Vaisala PR-43 User manual

Type
User manual
IM-EN-PR43ACAP v. 2.00
Instruction Manual
Vaisala K-PATENTS®
Sanitary Process Refractometer
PR-43-AC/AP
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärvene 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
General safety consideraons
The process medium may be hot or otherwise hazardous. Use shields and
protecve clothing adequate for the process medium - do not rely on avoid-
ance of contact with the process medium.
Precauons when removing a refractometer from the process line :
• Check rst that the process line is depressurized and drained.
• Ensure you stay clear of any possible spillage and you have a clear emer-
gency escape path.
It is the users responsibility to follow manufacturer’s safety and operang instrucons.
The client’s organizaon has the responsibility to develop and maintain occupaonal
safety and create a safety culture where individuals are expected to follow safety in-
strucons at all mes. Any negligence towards safety instrucons or failure to comply
with safe pracces should not be tolerated. It is the manufacturers responsibility to
produce goods that are safe to use when instrucons are followed.
Disposal
When wishing to dispose of an obsolete refractometer or any parts of a
refractometer, please observe local and naonal regulaons and require-
ments for the disposal of electrical and electronic equipment.
© Copyright Vaisala 2019. All rights reserved.
Symbols and terms used in this manual:
This indicates a warning. It provides safety precauon informaon needed to
avoid injury while operang the refractometer system.
This indicates that something is important for the operaon of the refractome-
ter system.
Note. Notes contain addional informaon and hints.
This product manual is delivered to the end user with a Vaisala K-PATENTS
®
product.
Informaon in this manual is subject to change without noce. When the manual is
changed, a revised copy is published at hp://www.kpatents.com/
Warranty
For standard warranty terms and condions, see www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of damage due to nor-
mal wear and tear, exceponal operang condions, negligent handling or installaon,
or unauthorized modicaons. Please see the applicable supply contract or Condions
of Sale for details of the warranty for each product.
Technical support
Contact Vaisala technical support at [email protected]. Provide at least the fol-
lowing supporng informaon:
• Product name, model, and serial number
• Name and locaon of the installaon site
• Name and contact informaon of a technical person who can provide further infor-
maon of the problem
For more informaon, see www.vaisala.com/support.
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
Table of contents
1 Refractometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Mounng . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Choosing the mounng locaon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Check list for pipe mounng . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.3 Check list for mounng in a tank, a vessel or a large pipe . . . . . . . . . 3
2.4 Electrical connecons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4.1 Wiring opons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Specicaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1 Compability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Refractometer rangeability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Sanitary process refractometer PR-43-AC . . . . . . . . . . . . . . . . . . . . . 10
3.3.1 PR-43-AC model code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.2 PR-43-AC mounng hardware model code . . . . . . . . . . . . . . . . . . 13
3.3.3 PR-43-AC specicaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.3.4 PR-43-AC parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3.5 PR-43-AC mounng specics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3.6 Mounng specics for EHEDG cered PR-43-AC conguraon . . 28
3.3.7 3-A Sanitary Standard compliance . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.4 Sanitary process refractometer PR-43-AP . . . . . . . . . . . . . . . . . . . . . 30
3.4.1 PR-43-AP model code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.4.2 PR-43-AP mounng hardware model code . . . . . . . . . . . . . . . . . . 35
3.4.3 PR-43-AP specicaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.4.4 PR-43-AP parts lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.4.5 PR-43-AP mounng specics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.4.6 Mounng specics for EHEDG cered PR-43-AP conguraon . . 42
3.4.7 3-A Sanitary Standard compliance . . . . . . . . . . . . . . . . . . . . . . . . . 42
4 Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.1 Prism coang . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.2 Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.2.1 Recommended wash pressures and mes . . . . . . . . . . . . . . . . . . . 45
4.2.2 Prism wash systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.2.3 Prism wash nozzles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5 Refractometer vericaon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
PR-43 refractometer vericaon form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
EU declaraon of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
1 Refractometer 1
© Copyright Vaisala 2019. All rights reserved.
1 Refractometer
The sanitary refractometers PR-43-AC and PR-43-AP provide a digital interface detec-
on system for alimentary industry.
The PR-43-AC/AP (gure 1.1) measures the refracve index n
D
and the temperature of
the process medium. The concentraon of the process liquid is calculated from these
values when the composion of the process medium is known.
The output values of the refractometer are transmied through mA output and digitally
through an Ethernet connecon by using a UDP/IP protocol (see the general manual for
specicaons). mA output is available with mA output cable (mA only) or with a split
cable (mA and Ethernet).
PR-43-AC PR-43-AP
Figure 1.1 Refractometer models
2 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
2 Mounng
The refractometer mounng locaon should be chosen with care to ensure reliable
readings from the process.
2.1 Choosing the mounng locaon
The mounng locaon needs to be such that sediments or gas bubbles cannot accumu-
late by the refractometer. Good ow velocity is essenal in keeping the prism clean.
Important: If the process pipe vibrates, support the pipe. A vibrang pipe might dam-
age the in-line refractometer mounted on it.
A Vaisala K-PATENTS
®
inline refractometer can be located either indoors or outdoors in
most climates. However, when a refractometer is located outdoors, some basic pro-
tecon against direct exposure to sunlight and rain should be provided. Special care
should be taken if the pipe wall is translucent (e.g. of berglass), as light from outside
reaching the prism through the pipe wall may disturb the measurement.
The refractometer cover should not be exposed to high temperature radiaon. In most
cases, dra and natural convecon provide sucient air cooling if the air gets to ow
freely around the refractometer head.
Addional cooling is necessary when the ambient temperature is higher than 45 °C
(113 °F) or when the process temperature is above 110 °C (230 °F) and the ambient
temperature is above 35 °C (95 °F). The air cooling is improved by blowing pressurized
air against the refractometer cover. The pressurized air can be supplied by the ven-
laon system. It is also possible to mount a PR-14038 cooling cover for cooling with
water.
Important: Always mount the refractometer so that the interconnecng cable points
downwards from the refractometer head.
2.2 Check list for pipe mounng
The PR-43-AC and PR-43-AP are mounted in a pipe. Vaisala recommends a ow velocity
between 1 and 3 m/s (3–10 /s). If the ow velocity exceeds 6 m/s (20 /s), there is
a risk of cavitaon. Cavitaon may damage the refractometer and the piping. Too
2 Mounng 3
© Copyright Vaisala 2019. All rights reserved.
slow a ow velocity may cause erraneous measurement readings due to coang of
stracaon (layering) of the sample on the prism.
The diameter and form of the pipe and the process temperature all aect the measure-
ment and need to be taken into account.
– If the process pipe diameter varies, select the posion with the smallest diameter
(and accordingly highest velocity). Then the prism stays clean beer.
– If the refractometer is used in a feed-back control loop, make the me lag short. E.g.
when a diluon valve is controlled, mount the refractometer close to the diluon
point. However, make sure complete mixing has occurred at mounng locaon.
– If the temperature varies along the process pipe, select the posion with the highest
process temperature. That minimizes the risk of coang, because higher tempera-
ture means higher solubility and also lower viscosity.
– Oen the posion with the highest process pressure (= aer pump + before valve)
has favorable ow condions without sedimentaon or air trapping risks.
– The refractometer should be conveniently accessible for service.
2.3 Check list for mounng in a tank, a vessel or a large pipe
A probe refractometer PR-43-AP can be inserted with a ange or clamp into tanks and
vessels which either don’t have a scraper or where the mixer doesn’t touch the vessel
wall. A probe refractometer can also be ush mounted in a cooker where the scraper
touches the wall.
– The inserted probe refractometer is mounted close to a srrer to ensure represen-
tave sample of the process liquid and to keep the prism clean.
– The refractometer should be conveniently accessible for service.
4 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
Prevent outside light
Easy access
Easy access
High pressure
High temperature
High pressure
Small diameter
High temperature
High velocity >1.5m/s
(>5ft/s)
Upper pipe bend
Ambient air
cooling
2 Mounng 5
© Copyright Vaisala 2019. All rights reserved.
2.4 Electrical connecons
The refractometer has an M12 connector in the refractometer for power supply, mA
output and Ethernet connecons.
Figure 2.1 The M12 connector
PR-43 refractometers are powered with 24 VDC. For connecng the refractometer to
Mulchannel user interface MI, see the manual for Mulchannel user interface. For
connecng the refractometer to Compact user interface CI, see the manual for Compact
user interface. See Figure 2.2 for instrucon how to connect the M12 refractometer
cable.
2.4.1 Wiring opons
The PR-43 refractometer provides both analogue (mA) and digital output signals. See
Figure 2.3 for the wiring when only analogue output is used.
Opons for connecng the refractometer with both analogue and digital outputs are
shown in Figure 2.4. Both Compact user interface CI and Mulchannel user interface
MI use the digital output signal.
6 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
Turn to tighten
Align groove and
guide
P4-031
Connector screw thread
Figure 2.2 Connecng the refractometer cable to the refractometer
10m (33ft)
90m (295ft)
180m (590ft)
200m (656ft)
P4-001
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
max. 90m (295ft) max. 90m (295ft)
max. 90m (295ft)
10m (33ft)
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8660
Platform 4 cable extender
M12-8 pin, A-code, F+M
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
max. 200m (656ft)
PR-8680
M12-8 pin, A-code, F
Field plug
PR-8432
Power supply and
mA output cable
Refractometer only Cables
M
M M
M MM
F
F
F
F
F
F F
Figure 2.3 Wiring with analogue output only
2 Mounng 7
© Copyright Vaisala 2019. All rights reserved.
10m (33ft)
90m (295ft)
180m (590ft)
200m (656ft)
P4-001
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
max. 90m (295ft) max. 90m (295ft)
max. 90m (295ft)
10m (33ft)
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8660
Platform 4 cable extender
M12-8 pin, A-code, F+M
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
max. 200m (656ft)
PR-8680
M12-8 pin, A-code, F
Field plug
PR-8432
Power supply and
mA output cable
Refractometer only Cables
M
M M
M MM
F
F
F
F
F
F F
Figure 2.4 Wiring opons when both analogue and digital outputs are used
8 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
10m (33ft)
90m (295ft)
180m (590ft)
P4-001/1
PR-43
REFRACTOMETER
M-12-8 pin, A code, M
Unclassified area
max. 90m (295ft) max. 90m (295ft)
max. 90m (295ft)
PR-8430-___
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430-___
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430-___
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8660
Platform 4 cable extender
M12-8 pin, A-code, F+M
PR-8444-001/010
Split cable M12
M12-8 pin, A-code, F
Ethernet (RJ 45)
mA, 24V
1m (3ft) or 10m (33ft)
Stand-alone refractometer with RJ45
M
MM
F
F
F
F
PR-8444-001
Split cable M12
M12-8 pin, A-code, F
Ethernet (RJ45)
mA, 24V
1m (3ft)
F
PR-8444-001
Split cable M12
M12-8 pin, A-code, F
Ethernet (RJ45)
mA, 24V
1m (3ft)
F
PR-43
REFRACTOMETER
M-12-8 pin, A code, M
Unclassified area
PR-43
REFRACTOMETER
M-12-8 pin, A code, M
Unclassified area
Figure 2.5 Wiring opons when both analogue and
digital outputs are used, with RJ45 ethernet connector
2 Mounng 9
© Copyright Vaisala 2019. All rights reserved.
Figure 2.6 Connecons in M12 and RJ-45 connectors
In split cables the power cable and the mA output cable are marked near the end of
the cable.
Power cable:
• white: +24 DC
• brown: GND
mA output cable:
• white: mA+
• brown: mA-
10 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
3 Specicaons
3.1 Compability
Electrically: The PR-43 refractometers are not interchangeable with any other refrac-
tometer model. All PR-43-AC/AP refractometers are however interchangeable with
each other as long as they have the same prism. The PR-43-AC/AP refractometers are
not compable with the indicang transmiers DTR, STR or IT-R.
Mechanically: The sanitary process refractometer PR-43-AC/AP ts the same sanitary
process connecons as PR-23-AC/AP.
3.2 Refractometer rangeability
The refracve index standard range of a PR-43-AC/AP is 1.320-1.530 (corresponds to
0-100 Brix), Figure 3.1, with a Sapphire prism. A PR-43-AC can also be equipped with a
Sapphire prism with a refracve index range 1.260-1.470.
Figure 3.1 PR-43-AC rangeability
3.3 Sanitary process refractometer PR-43-AC
The refractometer PR-43-AC is a 3-A and EHEDG cered sanitary process refractome-
ter for measuring concentraons in a pipeline. It is easy to install in any pipe size di-
rectly or using a ow cell. The Sanitary process refractometer is suitable for all food
and beverage processing applicaons where real-me monitoring and control can help
to improve product quality and reduce costs.
3 Specicaons 11
© Copyright Vaisala 2019. All rights reserved.
Figure 3.2 PR-43-AC dimensions
12 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
3.3.1 PR-43-AC model code
SANITARY COMPACT REFRACTOMETER for pipelines
Model and descripon Model
PR-43 = Refractometer PR-43
Refractometer model
-AC = Compact sanitary cered process refractometer, inseron length 14
mm
-AC
Prism material and Refracve Index range limit
-73 = R.I. 1.320-1.530 n
D
(0-100 Brix) Sapphire prism -73
-74 = R.I. 1.260-1.470 n
D
Sapphire prism -74
Connecon type and size
-E25-P15 = Varivent DN65, 15 bar (A) -E25-P15
-H25-P15 = Sanitary 3A-clamp, inch, 15 bar (A) -H25-P15
-H25-P40 = Sanitary 3A-clamp, inch, High Pressure, 40 bar @ 20°C (A) -H25-P40
-Z25-P15 = inch I-clamp, 15 bar (B) -Z25-P15
Refractometer weed parts material
-SS = AISI 316 L -SS
-HC = Alloy C276 -HC
Electrical classicaon
-UN = Unclassied area, general purpose, ordinary locaon -UN
-AX = EX and IECEx cered Ex II 3G, Ex nA IIC T4 Gc
(up to zone 2) (T
amb
-40...+65°C)
-AX
-IA = ATEX and IECEx cered Ex II 1G, Ex ia IIC T4 Ga
(up to zone 0) (T
amb
-40 ... +65°C)
-IA
High accuracy opon
-HAC = High accuracy version in range 0-30 Brix 4-30°C -HAC
EHEDG opon
-EH = EHEDG Type EL class I Cered Model -EH
Polishing opon
-EP = Electropolished refractometer weed parts (Ra 0.4 µm, 15 µ inch) -EP
(A) EHEDG cered version available
(B) Sanitary 3-A cered
3 Specicaons 13
© Copyright Vaisala 2019. All rights reserved.
3.3.2 PR-43-AC mounng hardware model code
Mounng hardware (without wash nozzles)
MODEL AND DESCRIPTION MODEL
AFC = Elbow ow cell AFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
10 = 25 mm (1 inch) (A) 10
15 = 40 mm (1 1/2 inch) 15
20 = 50 mm (2 inch) 20
25 = 65 mm (2 1/2 inch) (A) 25
30 = 80 mm (3 inch) (A) 30
40 = 100 mm (4 inch) (A) 20
Flow cell inlet type
-SI = Straight pipe -SI
-RI = Reduced pipe (cone) -RI
Opons
-EP = Electropolished process weed parts (RA 0.4µm, 15 µ inch) -EP
(A) with -SI opon only
EHEDG cered mounng hardware
MODEL AND DESCRIPTION MODEL
AFC = Elbow ow cell AFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
20 = 50 mm (2 inch) 20
Flow cell inlet type
-SI = Straight pipe -SI
EHEDG
-EH = EHEDG Type EL Class I Cered model -EH
Polishing opon
-EP = Electropolished process weed parts (Ra 0.4µm, 15 µ inch) -EP
For EHEDG cered refractometers use ISO2852 type EHEDG cered gasket
PR-9202-EH (2½ inch) or PR-9272-EH (4 inch).
14 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/2.00 Eecve: September 2, 2019
Mounng hardware with wash nozzles
MODEL AND DESCRIPTION MODEL
AFC = Elbow ow cell AFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
10 = 25 mm (1 inch) 10
15 = 40 mm (1 1/2 inch) 15
20 = 50 mm (2 inch) 20
25 = 65 mm (2 1/2 inch) (A) 25
30 = 80 mm (3 inch) (A) 30
40 = 100 mm (4 inch) (A) 40
Flow cell inlet type
-SI = Straight pipe -SI
-RI = Reduced pipe (cone) -RI
Wash nozzle connecon
-NC = Nozzle connecon -NC
Wash nozzles for 10/15 ow cells
-SN = Steam nozzle, threads G ¼ inch female -SN
-WN = Water nozzle, threads G ¼ inch female -WN
-WP = Pressurized water nozzle, threads G ¼ inch female -WP
-PG = Plug for nozzle connecon -PG
(A) with -SI opon only
Example: Refractometer:PR-43-AC-73-H25-P15
Flow cell: AFC-HSS-H20-SI
Note: Prism wash control and diagnoscs require the use of a Mulchannel user in-
terface MI (or customers own control system).
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65