630 Barrier-Free® Universal Procedures Table (-001 thru -009)

Midmark 630 Barrier-Free® Universal Procedures Table (-001 thru -009), 230 Universal Procedures Chair Installation guide

  • Hello! I am an AI chatbot trained to assist you with the Midmark 630 Barrier-Free® Universal Procedures Table (-001 thru -009) Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • What is the intended use of the Urology Drain System?
    What are the restrictions on using the Urology Drain System?
    What special tools are required for installation?
    How do I remove the seat and back upholstery?
    How do I install the hose guide?
    How do I route the hose through the chair?
© Midmark Corporation 2008
1
Urology Drain System - Field Installed
[9A329002 & 9A329004]
Contents:
Installation ..............1
Exploded View ........ 3
Special Tools:
none
003-1428-00 Rev. GA1 10/25/18
Intended Use:
To provide drainage for urology procedures.
Step 1: Run BASE, TILT, & FOOT functions all the way UP.
Unplug power cord.
Step 2: Remove seat / back upholstery.
Step 3: Remove mounting board
and support plate.
Restrictions:
Not for use with rotational base units.
Style A
NextNext
BackBack
© Midmark Corporation 2008
2
Wire Harness Cover
Cage Bracket
Step 5: Remove two screws from
wire harness cover.
Note: Keep screws to secure cage bracket.
Step 6: Position the cage bracket as desired (see Note).
Secure with two screws.
Note: The cage bracket can be positioned on either side of the chair.
Step 4: Remove four nuts and
pan holder assembly.
NextNext
BackBack
© Midmark Corporation 2008
3
Step 7: Install hose guide.
Step 8: Install washers on new pan holder.
Note: Install three washers on both front studs &
one washer on right rear stud.
Step 9: Position pan holder.
Secure with four nuts.
Step 10: Secure cage to
mounting bracket.
Step 11: Attach hose to hose clip.
Insert hose clip into bottle.
Install bottle into cage.
Step 12: Route hose thru chair.
Attach hose to drain pan w/ hose clamp.
NextNext
BackBack
© Midmark Corporation 2008
4
Serial Numbers / Dates:
all
17 041-0008-02
1 016-0885-01
2 030-0647-00
3 030-1513-00
5 053-0788-00
030-1524-01 9
050-6066-00 8
053-1453-01 7
016-0972-00 6
040-0250-17 10
a): 050-5966-01 11
b): 050-5966-03
040-0312-41 12
053-0050-12 16
050-5983-00 13
040-0010-47 14
15 053-0382-00
4 029-3296-00
9A329002 (includes items 1 thru 16)
9A329004 (includes items 1 thru 15)
Urology Drain System - Field Installed
Item Description Qty.
1 • Hose Clamp 1
2 • Screen 1
3 • Drain Pan 1
4 • Pan Holder Assembly 1
5 • Container 1
6 • Hose Clip 1
7 • Drain Hose 1
8 • Cage Bracket 1
9 • Cage Weldment 1
10 • Screw (1/4-20 x 1/2”) 2
11
• Spacer Cover
a) Midmark models
b) Ritter models
1
12 • Hex Screw (5/16-18 x 1") 3
13 • Hose Guide 1
14 • Screw (#10-24 x 3/8”) 2
15 • Nylon Washer 7
16 • Hole Plug 2
17 Nut (8-32) - not included 2
Rev. (2/18/15)
NextNext
BackBack
/