A.MENARINI GlucoMen Areo is a portable wireless device that provides blood glucose monitoring with wireless transmission via Near Field Communication (NFC).

It has a large display that shows the blood glucose level, time, and date. The device also stores up to 730 results with date, time, and marks, and allows users to view average results for the past 1, 7, 14, 30, 60, or 90 days.

The GlucoMen Areo comes with a sampler that uses a new sterile lancet for each sampling and a sliding sleeve to lock the sampler. The puncture depth can be adjusted for different skin types, and the device provides error codes and messages to help users identify and resolve any issues.

A.MENARINI GlucoMen Areo is a portable wireless device that provides blood glucose monitoring with wireless transmission via Near Field Communication (NFC).

It has a large display that shows the blood glucose level, time, and date. The device also stores up to 730 results with date, time, and marks, and allows users to view average results for the past 1, 7, 14, 30, 60, or 90 days.

The GlucoMen Areo comes with a sampler that uses a new sterile lancet for each sampling and a sliding sleeve to lock the sampler. The puncture depth can be adjusted for different skin types, and the device provides error codes and messages to help users identify and resolve any issues.

How to use the sampler
GlucoJect Dual PLUS
®
1.
Wash hands or
the alternative site
(AST). Unscrew the
sampler’s tip by tur-
ning it counterclock-
wise.
2.
Insert a new
sterile lancet into
the lancet holder and
push it fully.
Sampling AST
Place the sampler on the
skin at selected plug area.
Press and hold for a few
seconds, then press the
shutter release button
to take a blood test.
CAUTIO N!
NEVER use sampler on more than one per-
son. Take a new lancet for each sampling and
do not leave a used lancet in the sampler.
5.
Press the lancet ejec-
tor with your thumb
while pulling out the
sliding sleeve for dis-
posing of the used
lancet.
4.
Pull out the sliding
sleeve to lock the sam-
pler. Press and hold the
sampler against your
nger and press the
trigger.
3.
Adjust the puncture
depth by rotating the
tip. 1-2 for soft or thin
skin, 3 for normal skin,
4-5 for thick or harde-
ned skin.
NOTE! This is a concise quick reference guide. Detailed information can be found in the manual supplied with the meter.
Medeon Science Park Per Albin Hanssons väg 41, hus D SE-205 12 Malmö
Tel 040
32 12 70 Fax 040
32 12 71 www.menarinidiagnostics.se
Blood glucose measurement with wireless transmission
46302 - 10/15 GlucoMen areo_Quick guide - EN
46302 - 10/15 GlucoMen areo_Quick guide - EN
Quick Guide
Error codes and messages
ER 1
ER 2
Er 3
Er 4
Er 5
HI
LO
System error hardware. Remove and insert the batteries again and repeat the
test with a new test strip. If error code persists, please contact customer service.
Test strip is damaged or used, or have the blood sample applied before
”blood drop symbol” began to fl ash on the screen. Repeat the test
with a new strip.
Test strip was given the small amount of blood. Stick again to repeat
the test with a new test strip and the right amount of blood.
Test strip was removed or dislodged during the countdown. Repeat
the test with a new test strip.
NFC was disconnected during transmission. Repeat the NFC-transfer.
If the problem persists, contact customer service.
Your test result is higher than 33,3 mmol/L!
Your test result is lower than 1,1 mmol/L!
CAUTION! If HI or LO appears, contact your health care provider before
changing your treatment. Shown HI or LO repeatedly even though your
blood sugar levels are high or low, contact customer service.
Customer Service: 040-32 12 70 [email protected]
English
How to set the time and date.
1. Check that the meter is off. (If not, press for 3 seconds)
2. Press 4 seconds to enter the setting menu.
3. Press or to scroll the setup menus and press to access the
various menus.
4. Use or to select the value. Press to confirm and go to the
next step: year, month, day, time format, hour, minute.
5. After confirming the minutes (by pressing ) will return to
the Setup menu.
How you measure your blood sugar.
Time and date are preset, please check that this is correct the first time you use
the meter.
1. Wash hands with soap and water.
2. Take a new test strip from the bottle and replace the cap immediately.
3. Place the sole of the test strip that has a metallic contact surface, into
the meter opening.
4. The meter shows a blinking blood drop and is now ready for measurement.
5. Stick with the sampler.
6. Put the blood drop against the short side of the stick that absorbs the right
amount of blood. A beep sounds and countdown starts on the display.
7. Read the test results. A beep sounds (if enabled) when the result is
displayed on the screen.
8. Press the eject button to eject the used test strip. The meter turns off
automatically.
How to select a measured value.
1. When your test result appears, press until the selection symbols
starts flashing. With the arrow keys ( ), select the desired mark
(before meal), (after meal), (training) or (general).
2. Confirm your selection by pressing the button. You can select
multiple selections (but only when you select the or ),
repeat steps 1 and 2.
3.
6.
The opening
of the test
Contact Area
How to manage the measurement results in memory.
Your meter stores up to 730 results with date, time and marks.
1. Make sure the meter is off. (If not, press for 3 seconds).
2. Press for 2 seconds to start (”mem” appears on the screen).
3. The latest test results with the time, date and selections are shown.
4. Press or to scroll to the memory.
5. Press for more than 3 seconds to turn off the meter.
View average results for the 1, 7, 14, 30, 60 or 90 days.
1. Press for 2 seconds to turn on the meter. Press again (”AVG”
is displayed on the screen). The meter shows the day’s average.
2. Continue to press to display the average for 1, 7, 14, 30, 60 or
90 days.
3. Press for more than 3 seconds to turn off the meter.
How to replace the battery in GlucoMen
®
Areo.
The meter is powered by two CR2032 3V batteries. When the symbol low
battery is lit, it is time to change. You do not have to set the date and time if you
insert the new batteries within 2 minutes after the old is removed. How to use:
1. Make sure the meter is off.
2. Open the battery cover on the back of the meter and remove the batteries.
3. Insert the new batteries in the battery holder with the ”+”- side up.
4. Close the battery cover.
Wireless transmission via NFC
(Near Field Communication)
Your saved test results can be transferred to an NFC-enabled device (such as
a smartphone) with an NFC-compatible application installed (eg GlucoLog).
The first time, you must turn on your meter NFC function (found in the
setup menus) by pressing the or to select ON or OFF. Press to
confirm and leave.
1. Activate the NFC-transfer in the app in your smartphone.
2. Place your meter close to your smartphone (<1 cm).
3. The meter starts automatically and the data is transferred ( flashes
on the screen).
4. When the data transfer is completed, ”End” appears on the screen
along with the symbol , and a beep sounds.
5. Press the button for 3 seconds to turn off the meter.
Test Strip Port
Button for
On/Off/Enter
Buttons
Up/Down
Eject button
Antenna (NFC)
Battery Cover
Data Transfer
Port
  • Page 1 1
  • Page 2 2

A.MENARINI GlucoMen Areo Quick Manual

Type
Quick Manual

A.MENARINI GlucoMen Areo is a portable wireless device that provides blood glucose monitoring with wireless transmission via Near Field Communication (NFC).

It has a large display that shows the blood glucose level, time, and date. The device also stores up to 730 results with date, time, and marks, and allows users to view average results for the past 1, 7, 14, 30, 60, or 90 days.

The GlucoMen Areo comes with a sampler that uses a new sterile lancet for each sampling and a sliding sleeve to lock the sampler. The puncture depth can be adjusted for different skin types, and the device provides error codes and messages to help users identify and resolve any issues.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI