Precision Source Measure Unit
GSM-20H10
User Manual
GW INSTEK PART NO. 82SM320H10EB1
ISO-9001 CERTIFIED MANUFACTURER
Copyright Statement
This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
GSM-20H10 User Manual
1
Table of Contents
Table of Contents .......................................................... 1
SAFETY INSTRUCTIONS............................................. 5
Safety Notes ........................................................... 5
Symbols .................................................................. 6
Guidelines .............................................................. 8
OVERVIEW ................................................................. 11
Main Function ..................................................... 11
Overview ........................................................................................ 11
Main Function ............................................................................... 12
Key Features .................................................................................. 14
Front Panel .......................................................... 16
Display & Parameter .................................................................... 16
Control Panel ................................................................................. 22
Terminals ....................................................................................... 26
Rear Panel ............................................................ 26
Terminals ....................................................................................... 27
GETTING STARTED .................................................. 29
Safety Remind...................................................... 29
Prepare & Start Up .............................................. 30
Check The AC Voltage ................................................................. 30
Connect The AC Power Cord ...................................................... 30
Turn The Power On ...................................................................... 31
Turn The Power Off ...................................................................... 32
Terminal Connection ........................................... 33
Front/Rear Terminals .................................................................. 33
Connect To Load ........................................................................... 33
Sense Connect Methods ............................................................... 34
Sense And Guard Selections ........................................................ 37
Wires Selections ............................................................................ 39
Power On Settings ............................................... 40
Table of Contents
2
BASIC OPERATION ................................................... 41
Preparation .......................................................... 41
Operating Environment ............................................................... 41
Funtion Overview ......................................................................... 42
Parameter Setting Method ........................................................... 44
Source Function ................................................... 46
Circuit Configuration ................................................................... 46
V/I Output Range ......................................................................... 48
Sink Function ................................................................................. 52
Operation Precautions .................................................................. 54
Front Panel Programming Interface ........................................... 55
Parameter Description And Operation ...................................... 56
Output Operation ......................................................................... 78
State Description ........................................................................... 79
Measure Function ................................................ 80
Measurement Interface ................................................................. 80
V/I Meter ....................................................................................... 81
Ohms Meter ................................................................................... 83
Calculation ..................................................................................... 86
Parameter Description And Operation ...................................... 89
Parameter Settings of Calculation Function .............................. 92
Measurement Operations ............................................................ 98
State Description ........................................................................... 98
Limit Function ................................................... 100
Display Interface ........................................................................... 100
Parameter Description And Operation ...................................... 102
External I/O ................................................................................... 117
Limit Operation ............................................................................. 120
State Description ........................................................................... 122
Sequence Function ............................................. 123
Display Interface ........................................................................... 123
Features .......................................................................................... 124
Parameter Description And Operation ...................................... 128
Generating Waveform And Exporting File ............................... 135
Sequence Output ........................................................................... 138
TRIG Fuction ..................................................... 139
Programming Interface ................................................................ 139
Trigger Process .............................................................................. 140
Parameter Description And Operation ...................................... 141
Interface Requirements ................................................................ 146
GSM-20H10 User Manual
3
SYSTEM SETTINGS .................................................. 148
Range Limitation ............................................... 148
System Parameters ............................................ 151
Save/Recall ........................................................ 159
Factory Settings ................................................. 160
Software Upgrade.............................................. 165
System Clock ..................................................... 168
Mobile Disk Usage ............................................ 168
REMOTE CONTROL .................................................. 170
Setting Interface ................................................ 170
USB ................................................................................................. 170
RS-232 ............................................................................................. 173
GPIB ................................................................................................ 177
LAN ................................................................................................ 180
COMMAND SYNTAX ................................................ 184
SCPI Commands Overview ............................... 184
Command Format ......................................................................... 184
Symbol Description ...................................................................... 185
Parameter Types ........................................................................... 185
Command Abbreviations ............................................................. 186
Command Terminators ................................................................ 187
Command List ................................................... 188
Calculate Instructions ................................................................... 188
Display Commands ...................................................................... 190
Data Format Commands .............................................................. 190
Output Commands ....................................................................... 190
Source Commands ........................................................................ 191
Measurement Commands ............................................................ 194
Status Commands ......................................................................... 196
System Commands ....................................................................... 197
Trigger Commands ....................................................................... 198
IEEE488.2 Common Commands ................................................. 199
Command Details .............................................. 201
Calculate Commands ................................................................... 201
Display Commands ...................................................................... 229
Data Format Commands .............................................................. 230
Output Commands ....................................................................... 242
Table of Contents
4
Source Commands ........................................................................ 245
Measurement Commands ............................................................ 288
Status Commands ......................................................................... 314
System Commands ....................................................................... 319
Trigger Commands ....................................................................... 328
IEEE488.2 Common Command ................................................... 345
Status Registers ................................................. 350
The Structure Of Status Registers ............................................... 351
Programming And Reading Registers ....................................... 352
Status Byte And Service Request (SRQ) ..................................... 353
Status Register Sets ....................................................................... 355
Queues ............................................................................................ 359
Errors ................................................................. 360
Error Message ................................................................................ 360
Command Errors .......................................................................... 360
APPENDIX .................................................................. 366
Replacing The Fuse ........................................... 366
Battery Replacement ......................................... 367
Specification ...................................................... 368
Optional Accessories ......................................... 373
Certificate Of Compliance ................................. 374
GSM-20H10 User Manual
5
SAFETY INSTRUCTIONS
This chapter contains important safety instructions
that you must follow during operation and storage.
Read the following before any operation to insure
your safety and to keep the instrument in the best
possible condition.
Safety Notes
This product is intended for use by qualified personnel who
recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Read and follow
all installation, operation, and maintenance information carefully
before using the product.
Ensuring that the equipment is operated within its specifications
and operating limits, and ensuring that operators are adequately
trained. They must be trained in electrical safety procedures and
proper use of the instrumentand must be protected from
electric shock and contact with hazardous live circuits.
Exercise with extreme caution when a shock hazard is present.
Lethal voltage may be present on cable connector jacks or test
fixtures. A shock hazard exists when voltage levels greater than
30V RMS, 42.4V peak, or 60VDC.
For maximum safety, do not touch the product, test cables, or
any other instruments while power is applied to the circuit
under test.
SAFETY INSTRUCTIONS
6
Do not touch any object that could provide a current path to the
common side of the circuit under test or power line (earth)
ground.
The instrument and accessories must be used in accordance with
its specifications and operating instrument, or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and
accessories as defined in the specifications and operating
information (as shown on the instrument or test fixture panels).
When fuses are used in a product, replace with the same type
and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for
measuring circuits, NOT used as safety earth ground
connections.
If you are using a test fixture, keep the lid closed while power is
applied to the device under test. Safe operation requires the use
of a lid interlock.
Symbols
These symbols may appear in the manual or on the instrument.
ATTENTION
CAUTION
GSM-20H10 User Manual
7
WARING
DANGER
GROUND
SAFETY INSTRUCTIONS
8
Guidelines
General
Guideline
CAUTION
Do not place any heavy object on the unit.
Avoid severe impact or rough handling that
leads to damaging the unit.
Do not discharge static electricity to the unit.
Do not block the cooling fan opening.
Do not disassemble the GSM-20H10 unless you
are qualified.
EN 61010-1 specifies the measurement
categories and their requirements as follows.
The GSM-20H10 operates under category of Not
II, III, IV.
Measurement category IV is for measurement
performed at the source of low-voltage installation.
Measurement category III is for measurement
performed in the building installation.
Measurement category II is for measurement
performed on the circuits directly connected to the
low voltage installation.
0 is for measurements performed on circuits not
directly connected to Mains.
EN 61010-1 specifies pollution degrees and their
requirements as follows. Pollution refers to
“addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a
reduction of dielectric strength or surface
resistivity”.The GSM-20H10 operates under
degree 2.
Pollution degree 1: No pollution or only dry, non-
conductive pollution occurs. The pollution has no
influence.
GSM-20H10 User Manual
9
Pollution degree 2: Normally only non-conductive
pollution occurs. Occasionally, however, a temporary
conductivity caused by condensation must be
expected.
Pollution degree 3: Conductive pollution occurs, or
dry, non-conductive pollution occurs which becomes
conductive due to condensation which is expected. In
such conditions, equipment is normally protected
against exposure to direct sunlight, precipitation, and
full wind pressure, but neither temperature nor
humidity is controlled.
Power Supply
AC Input voltage range: 100~240VAC(±10%)
50/60Hz
WARNING: To avoid electrical shock,
connect the protective grounding conductor of the
AC power cord to an earth ground.
Fuse
Fuse type: T2.0A/250V
WARNING:
Make sure to use the correct fuse type before
starting up.
To prevent fire, replace the fuse only with the
specified type and rating.
Disconnect the power cord before replacing the
fuse.
Make sure the cause of fuse blowout is fixed
before replacing the fuse.
SAFETY INSTRUCTIONS
10
Cleaning the
instrument
WARNING:
Disconnect the power cord before cleaning the
instrument.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid
into the instrument. Clean the exterior of the
instrument only.
Do not use chemicals containing harsh products
such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: < 80%
Altitude: < 2000m
Temperature: 0°C to 40°C
Storage
environment
Location: Indoor
Relative Humidity: < 80%
Temperature: -20°C to 70°C
OVERVIEW
11
OVERVIEW
This chapter contains a brief introduction to GSM-
20H10, including the main features, as well as an
overview of the front and rear panel. Use the Getting
Started chapter on page 29 to start up instructions and
how to setup the appropriate operation environment.
Main Function
Overview
The GSM combines a precise, low-noise, highly stable DC power
supply with a low-noise, high-impedance multimeter. It has 0.012%
basic accuracy with 61/2 digit resolution. At 41/2 digits display
resolution (Medium), the GSM delivers 520 readings/second over
the IEEE-488 bus. At 31/2 digits display resolution (Fast), it can read
up to 2000 readings/second into its internal buffer.
Because of its built-in source function, the GSM can be used to
generate a set of current voltage (I-V) characteristic curves, which is
very useful for semiconductor device and material testing.
It is equivalent to the combination of voltage source, current
source, voltmeter, amperemeter and ohmmeter.
The digital GSM is designed for routine testing and high-speed
production testing applications. In the manufacturing of
components and modules of communication, semiconductor,
computer, automobile and medical industry, GSM is of great
practical valuable for their various characteristics analysis and
production process testing.It is also usually used in testing and R &
D laboratories.
OVERVIEW
12
Main Function
Four Quadrant Source Function (V/I)
GSM-20H10, as a conventional power supply, has the function of
automatic CC/CV crossover function. You can use the front panel to
set as voltage source or current source, and configure parameters
such as voltage testing, current testing, display digits, data sampling
period, power status, OVP, etc. The voltage and current testing
settings and the actual voltage/current testing value are displayed
on the LCD. V/I measurement, limit measurement, sequence output
and trigger functions are all extended on the basis of source
function. The relationship is shown in figure below:
For more source information, see page 46.
Measure Function (V/I/Ω)
GSM-20H10 can also be used as voltmeter/amperemeter and
ohmmeter to measure external voltage/current value and
resistance. The measurement function can also comply five
calculation functions through the built-in mathematical functions:
Power, CompOhms, VarAlpha, Vcoeff, DEV. The relevant
parameter settings can be set through the front panel keys and
displayed on the LCD. See page 80 for details.
Source
Limit Measure Sequence
GSM-20H10 User Manual
13
* The user can define up to 5 operations through the remote
command. For details, see page 201.
Limit Function
There are three types of limits: compliance, coarse limits and fine
limits. And summarizes the two operating modes: grading and
sorting. For details, see page 100. Covers pass/fail condition for the
grading and sorting modes. For details, see page 110.
Sequence Function
There are four basic sweep types: linear Staircase, Logarithmic
staircase, Custom and SRC-MEM sweep. For details, see page 123.
TRIG Function
The trigger model consists of two layers (Arm Layer and Trigger
Layer) to provide versatility. Programmable counters allow
operations to be repeated, and various input and output trigger
options are available to provide source-measure synchronization
between the GSM and other instruments (via the Trigger Link). For
details, see page 139.
Remote Control
To meet the various needs of customers, the GSM-20H10 is designed
for USB (TMC), RS-232, GPIB and LAN remote control. For details,
see page 170.
External I/O port
The GSM-20H10 has external control signals for customers.
Including input trigger signal and output control signal. For details,
see Limit function on page 117 and Trigger function on page 146.
OVERVIEW
14
Key Features
General
Low noise, the speed of fan is controlled by
thermostatic.
Compact, meets the 2U height and 1/2Rack width
requirement.
4.3 inch TFT LCD display.
Constant voltage and constant current operation
(CV/CC).
Front and Rear output control key
(FRONT/REAR).
Digital voltage and current settings by number
pad.
Alarm buzzer (BEEP).
Key lock function (LOCK).
Additional
˙ Source-measure sweep capabilities (Linear and
Logarithmic staircase sweeps, source sweep list
of up to 2500 points, memory sweep of up to
100 instrument setups).
˙ 6-wire Ω measurement with programmable I-
source or V-source.
˙ 4-quadrant source and sink operation.
˙ Up to 11 stages of limit testing with a built-in
comparator for pass/fail testing.
˙ Digital I/O interface for component handler.
˙ Five built-in math expressions, up to five user-
defined math expressions (remote command
only).
GSM-20H10 User Manual
15
˙ Reading and setup storage, up to 2500 readings
and six storage area (four user settings, two
factory defaults) can be stored and recalled.
Protection
˙ Overvoltage protection (OVP).
˙ Overtemperature protection (OTP).
Interface
˙ USB remote control.
˙ RS-232 remote control.
˙ GPIB remote control.
˙ LAN remote control.
OVERVIEW
16
Front Panel
LCD Display Numberpad and Power
Secondary function key on/off key
USB host port Auxiliary C/Pict key Front panel Input/
function key Output terminals
Display & Parameter
Power on
Display
(Source)
GSM-20H10 User Manual
17
Description
The display is used primarily to program source and
compliance values and display the real measured
readings, The area is divided as follows.
Status and Error messages
Status and error messages are displayed
momentarily, which are located on the top of the
display, indicate various states of operation. The
meaning of each status information is described
separately in each function introduction chapter.
The error message contains a variety of error
types, which can be read by instructions under
remote control. Typical error messages
description and types are listed in page 360.
Reading Data/Format
The readback value is displayed with a resolution
of up to 61/2 digits. The number of display digits
can be changed by directly setting Digits or
modifying the refresh Speed. Take the highest
resolution of the readback voltage as an example:
The readback display area is divided into 3 parts:
1: sign, 2: value, 3: units.
˙ The measurment values of voltage, current
and resistance can be positive or negative.
˙ The digital display has a total of 7 digits at
most, and the range is determined by the
integer digits combined with the unit.
When measuring V:
˙ 3 integer digits for 200V or 200mV range
˙ 2 integer digits for 20V range
˙ 1 integer digits for 2V range
OVERVIEW
18
When measuring I:
˙ 3 integer digits for 100mA or 100uA range
˙ 2 integer digits for 10mA or 10uA range
˙ 1 integer digits for 1A or 1mA or 1uA range
When measuring Ω
˙ 3 integer digits for 200MΩor 200kΩ or 200Ω
range
˙ 2 integer digits for 20MΩor 20kΩ or 20Ω
range
˙ 1 integer digits for 2MΩor 2kΩ or 2Ω range
The unit of voltage is V or mV; the unit of current
is A or mA or uA; the unit of resistance is MΩ or
KΩ or Ω; when performing Math, there may not
be the above unit.
When the output is turned on, the top (main)
display area is used for measurement. The read
information can be displayed in integer or
floating point format using engineering notation
or scientific notation. Example of engineering
notation: 1.23456μA, example of scientific
notation: 1.23456e-6. Using System->Control-
>Numbers option to select the display format.
Edit the source
Under the reading display area, is used for
programming source value (Vsrc or Isrc) and
compliance (Cmpl) limit. For the settings of OVP,
please refer to Source Operation.
Voltage source/Current source setting
operation:
Contains 3 aspects, as shown in the figure:
/