slots
(number 1~16 from left to
right)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
11
━ Module Position Description
Liquid handling
module
Work
platform
Heating module
Optic module
(Super Only)
Frame
Gantry structure
12
RBCBioscience
Unpacking
Prepare Unpacking Tools
If the cargo impacted, White Red If the cargo upside down, Sliver Red
2. Phillips Screwdriver
1. Hex Wrench (3mm)
3. Scissors
13
Install MagCore® HF16 Plus/Super
Proceed installation procedures after placing MagCore on a solid bench-top.
1.Use screw drivers to loose the screws as show in the picture.
2.Remove the red fixing plates.
3.Save all the materials for maintenance and shipment in the future.
Super Only
14
RBCBioscience
━ Installation settings and test
Step 1.
Please set up the local time.
Please refer to System Status chapter for the
information about the operation.
(Plase refer to P.28)
Step 2.
Test piston
Please refer to Piston Test chapter for the
information about the operation.
(Plase refer to P.23)
Step 3.
Test heater
Please refer to Heater Test chapter for the
information about the operation.
(Plase refer to P.26)
Step 4
Test optic module
(MagCore® Super Only)
Please refer to Optic Test chapter for the
information about the operation.
When the optic module test is finished, please
compare the result with the factory document to
make sure the function of optic module.
(Plase refer to P.82)
SystemTime Setup
Month
Day
Minute
Current Time
Year Hour
59
01
01
2014 23
2014/01/01 23:59
CancelUpdate
15
Touch Panel and User Interface
MagCore® Super / HF16 Plus provides 7-inch
full-color touch panel and easy-to-use user
interface. Operator only needs to pay attention
to the panel to know the current status, ending
time and test report of the machine. The above
figure shows the function list of the operation
screen and the description of the status icons.
Start Maintenance
Barcode Optics Data
Start Maintenance
Barcode Oprics Data
Start
Barcode
Maintenance
Optics
Execute the
extraction process Easy maintenance
and system setting
functions
Optics data browsing and
output functions
(MagCore Super only)
Barcode scanning,
editing, printing
functions
(MagCore HF16 Plus
optional)
16
RBCBioscience
━ Install Tube, Tip
Warning :
Please do not use the Tips and Tubes which are not provided by the original manufacturer.
The test result may be not correct and the machine may be damaged due to different Tips
and Tubes.
━ Install reagent cartridge━ Install MagCore® Cuvette
(MagCore Super Only / for optic test)
1. Please put MagCore® Cuvette into the
corresponding sample's well. The installation
of Cuvette is directional. It is impossible to put
it into the well if the direction is not correct.
2. Put the 200 μl SP tip into the W4 of the T-Rack.
Please insert the front end of the cartridge into the space
below the fixing plate of the Cartridge Rack.
1. Put the tip into the corresponding well according to the left
figure.
2. Put the T-Rack on the machine.
Please install the Tip and Tube
according to the instructions of
extraction kit user manual.
Please insert the Cartridge Rack before
the T-Rack.
W5 Elution Tube
W4 200μl SP Tip
W3 Pipette Tip
W2 DNase / Tip Holder set
W1 Sample Tube - FS Tip
When installing it,
please do not scratch
the optic test window.
17
Stop
━ Emergency Stop
Please read the security information to correctly stop the machine
when an accident takes place.
When an emergency situation occurs and the
machine is running, please press the stop button or
open the door of the machine to stop the machine
immediately.
18
RBCBioscience
Please confirm the number of the reagent set
which you install.
Select the number of the cartridge.
Select DNase Treatment. (Only for RNA kit.)
Select Sample Volume.
Select Elution Buffer. (Only for tissue kit.)
Please pretreat the sample according to the
instructions of the user manual of the MagCore®
Kit and put consumables into the machine.
Press Start to go to the next step.
System will ask you whether or not to scan the
barcode.
If there is a barcode record from previous test
processes, the system will ask you whether or
not to delete previous barcode record. If there is
no barcode record, the system will not ask you.
(MagCore Super only; HF16 Plus optional)
Start Maintenance
Barcode Optics Data
Select Cartridge Code
Cancel
101
102 202
201
Prev Next
610
Program Code :
DNase Treatment
Ye
sN
o
Back
201
Program Code :
Select Sample Volume
200 400
Back
401
Program Code :
Select Elution Buffer
Water Tris
Back
Start Barcode Scanning?
Yes
Barcode
No
Delete All Barcode Record ?
Yes
Barcode
No
━ Start Programs
!The number can be
found in the sticker on
the cartridge.
1
2
3
5
6
4
19
If the test result of the optic
measurement is over detection
Select to enable or disable the optic measurement
function. (MagCore Super Only)
( 201, 202, 211, 105 without the function)
Please confirm the parameters and press the Start
button to execute the program.
Automatic extraction process.
The extraction process is completed.
If you select the optic measurement funtion, you
can browse, print or output the test result to a USB
Flash Drive (please refer to the optic data chapter).
If the test result of the optic test shows the
concentration is over detection ( DNA >300ng/μl /
RNA > 240 ng/μl), the system will ask you to dilute.
A.Please record and save the
current measured data.
Please select YES.
Please check the value of each sample. If
there is a value showing DNA >300 ng/
μl or RNA > 240 ng/μl, please record
the number of the sample. It means the
sample needs to be diluted. After recording
the number of the slot, please insert the
USB Flash Drive and press the Save button
to save to current measured data. After
finishing the above steps, please press Next
Step.
10
7
8
9
102
Program Code :
Select Optic Measurement Function
Enable Disable
Back
Press Start to Process
Back
201
Program Code :
Start
N/A
200
N/A
300
Enable
DNase :
Sample Volume :
Elution Buffer :
Elution Volume :
Optical Measurement :
Program Completed
OK
Completed
Print Data Review Data OK
Program Completed
OK
Program Completed
10X Dilution ?
Print Data YesNo
Program Completed
OK
Program Completed
10X Dilution ?
Print Data YesNo
Note the samples which need dilution
Prev
Barcode
1234567890
> 300
2.05
2.364
1.154
ng / μL ( DNA )
260 / 280
A260
A280
2014 / 01 / 30 09:19:59
Date
NextNext StepSave
12345678910 11 12 13 14 15 16
Stop
Program Processing
201
Program Code :
Start
Remaining Time
Process Message2014/01/01 23:59:59
10 / 75
D2 Lamp OFF
1
The values are DNA >300 ng/μl /
RNA > 240 ng/μl
Please save the data in the USB Flash Drive
and transfer to computer, or the data in
the USB Flash Drive may be overwritten
later.
File name: OpticsSampleData.cvs
(MagCore Super Only)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96

RBCBioscience MagCore Super Training manual

Type
Training manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI