2
3. Insert the 32-gauge stainless steel needle through the
26-gauge needle that protrudes from the nozzle of the ESI probe
(see Figure 5). Then push the needle through the probe until it
protrudes from the sample inlet at the back of the probe
(see Figure 6).
Figure 5. Inserting the stainless steel needle into the electrospray needle
Step 2: Securing the Needle Position
YTo secure the needle position
1. Slip the adapter nut over the end of the 32-gauge stainless steel
needle that protrudes from the sample inlet port. Screw the
fitting, fingertight, into the port (see Figure 6).
Figure 6. Adapter nut shown installed in the sample inlet port
2. Wrap Teflon tape around the threads on one end of the natural
PEEK external union provided in the Metal Needle Kit
(see Figure 7). Take care to avoid taping over the thru hole.
Figure 7. External union and two-ferrule assembly
3. Slide the end of the needle that protrudes from the back of the
ESI probe through the Teflon-wrapped end of the external union.
4. Screw the Teflon-wrapped end of the external union into the
adapter, fingertight.
5. Slide a clear, 0.012 in. ID ferrule over the needle that protrudes
from the back of the ESI probe.
6. To ensure that you have installed the needle correctly, check the
following:
• One end of the 32-gauge stainless steel needle is flush with
the tip of the free ferrule (see Figure 8).
• The other end of the 32-gauge stainless steel needle does not
protrude from the front of the probe by more than 1 mm.
Figure 8. Back of the ESI probe with the needle shown in the correct position
For optimal spray quality, the tip of the 32-gauge stainless steel
needle must protrude by approximately 1 mm from the tip of
the 26-gauge electrospray needle (see Figure 9).
Figure 9. Front of the ESI probe with an enlarged view of the needle tip
Step 3: Connecting the Probe to the Grounding Union
YTo connect the ESI probe to the grounding union
1. With a 5/16 in. open-ended wrench, brace the external union as
you screw the internal union made of non-conductive, natural
PEEK onto the free end of the external union (see Figure 10).
Figure 10. Connecting the internal union
2. With red PEEK tubing provided in the MS Accessory Kit and the
high-pressure fingertight fittings and ferrules provided in the
Metal Needle Kit, connect the free end of the internal union to
the grounding union held by the grounding bar (see Figure 2 on
page 1).
Trademarks
Teflon is a registered trademark of E.I. du Pont de Nemours and
Company. PEEK is a trademark of Victrex PLC. Ion Max is a
trademark of Thermo Fisher Scientific Inc.
Tip Because you must insert the 0.009 in. OD stainless steel
needle through narrow orifices, Thermo Fisher Scientific
recommends the use of a magnifying glass for this procedure.
32-gauge, 0.009 in. OD
stainless steel needle
26-gauge, 0.01 in. ID electrospray needle that
protrudes from the nozzle of the ESI probe
External union, natural PEEK
End wrapped with
Teflon tape
Do not cover the
thru hole with tape.
CAUTION Only use the supplied, non-conductive,
natural PEEK union to connect the ESI probe to
the grounding union. Using a conductive union
instead of the supplied, non-conductive union can
create an electric shock hazard when the mass
spectrometer is in operation.
Needle is flush with the
tip of the ferrule.
26-gauge electrospray
needle tip
32-gauge
stainless steel
needle protruding
by 1 mm
Nozzle on
the front of the
ESI probe
Non-conductive, natural PEEK internal union
5/16 in. open-ended wrench