SICK Comtrol DeviceMaster Operating instructions

Type
Operating instructions
OPERATING INSTRUCTIONS
Comtrol DeviceMaster
PC driver for connecting a LMS211/221 Laser
Measurement System via Ethernet interface
Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
2 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Copyright
Copyright © 2008
SICK AG Waldkirch
Auto Ident, Reute Plant
Nimburger Strasse 11
79276 Reute
Germany
Trademark
Windows 2000
TM
, XP
TM
, Vista
TM
and Internet Explorer
TM
are registered trademarks or trade-
marks of the Microsoft Corporation in the USA and other countries.
Acrobat
TM
Reader
TM
is a trademark of Adobe Systems Incorporated.
Latest manual version
For the latest version of this manual (PDF), see www.sick.com.
Operating Instructions
Comtrol DeviceMaster Ethernet
Contents
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 3
Table of contents
1 Notes on this document ............................................................................................... 4
Purpose...................................................................................................................... 4
Target group .............................................................................................................. 4
Depth of information ................................................................................................. 4
Used symbols ............................................................................................................ 5
2 Install PortVision Plus ................................................................................................... 6
3 PortVision Connections to DeviceMaster................................................................... 8
4 Configure the DeviceMaster IP information .............................................................. 9
5 Driver installation ........................................................................................................10
6 Driver configuration.....................................................................................................15
Operating Instructions
Comtrol DeviceMaster Ethernet
Figures and Tables
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 4
1 Notes on this document
1.1 Purpose
This document instructs technical staff to install the driver of the interface converter Com-
trol DeviceMaster Ethernet on PC. The driver is used to connect a SICK laser measuring sy-
stem LMS211/221 to the PC via Ethernet.
The following types of laser measurement system LMS211/221 have an Ethernet interface:
LMS211-S26 (Nr. 1044296)
LMS221-S26 (Nr. 1044297)
LMS211-S27 (Nr. 1044298)
This document contains the following information:
Installation or update of the driver on PC
Configuration and test of the interface converter Comtrol DeviceMaster Ethernet
A step-by-step approach is taken for all tasks.
Important The interface converter Comtrol DeviceMaster Ethernet is named in the following simplified
"Device Master" and the PC driver for the Comtrol DeviceMaster Ethernet is named "driver".
1.2 Target group
The traget group of this document is persons assigned the following tasks:
1.3 Depth of information
This document contains all information to install the driver on PC.
Important Further information about the driver of the Comtrol DeviceMaster Ethernet is available from
SICK AG, Auto Ident division.
On the Internet at www.sick.com.
Tasks Target group
Installation and configuration of
the driver
Qualified staff, e.g. service technicians or engineers
Table 1-1: Target group
Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
5 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Figures and Tables
1.4 Used symbols
To gain easier access, some information in this dokumentation is emphasised as follows:
NOTICE
Notice!
Indicates a potential risk of damage or impair on the functionality of the field bus module or
the connected device.
WARNING
Warning notice!
A warning noctice indicates real or potential danger. This should protect you against acci-
dents.
The safety symbol next to the warning notice indicates why there is a risk of accident. e.g.
due to electricity. The warning levels (CAUTION, WARNING, DANGER) indicate the serious-
ness of the risk.
¾ Carefully read and follow the warning notices.
Reference Italic script denotes a reference to further information.
Important This important note informs you about specific features.
Explanation An explanation provides background knowledge of technical nature.
Recommendation A recommendation helps you to carry out tasks correctly.
TIPP A tip explains setting options in the SOPAS-ET configuration software.
C
OMMUNICATION This type of script denotes a term in the user interface in the SOPAS-ET configuration soft-
ware.
This symbol denotes a section in which the operation steps with the SOPAS-ET configuration
software are described.
This symbol indicates supplementary technical documentation.
¾ There is a procedure which needs to be carried out. This symbol indicates operational in-
structions which only contain one operational step or operational steps in warning notices
which do not have to be followed in any particular order.
Operational instructions comprising several steps are denoted using consecutive numbers.
Operating Instructions
Comtrol DeviceMaster Ethernet
Figures and Tables
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 6
2 Install PortVision Plus
This chapter describes the installation of PortVision Plus. The required file is located on the
LMS CD in the directory .. \Ethernet installation\PortVisionV3.02.
Double-click the file pvplus_v3.02.msi.
The pop-up window PortVision Plus Setup appears:
1. Select Next.
2. To continue with the installation, select Next.
Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
7 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Figures and Tables
3. Accept the license agreement and select Next.
4. To confirm the installation folder, select Next.
5. To start with the installation, select Install.
6. Close the installation with Finish.
You will now have a PortVision Plus icon on the desktop.
Chapter 3 Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
8 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
PortVision Connections to DeviceMaster
3 PortVision Connections to DeviceMaster
This chapter describes the connection between PortVision and DeviceMaster.
1. Connect an Ethernet patch cable from the NIC in the PC to the DeviceMaster.
If the DeviceMaster has both an UP and DOWN Ethernet port, connect the patch cable
to the DOWN port.
2. Start PortVision Plus.
The pop-up window Create Shortcuts appears.
3. Select No.
Explanation There is a problem with the WCOM-Tool once open, it blocks the COM Port.
4. Select Yes to scan the network.
PortVision Plus opens up.
Important The screenshot shows IP information that does not apply to your network.
5. Note the IP address.
The IP address must be compatible with the static IP address assigned to the PC.
6. To communicate directly with the MasterDevice, use Telnet.
Operating Instructions Chapter 4
Comtrol DeviceMaster Ethernet
Configure the DeviceMaster IP information
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 9
4 Configure the DeviceMaster IP information
This chapter describes the configuration of the DeviceMaster.
Requirement: PortVision and DeviceMaster are connected.
1. Double-click on the DeviceMaster.
The pop-up window Configure Device: Device ... appears.
2. Modify the IP address, Subnetz Mask and Gateway address for compatibility with your
network.
Important The screenshot shows IP information that does not apply to your network.
3. Select Apply Changes.
4. Select Close.
The DeviceMaster will now reboot to set the new IP information.
Chapter 5 Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
10 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Driver installation
5 Driver installation
This chapter describes the installation of the driver. The required file is located on the LMS
CD in the directory .. \Ethernet installation\Driver.
Important If installing multiple DeviceMasters, you may repeat the following process once for each De-
viceMaster. Once all of the DeviceMaster drivers have been installed you can then go into
DeviceManager to configure the different units. You may also decided to install the driver,
then configure the driver in DeviceManager and then install the next unit, configure, install,
configure, etc. Either method is acceptable.
Important You may not see all of the following screenshots based on the Operating System that you
are running.
1. Double-click the file csa_nslink-winxp-1.00.exe.
This will create a default directory at C:\Comtrol and extract its files to this folder.
2. Go to Control Panel and Add Hardware.
Hardware-Wizard opens up.
3. Select Next.
Operating Instructions Chapter 5
Comtrol DeviceMaster Ethernet
Driver installation
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 11
4. Select Next.
5. Select Next.
6. Select Next.
Chapter 5 Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
12 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Driver installation
7. Select the type of hardware.
8. Select the Have Disk... option.
9. Enter the path of C:\Comtrol.
10. Select the correct model type..
Operating Instructions Chapter 5
Comtrol DeviceMaster Ethernet
Driver installation
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 13
11. Select Next to install the hardware.
12. Select Finish.
Depending on the Windows version, the New Hardware Wizard opens up.
Chapter 5 Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
14 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Driver installation
13. Select Next.
14. Select Next.
15. Select Finish.
16. Repeat steps 2–15 to install the driver for further DeviceMaster.
Operating Instructions Chapter 6
Comtrol DeviceMaster Ethernet
Driver configuration
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 15
6 Driver configuration
This chapter describes the configuration of the driver.
Important In DeviceManager, do not open the properties pages for multiple DeviceMasters at the
same time!
1. Go to System Properties and System.
2. Select hardware.
3. Select DeviceManager.
4. Expand MultiPortSerial Adapters.
5. Double-click the DeviceMaster.
The properties pages opens up.
Chapter 6 Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
16 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Driver configuration
6. Go to the Port Settings tab.
7. Choose COM-Port and select Properties.
The pop-up window COM5 Properties opens up.
8. Change COM port RS mode to RS422 and select OK.
Operating Instructions Chapter 6
Comtrol DeviceMaster Ethernet
Driver configuration
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 17
9. Go to the Device Settings tab.
10. Set Keep Alive Timeout (seconds): 12.
11. Place a checkmark in the option Do NOT attempt to load firmware to device.
12. Select OK.
The pop-up window Warning appears.
13. Select Yes.
The pop-up window RpshSi appears.
14. Select Yes.
Driver configuration pages will now close.
15. Reopen the properties.
16. Go to the Network Connections tab.
Chapter 6 Operating Instructions
PC driver for Comtrol DeviceMaster Ethernet
18 © SICK AG · Division Auto Ident · Germany · All rights reserved 8012680/0000/2008-07-25
Driver configuration
17. Click the drop down arrow next to the MAC mode and select the appropriate MAC
address.
18. Select OK.
The proerties pages will close.
19. Reopen the properties page
20. Go again to the Network Connections tab.
21. Choose IP Mode and set the correct IP address.
Operating Instructions Chapter 6
Comtrol DeviceMaster Ethernet
Driver configuration
8012680/0000/2008-07-25 © SICK AG · Division Auto Ident · Germany · All rights reserved 19
22. Choose button Retrieve Network Addresses from the Device.
23. Go to the Advisor tab.
The summary should show Device is active and OK.
24. Select OK and test the DeviceMaster for proper operation.
SICK AG | Waldkirch | Germany | www.sick.com
8012680/2008-07-xx 5M/TR <PM6.5/FM7.0/PDF>/VD ∙ Printed in Germany ∙ Subject to change without notice ∙ The specified product features and technical data do not represent any guarantee ∙ 05 Pre int61
Australia
Phone +61 3 9497 4100
1800 33 48 02 – tollfree
Belgium/Luxembourg
Phone +32 (0)2 466 55 66
E-Mail
Brasil
Phone +55 11 3215-4900
E-Mail
Ceská Republika
Phone +420 2 57 91 18 50
E-Mail
sick@sick.cz
China
Phone +852-2763 6966
E-Mail
Danmark
Phone +45 45 82 64 00
E-Mail
sick@sick.dk
Deutschland
Phone +49 211 5301-270
E-Mail
España
Phone +34 93 480 31 00
E-Mail
France
Phone +33 1 64 62 35 00
E-Mail
Great Britain
Phone +44 (0)1727 831121
E-Mail
India
Phone +91–22–4033 8333
Israel
Phone +972-4-999-0590
E-Mail
Italia
Phone +39 02 27 43 41
E-Mail
Japan
Phone +81 (0)3 3358 1341
E-Mail suppor[email protected]
Nederlands
Phone +31 (0)30 229 25 44
E-Mail
Norge
Phone +47 67 81 50 00
E-Mail
Österreich
Phone +43 (0)22 36 62 28 8-0
E-Mail
Polska
Phone +48 22 837 40 50
E-Mail
Republic of Korea
Phone +82-2 786 6321/4
E-Mail
kang@sickkorea.net
Republika Slowenija
Phone +386 (0)1-47 69 990
E-Mail
România
Phone +40 356 171 120
E-Mail
Russia
Phone +7 495 775 05 34
E-Mail
info@sick-automation.ru
Schweiz
Phone +41 41 619 29 39
E-Mail
Singapore
Phone +65 6744 3732
E-Mail
Suomi
Phone +358-9-25 15 800
E-Mail
sick@sick.fi
Sverige
Phone +46 10 110 10 00
E-Mail
Taiwan
Phone +886 2 2365-6292
E-Mail
Türkiye
Phone +90 216 587 74 00
E-Mail
USA/Canada/México
Phone +1(952) 941-6780
1 800-325-7425 tollfree
E-Mail info@sickusa.com
More representatives and agencies
in all major industrial nations at
www.sick.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20