  • Hello! I am an AI chatbot trained to assist you with the Midmark Elevance® Dental Chair Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
003-10095-00 Revision AA2 04/23/2019
Installation Instructions for
153941 (002-2018-XX) Power Supply Fuse Field Installation Kit
P/Ns 002-10131-00 (115V) and 002-10132-00 (230V)
WARNING: Always disconnect the power supply from the power
source before removing any covers. Failure to do so may result in
personal injury.
Step 1: Disconnect the power supply from the power source by disconnecting the
power cord.
Step 2: Remove the single Phillips head screw and lock washer that are retaining
the fuse access cover. Retain the screw and lock washer for reuse.
Step 3: Remove the cover and discard.
Step 4: Locate fuse labelled F8 on the printed circuit board.
Step 5: Remove fuse F8 taking care not to bend the fuse holder clips. Discard
the fuse.
Step 6: Install the fuse, contained in the kit, in the location of the removed fuse.
Step 7: The power supply cover has an updated fuse label. Take the cover
provided in the kit and install on the power supply.
Step 8: Install the screw and lock washer and tighten to secure the cover.
Step 9: Connect the power supply to the power source by reconnecting the
power cord.
/