8
INTRODUCTION | GILSON CPC SYSTEMS
UNPACKING
Unpacking
Upon receipt of the instrument, inspect the exterior of the shipping box. It should arrive
unopened and undamaged. If examination reveals that damage has occurred in shipment,
notify the carrier and Gilson immediately. Refer to Customer Service on page 14.
Do not plug in the instrument if any damage is detected. Powering the instrument in a
damaged state may result in serious injury and may damage internal components of the
instrument.
The Gilson CPC Systems are delivered with most major components already assembled. Keep
the original container and packing assembly so the unit may be shipped safely, if necessary.
Carefully unpack the Gilson CPC System and its accessories. Verify that all parts are included
and undamaged. Do this now, even if the Gilson CPC System will not be used immediately.
Many carriers must receive concealed damage claims within seven days of delivery.
The CPC 100, CPC 250, and CPC 250 PRO can each weigh up to approximately
70kg(154lbs.) and an additional 10kg(22lbs.) with packaging. These systems are too heavy
to be lifted or moved by one person safely. To avoid personal injury and for general safety, if
moving or lifting the Gilson CPC System, always get another person to assist you.
The CPC 1000 and CPC 1000 PRO can weigh up to approximately 120kg(264lbs.) and an
additional 10kg(22 lbs.) with packaging. These systems are too heavy to be lifted or moved
by one or two persons safely. To avoid personal injury and for general safety, if moving or
lifting the Gilson CPC System, always get two additional persons (four in all) to assist you.
Do not attempt to lift the instrument from a valve. Always grip it from its base.
Gilson CPC Systems are shipped with rotor and fluid path filled with methanol and water.
Observe all laboratory safety precautions when handling solvents.
To unpack the instrument:
1. Open the box.
2. Remove the accessories packed inside the box.
3. Remove the packing material on top. The unit is not attached to the box, it is put on a
dedicated foam.
4. Lift the unit out of the box and place it at suitable location, such as a lab bench or cart, and
always on a plane surface.
It is recommended to operate the Gilson CPC Systems inside a fume hood to ensure
proper ventilation.
Allow sucient spacing around the system for proper cooling and for making
connections.
Allow at least 20 cm free space around the instrument at all times to allow for proper
tubing and cable connections and to provide adequate ventilation during operation.
Do not place item(s) on the Gilson CPC System.