Thermo Fisher Scientific HIGHPlate 2x5 Rotor User manual

Category
Fridge-freezers
Type
User manual
Thermo Scientific
Highplate Rotor 2 x 5
Instruction Manual
50134653-b • 07 / 2020
Visit us online to register your warranty:
thermofisher.com/labwarranty
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic
Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:
Thermo Scientific 75006444 Highplate Rotor 2 x 5 i
T
Preface ........................................................................................................................................ ii
Scope of Supply ........................................................................................................................ iii
Precautions ...............................................................................................................................iv
Chapter 1 Rotor Data ................................................................................................................................ 1-1
Chapter 2 Rotor Installation ................................................................................................................... 2-1
Rotor Installation ..................................................................................................... 2-2
Removing the Rotor ................................................................................................ 2-2
Chapter 3 Rotor Loading .......................................................................................................................... 3-1
Before a Run ............................................................................................................ 3-2
Proper Loading ........................................................................................................ 3-2
Improper Loading .................................................................................................... 3-2
Handling Micro Plates ............................................................................................. 3-3
Chapter 4 Aerosol-tight Applications ................................................................................................... 4-1
Basic Principles ........................................................................................................ 4-2
Fill Level .................................................................................................................. 4-2
Checking the Aerosol-Tightness .............................................................................. 4-2
Chapter 5 Maintenance and Care ......................................................................................................... 5-1
Cleaning Intervals .................................................................................................... 5-2
Cleaning .................................................................................................................. 5-2
Disinfection ............................................................................................................. 5-3
Decontamination ..................................................................................................... 5-4
Autoclaving .............................................................................................................. 5-5
Service of Thermo Fisher Scientific .......................................................................... 5-5
RCF-Values ...............................................................................................................A-1
Chemical Compatibility Chart ...............................................................................B-1
Table of Contents
Technical Data ........................................................................................................ 1-2
Thermo Scientific 75006444 Highplate Rotor 2 x 5 iii
P
Preface
Before starting to use the rotor, read through this instruction manual carefully and follow the
instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it
is forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the
expiration of the sellers warranty.
Scope of Supply
If any parts are missing, please contact your nearest Thermo Fisher Scientific representative.
Article Number Quantity Check
75006444 Highplate 2 x 5 1
76003500 Grease for rubber sealing 1
75003786 Bolt grease 1
50134653 Instruction manual 1
This symbol refers to general hazards.
WARNING means that injuries or material damage or contamination could occur.
CAUTION means that material damage could occur.
This symbol refers to biological hazards.
Observe the information contained in the instruction manual to keep yourself and your
environment safe.
Preface
iv75006444 Highplate Rotor 2 x 5 Thermo Scientific
Precautions
In order to ensure safe operation of the Highplate rotor 2 x 5, the following general safety
regulations must be followed:
Do not remove the magnet at the rotor bottom.
Do not use rotors which show any signs of corrosion and/or cracks.
Use only with rotors which have been loaded properly.
Never overload the rotor.
Operate the rotor always with the lid closed.
Use only accessories which have been approved by Thermo Fisher Scientific. Exceptions to this
rule are commercially available glass or plastic centrifuge tubes, provided they have been approved
for the speed or the RCF value of the rotor.
Please observe the safety instructions.
Please pay particular attention to the following aspects:
Rotor installation: Check that the rotor is locked properly into place before operating the
centrifuge.
Always balance the samples.
Maximum sample density at maximum speed: 1.2
g
cm3
-----------
Thermo Scientific 75006444 Highplate Rotor 2 x 5 1-1
1
Rotor Data
Contents
Technical Dataon page 1-2
1 Rotor Data
1-2 75006444 Highplate Rotor 2 x 5 Thermo Scientific
Technical Data
Technical Data
Table 1-1. 230 V, 50 / 60 Hz
Centrifuge Cryofuge 5500i Cryofuge 5500i Multifuge 4KR
Catalog Number 75004471 75004479 75004461
Weight of empty Rotor [ kg ]14 14 14
Maximum permissible Load [ g ] 2 x 500 2 x 500 2 x 500
Maximum Speed nmax [ rpm ] 5650 5650 5650
Maximum RCF-Value at nmax 5889 5889 5889
Radius max. / min. [ cm ] 16.5 / 10.5 16.5 / 10.5 16.5 / 10.5
Acceleration / Braking Time [ s ] 95 / 100 95 / 100 80 / 100
Sample Heating at nmax [ °C ] referred to Ambient
Temperature of 23-25 °C, Run Time 60 Minutes
00
*0
Maximum Speed nmax [ rpm ] at 4 °C 5650 5650 5650
Aerosol-tight** Yes Yes Yes
Autoclavable No No No
*at 200 V, 50 Hz up to 3 °C
**tested by TÜV Product Service GmbH - Hamburg
Table 1-2. 120 V, 60 Hz
Centrifuge Cryofuge 5500i
Catalog Number 75004475
Weight of empty Rotor [ kg ]14
Maximum permissible Load [ g ] 2 x 500
Maximum Speed nmax [ rpm ] 5650
Maximum RCF-Value at nmax 5889
Radius max. / min. [ cm ] 16.5 / 10.5
Acceleration / Braking Time [ s ] 95 / 100
Sample Heating at nmax [ °C ] referred to Ambient
Temperature of 23-25 °C, Run Time 60 Minutes
0
Maximum Speed nmax [ rpm ] at 4 °C 5650
Aerosol-tight*
*tested by TÜV Product Service GmbH - Hamburg
Yes
Permissible Temperature Range autoclavable °C No
1 Rotor Data
Thermo Scientific 75006444 Highplate Rotor 2 x 5 1-3
Table 1-3. 230 V, 50 / 60 Hz
Centrifuge Sorvall RC-4
Catalog Number 75004473 75004481
Weight of empty Rotor [ kg ]14 14
Maximum permissible Load [ g ] 2 x 500 2 x 500
Maximum Speed nmax [ rpm ] 5650 5650
Maximum RCF-Value at nmax 5889 5889
Radius max. / min. [ cm ] 16.5 / 10.5 16.5 / 10.5
Acceleration / Braking Time [ s ] 95 / 100 95 / 100
Sample Heating at nmax [ °C ] referred to Ambient
Temperature of 23-25 °C, Run Time 60 Minutes 00
*
Maximum Speed nmax [ rpm ] at 4 °C 5650 5650
Aerosol-tight** Yes Yes
Permissible Temperature Range autoclavable °C No No
*at 200 V, 50 Hz up to 3 °C
**tested by TÜV Product Service GmbH - Hamburg
Table 1-4. 120 V, 60 Hz
Centrifuge
Catalog Number 75004477
Weight of empty Rotor [ kg ]14
Maximum permissible Load [ g ] 2 x 500
Maximum Speed nmax [ rpm ] 5650
Maximum RCF-Value at nmax 5889
Radius max. / min. [ cm ] 16.5 / 10.5
Acceleration / Braking Time [ s ] 95 / 100
Sample Heating at nmax [ °C ] referred to Ambient
Temperature of 23-25 °C, Run Time 60 Minutes
3
Maximum Speed nmax [ rpm ] at 4 °C 5650
Aerosol-tight*
*tested by TÜV Product Service GmbH - Hamburg
Yes
Permissible Temperature Range autoclavable °C No
Sorvall RC-4
Sorvall RC-4
Technical Data
Thermo Scientific 75006444 Highplate Rotor 2 x 5 2-1
2
Rotor Installation
Contents
“Rotor Installationon page 2-2
“Removing the Rotoron page 2-2
2 Rotor Installation
Rotor Installation
2-2 75006444 Highplate Rotor 2 x 5 Thermo Scientific
Rotor Installation
Use only the rotors with approved accessories.
Proceed as follows:
1. Open the lid of the centrifuge and if necessary, remove any dust, foreign objects or residue from
the chamber.
Thread and O-ring must be clean and undamaged.
2. Hold the rotor over the centrifuge spindle.
3. Let the rotor slide slowly down the centrifuge spindle.
4. Thread the fastening tool into the centrifuge spindle clockwise. Hold the rotor with the other hand
into position.
5. Check if the rotor is properly installed by lifting it slightly on the handle.
6. If available close the rotor with the rotor lid.
Removing the Rotor
To remove the rotor, proceed as follows:
1. Open the centrifuge door.
2. Thread the fastening tool in the centrifuge spindle counterclockwise. Hold the rotor with the
other hand into position.
3. At the same time, pull the rotor directly upwards with both hands and remove it from the
centrifuge spindle. Make sure not to tilt the rotor while doing this.
CAUTION Unapproved or incorrectly combined accessories can cause serious damage to
the centrifuge.
CAUTION Do not force the rotor onto the centrifuge spindle.
WARNING Check for any damage to the rotor: Damaged rotors must not be used.
Keep the centrifuge spindle area of the rotor clear of objects.
CAUTION Check that the rotor is properly locked on the centrifuge spindle before each
use by pulling it at its handle.
Be sure to check all sealings before starting any aerosol-tight applications.
See the information in the rotor instruction manual.
2 Rotor Installation
Removing the Rotor
Thermo Scientific 75006444 Highplate Rotor 2 x 5 2-3
Aerosol-tight Rotors
When using an aerosol tight lid the rotor can only be removed with the lid closed. This is to protect
you and the samples.
Thermo Scientific 75006444 Highplate Rotor 2 x 5 3-1
3
Rotor Loading
Contents
“Before a Runon page 3-2
“Proper Loadingon page 3-2
“Improper Loadingon page 3-2
“Handling Micro Plateson page 3-3
3 Rotor Loading
Before a Run
3-2 75006444 Highplate Rotor 2 x 5 Thermo Scientific
Before a Run
1. Please read and observe the safety instructions contained in this operating instruction and in the
instructions for use.
2. Check the rotor and all accessory parts for damages such as cracks, scratches or traces of corrosion.
3. Check the rotor chamber, the centrifuge spindle and the fixation of the rotor for damages.
4. Check the rotors suitability using the chemical compatibility chart on page B-1.
5. Make sure the tubes or bottles do not touch the bucket lid.
6. Check the rotor bolts and apply grease 75003786 before using for the first time.
7. Check that each bucket or micro-plate carrier can swing freely by moving it carefully with your
hand. Weigh the bucket content (adapter and tube). Make sure you do not exceed the maximum
compartment load.
Proper Loading
To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times.
Improper Loading
CAUTION Always be sure to use buckets of the same weight class. Operate the rotor always
with the lid closed.
3 Rotor Loading
Handling Micro Plates
Thermo Scientific 75006444 Highplate Rotor 2 x 5 3-3
Handling Micro Plates
Always remove the plate holder (70057477) before loading and unloading it.
The rubber cushion (20056846) should be in the plate holder.
Let the plate holder (70057477) slide vertically into the bucket.
Deepwell plates can be inserted without plate holder (70057477).
Make sure the used micro plates can withstand the centrifugal forces.
Thermo Scientific 75006444 Highplate Rotor 2 x 5 4-1
4
Aerosol-tight Applications
Contents
“Basic Principleson page 4-2
“Fill Level” on page 4-2
“Checking the Aerosol-Tightnesson page 4-2
4 Aerosol-tight Applications
Basic Principles
4-2 75006444 Highplate Rotor 2 x 5 Thermo Scientific
Basic Principles
Check that the sample containers are well suited for the desired centrifugation process.
Fill Level
The tubes are only to be filled to a level which ensures that the sample is unable to reach the top of the
tube during centrifugation. Therefore, fill the tube only 2/3 of the rated level.
Checking the Aerosol-Tightness
The aerosol tightness testing of the rotors and buckets depend on the microbiological test process in
accordance with the EN 61010-2-020 Appendix AA.
Whether or not a rotor is aerosol-tight depends primarily on proper handling.
Check as needed to make sure your rotor is aerosol tight.
The careful inspection of the seals and seal surfaces for signs of wear and damage such as cracks,
scratches and embrittlement are extremely important.
Aerosol-tight applications are not possible if the lids are open.
Aerosol-tightness requires the correct operation when filling the sample vessels and closing the rotor lid.
Lubricate all seals lightly.
Always use the special grease 76003500 when lubricating the seals.
When you load a rotor before installing it in the centrifuge, make sure you do not tilt it when
installing.
CAUTION Aerosol-tight rotors and tubes may only be opened in an approved safety
work-bench when centrifuging dangerous samples.
Mind the maximum permissible load.
CAUTION Be sure to check all sealing before starting any aerosol-tight applications.
CAUTION This quick test is suited for validating the aerosol tightness of your
Highplate 2 x 5 rotor. Check the seals and sealing surfaces of the lid thoroughly.
CAUTION Prior to each use, the seals in the bucket are to be inspected in order to assure
that they are correctly seated and are not worn or damaged. Lubricate all seals lightly.
Damaged parts must be replaced immediately.
When loading the rotor, ensure that the rotor lid closes securely.
Damaged or clouded rotor lids are to be replaced immediately.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35

Thermo Fisher Scientific HIGHPlate 2x5 Rotor User manual

Category
Fridge-freezers
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI