Michell Instruments TDL6 User manual

Category
Measuring, testing & control
Type
User manual
OptiPEAK TDL600
Process Moisture Analyzer
User Manual
97319 Issue 5
August 2023
Please ll out the form(s) below for each instrument that has been purchased.
Use this information when contacting Michell Instruments for service purposes.
Product Name
Order Code
Serial Number
Invoice Date
Installation Location
Tag Number
Product Name
Order Code
Serial Number
Invoice Date
Installation Location
Tag Number
Product Name
Order Code
Serial Number
Invoice Date
Installation Location
Tag Number
© 2023 Michell Instruments
This document is the property of Michell Instruments Ltd and may not be copied or
otherwise reproduced, communicated in any way to third parties, nor stored in any Data
Processing System without the express written authorization of Michell Instruments Ltd.
OptiPEAK TDL600
For Michell Instruments' contact information please go to
www.ProcessSensing.com
OptiPEAK TDL600 User Manual
iv 97319 Issue 5, August 2023
1 INTRODUCTION ................................................................................................1
1.1 Application ......................................................................................................... 1
1.2 Features ............................................................................................................ 1
1.3 Theory of Operation ........................................................................................... 2
1.3.1 Measurement Using a Laser ........................................................................... 4
2 INSTALLATION ..................................................................................................6
2.1 Unpacking the Instrument ................................................................................... 6
2.2 Lifting and Handling ........................................................................................... 7
2.3 Laser Safety ....................................................................................................... 7
2.4 Hazardous Area Safety ........................................................................................ 8
2.5 Electrical Safety .................................................................................................. 9
2.5.1 Equipment Ratings and Installation Details ...................................................... 9
2.6 Pressure Safety ................................................................................................ 11
2.7 Basic Installation Guidelines .............................................................................. 11
2.8 Electrical Connections ....................................................................................... 15
2.8.1 Power Connection ....................................................................................... 15
2.8.2 Analog Outputs ........................................................................................... 16
2.8.3 Analog Inputs ............................................................................................. 16
2.8.4 Alarm Relays............................................................................................... 16
2.8.5 Modbus RTU / RS485 Connection ................................................................. 17
2.9 Environmental Requirements ............................................................................. 17
2.10 Sample Conditioning Requirements .................................................................... 18
2.10.1 Gas Connections ......................................................................................... 18
2.10.2 Sample Flow Gas Handling Components ....................................................... 18
2.11 Options ............................................................................................................ 21
2.11.1 Enclosure Heater Temperature Control (Outdoor systems ONLY) ..................... 21
2.11.2 Vortex Cooling (Outdoor systems ONLY) ....................................................... 21
2.11.3 Trace Heated Sample Line ........................................................................... 21
3 OPERATION ....................................................................................................22
3.1 Start-Up Procedure ........................................................................................... 22
3.2 Shut Down Procedure ....................................................................................... 23
3.3 User Interface .................................................................................................. 24
3.3.1 Interface Controls ....................................................................................... 24
3.3.2 ‘Up/Down Arrow’ Keys ................................................................................. 24
3.3.3 ‘ENTER’ Key ................................................................................................ 25
3.3.4 ‘ESC’ Key .................................................................................................... 25
3.4 Description of Measured Parameters .................................................................. 25
3.5 On Power Up .................................................................................................... 26
3.5.1 Advanced Menu default settings ................................................................... 26
3.6 Menu Structure ................................................................................................ 27
3.7 Main Menu Screen ............................................................................................ 28
3.7.1 Parameters Screen ...................................................................................... 29
3.7.2 Display Screen ............................................................................................ 30
3.7.3 Log Menu Screen ........................................................................................ 31
3.7.4 About Screen .............................................................................................. 32
3.7.5 Graph Screen .............................................................................................. 32
Contents
Safety ...............................................................................................................................vii
Electrical Safety ..........................................................................................................vii
Pressure Safety ...........................................................................................................vii
Toxic Materials ............................................................................................................vii
Repair and Maintenance ..............................................................................................vii
Calibration (Factory Validation) .....................................................................................vii
Safety Conformity .......................................................................................................vii
Abbreviations .................................................................................................................... viii
OptiPEAK TDL600 User Manual
Michell Instruments v
Figures
Figure 1 Beer Lambert Law ......................................................................................3
Figure 2 Laser Scan .................................................................................................4
Figure 3 System Block Schematic ..............................................................................5
Figure 4 Unpacking the TDL600 ................................................................................6
Figure 5 Earthing Stud And Nut Washer Assembly ...................................................10
Figure 6 OptiPEAK Sampling System – Typical Indoor Version ...................................12
Figure 7 OptiPEAK Sampling System – Typical Indoor Version ...................................12
Figure 8 OptiPEAK Sampling System – Typical Outdoor Version .................................13
Figure 9 OptiPEAK Sampling System – Typical Outdoor Version .................................14
Figure 10 User Interface ..........................................................................................24
Figure 11 Up/Down Arrow Keys ................................................................................24
Figure 12 'ENTER’ Key .............................................................................................25
Figure 13 ‘ESC’ Key .................................................................................................25
Figure 14 Initialization Screen ..................................................................................26
Figure 15 Menu Structure ........................................................................................27
Figure 16 Main Menu Screen ....................................................................................28
Figure 17 Parameters Screen ...................................................................................29
Figure 18 Display Setup Screen ................................................................................30
Figure 19 Data Logging Screen ................................................................................31
Figure 20 Contact/About Screen ...............................................................................32
Figure 21 Graph Screen ...........................................................................................32
Figure 22 Advanced Settings Screen .........................................................................33
Figure 23 Output Screens ........................................................................................34
Figure 24 Alarm Screens ..........................................................................................35
Figure 25 Typical Alarm Status Indication on the Run-Mode Screen .............................38
Figure 26 Input Screen ............................................................................................39
Figure 27 Line Pressure Setup Screen .......................................................................40
Figure 28 Spare Input Setup Screen .........................................................................41
Figure 29 Set Date/Time Screen ...............................................................................42
Figure 30 Modbus Settings Screen ............................................................................43
Figure 31 N2-Mode (Measurement Mode) Screen ......................................................44
Figure 32 Safe Mode (Laser Disabled) Screen ............................................................45
Figure 33 Dimensional Drawing – Indoor System Enclosure ........................................57
Figure 34 Dimensional Drawing – Outdoor System Enclosure ......................................58
Figure 35 Wiring Diagram Indoor System ..................................................................60
Figure 36 Wiring Diagram Outdoor System ................................................................62
3.7.6 Advanced Settings Screen ............................................................................ 33
3.7.6.1 Outputs Screen ..................................................................................... 34
3.7.6.2 Alarms Screen ...................................................................................... 35
3.7.6.3 Inputs Screen ....................................................................................... 39
3.7.6.4 Clock Screen ......................................................................................... 42
3.7.6.5 Modbus Screen ..................................................................................... 43
3.7.6.6 N2-Mode (Measurement Mode) Screen ................................................... 44
3.7.6.7 Safe Mode (Laser Disabled) Screen ........................................................ 45
3.8 Enclosure Cover and User Interface ................................................................... 45
4 MAINTENANCE ................................................................................................46
4.1 Inspection of the Enclosure Cover ...................................................................... 47
4.2 Replacement of the Micro SD Data Logging Card ................................................ 49
4.3 Membrane and Particulate Filter Element Replacement ........................................ 51
4.3.1 Service Intervals ......................................................................................... 51
4.3.2 Installing the Filter Element and Membrane .................................................. 51
4.3.3 Field measurement verication ..................................................................... 53
4.3.4 Long-term maintenance – Laser replacement ................................................ 54
OptiPEAK TDL600 User Manual
vi 97319 Issue 5, August 2023
Appendices
Appendix A Technical Specication ...............................................................................56
A.1 Dimensional Drawings .................................................................. 57
A.2 Dimensional Drawings .................................................................. 58
Appendix B Indoor Sampling System Wiring Diagram ....................................................60
Appendix C Outdoor Sampling System Wiring Diagram .................................................. 62
Appendix D Flow Diagram ............................................................................................64
Appendix E Modbus Holding Register Map ....................................................................66
Appendix F Quality, Recycling, Compliance & Warranty Information ................................ 71
Appendix G Hazardous Area Certication ......................................................................73
G.1 ATEX/UKCA ................................................................................. 73
G.2 IECEx .........................................................................................73
G.3 North American (cQPSus) .............................................................73
G.4 Special Conditions of Use .............................................................74
G.5 Maintenance and Installation ........................................................74
Appendix H Return Document & Decontamination Declaration ........................................ 76
Tables
Table 1 Parameters Screen Parameters ...................................................................29
Table 2 Display Setup Screen Parameters ................................................................ 30
Table 3 Data Logging Screen Parameters ................................................................31
Table 4 Output Screen Parameters .........................................................................34
Table 5 Line Pressure Setup Screen Parameters .......................................................40
Table 6 Spare Input Setup Screen Parameters ......................................................... 41
Table 7 Set Date/Time Screen Parameters .............................................................. 42
Table 8 Modbus Screen Parameters ........................................................................43
Table 10 N2-Mode Parameters .................................................................................44
OptiPEAK TDL600 User Manual
Michell Instruments vii
Safety
This manual contains all the required information to install, operate and maintain the OptiPEAK
TDL600 Process Moisture Analyzer. Prior to installation and use of this product, this entire manual
should be read and understood. Installation and operation of this product should be carried out by
suitably competent personnel only. The operation of this product must be in accordance with the
terms of this manual and associated safety certicates. Incorrect installation and use of this product
for other than its intended purpose will render all warranties void.
This product is intended for use in a Hazardous Area and is ATEX, IECEx and UKCA approved. It
is also approved for use in North America in accordance with the requirements for the USA and
Canada. The relevant certicates should be fully examined prior to installation or use of this product.
!
Where this hazard warning symbol appears in the following sections,
it is used to indicate areas where potentially hazardous operations
need to be carried out and where particular attention to personal and
personnel safety must be observed.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories
supplied by the manufacturer for use with the instrument. The input power supply voltage is
either 110 V AC or 230 V AC, 50/60 Hz (dependent on chosen options).
Pressure Safety
DO NOT permit pressures greater than the safe working pressure to be applied directly to the
instrument's sample cell. The specied working pressure is 0.7...1.4 bara (10...20.3 psia). Refer to
the Technical Specications in Appendix A.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During
normal operation it is not possible for the user to come into contact with any hazardous substance
which might be employed in the construction of the instrument. Care should, however, be exercised
during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent. Refer
to www.ProcessSensing.com for details of Michell Instruments’ worldwide oces contact information.
Calibration (Factory Validation)
Prior to shipment, the analyzer undergoes stringent factory calibration to traceable standards. Due to
the inherent stability of the instrument, regular eld calibration is not required under normal operating
conditions. The analyzer should perform reliably for many years with just basic maintenance and
housekeeping. Michell can provide a fully traceable factory calibration service for the instrument
when required. Please contact your local Michell oce or representative for further details (www.
ProcessSensing.com).
Safety Conformity
This product meets the essential protection requirements of the relevant UK and EU directives.
Further details of applied standards may be found in the product specication.
OptiPEAK TDL600 User Manual
viii 97319 Issue 5, August 2023
Abbreviations
The following abbreviations are used in this manual:
A ampere
AC alternating current
bara pressure unit (=100 kPa or 0.987 atm)
barg pressure unit (=100 kPa or 0.987 atm) gauge
°C degrees Celsius
°F degrees Fahrenheit
EU European Union
ft feet
hr hour
kg kilogram(s)
lbs pound(s)
lb/MMscf pounds per million standard cubic feet
LCD liquid-crystal display
Nl/min normal liters per minute
m meters
mA milliampere
max maximum
*mg/m3 milligrams per cubic meter
mm millimeters
nm nanometers
NPT(F) National pipe thread (female)
PCB printed circuit board
ppmV parts per million by volume
psia pounds per square inch absolute
psig pounds per square inch gauge
RH relative humidity
RS485 serial data transmission standard
scfh standard cubic feet per hour
sec seconds
TDL Tunable Diode Laser
V Volt
W Watts
% percentage
inch(es)
ø diameter
* mg/m3 refers to standard sm3 (i.e. 15 °C at atmospheric pressure)
OptiPEAK TDL600 User Manual
Michell Instruments 1
INTRODUCTION
1 INTRODUCTION
The OptiPEAK TDL600 Tunable Diode Laser Analyzer employs the latest techniques in laser
absorption spectroscopy and signal processing power to oer a robust high performance
analyzer, designed specically for the measurement of moisture in natural gas. The analyzer is
fully hazardous area certied and delivers class-leading measurement performance, stability
and detection sensitivity.
The complete OptiPEAK TDL600 Analyzer Sampling System can be located close to the gas
sample take-o point in a potentially explosive environment designated Zone 1 and Zone 2
hazardous area.
The indoor version Sampling System gas handling components are assembled on a 316 Stainless
Steel plate suitable for wall mounting within a temperature controlled analyzer house.
The outdoor version Sampling System is housed within a stainless-steel enclosure (304
or 316), with optional thermostatically controlled heating and cooling, for direct eld
installation in a 100% shaded location next to the process line (with overall environmental
protection to IP66).
All sample gas wetted metallic parts are in AISI 316L stainless steel with Viton® soft parts that
comply with the NACE standard MR-01-75 (latest edition). Tube ttings are type 316 Stainless
Steel. All gas and cable entries are located in the base of the enclosure.
1.1 Application
The measurement of moisture in natural gas streams is an essential and highly critical
analysis for the natural gas industry. Gas companies need to meet specic quality
standards for transmission, custody transfer and delivery. High levels of water in the gas
increase the cost of transportation and lower the caloric value of the gas. In addition,
excessive moisture content in the gas stream can lead to internal pipe corrosion and
hydrate formation, requiring expensive pipe cleaning or 'pigging'. In severe cases,
pipeline blockage can occur.
Although the analyzer is designed for the measurement of water in transmission quality
natural gas backgrounds it has been congured for use with almost any natural gas
stream. This provides full exibility if, for example, the analyzer is later re-deployed to
a dierent application. (See Section 3, Operation, for further details.)
1.2 Features
High Measurement Sensitivity
The OptiPEAK TDL600 features a lower detection limit (LDL) of 1 ppmV water content. This
high sensitivity, coupled with the inherent fast response of the TDL optical measurement,
provides an extremely fast, accurate and reliable non-contact gas measurement.
D-MET – Dynamic Methane Compensation. BioGas Ready
Moisture measurement is virtually independent* of changes in the methane composition
of the natural gas feed and accuracy is not reliant on manual software correction factors
being applied.
The analyzer can be used with a wide range of background gas compositions. With
increasing statutory requirements in many regions for Biomethane to be added to
natural gas streams, the analyzer has been future proofed by being Biomethane ready.
* Over a methane concentration range of 40...100% CH4
OptiPEAK TDL600 User Manual
2 97319 Issue 5, August 2023
INTRODUCTION
Laser Lock System
Tunable diode lasers can drift. This means that the laser wavelength may slowly change
with time and, eventually, may not precisely match the absorption peak of the water.
This can lead to a reduction in sensitivity and analyzer drift. This inherent property of
diode lasers is overcome in the OptiPEAK TDL600 by the built-in Laser lock system.
This system monitors the optical prole of the gas absorption peaks to ensure the
laser remains locked to the correct water absorption peak, maintaining a high integrity
measurement at all times.
Fast Response
Being a non-contact optical measurement, the analyzer oers fast response times,
meaning no long wet-up or dry-down times in contrast to traditional moisture sensors.
None of the sensing components are exposed to the gas stream, protecting them from
any aggressive components or harmful contamination.
HMI system
Provides a highly intuitive menu driven interface, utilizing a capacitive touch screen
system, oering stylus-free setup and operation without the need for a 'hot work' permit
to adjust settings or to perform validation checks.
1.3 Theory of Operation
The OptiPEAK TDL600 uses the technique of absorption spectroscopy to measure the
concentration of water vapor in the gas stream. Many gas molecules exhibit very specic
resonant vibrations in the infrared region of the electromagnetic spectrum. If infrared
energy, at the same resonant wavelength, is passed through these molecules, some of
this energy will be absorbed. If a suitable detector is used to measure the amount of
received energy, and the gas is contained within a cell of a known path length, then the
gas concentration can be calculated. This can be expressed mathematically and is often
referred to as the Beer-Lambert Law.
c =
A
ε
l
where:
A
= absorbance
ε
= extinction coecient (absorption strength of gas at a specic wavelength)
l
= sample cell path length
c
= gas concentration
This law states that, if the sample cell path length
(l)
is known, and the extinction
coecient of the water molecule (
ε
a constant that describes how strongly a particular
gas absorbs light at a specic wavelength) is known, then, if the absorbance of laser
energy by the water molecules is measured (
A
), the water concentration (
c
) of the
sample stream can be calculated. This gas law is the basis of all photometric gas
absorption measurement.
OptiPEAK TDL600 User Manual
Michell Instruments 3
INTRODUCTION
Beer Lambert law
Absorbent
concentration (c)
Path length (l)
IOIX
Figure 1
Beer Lambert Law
The Michell OptiPEAK TDL600 uses a tunable diode laser source to generate a narrow
and coherent beam of near infrared (NIR) energy at the precise resonant wavelength
of water vapor. Traditionally, infrared analyzers use broadband sources which generate
a wide gamut of wavelengths. To make these analyzers as selective as possible to
moisture only, optical lters have to be deployed to 'narrow' the range of wavelengths
that are nally passed through the sample. These lters do not oer very high selectivity
– they are quite broadband, which can lead to signicant spectral interference, as other
gas peaks close to the water absorption peaks are also detected, leading to cross
interference, drift and general degradation in measurement performance.
In contrast, the laser has a bandwidth of less than 0.0001 nanometers. This means the
laser is very selective in detecting only the water and not any other gases present in
the gas stream.
This optical technique also has the advantage that the analyzer uses a non-contact
method of measurement, i.e. there is no sensing element in contact with the gas
stream. This oers a highly robust and reliable measurement, as only photons of light
pass through the gas. This provides very fast response and no long wet-up or dry-down
times.
OptiPEAK TDL600 User Manual
4 97319 Issue 5, August 2023
INTRODUCTION
1.3.1 Measurement Using a Laser
The diagram below illustrates the advantages of using a laser source, compared to a
traditional broadband source.
The water absorption peak is shown in the center of the diagram (red area). The width
of the laser beam is very narrow and is represented by the yellow line.
The laser wavelength is varied in order to scan across the water absorption peak (yellow
area). By scanning the peak in this way, important information can be extracted, such
as changes to the absorption peak caused by variation in the sample gas. This very
precise scan range minimizes any overlap with nearby absorption bands, as would be
the case with conventional broadband infrared sources and optical lters (blue area).
Scan range
Potential interferant gases Potential interferant gases
Gas absorption
(peak of interest)
Bandwith
approximately 0.05nm
Conventional IR
source gives
very broad
transmission
Typical broadband source bandwidth
5 Hz ‘sawtooth’
laser scan
Figure 2
Laser Scan
OptiPEAK TDL600 User Manual
Michell Instruments 5
INTRODUCTION
The schematic of the laser measurement system is shown in
Figure 3
below. This
highlights the major control and signal processing sections of the analyzer. The analyzer
uses the WMS (Wavelength Modulation Spectroscopy) technique, in combination with
proprietary signal processing algorithms to provide a selective response and high
sensitivity to moisture.
Here, a single frequency, distributed feedback (DFB) laser diode is tuned by applying
a current ramp to the laser. A further sinusoidal modulation is then applied. Lock-in
detection is applied to the photo-detector signal obtained by passing the tuned laser
radiation through the gas cell. The second harmonic signal from the lock-in detection is
measured to recover the spectroscopic peak of interest. WMS oers a practical method
of recovering weak signal changes from a dilute trace gas sample. WMS is becoming
wide spread within the sector of natural gas monitoring and represents current state-
of-the-art technology.
Temperature
controlled
diode laser
Detector
Launch head
Diode
laser
power
Ramp
current
High
frequency
‘carrier’
modulation
Signal
processing
Mixer
Filter
Diode
current
Second harmonic
Direct signal
Fiber Optic cable
Laser scanning
Det. current
Figure 3
System Block Schematic
OptiPEAK TDL600 User Manual
6 97319 Issue 5, August 2023
INSTALLATION
2 INSTALLATION
2.1 Unpacking the Instrument
Open the crate and unpack carefully as follows:
!
WARNING:
The instrument is heavy and should not be lifted alone.
Mechanical lifting aids may be required for larger systems.
1
2
3
2
Figure 4
Unpacking the TDL600
1. Remove the accessories box (1).
2. Remove the spacer foam (2).
3. Remove the instrument enclosure (3) and set it down at the site of
installation.
It is recommended to save all the packing materials for the purpose of returning the
instrument for warranty claims.
The accessories box should contain the following items:
Calibration Certicate
CD containing System Documentation
OptiPEAK TDL600 User Manual
Michell Instruments 7
INSTALLATION
2.2 Lifting and Handling
!
WARNING:
This product is in excess of 75kg (165lbs).
Personnel must observe suitable lifting and handling
precautions.
The TDL600 is not designed as portable or transportable equipment. The product should
be rigidly xed in position as per the full installation instructions.
Appropriate lifting and handling techniques should be used during the installation process.
Before commencing any lifting or handling ensure that its intended location is suitable
and appropriately prepared. Make sure that mounting point design considerations have
employed locally approved safety factors.
When handling and installing this instrument (particularly after removal from its
packaging) ensure that it is not dropped, impacted or subjected to high levels of
vibration or environmental conditions that may impair its operation.
2.3 Laser Safety
This product contains a Diode Laser with an invisible beam, operating in the near
infrared range. The laser as used in this product classies it as a CLASS 1 product.
For the purposes of CDRH and FDA Registration the OptiPEAK TDL600 complies with
21CFR1040 with deviations pursuant to Laser Notice 50 and with IEC/EN 60825-1:2007.
WARNING:
This product is a CLASS 1 LASER PRODUCT.
Beware of Laser radiation.
Do not access the Laser.
Do not view the Laser directly.
!
WARNING:
Use of controls or adjustments, or performance of
procedures other than those specied herein, may result in
hazardous radiation exposure.
OptiPEAK TDL600 User Manual
8 97319 Issue 5, August 2023
INSTALLATION
2.4 Hazardous Area Safety
Refer to Appendix G for the Hazardous Area Certication of this product.
This product is tted with a marking label that contains Hazardous Area information
pertinent to the suitable location and installation.
During all installation and operation activities, local regulations and permitted working
routines must be observed. Installation should only be performed by competent
personnel and in accordance with IEC 60079-14:2007 and EN 60079-14:2008 or local
equivalent.
Cable glands / conduit seals shall be installed in accordance with the manufacturer’s
instructions.
Conduit seals used should be suitable for a reference pressure of 6.1 bar (89 psi).
Repair and servicing of this equipment must only be carried out by the manufacturer.
WARNING:
This product is certied safe for use in a Zone 1 and Zone 2
area only. This product must not be installed or used within
a Zone 0 area.
WARNING:
This product must not be operated within an explosive
atmosphere greater than 1.1 bara (16 psia).
!
WARNING:
This product must not be operated within an enriched
oxygen atmosphere (more than 21% oxygen content).
WARNING:
This product must not be operated outside of the
temperature range of -20 to +55 °C (-4 to +131 °F)
WARNING:
The analyzer enclosure of this product provides Exd
protection, partly through the threads used for mounting
the lid, stopping plugs and cable gland. At all times eort
should be made to ensure these threads are suitably
protected from damage and that only appropriately rated
mating parts are applied to them, in accordance with the
certifying requirements.
OptiPEAK TDL600 User Manual
Michell Instruments 9
INSTALLATION
2.5 Electrical Safety
DANGER
Electric
Shock Risk
WARNING:
During the installation of this product, ensure that all
applicable national and local electrical safety regulations are
observed.
WARNING:
Always ensure that power is switched o prior to accessing
the product for any purpose other than normal operation, or
prior to disconnecting any cables.
2.5.1 Equipment Ratings and Installation Details
The following mandatory statements refer to the Ex certified TDL600 Analyzer and
sampling system.
All electrical connections to the analyzer are made through junction boxes, mounted on
the panel of the sample system in accordance with Section 2.8.
Any power cable should be 3 core over sleeved, with minimum 0.5mm insulation
and rated at 300 V. Cables should have Live (L), Neutral (N) and Earth [Ground] (E)
conductors. Ensure suitably rated power supply cables and glands are used to ensure
that electrical safety is maintained. Ensure the power supply can deliver sucient power
to meet the consumption requirements.
Any power supply terminals and voltages must be suitably separated from the other I/O
requirements to this product.
Before applying power, perform a continuity test to ensure that the power supply cable
and the TDL600 are eectively connected to the protective Earth.
OptiPEAK TDL600 User Manual
10 97319 Issue 5, August 2023
INSTALLATION
The protective Earth terminal is mounted internally and the Earth wire connected to it
should never be disconnected. The analyzer enclosure is supplied with an external earth
stud at the lower right hand side. This earth stud is connected to the sampling system
earth using 4mm2 minimum earthing bonding.
Figure 5
Earthing Stud And Nut Washer Assembly
Fuse: A replacement fuse can be obtained by contacting Michell Instruments' technical
support. Fuse rating = 5 x 20mm 2.5 A anti-surge to IEC 60127-2.
This measuring product is designed, where applicable and possible, to be in compliance
with EN/BS/IEC61010 safety requirements or electrical equipment or measurement,
control, and laboratory use. This product is designed to be safe at least under the
following conditions: between a temperature range of -40...+60 °C (-40...+148 °F), in
maximum 80% relative humidity for temperatures up to +31 °C (+88 °F) decreasing
linearly to 50% rh at +50 °C (+122 °F). Supply voltages of ±10% and transient over
voltages up to Overvoltage Category II. Pollution Degree 2. Altitudes up to 2000m.
Outdoor mounting is permitted using suitably rated glands equivalent to NEMA 4 / IP66.
See Appendix A, Technical Specication, for full operating parameters.
NOTE: Do not remove or exchange any of the cables or electrical components
supplied with this product. Doing so will invalidate all warranties.
There are no additional or special electrical safety requirements other than those
referred to in this manual.
For location and mounting arrangements please refer to the relevant sections of this
manual.
Installation of this equipment must include the provision of a suitable and locally
positioned power isolation switch or circuit breaker. Indication of the purpose of the
switch or circuit breaker is strongly recommended. An over-current protection device
should be rated to a maximum of 3 A.
This equipment and all power isolation devices must be installed in a location and
position that allows safe and easy access to their operation and is able to rigidly support
the equipment.
OptiPEAK TDL600 User Manual
Michell Instruments 11
INSTALLATION
Do not install this equipment in a location that would expose it to impact or high levels
of vibration.
Operation of this equipment, other than in a manner specied by the manufacturer, may
impair the safety protections provided.
The safe installation of this equipment and any system incorporating this equipment is
the responsibility of the installer. Ensure local regulations and requirements are referred
to prior to any installation commencing.
2.6 Pressure Safety
!
WARNING:
This product is used in conjunction with pressurized gases.
Observe pressurized gas handling precautions.
Pressurized gas should only be handled by suitably trained
personnel.
The TDL600 measurement chamber requires pressurized gas to be connected to it.
Observe pressurized gas handling regulations. Only suitably trained personnel should
carry out tasks that include the use of pressurized gas media.
The TDL600 measurement cell accepts a maximum sample pressure of 1.4 bara (20.3 psia).
2.7 Basic Installation Guidelines
The OptiPEAK TDL600 Moisture Analyzer Sampling System gas handling components
are assembled onto a stainless steel mounting plate suitable for wall mounting.
The outdoor version Sampling System provides environmental ingress protection to IP66
and should be mounted vertically, free of any appreciable vibration, in a permanently
shaded position to prevent heating eects through sun radiation. The Sampling System
enclosure can be specied with optional thermostatically controlled heating (xed set
point). Optional enclosure cooling, using a compressed-air-driven vortex tube and xed
set point thermostat, is recommended for installation in hot climates (>+45 °C (>+113
°F)).
NOTE: Any TDL being installed within a plant where it cannot vent to open
atmosphere needs a are line connection that runs to the highest point and
enters the are system from a topside connection. This is to prevent liquids
present in the are stack from draining back into the analyzer system.
NOTE: The actual detailed conguration will be shown in the as-built drawings
provided with the shipped analyzer.
For start-up instructions refer to Section 3.
OptiPEAK TDL600 User Manual
12 97319 Issue 5, August 2023
INSTALLATION
1BV1 Ball Valve
2 F2 Particulate Filter
3PR1 Pressure Regulator
4PG1 Pressure Gauge
5PG2 Pressure Gauge
6 F2 Coalescing & Membrane Filter
7PR2 Pressure Regulator
8PG3 Pressure Gauge
9PRV1 Pressure Relief Valve
10 FM1 Flowmeter
11 AN1 Moisture Analyzer
12 MV1 Metering Valve
13 NV1 Needle Valve
14 FM2 Flowmeter
15 HT Enclosure Heater
16 TS1 Thermostat
17 SOV1 Solenoid Valve
18 TH Trace Heating
1
TP1
TP2
TP4 TP3
14
2
4
18
17
16
13
12
6
5
7
8
9
11
10
3
15
TP5
TP CONNECTIONS
11/4" NPT(F) Sample Gas Inlet
21/4" NPT(F) Sample Gas Outlet
31/4" NPT(F) Bypass Flow Gas Outlet
41/4" NPT(F) Drain
51/4" NPT(F) Vortex Cooler Inlet
A4 PAPER
1
TP1
TP2TP4 TP3
14
2
4
13
12
6
5
7
8
9
11
10
3
TP CONNECTIONS
11/4" NPT(F) Sample Gas Inlet
21/4" NPT(F) Sample Gas Outlet
31/4" NPT(F) Bypass Flow Gas Outlet
41/4" NPT(F) Drain
1BV1 Ball Valve
2 F2 Particulate Filter
3PR1 Pressure Regulator
4PG1 Pressure Gauge
5PG2 Pressure Gauge
6 F2 Coalescing & Membrane Filter
7PR2 Pressure Regulator
8PG3 Pressure Gauge
9PRV1 Pressure Relief Valve
10 FM1 Flowmeter
11 AN1 Moisture Analyzer
12 MV1 Metering Valve
13 NV1 Needle Valve
14 FM2 Flowmeter
1BV1 Ball Valve
2 F2 Particulate Filter
3PR1 Pressure Regulator
4PG1 Pressure Gauge
5 PR2 Pressure Gauge
6F2 Coalescing & Membrane Filter
7PR2 Pressure Regulator
8PG3 Pressure Gauge
9PRV1 Pressure Relief Valve
10 FM1 Flowmeter
11 AN1 Moisture Analyzer
12 MV1 Metering Valve
13 NV1 Needle Valve
14 FM2 Flowmeter
TP Connections
TP1 Sample Gas Inlet 1/4” NPT (F)
TP2 Sample Gas Outlet 1/4” NPT (F)
TP3 Bypass Flow Gas Outlet 1/4” NPT (F)
TP4 Letdown Gas Vent/Drain 1/4” NPT (F)
Figure 6
OptiPEAK Sampling System – Typical Indoor Version
MODEL: OPTIPEAK TDL600
TYPE: PROCESS MOISTURE ANALYZER
SERIA L NUMBER : nnnnnn /2017
WO NUMBER : BC45565
www.michell.com
MOUNTING HOLES
(10xO11mm)
316SS MOUNTING
PLATE
POWER INLET
CABLE ENTRY
(M20)
SIGNAL OUTPUTS
CABLE ENTRIES
(2 x M20)
316SS NAMEPLATE
TP02
TP04 TP01 TP03
12.0 726.0 CRS 12.0
25.0 700.0 CRS 25.0
750.0
10.0 10.0
22.0 22.0
970.0
250.0
MIN. CLEARANCE
12.0
475.0 CRS 475.0 CRS
463.0 CRS 463.0 CRS
Figure 7
OptiPEAK Sampling System – Typical Indoor Version
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86