Invacare HomeFill HF2PC9 User manual

Category
Oxygen Equipment
Type
User manual

This manual is also suitable for

Dealer: This manual MUST be given to the end
user.
User:
Before using this cylinder, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
Operator’s Manual
HomeFill
Oxygen Conserving
Cylinder
HF2PCL6
HF2PC9
HF2PCL4
WARNING
HomeFill™ 2 Part No 1112648
WARNING
DO NOT OPERATE THIS PRODUCT OR ANY
AVAILABLE OPTIONAL EQUIPMENT WITHOUT
FIRST COMPLETELY READING AND
UNDERSTANDING THESE INSTRUCTIONS AND
ANY ADDITIONAL INSTRUCTIONAL MATERIAL
SUCH AS OWNER’S MANUALS, SERVICE
MANUALS OR INSTRUCTION SHEETS SUPPLIED
WITH THIS PRODUCT OR OPTIONAL
EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS, OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNEL BEFORE ATTEMPTING TO USE
THIS EQUIPMENT - OTHERWISE INJURY OR
DAMAGE MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
NOTE:Updatedversionsofthismanualareavailableon
www.invacare.com.
TABLE OF CONTENTS
Part No 1112648 3 HomeFill™
TABLE OF CONTENTS
SPECIAL NOTES ............................................... 4
RECEIVING INSPECTION .................................. 5
INTENDED USE ................................................ 5
SECTION 1—GENERAL GUIDELINES .................. 6
Operating Information ....................................................................... 6
Maintenance.......................................................................................... 8
SECTION 2—FEATURES .................................... 9
SECTION 3—TYPICAL PRODUCT PARAMETERS ...
10
SECTION 4—INSPECTION AND COMPRESSOR
INTERFACE ................................................... 11
Cylinder Prefill Inspection ...............................................................11
Connecting/Disconnecting the Cylinder to/from the
Compressor........................................................................................14
SECTION 5—OPERATING INSTRUCTIONS ....... 19
SECTION 6—MAINTENANCE .......................... 21
SECTION 7—CYLINDER FILL TIMES ................. 22
Troubleshooting ................................................................................22
Replacement Parts - Accessories ..................................................22
LIMITED WARRANTY .................................... 23
SPECIAL NOTES
HomeFill™ 4 Part No 1112648
SPECIAL NOTES
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothefollowingtablefordefinitions
ofthesignalwords.
NOTICE
Theinformationcontainedinthisdocumentissubjectto
changewithoutnotice.
DANGER
DONOTSMOKEwhileusingthisdevice.Keepall
matches,litcigarettesorothersourcesofignitionoutofthe
roominwhichthisproductislocatedandawayfromwhere
oxygenisbeingdelivered.
NOSMOKINGsignsshouldbeprominentlydisplayed.
Textilesandothermaterialsthatnormallywouldnotburn
areeasilyignitedandburnwithgreatintensityinoxygen
enrichedair.Failuretoobservethiswarningcanresultin
severefire,propertydamageandcausephysicalinjuryor
death.
SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in
death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result
in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in
property damage or minor injury or both.
RECEIVING INSPECTION
Part No 1112648 5 HomeFill™
CAUTION
“Caution:Federallawrestrictsthisdevicetosalebyorrental
orontheorderofaphysician,oranyotherpractitioner
licensedbythelawofthestateinwhichhe/shepracticesto
useorordertheuseofthisdevice.”
RECEIVING INSPECTION
Removeproductfrompackageandinspectfordamage.If
thereisanydamage,DONOTuse.Contactyourdealer/
HomeCareequipmentprovider.
INTENDED USE
Tostore,regulate,anddeliversupplementaloxygento
patientswhomayhavedifficultyextractingoxygenfromthe
airtheybreathe.Itisforpatientswhowouldnormallyreceive
theoxygenviaanasalcannula.ThedevicedeliversUSP93%
oxygenatanumberofdifferentflow/pulsesettings.Itis
intendedtobeusedasanoxygensavingdevicethatmay
reducethedryingoutoftheairways.
SECTION 1—GENERAL GUIDELINES
HomeFill™ 6 Part No 1112648
SECTION 1—GENERAL GUIDELINES
Inordertoensurethesafeinstallation,assemblyandoperation
oftheHomeFillOxygenConservingCylinderthese
instructionsMUSTbe followed.
WARNING
Thissectioncontainsimportantinformationforthesafe
operationanduseofthisproduct.
Operating Information
Onlypersonnelinstructedandtrainedintheuseofthe
HomeFillO
2
ConservingCylindershouldoperatethisproduct.
Productcontainsmagnetic,ferrousmaterialthatmayaffectthe
resultsofanMRI.
Besureallconnectionsaretightandleakfree.
DONOTautoclav e.
DONOTgassterilizewithethyleneoxide.
DONOTcleanwitharomatichydrocarbons.
DONOTimme rse productinanykindofliquid.
Storeproductinacleanareawhennotinuse.
OnlyuseInvacareCorporationHF2PCL6carryingbag,
HF2PC9carryingbag,orHF2PCL4carryingbag.
NEVERsmokeinanareawhereoxygenisbeingadministered.
ALWAYSconfirmprescribed dosebeforeadministeringto
patientandmonitoronafrequentbasis.
DONOTuseifdirtorcontaminantsarepresentonoraround
cylinder,valve,productorconnectingdevices.
SECTION 1—GENERAL GUIDELINES
Part No 1112648 7 HomeFill™
NEVERusenearanytypeofflameorflammable/explosiv e
substances,vaporsoratmosphere.
ALWAYSfollowANSI andCGAstandardsforMedicalGas
ProductsandFlowmeters(E7)andOxygenHandling(G4).
NOOXYGENisdeliveredwhenthepointerisalignedwith
“Off”.
DONOTblocktheoutletfittingorkinkthecannulawhenthe
productisinuse,this maydamagetheunit.
Theuseofthisdeviceislimitedtotheoxygenpatient.
CylindersMUSTbeusedonlybytheoxygenpatientandare
nottobedistributedto anyotherindividual foranypurpose.
ThisproductisNOTtobeusedbypatientswhobreathe
throughtheirmouths.
DONOTuseoils,greases,organiclubricants orany
combustiblematerialsonornearthisproduct.Washanddry
handsproperlypriortousage.
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththecompressorcouplingorbottlefittings.
Thesesubstancesmaybecomeflammableinthepresenceof
oxygenandcauseinjury.Avoidtouchingthebottlefitting
whenever possible.Exampleofitemstoavoidarehandcreams
orlotions,cookingoils,suntanoilsorsunblocksandsimilar
oilyprodu ct s. Ifyouusethesetypesofproducts,washyour
handspriortousingtheHomeFillunit.Ifthebottlefitting
shouldcomeincontactwithsuchsubstancesitshouldbe
cleanedoffwithadampclothpriortoconnectingittothe
compressor.
NEVERoilor lubricatethecompressorcouplerorbottle
fittings.NEVERuseanypenetratingoilorlubricantsuchas
WD40
®
or3in1oil
®
.If connectiondifficultyisexperienced
verifythatthebottlefittingisbeinginsertedstraightdowninto
thecompressorcoupler.Ifcontinueddifficultyisexperienced
contactyoursupplierforassistance.
SECTION 1—GENERAL GUIDELINES
HomeFill™ 8 Part No 1112648
DONOTuseahumidifierwiththisproduc t.
DONOTallowcylinderstotiporfall.Securegascylindersso
theycannotfall.Ensurecylinderisalwaysinuprightposition.
DONOTstorecylindersnearsourcesofheatorflame.
Thisproductisequippedwithareliefvalve.Ifyouhearaloud
hissingorpoppingsoundcomingfromtheproduct,
discontinueuseofproduct,turntheflowselectortothe“Off
position,andcontact yourdealer/HomeCareequipment
provider.
DONOTusewhilesleepingwithoutconsultingyour
healthcareprovider.
Thisproductisdesignedtooperatewithastandardsingle
lumen,adultoxygennasalcannulawithamaximumlengthof
sev enfeet.
Thecannulais forsingle patientuseonly.
CannulaMUSTbeinstalledonoutletconne ctionpriortouse.
DONOTplacetheunitunderclothingwhileinuse.Whenthis
productisinuseasmallamountofoxygenisvented.Wearing
theunitunderclothingmaysaturatefabricswithoxygenand
causethemtoburnrapidlyifexposedtosparksorflame.Itmay
takesev era lhoursforoxygenlev elsin fabricstoreturnto
normal.
Maintenance
DONOTusealiquidleak detecto rtotestforleaks.
DONOTattempttorepairproduct.
SECTION 2—FEATURES
Part No 1112648 9 HomeFill™
SECTION 2—FEATURES
DANGER
Thisproductisnotintendedasalifesustainingor
lifesupportingdevice.
FIGURE 2.1 Product Diagram
Outlet
Connection
Cannula
Flow Dial
Cylinder
Gauge
Cylinder Fill
Port
SECTION 3—TYPICAL PRODUCT PARAMETERS
HomeFill™ 10 Part No 1112648
SECTION 3—TYPICAL PRODUCT
PARAMETERS
CONSERVER S/N RANGE 1-69,999 75,000 +
INLET PRESSURE RANGE: 300 - 2100 psig 300 - 2100 psig
INTERNAL REGULATED
PRESSURE: 50 psig 50 psig
DIMENSIONS:
HF2PCL6: Weight 4.3 lbs (1.95 kg) 4.1 lbs (1.86 kg)
HF2PCL6: Overall Length 11.1 in (28 cm) 10.9 in (27.7 cm)
HF2PCL6: Diameter 4.4 in (11.1 cm) 4.4 in (11.1 cm)
HF2PC9: Weight 5.3 lbs (2.4 kg) 5.1 lbs (2.31 kg)
HF2PC9: Overall Length 14.2 in (36 cm) 14 in (35.5 cm)
HF2PC9: Diameter 4.4 in (11.1 cm) 4.4 in (11.1 cm)
HF2PCL4: Weight N/A 3.6 lbs (1.63 kg)
HF2PCL4:Overall Length N/A 8.9 in (22.2 cm)
HF2PCL4:Diameter. N/A 4.4 in (11.1 cm)
PULSE SETTINGS: 1,2,3,4, & 5
FLOW SETTINGS: 2 LPM Continuous
SAVINGS RATIO: Up to 4.3:1 Up to 5.7:1
CANNULA
REQUIREMENT:
4 ft recommended (MS0204) (7 ft long max)
std. adult single lumen oxygen nasal cannula.
OPERATING CONDI-
TIONS TEMPERATURE:
32°F (0°C) to 105°F
(40.6°C)
35°F (1.7°C) to 105°F
(40.6°C)
ALTITUDE: 0 to 5280 ft (0
to1600 m)
0 to 10,000 ft (0 to
3071 m)
STORAGE CONDITIONS
TEMPERATURE:
-40°F (-40°C) to
140°F (60°C)
-40°F (-40°C) to 140°F
(60°C)
MAXIMUM HUMIDITY: 95%
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 11 HomeFill™
SECTION 4—INSPECTION AND
COMPRESSOR INTERFACE
NOTE:RefertotheHomeFillCompressorOwner’sManualfor
compressoroperatinginstructions.
Cylinder Prefill Inspection
WARNING
AllcylindersMUSTbeinspectedbeforeattemptingtofill;
otherwise,injuryordamagemayoccur.
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE4.1.
The Hydrostatic Testing Date
AluminumcylindersMUSTundergotestingeveryfiveyears.
WARNING
DONOTfillcylindersthathavenotbeentestedinthepast
fiveyears.Otherwise,seriouspersonalinjurymayresult.
Contactyourdealerforreplacement.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill™ 12 Part No 1112648
External Examination
1. Examinetheoutsideofthecylinderforthefollowing
conditions,andreplacethecylinderiftheyexist:
•Dentsordings
•ArcBurns
•OilorGrease
•Anyothersignsofdamagethatmightcauseacylinder
tobeunacceptableorunsafeforuse.
2. Examinethecylinderforevidenceoffireorthermal
damage.Evidenceincludescharringorblisteringofthe
paint,orotherprotectivecoatingorheatsensitive
indicator.Iffireorthermaldamageisfound,replacethe
cylinder.
3. Inspectthecylinderfillportforthefollowing,andreplace
iffound:
•Debris,oilorgrease
•Noticeablesignsofdamage
•Signsofcorrosioninsidethevalve
•Signsofexcessiveheatorfiredamage
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 13 HomeFill™
FIGURE 4.1 External Examination
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill™ 14 Part No 1112648
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,injuryor
damagemayoccur.
DONOTusealiquidleakdetectortotestforleaks.
NOTE:Forthisprocedure,refertoFIGURE4.2andFIGURE4.3.
1. Examinethecylindergauge.Ifthecylinderpressureisless
than1500psig.,proceedtoSTEP2(FIGURE4.3).
NOTE:Ifthecylinderpressureisgreaterthan1500psig.,DO
NOTattempttotopoffthiscylinder.Itmaynotfill.
2. SettheregulatorflowdialonthecylindertoOFF(FIGURE
4.3).
3. Removethecylinderandconnectorfillportcovers(if
present).
CAUTION
DONOTconnecttheoutlet(cannula)connectiontothe
HomeFillunit,otherwisetheconservingcylinderwillnot
workproperly.
4. MomentarilypushDOWNontheouterring(sleeve)ofthe
connectorfillporttoresettheconnectorasshownin
FIGURE4.2.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 15 HomeFill™
NOTE:Iftheouterring(sleeve)isintheUPposition
(GREENdotsnotvisible),theconnectorfillportwillnot
beabletoacceptthecylinderfillport.PushingDOWN
momentarilywillresettheconnectorfillport(GREEN
dotsvisible)toacceptthecylinderfillport.
FIGURE 4.2 Resetting Connector Fillport
5. Graspthecylinder/regulatorassemblyintheareabehind
thecylindergauge(FIGURE4.3).
6. Positionthecylinderinthecompressorcradle(FIGURE
4.3).
7. Alignthecylinderfillportwiththeconnectorfillport
(FIGURE4.3).
8. PullUPontheouterring(sleeve)oftheconnectorfillport
whilepushingDOWNonthecylinder/regulator assembly
tocouplethecylinderfillportintotheconnectorfillport
(FIGURE4.3).
NOTE:Thecylinderisproperlyconnectedwhenanaudible“click”
isheard.
BEFORE
COUPLING
CYLINDER,
PUSH DOWN
ON SLEEVE
UNTIL
GREEN DOTS
ARE VISIBLE.
Connector Fillport
Outer Ring (Sleeve)
PUSH
DOWN
GREEN Dots
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill™ 16 Part No 1112648
FIGURE 4.3 Connecting the Cylinder to the Compressor
Cylinder/Regulator
Assembly
Compressor
Cylinder
Gauge
Cylinder
Fillport
Connector
Fillport
PULL UP
Outer Ring
(Sleeve)
Cylinder
Compressor
Cradle
Cylinder
Gauge
Regulator
Flow Dial
Outer Ring
(Sleeve)
Connector
Fillport
Outlet (Cannula)
Connection
Power
Switch
Regulator
Cylinder
Fillport
NOTE:Fillportcoversnotshownforclarity.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 17 HomeFill™
Disconnecting the Cylinder from the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
NOTE:Forthisprocedure,refertoFIGURE4.4.
1. PressthecompressorpowerswitchtotheOff(O)position.
2. Graspthecylinder/regulatorassemblyintheareabehind
thecylindergauge.
3. Withtheotherhand,grasptheouterring(sleeve)ofthe
connectorfillportandpushDOWN.
4. Liftuponthecylinder/regulatorassemblytoremovefrom
theconnectorfillport.
FIGURE 4.4 Disconnecting the Cylinder from the Compressor
Cylinder/Regulator
Assembly
Compressor
Cylinder
Gauge
Cylinder
Fillport
Connector
Fillport
Outer Ring
(Sleeve)
PUSH DOWN
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill™ 18 Part No 1112648
5. Whenthecylinderfillportisdisconnectedfromthe
connectorfillport,releasetheouterring(sleeve)ofthe
connectorfillportandusetwohandstoremovethe
cylinderfromthecompressorcradle.
6. Placethefillportcoversontotheconnectorandcylinder
fillports.
NOTE:Thefillportcoversshouldbeusedwhenevercylindersare
notbeingfilled.
SECTION 5—OPERATING INSTRUCTIONS
Part No 1112648 19 HomeFill™
SECTION 5—OPERATING
INSTRUCTIONS
WARNING
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE5.1.
1. Priortoeachuse,inspectproductforvisibledamage.DO
NOTuseifanydamageisfound.
NOTE:Ifforanyreason,anylabelbecomesillegibleorlost,contact
yourdealer/HomeCareequipmentprovider.
2. Besureregulatorisinthe“Offposition.Align“Offwith
theindicatingpointer.
3. Positionthecylindersotheoxygencylindervalveoutletis
pointingawayfromtheuserandanyotherperson(s).
WARNING
DONOTdirectflowofoxygenatanyperson,orflammable
materialwhenadjustingtheflowdial.
4. Attachastandardadultsinglelumenoxygennasal
cannula,nolongerthansevenfeettotheproduct’soutlet
accordingtothecannulamanufacturersinstructions.DO
NOTusepediatric,lowflownasalcannulasoroxygen
maskswiththisproduct.
NOTE:Aunitwithoutcannula attachedmaypulseifsetatposition
15.
SECTION 5—OPERATING INSTRUCTIONS
HomeFill™ 20 Part No 1112648
5. Placethecannulaover
yourearsandposition
theprongsinyournose
asinstructedbyyour
HealthCareprovideror
cannulamanufacturer.
RefertoFIGURE5.1.
6. Turndialuntilindicating
pointerisalignedwith
theprescribedsetting.
FIGURE 5.1 Cannula
Placement
WARNING
Whenpointeroncontentsgaugeentersredzone,itis
recommendedtorefillthecylinderontheInvacareHomeFill
Compressor.
NOOXYGENisdeliveredinbetweensettings.Toobtain
desiredflow,theindicatingpointermustalignwithaspecific
numberonthedial.
Toavoidinjurytopatient,ALWAYSconfirmprescribed
settingbeforeadministeringtopatientandmonitorflowona
frequentbasis.
OnlyuseInvacareCorporationHF2PCL6carryingbag,the
HF2PC9carryingbag,ortheHF2PCL4carryingbag.
7. Breathethroughthenoseandfeelapulseofoxygenasyou
inhale.
NOTE:Whenonapulsesetting,thereisflowora pulseonlyatthe
beginningofeachbreath.Ifyoudonotfeelthepulseatthebeginning
ofeachbreath,checkthesetting.
NOTE:Becauseeachpatient’sbreathingpatternisdifferentandthe
environmentvaries,itmaybedifficulttofeelsomelowsetting
pulses.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24

Invacare HomeFill HF2PC9 User manual

Category
Oxygen Equipment
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI