  • I have reviewed the Mounting Instructions for the Preamplifier DAM Series 3484. This document details how to properly mount the preamplifier in a 19-inch rack and provides instructions for chloridizing silver wires. I am ready to assist you with any questions about the mounting process or chloridization steps outlined in the document.
  • How to mount the preamplifiers?
    When the holders with silver wires should be chloridized?
    How to chloridize the silver wires?
    How to clean tarnished wire before chloridizing?
World Precision Instruments, Inc.
International Trade Center, 175 Sarasota Center Boulevard, Sarasota FL 34240-9258 USA
Telephone: 941-371-1003 • Fax: 941-377-5428 • E-mail: [email protected]
UK: Astonbury Farm Business Centre, Astonbury, Stevenage, Hertfordshire SG2 7EG England • Tel: 0438-880025 • Fax: 0438-880026
Germany: Liegnitzer Str. 15, D-10999 Berlin, Germany • Tel: 030-6188845 • Fax: 030-6188670
Japan: 1-4-2 Naka-Meguro, Meguro, Tokyo 153, Japan • Tel: 03-3220-5200 • Fax: 03-3220-5201
10/91
DAM Series Rackmount
Model 3484
Mounting
Instructions
WPI’s DAM Series preamplifiers
may be easily mounted in a standard 19-
inch instrument rack.
Chloridization
Holders with silver wires should be
chloridized before use. This may be
done easily by immersing the wire in 1 M
hydrochloric acid and applying +1.5 V to the
holder vs. a return electrode, such as a
large platinum wire. If, before chloridizing,
the wire is tarnished (yellow to dark red) it
may be cleaned by passing it quickly over a
flame, being careful to avoid melting the
acrylic body, or washing in urea solution.
/