Roche cobas IT middleware User manual

Category
Software
Type
User manual
cobas
®
IT middleware
User Manual version 6
Software version 1.05.00
cobas IT middleware
Roche Diagnostics
2 User Manual · version 6
Document information
Editor’s note Every effort has been made to ensure that the information contained in this manual is
accurate at the time of printing. Not all functionality described in this manual may be
available to all users. Roche Diagnostics International Ltd reserves the right to make
any further required changes to software without prior notice. Such changes may not
immediately be reflected in this document.
Intended use
This document is intended for the users of the cobas
®
IT middleware system, version
1.05.00.
The cobas
®
IT middleware is a middleware solution (MWS) for the diagnostic
laboratory. Laboratories are located in hospitals or run in private ownership. The
cobas
®
IT middleware can be connected to one or several of the following systems:
o Roche pre-analytic instruments
o Roche post-analytical instruments
o Roche analytical instruments (biochemistry, immunology, urine analysis,
coagulation)
o Roche live view (customer dashboard displaying operational KPIs based on data
generated by cobas
®
IT middleware)
o Non-Roche analytical instruments (biochemistry, immunology, urine analysis,
coagulation, hematology, and molecular and tissue diagnostics)
o Laboratory information systems
o Hospital information systems
o Electronic health record systems
o Non-Roche work area management solutions for hematology, clinical chemistry,
or urine-analysis.
Connectivity between cobas
®
IT middleware and the instruments or systems
mentioned above endorse to exchange data uni- or bidirectionally via interface. Uni-
and bidirectionally transmitted data contain patient information, order data, and
result data. The solution manages the data in a central database.
Copyright © 2012-2015, F. Hoffmann-La Roche Ltd. All rights reserved.
Manual version Software version Revision dates Main change
1.00.00 (revision 1) 1.00.00 January 2012 First version of the
manual.
1.01.00 (revision1) 1.01.00 August 2012 Software and manual
updates.
1.02.00 (revision1) 1.02.00 June 2013 Software and manual
updates.
Version 4 1.03.00 February 2014 Software and manual
updates.
Version 5 1.04.00 July 2014 Software and manual
updates.
Version 6 1.05.00 January 2015 Software and manual
updates.
Table 1 Revision history
Roche Diagnostics
User Manual · version 6 3
cobas IT middleware
Trademarks Recognized trademarks:
o ROCHE, COBAS, and LIFE NEEDS ANSWERS
o Roche and the Roche symbol are registered trademarks of the Roche group.
o All other trademarks are trademarks of their owners.
Feedback Every effort has been made to ensure that this manual fulfils its intended purpose as
mentioned above. All feedback on any aspect of this manual is welcome and is
considered during updates. Contact your Roche representative, should you have any
such feedback.
Contact addresses
Manufacturer
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Made in Switzerland
cobas IT middleware
Roche Diagnostics
4 User Manual · version 6
Roche Diagnostics
User Manual · version 6 5
cobas IT middleware
Table of contents
Document information 2
Contact addresses 3
Table of contents 5
Using this manual 7
Introduction
1 System overview
About cobas® IT middleware 15
Creating/deleting filters 20
Configuring the columns 22
Reviewing the error log files 23
Viewing the details of an error log 24
Task overview 25
2 Understanding the user interface
Main screen structure 31
Navigation bar 34
Tabs and panels 35
Generic user interface elements 38
Routine
3 Sample distribution and sample management
Sample activities before distribution 45
Sample distribution 52
Offline workplaces 60
Sample archiving 63
Sample retrieval 68
Sample disposal 69
Late sample tracking 71
About closed samples 75
4 Test result processing
About result evaluation and release 79
5 Patient management
About patient management 91
QC
6 QC configuration and routine QC
About QC configuration 99
About multi-rules QC 106
About profiles 109
About QC study lots 116
About QC 117
About calibration 131
7 Consumption report
About the consumption report 135
Configuration
8 Configuration
Laboratory configuration 143
Test configuration 150
Reference ranges configuration 156
Validation status configuration 161
Rule engine 163
Serum indices 164
Instrument configuration 167
Host Administration 176
Sample configuration 177
User administration configuration 185
Forms 191
System configuration 194
9 Rule management
About rules 199
Test configuration rules 201
Rule engine rules 206
Working examples 219
Appendix
10 Glossary
Index
Index 229
Roche Diagnostics
6 Online Help · version 6
3 cobas IT middleware
Roche Diagnostics
User Manual · version 6 7
cobas IT middleware
Using this manual
Ensure the following:
Q
o Keep this manual in a place where it will not be damaged.
o Ensure that this document is available at all times.
Conventions used in this manual
You can find detailed information on each task in this User Manual or by choosing
the Assistance button in the global information area.
The following symbols and conventions are used in this manual.
Symbols The following symbols may be used to draw your attention to important information:
Abbreviations The following abbreviations are used:
What is new in software version <1.05>
live view module removed The description of the live view module has been removed. This feature is now
described in separate user documentation.
Symbol Meaning
U Cross-reference
P Procedure start
S Procedure end
o List item
Q Tip
Safety alert
Table 2 Information symbols
Abbreviations Definitions
H
HIS hospital information system
L
LIS laboratory information system
M
MPA MODULAR PRE- ANALYTICS
Q
QC quality control
S
SD standard deviation
STAT short turnaround time
Table 3 Abbreviations
cobas IT middleware
Roche Diagnostics
8 User Manual · version 6
About closed samples New section. Description of closed samples, which can lead to failed host messages.
U About closed samples (p. 75)
Ordering a repeat test New section. You order a repeat test to perform the same processing parameters on a
sample.
U Ordering a repeat test (p. 81)
Filtering QC Updated. New filtering criteria for QC results.
U Filtering QC (p. 16)
Unlocking QC results for a test-
instrument combination
New section. You can define a test-instrument combination when manually releasing
a QC lock.
U Unlocking QC results for a test-instrument combination (p. 123)
Filtering cumulative QC statistics New section. New filtering criteria are available for filtering cumulative QC statistic.
U Filtering cumulative QC statistics (p. 126)
Applying the calculated mean and
SD to cumulative QC statistics
across instruments
New section. You can now apply the calculated mean and SD to QC statistics over
multiple instruments.
U Applying the calculated mean and SD to cumulative QC statistics across instruments
(p. 127)
Viewing time-based rules profiles Updated. Description of terms used in the Time-based rule panel.
U Viewing time-based rules profiles (p. 128)
Creating time-based rules profiles New section. This function allows you to assign a new time-based rules profile.
U Creating time-based rules profiles (p. 128)
Editing time-based rules profiles New section. This function allows you to change an existing time-based rules profile.
U Editing time-based rules profiles (p. 129)
Safety classifications
The following safety information applies to the system and is listed in accordance
with ANSI Z535.6-2006. The Danger safety classification to indicate a situation that
will result in injury or death is not applicable to this system and is therefore not listed.
Safety alert symbol
r The safety alert symbol alone promotes awareness to hazards that are generic or direct
the reader to safety information provided elsewhere in the document.
These symbols and signal words are used for specific hazards:
WARNING
Warning
r Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
Caution
r Indicates a hazardous situation which, if not avoided, could result in minor or moderate
injury.
Roche Diagnostics
User Manual · version 6 9
cobas IT middleware
NOTICE
Notice
r Indicates a hazardous situation which, if not avoided, may result in damage to the
system.
Important information that is not safety relevant is indicated with the following icon:
Q
Tip
Indicates additional information on correct use or useful tips.
System safety information
Failure to observe the following safety information may result in incorrect results,
data corruption, and data losses.
NOTICE
Incorrect operation or the use of wrong components
The data in the system can be incorrect or corrupt.
r Use only computers, monitors, printers, and accessories recommended by the
manufacturer.
Service your computer regularly (defragment the hard disk; install, run, and update
antivirus software; check the event log for system error entries).
NOTICE
Corrupt data due to viruses
Software viruses can corrupt system data.
r Install a firewall, maintain up to date antivirus software, and keep your operating
system up to date.
CAUTION
Incorrect or corrupt data due to unauthorized access
Data security is breached if unauthorized users have access to your user ID and password.
r Always enter your password unobserved.
r Do not write down your password.
r Never write down the password in a contact form, in the address book, or in a file on
the computer.
r Do not disclose your password to anyone.
r Roche will never ask you for your password.
r If you ever disclose your password to anyone, change it immediately after.
r If you think anyone else might have access to your account, then contact your local
Roche affiliate.
NOTICE
The host interface does not support secure data transmission
When using host communication, your infrastructure must ensure that transported data is
protected appropriately. It is your responsibility that access to the network is restricted only
to authorized entities and/or the communication protocol is secured.
r This can be achieved by using system level tunneling protocols like TLS or other means
of network encryption.
NOTICE
Internet Explorer
Ensure that you use the 32 bit Internet Explorer.
r 64-bit Internet Explorer is not supported.
cobas IT middleware
Roche Diagnostics
10 User Manual · version 6
NOTICE
Data is not encrypted
Unauthorized access to data.
r Your infrastructure must ensure that access to the data storage and backup files is
permitted only to authorized entities.
NOTICE
Disaster recovery
Unexpected disasters, such as hard disk failures, system software failures or natural
disasters such as fires or floods can occur.
r Your IT department must assemble a disaster recovery kit to protect your entire system.
Introduction
1 System overview ............................................................................................................................... 13
2 Understanding the user interface .................................................................................................... 29
Roche Diagnostics
User Manual · version 6 13
cobas IT middleware 1 System overview
Table of contents
System overview 1
In this chapter you can find information about accessing the system, filtering options,
column configuration, and the opened tasks and actions that you need to address.
In this chapter
Chapter
1
About cobas® IT middleware ..............................................................................................15
Logging into the system ................................................................................................15
Changing the password.................................................................................................15
Filtering options .............................................................................................................16
Filtering samples ......................................................................................................16
Filtering QC..............................................................................................................16
Filtering multi-rules.................................................................................................17
Filtering profiles.......................................................................................................17
Filtering patients ......................................................................................................17
Filtering test results..................................................................................................18
Filtering the list of racks..........................................................................................18
Filtering the error log list ........................................................................................18
Filtering the list of consumption reports ..............................................................19
Filtering the late samples list ..................................................................................19
Creating/deleting filters.......................................................................................................20
Configuring the columns ....................................................................................................22
Reviewing the error log files ...............................................................................................23
Viewing the details of an error log.....................................................................................24
Task overview........................................................................................................................25
Displaying pending tasks ..............................................................................................26
Configuring missing sample results ......................................................................26
Roche Diagnostics
14 User Manual · version 6
1 System overview cobas IT middleware
Table of contents
Roche Diagnostics
User Manual · version 6 15
cobas IT middleware 1 System overview
About cobas® IT middleware
About cobas® IT middleware
This system is a middleware solution for central laboratories, which provides the
following functionality:
o Sample entry
o Sample distribution
o Sample validation
o Sample archiving and retrieval
o QC administration
o Patient management
o Test management
The functionality available is dependent on the level of access granted to each user
account.
The cobas® IT middleware can connect to an LIS (laboratory information system) or
HIS (hospital information system), and one or several Roche pre-analytic, analytic,
and Roche post-analytic instruments.
Logging into the system
You log into cobas® IT middleware to start using the system.
P To log into the system
1
In the global information area, choose the Log on button.
The Log on dialog box is displayed.
2
Enter your user name, password, select a language, and choose the Log on button.
You are logged into the system.
S
Changing the password
You change your password when the current one expires or is no longer safe to use.
Q
Password format
Here, you cannot enter a password that starts with a number or special characters.
U For details of password security, see Incorrect or corrupt data due to unauthorized access
(p. 9).
P To change your password
1
In the global information area, choose the Log on button.
The Log on dialog box is displayed.
2
Choose the Change password button.
The Log on dialog box expands to display the Change password options.
Roche Diagnostics
16 User Manual · version 6
1 System overview cobas IT middleware
About cobas® IT middleware
3
Enter the old password, the new password twice, and then choose the Save
button.
The new password is saved in the system.
S
Filtering options
There are several workplaces where you can filter the data which is displayed. You
can predefine filters to help you use preferred filtering options. Depending on the
panel you are in, the filtering criteria can display different selections.
You can group and sort the data that you filtered for to facilitate information search.
Q
You can sort data in ascending or descending order by clicking the column headers.
Q
When no date is defined in the filtering criteria, a range of ten days is used by default.
Therefore samples older than ten days are not shown in the filtered list.
Q
By default, the filters only show results that are 10 days old or less. This value is
configurable. If necessary, talk to your Roche Service Representative.
Filtering samples
P To filter samples
1
Choose Routine >Release tests by sample.
The Samples: {0}panel is displayed containing the list of samples.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering QC
Q
To aid in searching for specific QC results, the following options allow you to search by
multiple criteria in the Filter by drop-down list:
o QC lot
o QC material
o Result status
o Sample type
P To filter QC
1
Choose Routine >Review QC results.
The QC results: {0}panel is displayed containing the list of QC results.
Roche Diagnostics
User Manual · version 6 17
cobas IT middleware 1 System overview
About cobas® IT middleware
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering multi-rules
P To filter multi-rules
1
Choose Monitoring > Rules.
The Rules: {0} panel is displayed containing list of all defined rules.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering profiles
P To filter profiles
1
Choose Monitoring > Profiles.
The Profiles: {0} panel is displayed containing all profiles that fit the default filter
criteria.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering patients
P To filter patients
1
Choose Routine > Create and update patients.
The Patients: {0}panel is displayed containing a list of patients.
2
Choose the button.
The filtering panel is displayed.
Roche Diagnostics
18 User Manual · version 6
1 System overview cobas IT middleware
About cobas® IT middleware
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering test results
P To filter tests results
1
Choose Routine > View test results.
The Tests: {0}panel is displayed containing the list of tests.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering the list of racks
P To filter the list of racks
1
Choose Monitoring > Manage racks and dispose of samples.
The Rack management: {0} panel is displayed containing a list of racks that fit the
default filter criteria.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering the error log list
P To filter the error log list
1
Choose Administration > View error log files.
The Error log files: {0} panel is displayed, containing a list of error that fit the
default filter criteria.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Roche Diagnostics
User Manual · version 6 19
cobas IT middleware 1 System overview
About cobas® IT middleware
Filtering the list of consumption reports
P To filter the list of consumption reports
1
Choose Monitoring > Consumption report.
The Consumption report panel is displayed containing a list of consumption
reports that fit the default filter criteria.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria you want to filter by, and
then choose the Apply button.
The matching data is displayed.
S
Filtering the late samples list
P To filter the list of late samples
1
Choose Monitoring > Late sample monitor.
The Late samples list is displayed containing a list of all samples which have
violated a late sample rule.
2
Choose the button.
The filtering panel is displayed.
3
From the filtering drop-down lists, choose the criteria by which you want to filter,
and then choose the Apply button.
The matching data is displayed.
S
Roche Diagnostics
20 User Manual · version 6
1 System overview cobas IT middleware
Creating/deleting filters
Creating/deleting filters
You can create custom filters but you can only have one default filter which applies
every time you open a panel with a filter. The ADMIN user can change the filters
globally, but the Routine users can only modify their personal filters. If you delete
your default filter, the ADMIN default filter applies on your next logon.
The filtering criteria are available in the following panels:
o Samples: {0}
o Retrieve samples
o QC results: {0}
o View cumulative QC statistics
o Patients: {0}
o Tests: {0}
o Rules: {0}
o Profiles: {0}
o QC instrument assignment: {0}
o Test/instrument assignments: {0}
o Rack management: {0}
o Error log files: {0}
Filtering criteria which support multi-selection filtering is available for:
P To create a custom filter
1
Choose Routine > Create and update samples.
The Samples: {0}panel is displayed containing the list of samples.
Panel Filter
Samples: {0} Instrument
Orderer
Orderer group
Tests: {0} Instrument
Orderer
Orderer group
Te s t
QC results: {0} QC
QC lot
Instrument
Te s t
View cumulative QC statistics QC
QC lot
Instrument
Te s t
Test/instrument assignments: {0} Instrument
Retrieve samples Test
Table 1-1 Panels with multi-selection filtering.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234

Roche cobas IT middleware User manual

Category
Software
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI