NGA2000 Reference
Verify you know which analyzer is which.
Normally this is easy, but if you have sev-
eral examples of the same kind of analyzer,
you will need to verify they are named ap-
propriately.
You can do this even though all the analyz-
ers are hooked up at once, since you can
look at them and see what their serial num-
bers are.
From the main screen, press "MENUS"
Select Technical level configuration...
Select Service menus...
Select Manufacturing data...
Select Analyzer module data. Read Ana-
lyzer module s/n: and determine which
analyzer you are looking at.
Edit the User tag number: line to make the
tag whatever you want it to be. The tag is
the heading on the top of the screen for the
analyzer you are looking at. This line won't
change until you enter another menu, but
the analyzer will remember whatever you
enter, up to 31 characters.
Press "HOME".
Press "NEXT" (it selects the next analyzer
in its list.).
Repeat the procedure for the new analyzer.
An alternative is to start the Rosemount
Analytical DDE server program in a PC
connected to the system. When it has
completed its network interrogation, open
the DDE server program icon and click on
"File". Select "Nodelist". Click on each
module's TAG in turn, and edit them as you
like. Then exit and restart the DDE server
before attempting to use any application
with it.
d. Details
TAG Variable
All modules use the variable called TAG to
contain their name. Modules also have a
node name, and a program id, and these
are used internally in the LON. They are
not normally accessible externally, though
both may be used by non-NGA LON nodes
such as a PC setup to monitor the NGA sys-
tem.
Analyzer Modules also have a serial num-
ber in their variable AMSN. This should
match the serial number assigned to the
analyzer and visible on a label on their
sides.
The control module puts the contents of
TAG onto the main display screens, and
also onto the top line of the menu screens.
It regards menus as belonging to an ana-
lyzer, the one selected in the main screen
or selected by pressing the NEXT button in
higher level menus. The TAG and the pre-
sent time reading are always displayed on
the top line of every menu screen.
NOTE
TAGs should be kept as short as possi-
ble, and should not contain blanks,
dashes or other mathematical signs.
This will avoid problems when using a
PC to interface with the module.
The gas measured is shown in the variable
GAS. This is normally straightforward, but
in the case of the FID you should verify that
this variable is matched with the response
factor, whether the analyzer is measuring in
C1 or C3 (methane or propane) units, for
example.
The analyzer sensitivity is described in
coded form in its bench code (AMBC). Par-
ticularly in the case of NDIR analyzers, the
bench has to be setup for the analysis de-
sired.
Individual analyzers may also have internal
serial numbers reported in other variables
visible in their Manufacturing data screen.
Network variables: TAG, AMSN, AMBC,
GAS, CARBON_ATOMS.