Acculab 98648-013-61 Operating instructions

Category
Kitchen scales
Type
Operating instructions
English page 2
In cases involving questions of
interpretation, the German-language
version shall prevail.
Deutsch Seite 17
Im Auslegungsfall ist die
deutsche Sprache maßgeblich.
Français page 32
En cas de questions concernant
l’interprétation, la version
en langue allemande fera autorité.
Italiano pagina 47
In caso di interpretazione,
fa testo la versione in lingua tedesca.
Español página 62
En caso de interpretación,
la versión en lengua alemana
será determinante.
2 Warnings and Safety
Precautions
3 Getting Started
5 Operation
5 Basic Weighing Function
5 Description of the Keys
Application Programs
6 Toggling between Weight Units
7 Counting
8 Weighing in Percent
9 “Hold” Display
10 Totalizing
11 Specific Gravity
12 Calibration/Span Adjustment
13 Configuration (Setup Menu)
14 Error Codes
Overview
15 Specifications
16 Accessories (Options)
16 C Marking
2
Contents Warnings and
Safety Precautions
Safety Information
§ To prevent damage to the equip-
ment, please read these operating
instructions carefully before using
your balance/scale.
! Do not use this equipment in
hazardous areas.
! Make sure the voltage rating printed
on the power supply is identical to
your local line voltage.
! Use only commercially available 9V
batteries. If desired, you can use a
rechargeable battery (not included).
The balance is energized at all
times unless you disconnect the AC
adapter and, if connected, the
battery.
Protect the AC adapter from contact
with liquid
! Exposure to excessive electromag-
netic interference can cause the
readout value to change. Once the
disturbance has ceased, the instru-
ment can be used again in accor-
dance with its intended purpose.
Installation
It is recommended to connect
Acculab accessories and options,
as these are optimally designed for
use with your balance/scale.
Do not open the balance/scale
housing as this may void the
manufacturer’s warranty.
3
Equipment Supplied
Balance/scale with in-use cover
Weighing pan
Plug-in AC adapter
Additional equipment with models VIC-612, VIC-412,
VIC-212, VIC-711, VIC-511:
Calibration weight
Additional equipment with models VIC-303, VIC-123,
VIC-4MG, VIC-2MG:
Calibration weight
Round glass draft shield
Level indicator and adjustable feet
Storage
$ Do not stack more than 3 balances on top of one another
at a time.
Installation
Choose a location that is not subject to the following nega-
tive influences:
Heat (heater or direct sunlight)
Drafts from open windows and doors
Extreme vibrations during weighing
Excessive moisture
Setting Up the Balance/Scale
§ Place the components on the balance/scale in the
following order:
Reversible round weighing pan
Round glass draft shield on models VIC-303, VIC-123,
VIC-4MG, VIC-2MG
Getting Started
4
Connecting the
Balance/Scale to AC Power
§ It is recommended to use only the
included AC adapter for optimal
performance and safety.
§ Insert plug into the jack
(located on back of scale)
§ Plug the AC adapter into an electrical
outlet
! LISTED power supply 11 V21 V
compliant with NEC Class 2 output.
Below-Balance/Under-Scale
Weighing
A port for a below-balance weighing
hanger is located on the bottom of
the balance/scale.
§ Open the cover plate on the bottom
of the balance/scale.
§ Attach the sample (e.g., using a sus-
pension wire) to the hanger.
$ Install a shield for protection against
drafts if necessary.
Installing the Battery
(not for models VIC-303, VIC-123,
VIC-4MG, VIC-2MG)
$ Batteries are not included with the
equipment supplied
! Use only commercially available 9V
batteries.
! If you use a rechargeable battery,
recharge it with an external battery
charger.
§ Lay the balance/scale on its side
§ Open the battery compartment:
remove the compartment cover
§ Install the battery in the compart-
ment
$ Make sure the polarity is correct.
! Close the battery compartment: slide
the cover into position until it snaps
into place
! Do not throw away used batteries
with normal household waste.
Rechargeable batteries contain toxic
materials and must be disposed of in
accordance with local waste disposal
regulations.
Leveling the Balance/Scale
(only for models VIC-303,
VIC-123, VIC-4MG, VIC-2MG)
§ Always level the balance/scale
again any time after it has been
moved to a different location.
Example: moving bubble from R to L
§ Turn the feet as shown in the dia-
gram until the air bubble is centered
within the circle of the level indicator.
> In most cases this will require several
adjustment steps.
Removing Weights for
Calibration/Span Adjustment
(only for models VIC-612, VIC-212,
VIC-711, VIC-511, VIC-303,
VIC-123, VIC-4MG, VIC-2MG)
§ Grasp the tab to turn and remove
the weight compartment.
$ Follow instructions on page 12 for
calibration/span adjustment.
5
Basic Weighing Function
Features
Zeroing the balance
You can zero the balance/scale within the entire weighing
range, up to the maximum capacity.
Preparation
§ Switch on the balance/scale: press the (ON/OFF) key
§ If necessary, zero the balance/scale: press the (ZERO) key
$ If necessary, change the configuration settings:
see the chapter entitled “Configuration”
$ If desired, load the factory settings:
see the chapter entitled “Configuration,” menu item 9.– 1
Additional functions:
$ Switching off the balance/scale: press the (ON/OFF) key
$ Battery operation: automatic shut-off after 2,5 or 10 min-
utes; see chapter on configuration. Example: 2 minutes.
If the weight readout remains unchanged and no keys are
pressed for at least two minutes, the battery symbol
“ starts flashing. After another 2 seconds, the
balance shuts off automatically, unless a key is pressed.
Description of the Keys
(ON/OFF) On/off key: switches the balance/scale on and
off or switches it to the standby mode.
Battery operation: on; turns backlight on; off
(ZERO) Zeros the balance; press and hold 2 seconds:
opens the application menu
(CAL) Starts calibration/adjustment
(F) Starts an application program;
Scrolling in application menu, configration
menu and calibration menu
(ENTER) Confirms the selected setting;
Exits application, configuration & calibration
menu if key is pressed and held for more than
2 seconds.
(PRINT) Generates a printout or data output
Operation
Example: Determine weight of sample
Step Key (or instruction) Display
1. Switch on the balance/scale (ON/OFF)
Self-test runs
Display: Software version
r 31.01
2. Open the in-use cover and leave open
while weighing
3. Place container on the balance/scale 52.0 g
(in this example, 52 g)
4
. Zero the balance/scale (ZERO)
0.0 g
5. Place sample in container on balance/scale 150.2 g
(in this example, 150,2 g).
6
Toggling between Weight Units
With this application program you can toggle the display of a weight value back and forth between two weight units (see table below).
Example: Toggle weight unit from pounds [lb] (application) to grams [g] (basic unit)
Step Key (or instruction) Display
1. Select application program (ZERO) > 2 sec
2. Select Toggling between Weight Units (F)
3. Confirm unit (ENTER)
4. Select weight unit; in this example: “5. Pound” (F) repeatedly
(see table below)
5. Confirm weight unit (pounds) (ENTER)
6. Place sample on balance/scale
7. Toggle weight unit (F)
Menu code Unit Conversion factor Display
1. user* Grams 1.00000000000 o
2.grams (factory setting) Grams 1.00000000000 g
4.carat Carats 5.00000000000 o
5.Pound Pounds 0.00220462260 lb
6.ounce Ounces 0.03527396200 oz
7.tryo Troy ounces 0.03215074700 ozt
8.tl.hon Hong Kong taels 0.02671725000 tlk
9.tl.sin Singapore taels 0.02645544638 tl
10.tl.ta Taiwanese taels 0.02666666000 tl
11.grai Grains 15.4323583500 GN
12.peny Pennyweights 0.64301493100 dwt
15.tl.ch Chinese taels 0.02645547175 tl
22.pdoz lb/oz 0.03527396200 lb:oz
23.newt Newtons 0.00980665000 N
* User-defined conversion is customer selectable with RS-232 or USB program option.
Application Programs
7
Purpose
With the Counting program you can determine the number of parts or items.
Example: Determine the number of uncounted parts; weigh in the selectable reference sample quantity (in this example: 20)
Step Key (or instruction) Display
1. Select application program (ZERO) > 2 sec
2. Select Counting (F) repeatedly
3. Confirm setting (ENTER)
Symbol “ “ on the display: application is active
4. Place empty container on the balance/scale
5. Zero the balance/scale (ZERO)
6. Place reference sample quantity (20)
on the balance/scale
7. Select reference sample quantity:
in increments of 1 (1, 2, 3, ..., 99) or (F) repeatedly (briefly) or
in increments of 10 (10, 20, 30, ...,100) (F) > 2 sec.
8. Confirm selected reference sample quantity (ENTER)
9. Place uncounted parts on balance/scale
10. Toggle display between mean piece weight,
total weight, and quantity (F) repeatedly
11. Unload the balance/scale
12. Counting application: clear the reference value (ENTER) > 2 sec
13. Reactivate Counting (if no other (F)
application program has been selected)
14. Repeat procedure starting from Step 5.
Counting
8
Purpose
This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight.
Example: Determine an unknown percentage; store the weight on the balance/scale as the reference percentage (100%)
Step Key (or instruction) Display
1. Select application program (ZERO) > 2 sec
2. Select Weighing in Percent (F) repeatedly
3. Confirm setting (ENTER)
Symbol “ “ on the display: application is active
4. Place empty container on the balance/scale
5. Tare the balance (ZERO)
6. Place the reference weight for 100% on the
balance/scale
7. If desired, change the number of decimal places
displayed:100.0%,100.00% or 100% (factory setting) (F) repeatedly
8. Confirm selected no. of decimal places (ENTER)
9. Place unknown weight on the balance/scale
10. Toggle display between weight and percentage (F) repeatedly
11. Unload the balance/scale
12. Weighing in Percent application:
clear the reference percentage (ENTER) > 2 sec
13. Reactivate Weighing in Percent (if no other (F)
application program has been selected)
14. Repeat procedure starting from Step 5.
Weighing in Percent
9
Purpose
“Holds” the displayed value; also, the display will be locked for 5 seconds after removing the sample from the pan.
Example: Determine weight of oversized sample
Step Key (or instruction) Display
1. Select application program (ZERO) > 2 sec
2. Select Hold Display (F) repeatedly
3. Confirm setting (ENTER)
Symbol “ “ on the display: application is active
4. If necessary: zero the balance/scale (ZERO)
5. Place oversized sample on balance/scale
6. Start application program (F)
Symbol “ “ flashes on the display:
the weight value is locked
7. Unload the balance/scale: the weight value
remains displayed for a further 5 seconds; or
8. Zero the balance/scale (ZERO)
9. End the Display Hold application (ENTER) > 2 sec
10. Reactivate Display Hold (if no other (F)
application program has been selected)
11. Repeat procedure starting from Step 5.
“Hold” Display
10
Purpose
With this application program you can add up successive weight values exerding capacity of balance/scale.
Example: Totalizing weight values
Step Key (or instruction) Display
1. Select application program (ZERO) > 2 sec
2. Select Totalizing (F) repeatedly
3. Confirm setting (ENTER)
Symbol “ “ on the display: application is active
4. If necessary: zero the balance/scale (ZERO)
5. Place sample on balance/scale (in this example, 380 g)
6. Store value in memory. Total weight is displayed (ENTER)
steadily; symbol flashes.
7. Remove sample from balance/scale
8. Place the next sample on the balance/scale
(in this example, 575 g)
9. Store value in memory. Totalized stored weight (ENTER)
is displayed; symbol flashes.
Note: symbol remains on indicating stored
value in memory until cleared
10. To view the current weight for 2 seconds (F)
(if a printer is connected, a printout is generated)
11. Clear totalizing memory (F) > 2 sec
(if a printer is connected, total is printed)
12. End Totalizing (ENTER) > 2 sec
13. Reactivate Totalizing (if no other (F)
application program has been selected)
(if a printer is connected, total is printed).
14. Repeat procedure starting from Step 6.
Totalizing
11
Purpose
Use this application program to determine the specific gravity of a sample. The result is displayed with one decimal place.
Beaker and wire not included with balance/scale.
Example: Determine the specific gravity of a solid.
Step Key (or instruction) Display
1. Select application program (ZERO) > 2 sec
2. Select Specific Gravity (F) repeatedly
3. Confirm setting (ENTER)
Symbol “ “ on the display: application is active
4. If necessary, zero the balance/scale (ZERO)
5. Start application program (F)
6. Confirm the display, “airval (ENTER)
7. Determine weight of the sample in air:
place sample on the balance/scale
8. Store weight-in-air value (ENTER)
9. Remove sample from balance/scale
10. Determine weight of sample in liquid:
connect wire and set up beaker
11. Confirm the display “Water (ENTER)
12. Place sample in liquid
13. Store the weight-in-liquid value and view the result (ENTER)
14. Clear the display (ZERO)
15. Exit the Specific Gravity application (ENTER) > 2 sec
16. Reactivate Specific Gravity (if no other (F)
application program has been selected)
17. Repeat procedure starting from Step 5.
Specific Gravity
12
Calibration is recommended after initial installation and each time the balance/scale is moved.
Features
Calibration/adjustment can be performed only when:
there is no load on the balance/scale,
the balance/scale is zeroed, and
the internal signal is stable.
If these conditions are not met, an error message
is displayed.
Example: Calibrate/adjust span of the balance/scale (here: model VIC-5101)
Step Key (or instruction) Display
1. Switch on the balance/scale (ON/OFF)
2. Zero the balance/scale (ZERO)
3. Start calibration (CAL)
The preset calibration weight is displayed without
the weight unit (in this example, 5000 g)
4. To select a different calibration weight value (F) repeatedly
5. Confirm calibration weight value and (ENTER)
start calibration/span adjustment
After the zero point is stored, the required
calibration weight flashes on the display.
6. Place the required calibration
weight on the balance/scale
The readout stops flashing if the weight is
applied within the defined time limit and tolerance.
If the weight value is accepted, the display stops
flashing and the stability symbol appears
on the display.
7. Remove the calibration weight
8. Calibration has been completed
Calibration/Span Adjustment
The weight required for calibration/adjustment is displayed.
Standard calibration weights for selected models: to remove,
see instructions on page 4.
Press (F) to select a different weight value.
To cancel the procedure: press and hold the (ENTER) key
(> 2 sec.).
13
To configure the balance/scale; i.e., adapt the balance/scale to individual requirements.
Step Key (or instruction) Display
1. Switch off the balance/scale (ON/OFF)
2. Switch on the balance/scale and (ON/OFF)
while all segments are displayed (ZERO) briefly
Navigation in the Setup Menu
Key Press briefly Press and hold
(ENTER) Menu level: move to the right (cyclical) Confirm setting
(F) Menu item: Scrolling
(ZERO) Menu level: Move to the left Save settings and exit Setup
Parameter Settings (Overview) ο Factory setting User-defined setting
Setup 1 Weighing 1.1. Adapt filter 1.1.1 Very stable conditions
menu 1.1.2 ο Stable conditions
1.1.3 Unstable conditions
1.1.4 Very unstable conditions
1.2. Application filter 1.2.1 ο Final readout
1.2.2 Filling
1.3. Stability range 1.3.1 1/4 digit
1.3.2 1/2 digit
1.3.3 1 digit
1.3.4 ο 2 digits
1.3.5 4 digits
1.5. Calib./adjust./linearization: 1.5.1 ο Calibration/adjustment
Function of the (CAL) key 1.5.2 Linearization: for service personnel only
1.5.3 Key blocked
1.6. Auto zero 1.6.1 ο On
1.6.2 Off
1.7. 1st weight unit, or 2nd unit in 1.7.1 User-defined unit; see
Toggle Weight Units app. to 1.7.23 “Toggling between Weight Units”
5. and 6. Only relevant with built-in data interface: see corresponding interface description
8. Additional 8.1. Block key functions 8.1.1 All keys blocked except for (ON/OFF)
functions and (ZERO)
8.1.2 ο All keys unblocked
8.2. Automatic shut-off 8.2.1 ο After 2 minutes
8.2.2 After 5 minutes
8.2.3 After 10 minutes
9. Reset menu 9.1. Factory settings 9.1.1 Restore
9.1.2 ο Do not restore
Configuration (Setup Menu)
14
Error codes are shown on the main display for approx. 2 seconds. The program then returns automatically to the previous mode.
Display/Problem Cause Solution
No segments appear No power available Check the power supply
on the display
The AC adapter is not plugged in Plug in the AC adapter
Battery is drained Replace battery; recharge battery using
external charger
oL The load exceeds the balance/scale Unload the balance/scale
capacity
UL Weighing pan not in place Place the weighing pan
on the balance/scale
Something is touching Move the object that is
the weighing pan touching the weighing pan
disErr Display overflow: Reduce load on the balance/scale
Value cannot be shown
on the display
calErr Calibration parameter not met; e.g.: Calibrate only when zero is displayed
– Balance/scale not zeroed Press (ZERO) to tare the balance/scale
– Balance/scale is loaded Unload the balance/scale
appErr Weight is too light or Increase the weight on the balance/scale
there is no sample on the balance/scale
with application in use
PrtErr Data interface for Contact the Acculab
printing is blocked customer service center
bal.Err Balance/scale loaded or defecteive Unload balance/scale before switching on
when power was turned on or contact Acculab custumer service
sys.Err Balance/scale defective Contact Acculab custumer service
Max. weighing capacity is less The balance/scale was switched on Place the weighing pan on the
than indicated under “Specifications” without the weighing pan in place balance/scale and press (ON/OFF)
The weight readout is obviously wrong The balance/scale was not calibrated/ Calibrate/adjust the balance/scale
adjusted before weighing
Balance/scale not zeroed Zero the balance/scale
If any other errors occur, contact your local Acculab customer service center.
Error Codes
15
Specifications
Model VIC-303 VIC-123 VIC-4MG VIC-2MG VIC-612 VIC-412 VIC-212
Weighing capacity g 300 120 410 210 610 410 210
Readability g 0.001 0.001 0.005 0.005 0.01 0.01 0.01
Tare range (subtractive) g 300 120 410 210 610 410 210
Linearity ±g 0.004 0.003 0.01 0.01 0.03 0.03 0.035
Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Stabilization time (average) s 2.5 2.5 2.5 2.5 2 2 2
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)
Calibration weight g 200 (F1) 100 (F1) 200 (F1) 200 (F2) 200 (F2) 200 (F2) 200 (M1)
Net weight, approx. kg 1.3 1.2 1.3 1.3 1.35 1.35 1.2
Pan size mm 97 d 97 d 97 d 97 d 142x130 142x130 97 d
Power source/voltage/frequency AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average) W 1 1 1 1 0.75 0.75 0.75
Hours of operation w/ 9V battery:
Alkaline (approx.) h 11 11 14
Rechargeable, fully ch.,
(NiMH), avg. h 2.5 2.5 4
Model
VIC-5101 VIC-3101 VIC-1501 VIC-711 VIC-511 VIC-10KG VIC-6KG VIC-4KG
Weighing capacity g 5100 3100 1500 710 510 10100 6100 4100
Readability g 0.1 0.1 0.1 0.1 0.1 1 1 1
Tare range (subtractive) g 5100 3100 1500 710 510 10100 6100 4100
Linearity ±g 0.2 0.2 0.2 0.2 0.2 2 2 1
Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Stabilization time (average) s 2 2 1.5 1.5 1.5 1.5 1.5 1.5
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)
Calibration weight kg 5 (F2) 2 (F2) 1 (M1) 0,2 (M2) 0,2 (M2) 5 (M1) 5 (M2) 2 (M2)
Net weight, approx. kg 1.1 1.1 1.1 1.25 1.25 1.1 1.1 1.1
Pan size mm 142x130
Power source/voltage/frequency AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average) W 1 1 0.75 0.75 0.75 0.75 0.75 0.75
Hours of operation w/ 9V battery:
Alkaline (approx.) h 11 11 14 14 14 14 14 14
Rechargeable, fully ch.,
(NiMH), avg. h 2.5 2.5 4444 44
Overview
16
Product Order No.
Data interface, mounting kit
RS-232 interface with cable YADAP-RS
USB interface with cable YADAP-USB
Data printer YDP03-0CE
Lock-down capability LC-1
(for anti-theft locking device)
Calibration weights
for VIC-5101 (5 kg; F2) YCW6548-00
for VIC-3101 (2 kg; F2) YCW6248-00
for VIC-1501 (1 kg; M1) YCW615-00
for VIC-10KG (5 kg; M1) YCW655-00
for VIC-6KG (5 kg; M2) YCW656-00
for VIC-4KG (2 kg; M2) YCW626-00
for weight compartment,
(right side), (100 g; F1) 69V00006
Product Order No.
In-use cover:
for models without glass draft shield 69V00001
for models with glass draft shield 69V00002
Round glas draft shield 69V00003
(25 mm high)
Weighing pan: Round 69V00004
Rectangular 69V00005
Leveling feet (set of one
adjustable foot and one fixed foot) 69V00007
Covers: (set of small parts) 69V00008
Battery compartment
Interface port
Weight compartment
Accessories (Options)
C Marking
The balance/scale complies with the following EC Directives
and European Standards:
Council Directive 89/336/EEC “Electromagnetic
compatibility (EMC)”
Applicable European Standards:
Limitation of emissions:
In accordance with product standard EN 61326-1 Class B
(residential area)
Defined immunity to interference:
in accordance with product standard EN 61326-1
(minimum test requirements, non-continuous operation)
Important Note:
The operator shall be responsible for any modifications to
Acculab equipment and must check and, if necessary, correct
these modifications.
On request, Acculab will provide information on the minimum
operating specifications (in accordance with the Standards
listed above for defined immunity to interference).
73/22/EEC “Electrical equipment designed for
use within certain voltage limits”
Applicable European Standards:
EN 60950
Safety of information technology equipment including electrical
business equipment
EN 61010
Safety requirements for electrical equipment for measurement,
control and laboratory use
Part 1: General requirements
If you use electrical equipment in installations
and under ambient conditions requiring higher safety standards,
you must comply with the provisions as specified in the
applicable regulations for installation in your country.
Programas de aplicación
Conmutación de unidades
Con este programa de aplicación puede indicarse el valor de peso en una unidad seleccionada (ver tabla) y en gramos.
Ejemplo: conmutar unidad: de libras [lb] (aplicación) a gramos [g] (unidad básica)
Paso Pulsar tecla Lectura
1. conmutar programa aplicación (ZERO) pulsión larga
2. seleccionar unidad a conmutar (F)
3. confirmar ajuste (ENTER)
4. seleccionar unidad peso; aquí “5. libras” repetidam (F)
(ver tabla, abajo)
5. confirmar unidad libras (ENTER)
6. poner muestra
7. conmutar unidad peso (F)
Item de menú Unidad Conversión Lectura
1. user* Gramos 1,00000000000 o
2.grams (ajuste fábrica) Gramos 1,00000000000 g
4.carat Quilates 5,00000000000 o
5.Pound Libras 0,00220462260 lb
6.ounce Onzas 0,03527396200 oz
7.tryo Onzas Troy 0,03215074700 ozt
8.tl.hon Taeles Hongkong 0,02671725000 tl
9.tl.sin Taeles Singapur 0,02645544638 tl
10.tl.ta Taeles Taiwan 0,02666666000 tl
11.grai Granos 15,4323583500 GN
12.peny Pennyweights 0,64301493100 dwt
15.tl.ch Taeles China 0,02645547175 tl
22.pdoz lb/oz 0,03527396200 lb:oz
23.newt Newton 0,00980665000 N
* = Unidad definida por el usuario, cargable con progr. PC en la balanza mediante interfaz opcional RS-232, o bien USB.
66
Copyright by Acculab.
All rights reserved. No part of this publication
may be reprinted or translated in any form
or by any means without the prior written
permission of Acculab.
The status of the information, specifications
and illustrations in this manual is indicated
by the date given below.
Acculab reserves the right to make changes
to the technology, features, specifications
and design of the equipment without notice.
Status:
July 2005, Acculab
W1A000 · KT
Publication No.: WVI6006-p05071
ACCULAB North America
131 Heartland Blvd.
Edgewood, NY 11717
USA
800-656-4400
International
+1.631.254.4299
ACCULAB Europe
Weender Landstrasse 94-108
37075 Goettingen, Germany
+49.551.308.1667
ACCULAB UK c/o Sartorius Ltd
Longmead Business Centre
Blenheim Road, Epsom
Surrey, KT19 9QQ
United Kingdom
+44.1372.737100
ACCULAB China
No.8 Pingeui West Road, Wang Jing
Industrial Zone
Chao Yang District
100102 Beijing
+86.10.6439.2552
ACCULAB ASIA-PACIFIC
Unit 1110-12, LU Plaza
2 Wing Yip Street, Kwon Tong
Kowloon, Hong Kong
www.acculab.com
www.acclulab-europe.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78

Acculab 98648-013-61 Operating instructions

Category
Kitchen scales
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI