Thermo Fisher Scientific 32-Gauge Stainless Steel Needle Installation guide

Type
Installation guide
© 2012 Thermo Fisher Scientific Inc.
All rights reserved.
Thermo Fisher Scientific supplies stainless steel needles in two
sizes, 32-gauge and 34-gauge. This guide describes how to install
the 32-gauge stainless steel needle, which supports flow rates
from5to100L.
Overview
A fused-silica capillary typically connects the solvent path between
the grounding union held by the grounding bar of the Ion Max™ or
Ion Max-S API source and the tip of the electrospray needle that
protrudes from the nozzle of the ESI probe.
Because acetonitrile adversely affects a fused-silica capillary,
Thermo Fisher Scientific recommends the use of a stainless steel
needle and red PEEK™ tubing for applications that use acetonitrile
in the mobile phase.
The stainless steel needle connects the solvent path inside the probe
(see Figure 1). The red PEEK tubing connects the probe to the
grounding union held by the grounding bar of the Ion Max or
Ion Max-S API source (see Figure 2).
Figure 1. Solvent path inside the ESI probe
Figure 2. Connection between the probe and the grounding union
Required Parts and Tools
To install the 32-gauge stainless steel needle, you must have the parts
and tools listed below. The MS Accessory Kit includes red PEEK
tubing. The Metal Needle Kit, OPTON-53003, contains the
additional required parts with the exception of the wrench and the
Teflon tape.
Step 1: Inserting the Needle into the Probe
YTo insert the 32-gauge stainless steel needle into the ESI probe
1. Wearing talc-free gloves, straighten one end of the thread
that protrudes from the bore of the stainless steel needle
(see Figure 3), and then pull the thread out of the needle from
the opposite end.
Figure 3. Stainless steel needle with internal thread
2. Remove the adapter from the sample inlet of the ESI probe
(see Figure 4).
Figure 4. Adapter nut shown removed from the probe
32-Gauge Stainless Steel Needle Installation Guide
Contents
Overview
Required Parts and Tools
Step 1: Inserting the Needle into the Probe
Step 2: Securing the Needle Position
Step 3: Connecting the Probe to the Grounding Union
32-gauge
stainless steel needle
Tip of the 26-gauge
electrospray needle
Adapter nut
External union
and two-ferrule
assembly
5.65 in. (140 mm)
Grounding bar
of the API
source
Fingertight
fittings
Red PEEK tubing
Grounding
union
Non-conductive,
natural PEEK
internal union
Description Qty Part Number
Ferrule, 0.012 in. ID, clear 6 00101-18116
External union,
10-32, natural PEEK
2 00101-18206
Internal union,
10-32, natural PEEK
1 00109-00305
10-32 × 1/4-28, Kel-F, adapter
for sample inlet of ESI probe
1 00101-18080
High-pressure, fingertight nut
and ferrule for 1/16 in. OD
tubing
2 00101-18195
Stainless steel needle, 32-gauge,
blunt-tipped, 5.65 in. length,
0.004 in. ID × 0.009 in. OD
3 00950-00954
0.005 in. ID, red PEEK tubing 00301-22912
Teflon t ape no t sup plied
5/16 in. open-ended wrench not supplied
Internal thread
5.65 in. stainless steel needle
10-32 ×1/4-28,
Kel-F, adapter nut
Sample inlet port
in the back of the
ESI probe
Nozzle on the front
of the ESI probe
Revision E 70001-97112
2
3. Insert the 32-gauge stainless steel needle through the
26-gauge needle that protrudes from the nozzle of the ESI probe
(see Figure 5). Then push the needle through the probe until it
protrudes from the sample inlet at the back of the probe
(see Figure 6).
Figure 5. Inserting the stainless steel needle into the electrospray needle
Step 2: Securing the Needle Position
YTo secure the needle position
1. Slip the adapter nut over the end of the 32-gauge stainless steel
needle that protrudes from the sample inlet port. Screw the
fitting, fingertight, into the port (see Figure 6).
Figure 6. Adapter nut shown installed in the sample inlet port
2. Wrap Teflon tape around the threads on one end of the natural
PEEK external union provided in the Metal Needle Kit
(see Figure 7). Take care to avoid taping over the thru hole.
Figure 7. External union and two-ferrule assembly
3. Slide the end of the needle that protrudes from the back of the
ESI probe through the Teflon-wrapped end of the external union.
4. Screw the Teflon-wrapped end of the external union into the
adapter, fingertight.
5. Slide a clear, 0.012 in. ID ferrule over the needle that protrudes
from the back of the ESI probe.
6. To ensure that you have installed the needle correctly, check the
following:
One end of the 32-gauge stainless steel needle is flush with
the tip of the free ferrule (see Figure 8).
The other end of the 32-gauge stainless steel needle does not
protrude from the front of the probe by more than 1 mm.
Figure 8. Back of the ESI probe with the needle shown in the correct position
For optimal spray quality, the tip of the 32-gauge stainless steel
needle must protrude by approximately 1 mm from the tip of
the 26-gauge electrospray needle (see Figure 9).
Figure 9. Front of the ESI probe with an enlarged view of the needle tip
Step 3: Connecting the Probe to the Grounding Union
YTo connect the ESI probe to the grounding union
1. With a 5/16 in. open-ended wrench, brace the external union as
you screw the internal union made of non-conductive, natural
PEEK onto the free end of the external union (see Figure 10).
Figure 10. Connecting the internal union
2. With red PEEK tubing provided in the MS Accessory Kit and the
high-pressure fingertight fittings and ferrules provided in the
Metal Needle Kit, connect the free end of the internal union to
the grounding union held by the grounding bar (see Figure 2 on
page 1).
Trademarks
Teflon is a registered trademark of E.I. du Pont de Nemours and
Company. PEEK is a trademark of Victrex PLC. Ion Max is a
trademark of Thermo Fisher Scientific Inc.
Tip Because you must insert the 0.009 in. OD stainless steel
needle through narrow orifices, Thermo Fisher Scientific
recommends the use of a magnifying glass for this procedure.
32-gauge, 0.009 in. OD
stainless steel needle
26-gauge, 0.01 in. ID electrospray needle that
protrudes from the nozzle of the ESI probe
Needle
Adapter nut
External union, natural PEEK
End wrapped with
Teflon tape
Do not cover the
thru hole with tape.
CAUTION Only use the supplied, non-conductive,
natural PEEK union to connect the ESI probe to
the grounding union. Using a conductive union
instead of the supplied, non-conductive union can
create an electric shock hazard when the mass
spectrometer is in operation.
Needle is flush with the
tip of the ferrule.
26-gauge electrospray
needle tip
32-gauge
stainless steel
needle protruding
by 1 mm
Nozzle on
the front of the
ESI probe
Non-conductive, natural PEEK internal union
5/16 in. open-ended wrench
  • Page 1 1
  • Page 2 2

Thermo Fisher Scientific 32-Gauge Stainless Steel Needle Installation guide

Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI