Co2meter RAD-0102SX2 Calibration Procedure

  • Hello! I am an AI chatbot trained to assist you with the Co2meter RAD-0102SX2 Calibration Procedure. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Calibr a tionPr ocedure:RADCO2Saf etyAlarms
Thisisamoredetailedexplanationofthecalibrationprocedurefoundinthe
manual.Checkyourstateorlocalcodeforcalibrationschedulerequirements.
Ataminimum,youshouldcalibrateatleastonceayear.
Thereare2differentcalibrationproceduresdependingonthealtitudeatwhich
thedeviceisto
beused:
1. 0‐3,000ft.(900m):PerformZeroCalibrationandSpanCalibration.
2. 3,000+ft.:PerformZeroCalibrationonly.
BeforeYouBegin
1. ForZeroCalibration,youwillneedagascylinderofpurenitrogen(0ppm
CO2).ForSpanCalibration,youwillneeda2
nd
cylinderof4%CO2.
2. Temporarilycoverthebuzzerholeintheupperrighthandcornerofthe
SEU(sensorunit)andRDU(remotedisplayunit)witha smallpieceof
maskingtapetominimizethealarmsoundduringcalibrat ion.
3. Ifcalibrationistobeperformedinsitu,first
informbuildingoccupantsthat
youwillbetestingtheCO2alarmforseveralminutes.
4. AttachincludedtubingbetweenthegascylindersandtheSEUcalibration
portasfollows:
a. RAD0102SX2Connectthetubingtothesuppliedplugandsecureit
overtheairentry/calibrationporton
thefrontoftheSEU.
b. RAD0102/RAD0200Connectthetubingtothecalibrationporton
thebottomoftheSEU.
A:RAD0102SX2 B:RAD0102/RAD0200
2
ZeroCalibration
1. Attachtubingbetweenthenitr ogen(0ppmCO2)gas
cylinderandthecalibrationport.
2. Turnonthegasregulator.Flowrateshouldbe
between0.150.3litersperminute.
3. WaituntiltheCO2readingontheSEUsettles
consistentlyatorne ar0ppmCO2onthescreen.
4.
PresstheModebuttonuntilthe“CALI”iconflashes.
5. PressEntertoenterthe“Zero”calibrationmode.The“CALI”iconand
“ZEro”shouldbedisplayedonthescreen.
6. PressandholdtheModebuttonforatleast10secondsuntil“Calibrating”
and“CO2”areflashingsimultaneously.
7. Zerocalibration
willbegin.Afterapproximately5minutes,theLCDwill
displayeither“PASS”or“FAIL”.IftheLCDreads“FAIL”,repeatthesteps
again.
8. PressEnter.Thedisplayshouldnowshow0ppm.Removethenitrogen
cylinder.

3
SpanCalibration
Warning:YoumustbeinawellventilatedareabeforestartingSpanCalibration.
Breathing4%CO2gasinanenclosedareacanresultindizziness,shortnessof
breath,orunconsciousness.
Devicecannotbespancalibratedabove3,000ft.(900m)altitude.
1. Attachtubingbetweenthe4%CO2gascylinderand
the
calibrationport.
2. Turnonthegasregulator.Flowrateshouldbebetween
0.150.3litersperminute.
3. WaituntiltheCO2readingontheSEUsettles
consistentlyatorne ar4%CO2onthescreen.
4. BothAlarm1andAlarm2willsound.
5. PresstheMode
buttonuntilthe“CALI” iconflashes.
6. PressEnter2timestoenterthe“SPAn”calibrationmode.The“CALI”iconand
“SPAn”shouldbedisplayedonthescreen.
7. PressandholdtheModebuttonforatleast10secondsuntil“Calibrating”and
“CO2”areflashingsimultaneously.
8. Spancalibration
willbegin.Afterapproximately5minutes,theLCDwilldisplay
either“PASS”or“FAIL”.IftheLCDreads“FAIL”,repeattheste psagain.
9. PressEntertocompletecalibration.RemovetheCO2cylinder.
10. Toresetthealarm,youmustdisconnecttheSEUfromthepowersource.Wait
afewseconds,
thenplugitbackinagain.
Ifyourunitfailseither“ZEro”or“SPAn”calibration,please
followtheprocedureagainbeforecallingsupport.
/