Invacare HomeFill HF2ECE6 User manual

Type
User manual

This manual is also suitable for

Invacar1.0L/1.7LCylinderwith
DualLumenConserver
ForHomeFill®OxygenSystem
enHF2ECE6,HF2ECE9
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
2014/68/EU
93/42/EEC
©2017InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.Trademarksareidentiedbyand®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A.
WD-40and3–IN-ONEareregisteredtrademarksoftheWD-40Company.
Contents
1General.........................................4
1.1Symbols.....................................4
1.2IntendedUse.................................5
1.3UserProle...................................5
1.4ServiceLife...................................5
1.5Warranty.....................................5
1.6ComponentIdentication........................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................8
3Setup...........................................14
3.1DealerSetup..................................14
3.2UserSetup...................................14
3.3ReceivingInspection............................14
3.4CylinderPrellInspection........................14
3.4.1TheHydrostaticTestingDate...................15
3.4.2ExternalExamination.........................16
4Usage..........................................17
4.1BeforeUse...................................17
4.2Connecting/DisconnectingtheCylinderto/fromthe
Compressor.................................17
4.3CylinderAssemblyOperation......................17
5Maintenance.....................................21
5.1Maintenance..................................21
5.2EmptyingtheCylinderAssembly...................22
5.3Cleaning.....................................23
5.4Disinfecting...................................23
5.5Reuse.......................................23
5.6Storage......................................24
5.7OptionalAccessories............................24
6AfterUse........................................25
6.1Disposal.....................................25
6.2WearandTear................................25
7Troubleshooting...................................26
7.1Troubleshooting...............................26
8Technicaldata....................................27
8.1TechnicalDescription............................27
8.2Specications.................................28
Invacar1.0L/1.7LCylinderwithDualLumenConserver
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazards
orunsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Seetheinformationbelowfordenitions
ofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin
deathorseriousinjury.
WARNING!
Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin
deathorseriousinjury.
CAUTION!
Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin
propertydamageorminorinjuryorboth.
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformation
forefcient,trouble-freeuse.
Readthemanual
Noopename
DONOTSmoke
Oxidizer
Recycle
DONOTdisposeofinhouseholdwaste
Manufacturer
EURepresentative
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
Thelaunchdateofthisproductisstatedinthe
CEdeclarationofconformity.
4
1176148-D
General
1.2IntendedUse
Thiscylinderisintendedtostore,regulate,anddeliver
supplementaloxygentopatientswhomayhavedifculty
extractingoxygenfromtheairtheybreathe.Itisforpatients
whowouldnormallyreceivetheoxygenviaanasalcannula.
ThedevicedeliversUSP93±3%oxygenatanumberof
differentowsettingsandmayincreasethedryingoutof
airways.Thisproductistobeusedasanoxygensupplement
andisnotintendedtobelifesupportingorlifesustaining.
DANGER!
RiskofInjuryorDeath
Thisproductistobeusedasanoxygen
supplementandisnotintendedtobe
life-supportingorlife-sustaining.ONLYusethis
productifthepatientiscapableofspontaneous
breath,abletoinhaleandexhalewithouttheuse
ofamachine.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse
andspecicationshasnotbeentestedandmay
leadtoproductdamage,lossofproductfunction,
orpersonalinjury.
DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended
usesectionsofthismanual.
1.3UserProle
Nospecicproductknowledgeortrainingisrequired,other
thanwhatiscontainedinthismanual,tooperatethe
product.Itisnotintendedforsupplyingoxygentherapy
duringrelling.Thisproductisintendedtobeusedbyor
underadultsupervisiononlyafterreadingandunderstanding
theinstructionsandwarningsofthisusermanual.
Onlypersonnelinstructedandtrainedintheuseofthe
HomeFill®ConservingCylindershouldoperatethisproduct.
Theuseofthisproductislimitedtotheoxygenpatient.
CylindersMUSTbeusedonlybytheoxygenpatientandare
nottobedistributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathe
throughtheirmouths.
1.4ServiceLife
Theexpectedservicelifeoftheconserverisveyearswhen
usedinaccordancewiththesafetyinstructions,maintenance
intervalsandcorrectusestatedinthismanual.Refertothe
Maintenancesection.
Theexpectedservicelifeofthecylinderisbasedontesting
thecylindereverytenyearsandusingitinaccordancewith
thesafetyinstructions,maintenanceintervalsandcorrect
usestatedinthismanual.Refertothefollowingprocedures:
TheHydrostaticTestingDate,ExternalExamination,
Maintenance
1.5Warranty
Termsandconditionsofthewarrantyarepartofthegeneral
termsandconditionsparticulartotheindividualcountriesin
whichthisproductissold.
1176148-D5
Invacar1.0L/1.7LCylinderwithDualLumenConserver
1.6ComponentIdentication
Item
Description
A
CylinderAssembly
B
Cylinder
C
ContinuousFlowConserver
D
FlowDial
E
IndicatorWindow
F
Switch
G
OutletConnection
H
CylinderFillport
I
ShutoffValve
J
CylinderGauge
K
CylinderGaugeNeedle
61176148-D
Safety
2Safety
2.1LabelLocations
Serialnumberlocatedaboveproductlabelshown
here.
1176148-D
7
Invacar1.0L/1.7LCylinderwithDualLumenConserver
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand
operationoftheproducttheseinstructionsMUSTbe
followed.
DANGER!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycause
death,injuryordamage.Thissectioncontains
importantinformationforthesafeoperation
anduseofthisproduct.
DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnel
beforeattemptingtousethisequipment.
CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskofDeath,InjuryorDamage
Usethiscylinderinaccordancewiththe
HomeFillcompressorusermanual.
DONOThandlethecylinderassemblyor
usethecontentsuntilyouareprofessionally
trained,includingemergencyprocedures.
Trainingrequiredtooperatetheproductis
containedintheusermanual.
Thisproductisnotintendedforsupplying
oxygentherapyduringrelling.
Thisproductisintendedtobeusedbyorunder
adultsupervision.
OxygenCylinderslledbytheInvacareHomeFill
oxygensystemshallbeusedforpersonaluse
only.“Nottobelledforresaleoruseby
professionalusers.
ThiscylinderistobeusedwiththeInvacare
HomeFilloxygensystem.Thiscylinder
contains93±3%oxygenproducedbyan
oxygenconcentratorusingthepressureswing
adsorptionprocess.
ALWAYSopenthevalveslowlyandclosethe
valveaftereachuseandwhenempty.
81176148-D
Safety
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
DONOTSMOKEwhileusingthisproduct.
DONOTusenearOPENFLAMEorIGNITION
SOURCES.
DONOTuseanylubricantsonthecylinder
unlessrecommendedbyInvacare.
NOSMOKINGsignsshouldbeprominently
displayed.
Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutof
theroominwhichthiscylinderislocatedand
awayfromwhereoxygenisbeingdelivered.
Keeptheoxygentubing,cord,andcylinder
outfromundersuchitemsasblankets,
bedcoverings,chaircushions,clothing,and
awayfromheatedorhotsurfacesincluding
spaceheaters,stoves,andsimilarelectrical
appliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesale
toorontheorderofamedicalpractitioner
licensedbyagovernmentalagencywherehe/she
practices.
ONLYalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
1176148-D9
Invacar1.0L/1.7LCylinderwithDualLumenConserver
OperatingInformation
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausere,injuryordeath.
DONOTuseifdirtorcontaminantsarepresent
onoraroundcylinderorconserver .
DONOTuseorllcylinderassemblyifthe
cylinderllportisdirtyorcontaminated.
Contactyourproviderforserviceorrepair.
DONOTallowoilfromyourhandsorother
sourcestocomeintocontactwiththe
compressorcouplingorcylinderassembly
ttings.Thesesubstancesmaybecome
ammableinthepresenceofoxygenandcause
injury.Examplesofitemstoavoidarehand
creamorlotion,cookingoil,suntanoilor
sunblockandsimilaroilyproducts.Ifyouuse
thesetypesofproductsInvacaresuggestsyou
washyourhandspriortousingtheHomeFill
system.Ifthecylinderllportshouldcome
incontactwithsuchsubstances,contact
yourproviderforservice.Avoidtouchingthe
cylinderllport.
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausere,injuryordeath.
DONOTcleanwitharomatichydrocarbons,
including,butnotlimitedtobenzeneand
toluene.
NEVERoilorlubricatethecompressorcoupler
orcylinderllport.NEVERuseanypenetrating
oilorlubricantsuchasWD-40®or3-IN-ONE®.
Ifconnectiondifcultyisexperienced,verify
thatthecylinderllportisbeinginserted
straightdownintothecompressorcoupler .
Ifcontinueddifcultyisexperienced,contact
yourproviderforassistance.
101176148-D
Safety
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausere,injuryordeath.
DONOTusecleaning/disinfectingsolutionson
thellporttting.
DONOTimmerseproductinanykindofliquid.
DONOTallowanykindofliquidtogetinside
openingsoftheconserver .
DONOTspraycleaning/disinfectingsolutions
onproduct.
DONOTautoclave.
DONOTgassterilizewithethyleneoxide.
DONOTattempttorepaircylinderassembly.
AllrepairsMUSTbeperformedbyInvacare
CorporationorinanInvacareapprovedrepair
center .
Storecylinderassemblyinaclean,dryandcool
areawhennotinuse.
DANGER!
RiskInjuryorDeath
Toavoidchokingoringestionofchemicalsfrom
airwaycontamination:
DONOTusetheproductinthepresence
ofpollutants,smoke,fumes,ammable
anesthetics,cleaningagents,orchemical
vapors.
WARNING!
RiskofInjuryorDeath
Contaminantsintheoxygenmaycauseinjuryor
death.
DONOTuseiftheoxygenhasanunusualodor.
DONOTuseifparticulatesareinhaled.
ContactInvacareoradealerforservice.
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromproductmisuse:
Theowdialshouldbeturnedwithnormal
pressurefromone’shand.Iftheowdialis
stifftoturn,donotforce,otherwisedamage
orinjurycouldoccur .
DONOTuseaowoutletfordrivingany
medicalequipment.
1176148-D
11
Invacar1.0L/1.7LCylinderwithDualLumenConserver
CAUTION!
RiskofInjuryorDamage
Toreducetheriskofinjuryordamagefrom
droppedorimpactedcylinderassemblies:
Usetwohandswhenhandling/transporting
oxygencylinders.
OnlyuseInvacareCorporationcarryingbag.
DONOTallowcylinderassembliestotiporfall.
DONOTdropcylinderassemblies.
Securegascylinderssotheycannotfall.
Ensurecylinderisalwaysinuprightposition
whennotconnectedtotheHomeFill
compressor .
DONOTallowcylinderstoimpactinanimate
objects.
CAUTION!
RiskofDamage
Toavoidproductdamage:
DONOTblocktheoutletttingorkinkthe
cannulawhentheproductisinuse,asthis
maydamagetheunit.
DONOTllthecylinderassemblyoutdoorsor
anywherebelow13°C(55°F).
DANGER!
RiskofInjuryorDeath
Thecannulacancausetripping,falling,orother
injuryifimproperlypositionedandsecured.To
avoidinjuryordeath:
ThecannulaMUSTberoutedandsecured
properly.
DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhile
wearingthecannula.
Avoidpositioningcannulasacrossareasofhigh
foottrafc(i.e.aisles,doorways,hallways,etc.)
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
Keepchildrenandpetsawayfromnasal
cannulaandtubing.
Closesupervisionisnecessarywhenthenasal
cannulaisusedbyornearchildrenand/or
disabledpersons.
12
1176148-D
Safety
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
Beforetherstpatientuse,theproviderMUST
disinfecttheproduct.Refertothedisinfecting
section.
Replacethenasalcannulaonaregularbasis.
Checkwithyourequipmentprovideror
physiciantodeterminehowoftenthecannula
shouldbereplaced.
DONOTsharecannulasbetweenpatients.
Thecylinderposesnobiocompatibilityhazardsand
islatexandphthalatefree.Thebiocompatibility
ofcylinderaccessoriesmayvarybymanufacturer .
Invacareaccessoriessuchasthecylinderbagand
Invacareoxygentubingarelatexandphthalatefree.
Refertothepackagingonanycylinderaccessoriesto
determinethebiocompatibilityhazard.
ThecylinderassemblyMUSTbeemptiedofgaspriorto
shippingbycourierorpostalservice.RefertoEmptyingthe
CylinderAssembly.
Becausethecylindercontainsmagnetic,ferrousmaterial,it
isnotsuitabletobeinanMRIroom.
Besureallconnectionsaretightandleakfree.DONOTuse
aliquidleakdetectortotestforleaks.Leaksareindicated
bylossofpressurewhenthecylinderassemblyisnotinuse
andbyhissingsounds.Thisproductisequippedwitharelief
valve.Ifyouhearaloudhissingorpoppingsoundcoming
fromtheproduct,discontinueuseofproduct,turntheow
dialtothe“0”setting,andcontactyourdealer/HomeCare
equipmentprovider .
ALWAYSconrmprescribeddosebeforeadministeringto
patientandmonitoronafrequentbasis.
NOOXYGENisdeliveredwhentheconserverowdial
settingis“0”.
DONOTuseahumidierwiththisproduct.
Keepcylinderinawellventilatedplace.
Thisproductisdesignedtooperatewithastandardor
highowduallumen,adultoxygennasalcannulawitha
maximumlengthof2.1m(7ft).
Thecannulaisforsinglepatientuseonly.
CannulaMUSTbeinstalledonoutletconnectionpriortouse.
RefertotheHomeFillUserManualforadditionalwarnings,
cautionsandsafetyinstructions.
1176148-D13
Invacar1.0L/1.7LCylinderwithDualLumenConserver
3Setup
3.1DealerSetup
WARNING!
Beforetherstpatientuse,theproviderMUST
disinfecttheproduct.RefertotheDisinfecting
section.
Beforepatientuse,theproviderMUSTcheck
thehydrostatictestingdate.Refertothe
HydrostaticTestingDatesection.
WARNING!
RiskofInjury
Formedicalapplications,itisveryimportant
toselecttheprescribedlevelofoxygenow.
Uninterrupteduseofhighconcentrationsof
oxygenoveralongduration,withoutmonitoring
itseffectonoxygencontentofarterialblood,
maybeharmful.
UseONLYasdirectedbyalicensedpractitioner.
DONOTincreaseordecreasetheowunlessa
changehasbeenprescribedbyyourphysician
ortherapist.
3.2UserSetup
WARNING!
Whenreceivingthecylinderfromtheprovider,
theuser/patientMUSTperformareceiving
inspection.RefertotheReceivingInspection
section.
3.3ReceivingInspection
Removeproductfrompackageandinspectfordamage.
Ifthereisanydamage,DONOTuse.Contactyour
dealer/HomeCareequipmentprovider .
3.4CylinderPrellInspection
WARNING!
RiskofInjuryorDamage
Oxygencylindersshallbeusedforpersonal
useonly.Nottobelledforresaleoruseby
professionalusers.
FillingofcylindersisperformedbyInvacare
HomeFilloxygensystemonly.
ThiscylinderistobeusedwiththeHomeFill
system.Thiscylindercontains93±3%oxygen
producedbyanoxygenconcentratorusingthe
pressureswingadsorptionprocess.
DONOThandlethecylinderorusecontents
untilyouareprofessionallytrained,including
emergencyprocedures.
Usethiscylinderinaccordancewiththe
HomeFillusermanual.
Openandclosetheshutoffvalveslowly.Close
theshutoffvalveaftereachuseandwhen
empty.
Securethecylinderduringstorageanduse.DO
NOTdrop.
Nosmokingincylinderarea.Keepawayfrom
heat,ameorspark.
Keepoutofthereachofchildren.
14
1176148-D
Setup
DANGER!
RiskofInjuryorDeath
Improperllingorabuseofthiscylinder,orfailure
tofollowthesesafetyinstructionsmaycause
seriousbodilyinjuryordeath.Highpressure
oxidizinggasvigorouslyacceleratescombustion.
DONOTalterormodifycylinderorrelated
components.
DONOTuseoilorgrease.Keepcylinder,
shutoffvalve,allequipmentandconnections
freeofoilandgreasetoavoidviolentignition
resultinginseriousbodilyinjuryordeath.
3.4.1TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeverytenyears.
DANGER!
RiskofDeath,InjuryorDamagefromFire
Toavoidre,death,injuryordamage:
Cylindersthathavebeenrenishedand/or
exposedtoelevatedtemperaturesMUSTbe
hydrostaticallytestedbeforelling.Cylinder
temperatureshouldnotexceed52°C(125°F).
Aluminumcylinderssubjecttotheactionof
reorheatedtemperaturesinexcessof177°C
(350°F)MUSTbewithdrawnfromserviceand
condemnedbytrainedpersonnel.DONOT
usecausticpaintstrippers.Valveandsafety
reliefdevicesMUSTberemovedandreplaced
bytrainedandauthorizedpersonnel.DONOT
alterorchangevalveassembly.
AllcylindersMUSTbeinspectedbefore
attemptingtoll;otherwise,injuryordamage
mayoccur.
WARNING!
RiskofInjury
Toreducetheriskofinjury:
DONOTllcylindersthathavenotbeentested
inthepasttenyears.Otherwise,serious
personalinjurymayresult.Contactyour
dealer/HomeCareequipmentproviderfor
replacement.
1176148-D15
Invacar1.0L/1.7LCylinderwithDualLumenConserver
3.4.2ExternalExamination
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamage:
DONOTusethecylinderifanyofthefollowing
itemsarefound.Contactyourdealer/Home
Careequipmentproviderforreplacement.
1.Examinetheoutsideofthecylinderforthefollowing
conditions.Iftheyexist,contactyourproviderfor
service.
Dentsordings,arcburns,oilorgrease,anyother
signsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse
2.Examinethecylinderforevidenceofreorthermal
damage.Evidenceincludescharringorblisteringof
thepaint,orotherprotectivecoatingorheatsensitive
indicator .Ifreorthermaldamageisfound,contact
yourdealer/HomeCareequipmentproviderfora
replacement.
3.Inspectthecylinderllportforthefollowing.Ifthey
exist,contactyourproviderforareplacement.
Debris,oilorgrease,noticeablesignsofdamage,
signsofcorrosioninsidethevalve,signsofexcessive
heatorredamage,signsofpressureleaks
161176148-D
Usage
4Usage
4.1BeforeUse
WARNING!
RiskofInjuryorDamage
Defectiveorinoperablecylindersmaycause
injuryordamage.
AllcylindersMUSTbeinspectedbefore
attemptingtoll.
Ifthecylinderisfoundtobedefectiveor
doesnotworkproperly,contactanInvacare
dealerorqualiedtechnicianforserviceor
replacement.
Beforellingacylinder,readandfollowtheinformationin
thefollowingsections:
qCylinderPrellInspection
qExternalExamination
4.2Connecting/DisconnectingtheCylinder
to/fromtheCompressor
RefertotheHomeFillCompressorusermanualfor
operatinginstructionstoconnectanddisconnectthe
cylindertotheHomeFillCompressor.
4.3CylinderAssemblyOperation
WARNING!
RiskofFire
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofameor
ammable/explosivesubstances,vaporsor
atmosphere.
1176148-D
17
Invacar1.0L/1.7LCylinderwithDualLumenConserver
1.Priortoeachuse,inspectproductforvisibledamage.
RefertotheExternalExaminationsection.DONOTuse
ifanydamageisfound.
Ifforanyreason,anylabelbecomesillegibleor
lost,contactyourdealer/HomeCareequipment
provider .
2.InspectthecylindergaugeJtoensurethepressure
isatleast1500psig(10342kPa).Ifbelow1500
psig(10342kPa),thenrellthecylinderBusingthe
HomeFillsystem.
WARNING!
Whencylindergaugeneedleoncylinder
gaugeentersredzone,itisrecommended
torellthecylinderontheInvacare
HomeFillCompressor .
3.Positionthecylindersotheoxygencylindervalve
outletGispointingawayfromtheuserandanyother
person(s).
WARNING!
RiskofInjury
Toreducetheriskofinjury:
DONOTdirectowofoxygenatany
person,orammablematerialwhen
adjustingtheowdial.
4.TurntheowdialDtothe“0”setting.
5.Attachaduallumenadultoxygennasalcannulawith
amaximumlengthof2.1m(7feet)totheproducts
twooutletconnectionsaccordingtothecannula
manufacturersinstructions.
Itdoesnotmatterwhichtubeisinstalledon
whichoutletconnection.
DONOTusepediatric,lowownasalcannulas
oroxygenmaskswiththisproduct.
DANGER!
RiskofDeath
Thecannulacancausetripping,falling,or
otherinjuryifimproperlypositionedand
secured.Toavoidinjuryordeath:
ThecannulaMUSTberoutedandsecured
properly.
DONOTpositionthecannulaaroundthe
neck.Ensurethepatientcanmovefreely
whilewearingthecannula.
Avoidpositioningcannulasacrossareasof
hightrafc(i.e.aisles,doorways,hallways,
etc.)
181176148-D
Usage
6.Placethecannulaoveryourearsandpositionthe
prongsinyournoseasinstructedbyyourHealthCare
providerorcannulamanufacturer.
7.Ensuretherearenokinksortwistsinthecannulatubing.
8.OpentheShutoffvalveIknobslowlytowardstheON
direction(counterclockwise)
Thiswillallowcylinderpressuretotravelthrough
theconserver.
9.UsetheswitchFtoselectoneofthefollowingmodes:
ContinuousFlowL
ConservingPulseM
Theconserverislabeled“CONTINUOUS”and
“PULSE”.
Therstthreeows(0.25,0.50and0.75L/min)
areincontinuousmodeatalltimes,regardless
oftheswitchposition.
10.TurnowdialuntiltheindicatorwindowEisaligned
withtheprescribedsetting.
Whentherearetwonumbersdisplayedinthe
indicatorwindow,thelargernumberisthe
PULSEowsettingandthesmallernumberis
theCONTINUOUSowsetting.
WARNING!
Toavoidinjurytopatient,ALWAYSconrm
prescribedsettingbeforeadministeringto
patientandmonitorowonafrequent
basis.
OnlyusetheInvacareCorporationcarrying
bag.
1176148-D19
Invacar1.0L/1.7LCylinderwithDualLumenConserver
11.Breathethroughthenoseandfeelapulseofoxygen
asyouinhale.
Whenonapulsesetting,thereisow,ora
pulse,onlyatthebeginningofeachbreath.If
youdonotfeelthepulseatthebeginningof
eachbreath,checkthesetting.
Becauseeachpatient’sbreathingpatternis
differentandtheenvironmentvaries,itmaybe
difculttofeelsomelowsettingpulses.
Ifoperatingthecylinderbelow1.7ºC(35ºF)
usethe2L/Mincontinuoussettingonly.
Whentheunitisrstswitchedfrompulseto
continuousow,asingleinhalationisrequired
toinitiatethecontinuousow.
12.Whennishedusingthecylinder,performthefollowing
steps:
a.Turntheowdialtothe“0”setting.
b.Slowlyclosetheshutoffvalve(clockwise)fully.
201176148-D
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Invacare HomeFill HF2ECE6 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI