Sirona DAC PROFESSIONAL Operating Instructions Manual

Type
Operating Instructions Manual
a^`=molcbppflk^i
léÉê~íáåÖ=fåëíêìÅíáçåë
ïáíÜ=ìåáí=ëçÑíï~êÉ=îÉêëáçå=QKMT
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
2D 3554.201.01.04.02
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 3
Contents
1Warning and safety information ................................................................................ 7
2Electromagnetic compatibility ................................................................................... 11
3System description ..................................................................................................... 13
3.1 Device views..................................................................................................................................... 13
4Technical data ............................................................................................................. 14
4.1 Primary technical data ...................................................................................................................... 14
4.2 Capacity / weights............................................................................................................................. 14
4.3 Operating media ............................................................................................................................... 14
4.4 Operating parameters....................................................................................................................... 14
4.5 Installation example .......................................................................................................................... 15
5Control panel ............................................................................................................... 16
6Performance features of unit ..................................................................................... 17
6.1 Fractioned pre-vacuum procedure.................................................................................................... 17
6.2 Sterilization categories...................................................................................................................... 17
6.3 Separate steam generation .............................................................................................................. 17
6.4 One-way system / conductivity measurement / water supply ........................................................... 17
6.5 Electronic parameter control............................................................................................................. 18
6.6 Vacuum drying.................................................................................................................................. 18
6.7 Preheating ........................................................................................................................................ 18
6.8 Documentation of batch results ........................................................................................................ 18
7Installation ................................................................................................................... 19
7.1 Installation requirements................................................................................................................... 19
7.2 Transport straps................................................................................................................................ 20
7.3 Alignment.......................................................................................................................................... 20
7.4 Power supply connection.................................................................................................................. 20
7.5 Outlet connection for one-way water ................................................................................................ 20
7.6 Internal water supply with feed water ............................................................................................... 20
7.7 Water treatment system.................................................................................................................... 21
7.8 Emptying the water storage tank ...................................................................................................... 21
8Startup.......................................................................................................................... 22
8.1 Printer connection and initialization (optional) .................................................................................. 22
8.2 Vacuum test...................................................................................................................................... 22
8.3 Test run............................................................................................................................................. 23
8.4 Installation log................................................................................................................................... 24
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
4D 3554.201.01.04.02
9For each sterilization cycle ........................................................................................ 25
9.1 Operating media .............................................................................................................................. 25
9.2 Re-processing of instruments .......................................................................................................... 26
9.3 Loading the autoclave...................................................................................................................... 27
9.4 Closing the door............................................................................................................................... 29
9.5 Program selection ............................................................................................................................ 29
9.6 Program start ................................................................................................................................... 30
9.7 Program cycle .................................................................................................................................. 31
9.8 Record printout ................................................................................................................................ 33
9.9 Removing the sterilized items .......................................................................................................... 34
9.10 Sterile storage.................................................................................................................................. 34
9.11 Sterilization frequency / interval periods .......................................................................................... 35
9.12 Manual interruption of the program .................................................................................................. 35
9.12.1 Interruption of sterilization............................................................................................ 35
9.12.2 Interruption of drying.................................................................................................... 36
9.13 Response in case of warnings and error messages ........................................................................ 38
9.14 Idle time ........................................................................................................................................... 38
10 Putting the autoclave out of operation / transport / putting the autoclave back into
operation ................................................................................................................................. 39
11 Special functions ........................................................................................................ 40
11.1 Display of water quality (conductivity) / preheating temperature of chamber .................................. 40
11.2 Selection of additional drying ........................................................................................................... 40
11.3 Logging/Documentation of batch results.......................................................................................... 41
11.3.1 Output of records .............................................................................................................. 41
11.3.2 Connection to a computer............................................................................................ 43
11.3.3 Immediate printout yes/no............................................................................................ 43
11.4 Setting the date and time ................................................................................................................. 48
11.5 Automatic preheating ....................................................................................................................... 49
11.6 Total cycle counter........................................................................................................................... 50
11.7 Feed water supply............................................................................................................................ 51
11.8 Program modifications ..................................................................................................................... 52
12 Operating errors/ Malfunctions.................................................................................. 53
12.1 What do do in case of malfunctions ................................................................................................. 53
12.2 Malfunctions without an error message ........................................................................................... 53
12.2.1 Empty display window.................................................................................................. 53
12.2.2 The autoclave uses too much feed water .................................................................... 53
12.2.3 Poor drying................................................................................................................... 54
12.3 Warning messages .......................................................................................................................... 55
12.4 Error messages................................................................................................................................ 57
13 Preserving the value of this device ........................................................................... 62
13.1 Preparation of instruments............................................................................................................... 62
13.2 Rust formation = drag-in rust ........................................................................................................... 63
13.3 Regular care of the DAC PROFESSIONAL..................................................................................... 63
13.4 Functional testing of the autoclave .................................................................................................. 65
14 Appendix...................................................................................................................... 67
14.1 Drying instructions ........................................................................................................................... 67
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 5
Dear Customer, Thank you for purchasing the DAC PROFESSIONAL
class B sterilizer from Sirona. This autoclave
corresponds to the requirements of all valid safety
standards. The manufacturer has checked the unit using
the parameters for sterilization with the unit properly
loaded.
It is extremely important to read the following Operating
Instructions very carefully prior to startup, since
improper use may impair the sterilization effect
considerably.
Thank you for your trust in us.
Your Hygiene Team
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
6D 3554.201.01.04.02
Maintenance of the unit’s operational reliability
and value depend on:
1. Correct processing of the instruments being sterilized
2. Careful care of the unit
3. The use of high-quality feed water
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 7
1 Warning and safety information
Retention of documentation It is very important to make sure that this manual is
stored together with the unit so that it will also be
available at any time in the future. In case the unit is sold
or the right to use it is transferred to a different owner,
make sure that the manual always remains located near
the unit. This is necessary to ensure that the new owner
can use the unit properly and has all required
information.
Read carefully before starting! These instructions serve to promote safety. Be sure to
read them carefully before installing and using the unit.
This symbol appears on the rating plate to remind the
user to observe the manual.
Medical electrical devices may be used only with
reference to the risk of electrical shock, fire or
mechanical hazards according to IEC 60601-1.
Intended use DAC PROFESSIONAL may be used only by qualified
dental personnel.
Disclaimer
The manufacturer assumes no liability whatsoever in
cases involving:
work or repairs performed by any personnel not
authorized by the manufacturer or importer.
an application which does not correspond to the
ones described in this manual.
the use of non-original components or any
components not described in the section entitled
Parts list.
Electromagnetic compatibility In order to prevent possible risks caused by
electromagnetic interference, no medical electrical
or other equipment should be used near the
DAC PROFESSIONAL. The unit complies with the
currently valid directives concerning electromagnetic
compatibility (IEC 60601-1-2).
Only medical devices which comply with IEC 60601-1
may be connected to the DAC PROFESSIONAL. Any
non-medical devices connected also must comply with
the IEC 60601-1-1 standard for medical devices.
This unit does not cause any harmful interferences.
It can, however, pick up interferences from other
devices, including interferences which cause
undesirable operation.
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
8D 3554.201.01.04.02
Operating environment Never use this unit in the presence of flammable
anesthetics or anesthetic agents mixed with air, oxygen
or nitrogen.
Inspection and maintenance This point should be implemented according to the
relevant Sirona maintenance documents.
Power supply DAC PROFESSIONAL must be operated with a power
supply of 230 V, 50 - 60 Hz.
As the manufacturer, Sirona Dental Systems GmbH
provides on written request qualified technical
personnel, wiring diagrams, parts lists, and other
useful information for the repair and maintenance of
those parts which the manufacturer considers to be
repairable.
Disposal It generally applies that any disposal of this product must
comply with the relevant national regulations.
Please observe the regulations applicable in your
country.
Within the European Economic Community, Council
Directive 2002/96/EU (WEEE) requires environmentally
sound recycling/disposal of electrical and electronic
devices.
Your product is marked with the adjacent symbol.
Disposal of your product with domestic refuse is not
compatible with the objectives of environmentally sound
recycling/disposal.
The black bar underneath the "garbage can" symbol
means that it was put into circulation after Aug. 13,
2005. (see EN 50419:2005)
Please note that this product is subject to Council
Directive 2002/96/EU (WEEE) and the applicable
national law of your country and must be recycled or
disposed of in an environmentally sound manner.
Please contact your dealer if final disposal of your
product is required.
ESD protective measures ESD stands for ElectroStatic Discharge.
ATTENTION!
Connector pins or sockets bearing ESD warning labels
must not be touched or interconnected without ESD
protective measures.
Humidity: 30% - 60%
Ambient temperature: 16°C - 26°C
60°F - 80°F
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 9
ESD protective measures include:
Procedures for preventing electrostatic charge
build-up (e.g. air conditioning, air moistening,
conductive floor coverings and non-synthetic
clothing)
Discharging the electrostatic charges of your own
body on the frame of the UNIT, the protective ground
wire or large metallic objects
Connecting yourself to ground using a wrist band.
We therefore recommend that all persons working with
this system be instructed on the significance of this
warning label. Furthermore, they also should receive
training in the physics of electrostatic discharges which
can occur in the practice and the destruction of
electronic components which may result if such
components are touched by electrostatically charged
USERS.
ESD stands for ElectroStatic Discharge. Electrostatic
discharge must be preceded by electrostatic charging.
Static electric charges generally build up whenever two
bodies are rubbed against each other, e.g. when walking
(shoe soles against the floor) or driving a vehicle (tires
against the street pavement). The amount of charge
depends on several factors:
Thus the charge is higher in an environment with low air
humidity than in one with high air humidity; it is also
higher with synthetic materials than with natural
materials (clothing, floor coverings).
The following rule of thumb can be applied to assess the
transient voltages resulting from an electrostatic
discharge.
An electrostatic discharge is:
perceptible at 3,000 V or higher
audible at 5,000 V or higher (cracking, crackling)
visible at 10,000 V or higher (arc-over)
The transient currents resulting from these discharges
have a magnitude of 10 amperes. They are not
hazardous for humans because they last for only several
nanoseconds.
Voltage differentials exceeding 30,000 volts per
centimeter may lead to a charge transfer (electrostatic
discharge, lightning, arc-over).
Integrated circuits (logical circuits and microprocessors)
are used in order to implement a wide variety of
functions in dental/X-ray/CEREC systems. The circuits
must be miniaturized to a very high degree in order to
include as many functions as possible on these chips.
This leads to structure thicknesses as low as a few ten
thousandths of a millimeter.
1 nanosecond
= 1/1,000,000,000 second
= 1 billionth of a second
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
10 D 3554.201.01.04.02
It is obvious that integrated circuits which are connected
to plugs leading outside of the unit via cables are
sensitive to electrostatic discharge. Even voltages which
are imperceptible to the user can cause breakdown of
the structures, thus leading to a discharge current which
melts the chip in the affected areas. Damage to
individual integrated circuits may cause malfunction or
failure of the system.
To prevent this from happening, the ESD warning label
next to the plug warns of this hazard.
Safety information Be careful when opening the door of the autoclave,
especially if you are interrupting a drying cycle.
Small amounts of residual steam can escape
through the door and burn operating personnel.
After opening the autoclave door, do not touch
exposed hot metal parts. – burning hazard! Be
sure to use the tray handle when removing still-hot
trays from the autoclave. Be sure to use a suitable
hand protection (e.g. potholders or gloves) when
removing other sterilizing containers.
It is recommended to install a leak detector in the
water inlet (see installation instructions, chapter 7).
We also recommend using the
SIRODEM water treatment system from Sirona.
This unit is not suitable for the sterilization of
liquids.
According to the currently valid VDE regulations,
this device is not suitable for operation in areas
subject to explosion hazards.
Empty the water supply tank before transporting
the autoclave. If the autoclave has water inside, do
not tilt it.
This device may be repaired only by an authorized
dealer or service department. The persons
performing such repairs must use original spare
parts and observe the relevant servicing
instructions.
Disconnect the power plug before opening the
housing!
This autoclave will properly sterilize only if the
operating personnel observe these Operating
Instructions, and especially, follow the instructions
regarding program-specific loading of the autoclave.
The autoclave is designed for use outside of the
patient’s immediate vicinity (radius of 1.5 meters
around the treatment center).
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 11
2 Electromagnetic compatibility
The UNIT is intended for operation in the
electromagnetic environment specified below.
The customer or user of the UNIT should make sure that
it is used in such an environment.
Interference immunity The UNIT is intended for operation in the
electromagnetic environment specified below.
The customer or user of the UNIT should make sure that
it is used in such an environment.
Emission test Conformity Electromagnetic environment
guidelines
RF emissions according to CISPR 11 Group 1 The UNIT uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions according to CISPR 11 Class B
The UNIT is intended for use in all
facilities, including residential areas
and in any facilities connected directly
to a public power supply providing
electricity to buildings used for
residential purposes.
Harmonics
according to IEC 61000-3-2
not applicable
Power output < 50 W
Voltage fluctuations/flicker according
to IEC 61000-3-3
not applicable
no significant
flicker
Interference
immunity tests
IEC 60601-1-1-2
test level
Conformance
level Electromagnetic
environment guidelines
Electrostatic discharge
(ESD) according to
IEC - 61000-4-2
± 6kV contact discharge
± 8 kV air discharge
± 6kV contact discharge
± 8 kV air discharge
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient/
burst according to
IEC 61000-4-4
± 1kV for input and output
lines
± 2kV for power cables
± 1kV for input and
output lines
± 2kV for power cables
The quality of the line power
supply should be that of a
typical commercial or hospital
environment.
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
12 D 3554.201.01.04.02
Device views: DAC PROFESSIONAL
Surge voltages
according to
IEC 61000-4-5
± 1kV differential mode
± 2kV common mode
± 1kV differential mode
± 2kV common mode
The quality of the line power
supply should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
variations of the power
supply according to
IEC 61000-4-11
<5% UT for ½ period
(>95% dip of UT)
40% UT for 5 periods
(60% dip of UT)
70% UT for 25 periods
(30% dip of UT)
<5% UT for 5 sec.
(>95% dip of UT)
<5% UT for ½ period
(>95% dip of UT)
40% UT for 5 periods
(60% dip of UT)
70% UT for 25 periods
(30% dip of UT)
<5% UT for 5 sec.
(>95% dip of UT)
The quality of the line power
supply should be that of a
typical commercial
or hospital environment.
If the user of the UNIT
requires it to continue
functioning following
interruptions of the power
supply, it is recommended to
have the UNIT powered by an
uninterruptible power supply
or a battery.
Magnetic field of power
frequencies (50/60 Hz)
according to
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Remarks: UT is the AC supply voltage prior to application of the test level.
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 13
3 System description
3.1 Device views
2
1
3
4
5
6
79 8
Seitenansicht
10
11
12
13
14
15
16
17
18
Rückseite
lateral view
rearview
1. Control and display panel 11. Elongated hole for optional upgrading with the safety
11. combination EN1717
2. Hinged door (opens to the left) 12. Sterile filter
3. Sliding closing handle 13. Spring-operated safety valve
4. Power switch 14. Without function on DAC PROFESSIONAL (sealed)
14. Option: one-way outlet
5. Adjustable front feet of unit 15. Emergency overflow (OUT)
6. Serial data and printer port (RS 232) 16. Cooler
7. Device fuses - 2 x 16 A / FF 17. Feed water connection (IN) (quick connector f.
17. hose 8x1)
8. Drain hose for water tank, feed water side (OUT) 18. Power cable
9. Drain hose for water tank, drain side (OUT)
10. Tank cover
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
14 D 3554.201.01.04.02
4 Technical data
4.1 Primary technical data
4.2 Capacity / weights
4.3 Operating media
4.4 Operating parameters
Programs / operating times
Model DAC PROFESSIONAL
Sterilization chamber (dia. x depth) 25 cm x 36 cm
Power ratings 2100 W / 230 V AC / 9.2 A / 50....60 Hz
Sterilization pressure / temperature 2.1 bar/134°C; 1.1 bar/121°C
Maximum load: 5 kg of instruments or 1.8 kg of textiles
Outside dimensions (HxWxD) 48.5 cm x 42.5 cm x 59.5 cm
Weight (without load) 43 kg
Volume of chamber 17.6 liters
Maximum loading 5 kg of instruments or 1.8 kg of textiles
Water tank (feed water) Fresh-water side: 5 liters (approx.7 cycles); drain side:3 liters
Power ratings
Power supply 230 V AC / 9.2 A / 50...60 Hz
Power ratings 2100 W; fuse 16 A, RCCB (resid. current circuit breaker); 30 mA
Feed water Demineralized / steam distilled water in compliance with EN 13060
Program Operating time
(not incl. drying period) Drying period Max. loading
“Universal program“ (134°C, 2.1 bar) 30 min 20 min 5 kg
“Quick program S“ (134°C, 2.1 bar) 15 min 5 min 5 kg
“Quick program B“ (134°C, 2.1 bar) 28 min 10 min 5 kg (unwrapped)
1.5 kg (wrapped)
“Gentle program“ (121°C, 1.1 bar) 45 min 20 min 1.8 textiles
“Prion program“ (134°C, 2.1 bar) 45 min 20 min 5 kg
"Bowie&Dick" (134°C, 2.2 bar) 33 min 40 min 5 min
“Vacuum test” -- 18 min (empty) --
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 15
Consumption of energy / water *)
*) These values apply to a constant 230V power supply
4.5 Installation example
2 x 20-character alphanumeric LED display for indication
of program status and parameters and for user guidance
of basic and special functions
Power consumption
"Preheating"
One-time heating up to preheating
temperature (134°C)
approx. 0.12 kWh
"Standby" operation per hour approx. 0.2 kWh
"Program cycle" (incl. drying) 0.3 kWh for "Quick program S", warm start, low load until
1.1 kWh for "Gentle program", cold start, full load
2.1 kWh "Quick program B"
Feed water consumption 300 ml for "Quick program S"
500 ml for "Universal program" and "Prion program"
600 ml for "Gentle program"
DAC PROFESSIONAL rear view
Standard version = version as supplied
(with internal water supply)
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
16 D 3554.201.01.04.02
5 Control panel
A) Function keys
for selecting, setting and displaying special
functions: printing, date/time, preheating, sum batch
numbers, conductivity, error report
acknowledgement
B) 2 x 20-character alphanumeric LED display
for indication of program status and parameters and
for user guidance of basic and special functions
C) Program selection key
for selection of the sterilization and test programs
and for selecting/setting options (submenus) for the
special functions
D) Start - Stop key
for starting and stopping programs, for drying, and
for control of special functions
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 17
6 Performance features of unit
6.1 Fractioned pre-vacuum procedure
Fractioned pre-vacuum repeatedly evacuates the air
from the autoclave and alternately introduces steam
between evacuations, thus providing the required
penetration of the sterilized objects with saturated
steam.
This ensures fast and effective execution of difficult
sterilization tasks: e.g. instruments with narrow bores
and large amounts of textiles.
Steam penetration can be confirmed via the Bowie &
Dick test (a recognized procedure used in large-scale
sterilization). A special test program has been
implemented for this purpose.
6.2 Sterilization categories
DAC PROFESSIONAL features 5 sterilization programs
for temperatures of 134°C: the "Universal program" for
wrapped objects, the "Fast program B", the "Prion
program" (a special Universal program) and the "Quick
program S" for unwrapped items. The “Gentle program”
is a program for sterilizing textiles and rubber articles at
121°C. The user can at any time perform additional
functional checks of the autoclave by running the Bowie
& Dick test for steam penetration and the Vacuum test
for leak testing.
6.3 Separate steam generation
The powerful steam generator outside the actual
sterilization chamber makes it possible to sterilize large
loads of instruments or textiles in a short time. No waiting
periods are necessary between sterilization cycles, and
excess temperatures cannot occur within the
sterilization chamber.
6.4 One-way system / conductivity measurement / water supply
DAC PROFESSIONAL operates with the tried and
proven one-way system: this means that all evaporated
water and all impurities contained in this water are
removed from the autoclave upon pressure release.
Clean feed water is therefore used for the following
program cycle.
An integrated conductivity measurement system
monitors the quality of the feed water used in steam
generation. We recommend using a water supply with
feed water.
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
18 D 3554.201.01.04.02
If the instruments are carefully prepared, this
arrangement prevents spots on the items sterilized and
avoids fouling of the autoclave.
6.5 Electronic parameter control
The microprocessor installed in the
DAC PROFESSIONAL implements an electronic
parameter control system which continuously monitors
pressure, temperature and time in the programs. The
total operating times can thus be optimized according to
the load and the temperature of the autoclave.
The process evaluation and monitoring system
contained in the program control system compares
momentary process parameters with standard
processing data and monitors the process with respect
to limit temperatures, limit times and limit pressures.
This function detects malfunctions in the program cycle
and guarantees the security of the sterilization result.
6.6 Vacuum drying
The vacuum drying feature ensures good drying results,
even for wrapped items.
6.7 Preheating
Activating the preheating function warms up the cold
autoclave chamber before sterilization or maintains the
temperature between sterilization cycles. This shortens
the duration of cycles and considerably reduces the
formation of condensation, thus optimizing drying
results.
6.8 Documentation of batch results
Batch results can be documented via the following
output media:
• NITRAprint
• NITRAflash
Computer with NITRAwin
• Modem
The memory of the electronic control system
permanently stores records of the last 40 programs.
For effective hard-copy documentation of batches and
for checking previously executed programs, a
NITRAprint printer can be connected to print out a record
immediately after completion of a program, or to print out
records from the memory. If a PC is connected, the
NITRAwin program can also transfer the records to the
PC for permanent archiving of all results and print out
any results as required.
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
D 3554.201.01.04.02 19
7 Installation
For the steps required to prepare for and carry out the
setup and installation of the autoclave, see the separate
instructions contained in "Setting up the
DAC PROFESSIONAL".
7.1 Installation requirements
The autoclave should be set up in a dry place which is
protected against dust.
The base should be stable and flat and capable of
supporting the weight of the unit
(unloaded weight = 43 kg).
The space required by the autoclave can be seen from
the external dimensions (see Section 4.1). A minimum
additional space of 10 cm should be allowed on both
sides and above the autoclave in order to ensure that
heat can escape.
The electrical power supply should be a separate
230V AC circuit with a 16A fuse.
This autoclave operates with a blower for the cooling
system, on the rear side of the unit. If there is a
restriction of the escape of heat above the unit, this can
impair the functioning and shorten the service life of the
autoclave. We therefore urgently advise our users not to
install the autoclave in an area where there is not
sufficient circulation of air.
If the user decides to install an automatic one-way
system, it is necessary to have a connection to the waste
water plumbing near the autoclave, preferably a wall
outlet (NW 40) or a sink drain (standard length of drain
hose = 2m, with an outer hose diameter = 15 mm).
A 2 meter long drain hose (Art. No. 61 26 556) is
supplied with the unit. The required connectors must be
ordered according to the given practice situation. The
work surface on which the autoclave stands must be
higher than the outlet so that the drain hose can be run
to the drain continuously downward and without bends
and twists which could prevent the water flowing out
freely. At the same time, the work surface must provide
convenient access to the autoclave and the display must
be clearly visible.
The autoclave must be supplied with feed water from the
integrated dual-chamber water tank, with fresh water
and waste water chambers. However, please note that
this will require additional space for the water treatment
system.
Sirona Dental Systems GmbH
Operating Instructions DAC PROFESSIONAL
61 33 826 D 3554
20 D 3554.201.01.04.02
7.2 Transport straps
Lift the autoclave out of the packaging using the
transport straps. The straps themselves are each
removed by unscrewing the two retaining screws, which
must then be screwed firmly back in place without
washers.
7.3 Alignment
In order to operate properly, and to allow unimpeded
return flow of condensate into the steam generator, the
autoclave must be installed so that it definitely slopes
toward the rear. First install the autoclave in a horizontal
(level) position and check that this position is in fact
horizontal with a spirit level at the chamber flange. Then
extend the front feet to lift up the front side of the
autoclave by screwing them out by at least three turns.
7.4 Power supply connection
The unit power cord must be connected to a
230 V, 50 Hz electric outlet. The power rating of the
autoclave is 2100 W. In order to avoid overloading the
building electricity supply, we recommend using a
separate electrical circuit fitted with a 16 A fuse and
optionally protected with a type RCCB (residual current
circuit breaker) with a tripping current of 30 mA.
7.5 Outlet connection for one-way water
DAC PROFESSIONAL operates only in the one-way
water system, i.e. used feed water (hereinafter referred
to as "waste water") is not recycled. Waste water is
collected in the drain tank. As an option, the autoclave
can be connected directly to the waste water plumbing
via the one-way outlet. A retrofit kit is connected to the
rear wall of the autoclave for this purpose. Its drain hose
must be run to the drain continuously downward and
without bends or twists and must not exceed a max.
length of 2 meters. The outlet must be connected to an
existing sink drain, or to a separate NW40 drain pipe with
use of the wall-mounted drain.
7.6 Internal water supply with feed water
If the user decides for internal supply of the autoclave
with feed water, this water is taken from the right
chamber of the built-in water storage tank. To install the
internal feed water supply system, connect the hose to
the feed water connection of the water tank on the rear
side of the autoclave. For this variation of water supply,
the autoclave is delivered ready for installation. To fill the
feed water tank, remove the cover of the tank and fill the
right chamber with feed water up to the maximum mark.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74

Sirona DAC PROFESSIONAL Operating Instructions Manual

Type
Operating Instructions Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI