Cellient
™
System Operator’s Manual
1.3
INTRODUCTION
Cytologic samples preserved in any fixative other than CytoLyt
or PreservCyt
™
Solution should be
washed in CytoLyt Solution and placed in a ThinPrep
™
vial containing PreservCyt Solution prior to
processing on the Cellient
™
Automated Cell Block System.
Do NOT use formalin on the Cellient Processor.
Samples that will be made into a cell block on the Cellient
™
System should have all residual material
saved from the routine cytology process. Quantity of sample will vary. Use the following table as a
guideline.
CAUTION:
Cytology samples
should utilize the Cellient System’s Auto Sample Dispense process.
Tissue fragments and core specimens
(maximum 14 gauge* needle) should be processed via the Cellient
System’s Manual Sample Dispense process.
* 14 g biopsy needle has an interior diameter of approximately 1.6 mm.
1. Hologic
ThinPrep
™
2000 Processor Operator’s Manual
. P/N MAN-09270-001.
see also:
Hologic
ThinPrep
™
Mucoid Specimens Quick Reference Guide.
P/N DS-05929-001.
Hologic
ThinPrep
™
Fine Needle Aspirates (FNA) Quick Reference Guide.
P/N DS-05928-001.
Hologic
ThinPrep
™
Body Fluids Specimen Collection Quick Reference Guide.
P/N DS-09699-001.
Table 1.1 Sample Quantity
Centrifuge Tube
Quantity
PreservCyt™ Solution
Vial Quantity Process
Centrifuge tube empty Vial with residual sample Run vial on the Cellient System
Tube with cell pellet Vial empty Put a maximum of 10 drops (approxi-
mately ¼ ml) of cell pellet into a vial con-
taining 20 ml of PreservCyt Solution and
allow to stand for a minimum of 15 min-
utes prior to processing to allow the
PreservCyt Solution to render the sample
non-infectious
Tube with cell pellet Vial with residual sample Put a maximum of 10 drops (approxi-
mately ¼ ml) of cell pellet into a vial con-
taining 20 ml of PreservCyt Solution. Top
off vial with PreservCyt Solution if needed
and allow to stand for a minimum of 15
minutes prior to processing to allow the
PreservCyt Solution to render the sample
non-infectious
Tube empty Vial empty Sample insufficient for cell block