Endres+Hauser Liquiline M CM330-B Operating instructions

Type
Operating instructions
BA415C/07/en/02.09
71070992
Valid as of:
Software version 04.00.00
Operating Instructions
Liquiline M CM330
Four-wire transmitter for two-channel measurement of
pH value and disinfection parameters with digital sensors
Part 2: Operation
Operating concept
a0007005
Fig. 1: Press soft key: select menu directly
a0003809-en
Fig. 2: Turn navigator: move cursor in the menu
a0003810-en
Fig. 3: Press navigator: select a function
a0003811-en
Fig. 4: Turn navigator: change value
a0003813-en
Fig. 5: Press navigator: select new value
Operating concept
1. You select a menu directly by pressing the
soft key in question.
2. You move the cursor in the menu by
turning the navigator.
3. Press the navigator and select the desired
function.
4. Change the value by turning the
navigator.
5. Press the navigator to accept the new
value.
Endress+Hauser
Liquiline M CM330
Table of contents
1 Display . . . . . . . . . . . . . . . . . . . 4
1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Status messages . . . . . . . . . . . . . . . . . . . . 4
2 Notes on software description . 5
2.1 Types of setting . . . . . . . . . . . . . . . . . . . . 5
2.2 Editing tables . . . . . . . . . . . . . . . . . . . . . . 5
3 Measure (MEAS) . . . . . . . . . . . 6
4 Specifying the parameters
(PARAM) . . . . . . . . . . . . . . . . . 6
4.1 Menu structure, top hierarchy level . . . . . 6
4.2 Sensor DI . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.3 Sensor pH . . . . . . . . . . . . . . . . . . . . . . . 10
4.4 Current output . . . . . . . . . . . . . . . . . . . . 14
4.5 General settings . . . . . . . . . . . . . . . . . . . 15
4.6 Display . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Device diagnosis (DIAG) . . . . 19
5.1 Menu structure, top hierarchy level . . . . 19
5.2 Errors/messages . . . . . . . . . . . . . . . . . . 19
5.3 Sensor state . . . . . . . . . . . . . . . . . . . . . . 19
5.4 Output state . . . . . . . . . . . . . . . . . . . . . . 20
5.5 Sensor information . . . . . . . . . . . . . . . . . 20
5.6 Device information . . . . . . . . . . . . . . . . 22
5.7 Service . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6 Calibration (CAL) . . . . . . . . . . 25
6.1 Types of calibration . . . . . . . . . . . . . . . . 25
6.2 Calibrate . . . . . . . . . . . . . . . . . . . . . . . . 25
6.3 Calibration menu . . . . . . . . . . . . . . . . . . 26
7 Maintenance. . . . . . . . . . . . . . 30
8 Troubleshooting . . . . . . . . . . . 30
8.1 Troubleshooting instructions . . . . . . . . . 30
8.2 Diagnosis messages . . . . . . . . . . . . . . . . 31
8.3 Process errors without messages . . . . . . . 35
8.4 Spare parts . . . . . . . . . . . . . . . . . . . . . . . 38
8.5 Return . . . . . . . . . . . . . . . . . . . . . . . . . . 40
8.6 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . 40
8.7 Software history . . . . . . . . . . . . . . . . . . . 40
9 Replacing modules . . . . . . . . .41
Index . . . . . . . . . . . . . . . . . . .42
Display Liquiline M CM330
4Endress+Hauser
1Display
1.1 Overview
1.2 Status messages
a0007008
Fig. 6: Local display (measuring mode)
1
2
3
Status line
Display and setting field
Assignment of the soft keys
Symbol for the measuring menu (MEAS)
Device status is OK. No alarms or warnings are present.
Service warning. Maintenance is recommended.
The device is operational; the reliability of the measured value may be restricted.
Alarm that points to an error.
The device is no longer operational or the measured value is no longer reliable.
Measured value display in the status line if a menu other than MEAS is launched.
The main measured values for channel 1 and channel 2 are displayed.
Hold is set to "ON", the measured value is "frozen".
Symbol for the simulation mode
Symbol for the calibration type "Sample calibration"
The sample is taken, laboratory analysis is running. The device is waiting for input.
I
II
I
II
7.68 pH
450 mV
7.68 pH
450 mV
Liquiline M CM330 Notes on software description
Endress+Hauser 5
2 Notes on software description
2.1 Types of setting
Display fields
You can only read the values, not change them.
Selection fields
You receive a list with options.
You select one of these options.
Input fields
There are value ranges with upper and lower range limits that depend on the measured
variable configured and its units.
There are also menu functions where you can enter arbitrary text. The number of
characters is then limited.
Set a value with the navigator:
Turn to increase/reduce a value/letter/special character
Press to confirm or to enter the next character for arbitrary text.
2.2 Editing tables
You can set some software functions using a table:
The number of columns depends on the menu function that is set via the table.
You can add lines ("Insert") or delete lines ("Del").
The maximum number of lines also depends on the menu function in question.
You can press "Esc" at any time to exit the table and stop entering information.
If the values entered result in a valid table, you receive the message "Table is valid" and the
options:
–Save table
Continue edit table
Discard table
If the values entered are invalid, an error message is output with the options:
Continue edit table
Discard table
Measure (MEAS) Liquiline M CM330
6Endress+Hauser
3Measure (MEAS)
!Note!
In the measuring menu, you can switch between four different types of display. To do so, simply
press the enter button of the navigator.
4 Specifying the parameters (PARAM)
4.1 Menu structure, top hierarchy level
4.2 Sensor DI
4.2.1 Menu structure
Function name of local operation (Display)
Meas
Main values channel 1 + channel 2
Main value channel 1 and temperature
Main value channel 2
All measured values
DI value
pH value
Temperature
Sensor current
Current output 1
Current output 2
PARAM
Sensor DI
Sensor pH
Current output
General settings
Display
Function name of display (local operation)
PARAM
Sensor DI
Measured value
Concentration (Measured value="Concentration")
Damping
Liquiline M CM330 Specifying the parameters (PARAM)
Endress+Hauser 7
4.2.2 Configuration options
Compensation pH
Switch
Mode
Preset pH (Mode="Fixed value")
Sensor pol. voltage
Switch
Adjust (Switch="off")
Clean timer (Switch="off")
Switch
Pol. voltg.
Repetition rate
Power on time
Hangover time
Temp. compensation
Medium Temperature
(Temp. compensation="Man. comp.")
Sensordiagnosis
Diagnosis limits
Slope DI
Upper alarm value
Upper warning value
Lower warning value
Lower alarm value
Zero pnt. DI
Upper alarm value
Upper warning value
Lower warning value
Lower alarm value
Function Options Info
Measured value Options
• concentration
•nA
Factory setting
concentration
The measured value can be displayed as concentration of
the disinfection reagent or as sensor current.
Concentration Options
•mg/l
μg/l
• ppm
• ppb
Factory setting
mg/l
Function name of display (local operation)
Specifying the parameters (PARAM) Liquiline M CM330
8Endress+Hauser
Damping 0.0 to 20.0 s
Factory setting
0.0 s
The damping causes a floating average curve of the
measured values over the time specified.
Compensation pH Switch=off
The concentration value is calculated as HClO.
Switch=on
The pH value is used to calculate a cumulative
concentration value of HClO and ClO- (total DI).
You can choose either to use the measured value from the
connected pH sensor or a fixed value, e.g. when no pH
sensor is connected.
Switch Options
•on
•off
Factory setting
on
Mode Options
Fixed value
Measured value
Factory setting
Fixed value
Preset pH 0.0 to 9.0
Factory setting
7.2
only when Mode="Fixed value"
Check the measured water for the right pH value here.
Sensor pol. voltage Select the sensor polarization voltage source and the value.
Switch Options
•Internal
• External
•off
Factory setting
Internal
Use "Internal" to use the factory calibration value.
Use "External" for special applications like sensor
regeneration. "External" will activate the automatic cleaning
option "Clean timer".
Adjust -750 to 0 mV
Factory setting
-20 mV
Preset value for continous measurement
Function Options Info
Liquiline M CM330 Specifying the parameters (PARAM)
Endress+Hauser 9
Clean timer The sensor will be cleaned periodically by applying a lower
polarization voltage.
Clean timer sequence:
normal mesurement -->
hold on (if PARAM/General settings/Hold
settings/"Clean active"="set value") -->
cleaning -->
•hangover time -->
hold delay ("Hold settings") -->
hold off -->
normal measurement
Switch Options
•on
•off
Factory setting
off
Pol. voltg. -750 to 0 mV
Factory setting
-400 mV
Repetition rate 1 to 48 h
Factory setting
3 h
Power on time 10 to 180 s
Factory setting
30 s
Hangover time 0 to 180 s
Factory setting
0 s
Temp. compensation Options
•Off
•Auto. comp. (ATC)
•Man. comp.
Factory setting
Auto. comp. (ATC)
Medium temperature –50 to + 250 °C
(–58 to + 482 °F)
Factory setting
20 °C (68 °F)
Only for "Temp. compensation" = "Man. comp.
(MTC)".
Function Options Info
Specifying the parameters (PARAM) Liquiline M CM330
10 Endress+Hauser
4.3 Sensor pH
4.3.1 Menu structure
Sensordiagnosis Adjustment of diagnosis messages
Diagnosis limits
Slope DI Selection and entry
Upper alarm value
Factory setting
400 %
Upper warning value
Factory setting
200 %
Lower warning value
Factory setting
25 %
Lower alarm value
Factory setting
10 %
General range of
adjustment
0to500 %
!Note!
Set values affect the range of the corresponding variable
(maint. value > alarm value).
Zero point DI Selection and entry
Upper alarm value
Factory setting
2.00 nA
Upper warning value
Factory setting
1.00 nA
Lower warning value
Factory setting
-1.00 nA
Lower alarm value
Factory setting
-2.00 nA
General range of
adjustment
-3.2 to 3.2 nA
!Note!
Set values affect the range of the corresponding variable
(maint. value > alarm value).
Function name of display (local operation)
PARAM
Sensor pH
Measured value
Damping
Cal. settings
Type of calibration
Function Options Info
Liquiline M CM330 Specifying the parameters (PARAM)
Endress+Hauser 11
4.3.2 Configuration options
Temp. compensation
Medium Temperature
(Temp. compensation="Man. comp.")
Buffer recognition
Buffer manufacturer
(Buffer recognition="Automatic", "Fixed")
Calib.-buffer1
(Buffer recognition="Automatic", "Fixed")
Calib.-buffer 2
(Buffer recognition="Automatic", "Fixed")
Isotherm pnt.
pH
Temp. compensation
Medium Temperature
(Temp. compensation="Man. comp.")
Sensordiagnosis
Diagnosis limits
Slope pH
Alarm value
Maint. value
Zeropnt. Glass
Upper alarm value
Upper warning value
Lower warning value
Lower alarm value
Stable criteria
Bandwidth
Timeframe
SCS glass 1
Function Options Info
Measured value Options
•pH
Damping 0.0 to 10.0 s
Factory setting
0.0 s
The damping causes a floating average curve of the
measured values over the time specified.
Cal settings
Type of calibration Options
2-point cal.
Factory setting
2-point cal.
Function name of display (local operation)
Specifying the parameters (PARAM) Liquiline M CM330
12 Endress+Hauser
Temp. compensation Options
•Off
•Auto comp. (ATC)
•Man. comp.
Factory setting
Auto comp. (ATC)
!Note!
This setting only refers to compensation during calibration,
not in measuring mode.
Medium
Temperature
–50 to + 250 °C
(–58 to + 482 °F)
Factory setting
20 °C (68 °F)
Enter the temperature of your calibration buffer.
Buffer recognition Options
•Automatic
•Fixed
•Manual
Factory setting
Fixed
If you choose "Fixed" , select fixed values from a list
depending on the buffer manufacturer.
If you chose "Manual," you can specify any two buffer
values of your choice. These must differ in terms of their pH
value.
Buffer manufacturer Options
•E+H (NIST)
•Ingold/Mettler
DIN 19267
DIN 19266
• Merck/Riedel
Factory setting
E+H (NIST)
Temperature values are stored internally in the unit for the
following pH values:
E+H: 2.00 / 4.00 / 7.00 / 9.00 / 9.20 / 10.00 / 12.00
Ingold: 2.00 / 4.01 / 7.00 / 9.21
DIN 19266: 1.68 / 4.01 / 6.86 / 9.18
DIN 19267: 1.09 / 4.65 / 6.79 / 9.23 / 12.75
Merck: 2.00 / 4.01 / 6.98 / 8.95 / 12.00
Calib. buffer 1 –2.0 to 16.0 pH
Factory setting
4.01 pH
Buffer recognition = Manual:
Configure the value of the buffer you use.
Buffer recognition = Fixed:
Select a value from the list.
Calib. buffer 2 –2.0 to 16.0 pH
Factory setting
6.98 pH
Isotherm pnt. –2.0 to 16.0 pH
Factory setting
7.0 pH
Isotherm intersection
Take the value from the manufacturer's certificate for your
sensor.
Temp. compensation Options
•Off
Auto. comp. (ATC)
•Man. comp.
Factory setting
Auto. comp. (ATC)
Automatically using the temperature sensor of your
sensor (ATC)
Manually by entering the process temperature (Man.
comp.) or
Not at all (Off).
Medium temperature –50 to + 250 °C
(–58 to + 482 °F)
Factory setting
20 °C (68 °F)
Enter the process temperature.
Function Options Info
Liquiline M CM330 Specifying the parameters (PARAM)
Endress+Hauser 13
Sensordiagnosis
Diagnosis limits
Slope pH (1) Selection and entry
Alarm value
5.00 to 99.99 mV/pH
Factory setting
53.00 mV/pH
•Maint. value
5.00 to 99.99 mV/pH
Factory setting
55.00 mV/pH
!Note!
Set values affect the range of the corresponding variable
(maint. value > alarm value).
The slope characterizes the sensor condition. The bigger the
deviation from the ideal value (59 mV/pH) the poorer the
condition of the sensor. You can also track this variable
graphically in the "DIAG/Sensor state" menu.
Zeropnt. glass Selection and entry
Upper alarm value
Factory setting
9.00 pH
Upper warning value
Factory setting
8.00 pH
Lower warning value
Factory setting
6.00 pH
Lower alarm value
Factory setting
5.00 pH
General range of
adjustment
–2 to 16.00 pH (glass)
Enter specific values for the alarm and warning limits for
your process.
Stable criteria Selection and entry
•Bandwidth
1to10mV
Factory setting
1mV
•Timeframe
5to60s
Factory setting
15 s
With the bandwidth, you define the permitted measured
value fluctuation which must not be exceeded in a certain
timeframe during calibration.
Calibration is not permitted if the permitted difference is
exceeded. In this case, calibration is aborted with an error
after maximum 60 seconds.
SCS glass 1 Options
•Off
•On
Factory setting
Off
The SCS (Sensor Check System) monitors the pH and
reference electrode for incorrect measurement and total
failure. --> CD-ROM
Function Options Info
Specifying the parameters (PARAM) Liquiline M CM330
14 Endress+Hauser
4.4 Current output
4.4.1 Menu structure
4.4.2 Configuration options
Function name of display (local operation)
PARAM
Current output
Current output 1
Output value
Low value (4 mA)
Upper value (20 mA)
Current output ´2
Output value
Low value (4mA)
Upper value (20mA)
Function Options Info
Current output 1
Output value Fix
•DI
The DI value is permanently assigned as an output variable
at current output 1.
Low value (4mA) 0 to 50 mg/l
Factory setting
0 mg/l
Using the 4 mA and 20 mA value, you map the measuring
range in the corresponding engineering unit of the
measured variable. The smaller the span you select between
the 4 mA and 20 mA value, the more accurately you map
this measuring range.
!Note!
The following minimum spans are settings that the software
allows but might not necessarily be useful for your process.
For this reason, use values that are plausible for your
process, within the limits of which you can generally expect
your measured values.
The smallest possible spread between 4 mA and 20 mA
value is:
0.1 mg/l
Upper value (20mA) 0 to 50 mg/l
Factory setting
5 mg/l
Current output 2
Output value Options
•pH
•Temperature
Factory setting
pH
Specify the measured variable assigned at current output 2.
Liquiline M CM330 Specifying the parameters (PARAM)
Endress+Hauser 15
4.5 General settings
4.5.1 Menu structure
Low value (4mA) –2 to 16 pH
–10 to 140 °C (14 to 95 °F)
Factory setting
2 pH
0 °C (32 °F)
Using the 4 mA and 20 mA value, you map the temperature
measuring range in °C or °F. The smaller the span you
select between the 4 mA and 20 mA value, the more
accurately you map the temperature measuring range.
The smallest possible spread between 4 mA and 20 mA
value is:
•1 pH
5 °C (9 °F).
Upper value (20mA) –2 to 16 pH
–10 to 140 °C (23 to 284 °F)
Factory setting
12 pH
40 °C (104 °F)
Function name of display (local operation)
PARAM
General settings
TAG number
Date format
Set date
Time format
Set time
Hold settings
Calib active
Calib value ("Calib active"="set value")
Param active
Param value ("Param active"="set value")
Diag active
Diag. value ("Diag active"="set value")
Clean active
Clean value ("Clean active"="set value")
Hold delay
User admin.
Login
(not if "Specialist" is already logged on)
Password protect.
Function Options Info
Specifying the parameters (PARAM) Liquiline M CM330
16 Endress+Hauser
4.5.2 Configuration options
Function Options Info
TAG number Can be edited at random Max. 32 characters
Date format Options
• dd.mm.yyyy
• mm.dd.yyyy
Factory setting
mm.dd.yyyy
Editing mode:
DD (day): 1 to 31
MM (month): 1 to 12
YYYY (year): 2005 to 2100
Set date Depends on the format
Time format Options
hh.mm (24 h)
•hh.mm (am/pm)
Factory setting
hh.mm (am/pm)
24-hour display or 12-hour display
Editing mode:
hh (hour): 0 to 23 / 0 am to 12 pm
mm (minutes): 0 to 59
ss (seconds): 0 to 59
Set time Depends on the format
Hold settings Activation of Hold functionality valid for all current outputs
Calib active Options
•No hold
• Freeze
•Fixed
Factory setting
No hold
• "Freeze":
The device keeps the last measured value.
"Set value":
You define a set display value.
Calib value
Calib active="Set
value"
3.6 to 22 mA
Factory setting
12 mA
Enter the value which you want to have displayed during
calibration.
Param active Options
•No hold
• Freeze
•Fixed
Factory setting
No hold
• "Freeze":
The device keeps the last measured value.
"Set value":
You define a set display value.
Param value
Setup active="Set
value"
3.6 to 22 mA
Factory setting
12 mA
Enter the value which you want to have displayed during
configuration.
Diag active Options
•No hold
• Freeze
•Fixed
Factory setting
No hold
• "Freeze":
The device keeps the last measured value.
"Set value":
You define a set display value.
Diag. value
Diag active="Set
value"
3.6 to 22 mA
Factory setting
12 mA
Enter the value which you want to have displayed during
diagnosis.
Liquiline M CM330 Specifying the parameters (PARAM)
Endress+Hauser 17
Clean active Options
•No hold
• Freeze
•Fixed
Factory setting
No hold
• "Freeze":
The device keeps the last measured value.
"Set value":
You define a set display value.
Clean value
Clean active="Set
value"
3.6 to 22 mA
Factory setting
12 mA
Enter the value which you want to have displayed during
sensor cleaning.
Hold delay 0 to 60 s
Factory setting
5 s
After changing to the measuring mode, the specified hold is
maintained for the hold delay period.
User administration
Login Options
• Specialist
• Maintenance
Factory setting
Maintenance
Once the user is selected, you are prompted to enter a
6-digit code of your choice.
Factory setting is empty.
!Note!
More information is provided on the user roles under
"Commissioning" in the first part of the Operating
Instructions.
Password protect. Options
•None
•Enter code
Factory setting
None
"Caution!
The "Enter code" option is only visible if you are logged on
as a "Specialist"!
Enter code Here, you can enter a 6-digit code for the user roles "Specialist" and "Maintenance" (you
have to be logged on as a "Specialist"!).
Specialist Options
Password
Factory setting
Empty (nothing entered)
•Store
Once you have specified a password of your choice, go to
Store (= Enter).
The message "Stored new password" and "OK" is displayed.
Maintenance
Function Options Info
Specifying the parameters (PARAM) Liquiline M CM330
18 Endress+Hauser
4.6 Display
4.6.1 Menu structure
4.6.2 Configuration options
Function name of display (local operation)
PARAM
Display
Language
DI format
pH format
Temperature unit
Temperature format
Function Options Info
Language Options
• English
•German
Factory setting
English
If you select the other language in question, all the other
settings remain intact.
DI format Options
•x
•x.x
•x.xx
•x.xxx
Factory setting
x.xx
You select how many commas should appear after the
decimal point in the measured value display.
pH format Options
•x
•x.x
•x.xx
Factory setting
x.xx
Temperature unit Options
•°C
•°F
Factory setting
° F
Temperature format Options
•x
•x.x
Factory setting
x.x
You select how many commas should appear after the
decimal point in the temperature display.
Liquiline M CM330 Device diagnosis (DIAG)
Endress+Hauser 19
5 Device diagnosis (DIAG)
!Note!
In the DIAG menu you will find information about the device state, in particular detailed error
and maintenance messages.
In addition to this, there are various service functions available1.
5.1 Menu structure, top hierarchy level
5.2 Errors/messages
!Note!
An explanation of the causes of the messages displayed here, and possible remedial measures,
can be found in the "Troubleshooting"/"Diagnosis messages" section.
5.3 Sensor state
1) Depending on the device version
DIAG
Errors/messages
Sensor state
Output state
Sensor information
Device information
Service
Function name of display (local display)
DIAG
Errors/messages
Function name of display (local display)
DIAG
Sensor state
Sensor DI
Sensor pH
Device diagnosis (DIAG) Liquiline M CM330
20 Endress+Hauser
5.4 Output state
5.5 Sensor information
5.5.1 Sensor channel 1
Function name of display (local display)
DIAG
Output state
Current output 1
Output source
Current output 2
Output source
Output range
Function name of display
DIAG
Memosens data: Sensor types channel 1 and channel 2
CCS142D
Identification
Hardware identifier
Serial number
Order code
Hardware version
Software version
Type of calibration
<Type of calibration> (last used)
Zero
Slope
Date of calibration zero
Time of calibration zero
Date of calibration slope
Time of calibration slope
No. cal. sensor
No. cal. cap
Delta zeropnt.
Delta slope
SN Transmitter
(= with which the last calibration was)
Temp. Calibration
Temperature offset
Temp. cal. date
Temp. cal. Time
Operating hours
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44